K122483 · Asante Solutions, Inc. · LZG · Jan 3, 2013 · General Hospital
Device Facts
Record ID
K122483
Device Name
PEARL DIABETES MANAGEMENT SYSTEM
Applicant
Asante Solutions, Inc.
Product Code
LZG · General Hospital
Decision Date
Jan 3, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Pearl Diabetes Management System is intended for continuous, subcutaneous delivery of insulin at programmable basal and bolus rates for the management of diabetes mellitus in adult patients requiring insulin.
Device Story
Pearl Diabetes Management System is a continuous, programmable insulin delivery system for adult patients with diabetes. System components include a controller unit with a user interface for programming basal and bolus delivery parameters, a single-use pump body containing the drive mechanism and battery, and an infusion set adapter. The device operates by applying controlled force to a pre-filled insulin cartridge to deliver insulin via an infusion set. Used in clinical or home settings, the device is operated by the patient or healthcare provider. Output consists of automated insulin delivery based on programmed settings, allowing for precise glycemic management. The system is a modification of the previously cleared Pearl System, maintaining the same fundamental drive mechanism and microprocessor-based controller.
Clinical Evidence
Bench testing only. No clinical data provided. Performance verification included accuracy testing, altitude and altitude shock, free fall, liquid ingress, positive and negative elevation, system occlusion, and software validation.
Indicated for continuous subcutaneous insulin delivery at programmable basal and bolus rates for management of diabetes mellitus in adult patients requiring insulin.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
{0}------------------------------------------------
# K 122483
Image /page/0/Picture/1 description: The image shows a logo with a stylized design. The logo features a circular element with abstract shapes inside, resembling a stylized letter or symbol. Below the logo, there is some text, but it is too blurry to read clearly. The overall design appears modern and minimalist.
Asante Solutions Inc. Pearl™ Diabetes Management System
Special 510(k) August 10, 2012
510(k) Summary Section 1.2
## Submitter Information
# JAN 0 3 2013
| Company: | Asante Solutions |
|---------------|----------------------------|
| | 352 East Java Drive |
| | Sunnyvale, CA 94089 |
| | (408) 716-5600 |
| | Robin Bush |
| | Regulatory Project Manager |
| Summary Date: | August 10, 2012 |
# Name and Classification
| Common Name: | Insulin Infusion Pump |
|----------------------|-----------------------------------|
| Proprietary Name: | Pearl™ Diabetes Management System |
| Classification Name: | Pump, Infusion, Insulin |
| Product Code: | LZG |
| Regulation Number: | 880.5725 |
| Class: | II |
## Predicate Devices
The Pearl Diabetes Management System has the same intended use as a