K100143 · CAO Group, Inc. · GEX · Aug 13, 2010 · General, Plastic Surgery
Device Facts
Record ID
K100143
Device Name
PILOT DIODE LASER, MODEL 002-00139
Applicant
CAO Group, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Aug 13, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Pilot Diode Laser is indicated for the procedures of 1) Removal of lesions, excision, incision, vaporization, ablation, hemostasis, and photocoagulation on soft tissue in the medical fields of otolaryngology (ear, nose, and throat), dentistry and oral surgery, arthroscopy, gastroenterology, dermatology, podiatry, general surgery, urology, gynecology, and plastic surgery. And 2) Temporary relief of minor muscle and joint pain, stiffness, minor arthritis pain, muscle spasm, temporary increase in local blood circulation, and temporary relaxation of muscles by means of topical elevated tissue temperature from infrared spectral emissions.
Device Story
Pilot Diode Laser delivers 810nm laser energy to human soft tissue for surgical procedures and therapeutic pain relief. Device utilizes solid-state diodes to generate energy, transmitted via proprietary optical fiber system to surgical site. Handpiece includes disposable single-use tip; activation controlled by foot-actuated switch. Features include 630nm aiming beam, adjustable power output, and continuous or pulse delivery modes. Used by clinicians in various surgical specialties (e.g., dentistry, dermatology, general surgery). Healthcare providers use laser to perform tissue procedures or apply topical heat for pain relief; output affects tissue via thermal energy. Benefits include precise tissue interaction and non-invasive pain management.
Clinical Evidence
Bench testing only. Device demonstrated compliance with 21 CFR 1040.10 and 1040.11, and IEC 60601-1 constructional requirements. Performance testing confirmed delivery of laser energy levels required for intended surgical and therapeutic pain relief applications. No clinical data presented.
Technological Characteristics
Solid-state diode laser; 810nm wavelength; max 9W output. Fiber-optic delivery system with disposable single-use tips. 630nm aiming beam. Foot-actuated switch. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1, UL60601-1, 47 CFR 15/18, ETSI 301-489-1. Continuous and pulse delivery modes.
Indications for Use
Indicated for patients requiring soft tissue surgical procedures (lesion removal, excision, incision, vaporization, ablation, hemostasis, photocoagulation) across various medical specialties, or patients seeking temporary relief of minor muscle/joint pain, stiffness, arthritis pain, muscle spasms, and increased local blood circulation.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
THOR VR Single Diode Laser Treatment Probe (K070024)
Submission Summary (Full Text)
{0}------------------------------------------------
K 100143
## 510(k) Summary of Safety and Effectiveness
CAO Group, Inc. 4628 West Skyhawk Drive West Jordan, UT 84084 Tel: 801-256-9282 Fax: 801-256-9287 AUG 1 3 2010
く
Robert Larsen, Preparation Date: July 16, 2010
Device Name:
"ﺑ "
| Trade Name: | Pilot Diode Laser |
|-------------------------|-----------------------------------|
| Common Name: | 810nm Diode Laser |
| Product Classification: | Powered Laser Surgical Instrument |
# Legally Marketed Predicate Devices for Substantial Equivalence:
Velure Family of Diode Lasers, manufactured by Lasering S.R.L. (K063396)
DenLaser 800 Plus, manufactured by CAO Group, Inc. (K062619)
THOR VR Single Diode Laser Treatment Probe, manufactured by Thor International (K070024)
# Rationale for Substantial Equivalence:
The aforementioned laser devices and their accompanying delivery systems share similar indications for use on soft tissue with the submitted device for cutting soft tissue, similal indications for use on son took too will are therapy for soft tissue relaxation and temporary relief of pain. The predicate devices and submitted device share similar allo telliporary rener of pain. The prounded and laser delivery method. The design leatures including warelengan, of control systems, safety features, and performance monitoring. The devices share similar performance specifications including power output montoning. The devices share similar applications to soft tissue and indications for use.
# Description of Submitted Device:
The Pilot Diode Laser is a device for delivering laser energy to human soft tissue The Pliot Drocedures and treatments. This energy is generated by solidfor a variety of surgical procedures and treaments. The councy of laser energy at 810 ±
{1}------------------------------------------------
20nm for a maximum of 9 watts of energy output. The laser energy is delivered to surgical site by means of a proprietary optical fiber system, which allows for the safe transmission of laser energy to the site without creating undue risk to the patient or operatory staff by errant or collateral laser emissions. The device features some user definable settings, including a switchable 630nm aiming beam, adjustable power output for both the working beam and aiming beam, and continuous delivery or pulse delivery option.
The working end of the delivery fiber is contained within a metal handpiece with a disposable single-use tip. This handpiece system is provided with the device. The activation of the working beam diodes is completed by use of a foot-actuated switch.
# Intended Uses of the Pilot Diode Laser system:
The Pilot Diode Laser is indicated for the procedures of
- 1) Removal of lesions, excision, incision, vaporization, ablation, hemostasis, and photocoagulation on soft tissue in the medical fields of otolaryngology (ear, nose, and throat), dentistry and oral surgery, arthroscopy, gastroenterology, dermatology, podiatry, general surgery, urology, gynecology, and plastic surgery.
And
- 2) Temporary relief of minor muscle and joint pain, stiffness, minor arthritis pain, muscle spasm, temporary increase in local blood circulation, and temporary relaxation of muscles by means of topical elevated tissue temperature from infrared spectral emissions.
# Technological Characteristics and Substantial Equivalence:
The Velure Family of Diode Lasers uses solid state diodes to generate laser energy in the 810mm range. This system uses a fiber delivery system to transmit laser energy to the surgical site. The system also features a 635nm aiming beam and features oners's that allow for adjusting the output of the working beam, and switching between a continuous or pulsed-mode laser emission with pulse emissions adjustable from 0.1 - 20 Hz. The maximum output of the working beam is 15 watts.
The DenLaser 800 Plus uses solid state diodes to generate laser energy in the 810nm range. This system uses a fiber delivery system to transmit laser energy to the surgical site. The system also features a 630nm aiming beam and features controls that Surgical site: The system and Isation beam, and switching between a continuous anow for adjusting the carpar of the device features a wireless foot switch for actuating of pursed mode not The maximum output of the working beam is 5 watts.
{2}------------------------------------------------
The THOR VR Single Diode Laser Treatment Probe uses solid state diodes to generate laser energy in the 810nm range. This system employs direct emissions from a handpiece containing the laser diode to transmit laser energy to the surgical site. The system features controls that allow for adjusting the pulse rate of the beam from 2.5 -20,000 Hz. The maximum output of the unit is 0.45 watts.
#### Conformity to Standards:
The Pilot Diode Laser complies with the performance requirements of 21 CFR 1040.10 and 1040.11, with permissible deviations relative to Laser Notice 50, dated July 26, 2001. The device also complies with the entirety of IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1, UL60601-1, 47 CFR 15 and 18, and ETSI 301-489-1.
### Performance Data
The Pilot Diode Laser tested as compliant with 21 CFR 1040.10 and 1040.11, with permissible deviations relative to Laser Notice 50, dated July 26, 2001 and constructional requirements of IEC 60601-1. This device was tested and found to deliver laser energy necessary for the temporary relief of minor muscle and joint pain, etc. as described in the device's Indications for Use. The device was tested and passed CAO Group internal requirements for construction and performance.
### Conclusion
The Pilot Diode Laser is substantially equivalent to the listed laser surgical devices without raising any issues of safety or effectiveness. This device shares similar intended uses, and similar functional and performance characteristics. The device is designed to comply with relevant federal and international safety and performance standards.
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized image of an eagle or bird-like figure, with its wings spread and head turned to the right.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
CAO Group, Inc. % Mr. Robert K. Larsen Regulatory Affairs Manager 4628 West Skyhawk Drive West Jordan, Utah 84084
AUG 1 3 2010
Re: K100143
Trade/Device Name: Pilot Diode Laser Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX, ILY Dated: August 09, 2010
Received: August 13, 2010
Dear Mr. Larsen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Robert K. Larsen
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Sincerely yours,
Mark M. Millhusan
Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K100143
### Indications for Use
AS 1 8 2010
510(k) Number (if known):
Device Name: Pilot Diode Laser
Indications For Use:
The Pilot Diode Laser is indicated for the procedures of
- 1) Removal of Iesions, excision, incision, vaporization, ablation, hemostasis, and photocoagulation on soft tissue in the medical fields of otolaryngology (ear, nose, and throat), dentistry and oral surgery, arthroscopy, gastroenterology, dermatology, podiatry, general surgery, urology, gynecology, and plastic surgery.
And
- 2) Temporary relief of minor muscle and joint pain, stiffness, minor arthritis pain, muscle spasm, temporary increase in local blood circulation, and temporary relaxation of muscles by means of topical elevated tissue temperature from infrared spectral emissions.
x Over-The-Counter Use AND/OR Prescription Use (21 CFR 801 Subpart C) (Per 21 CFR 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil RP Ryden formxm
(Division of Surgical, Orthopedic, and Restorative Devices
Page I of of 1 -
510(k) Number K100143
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.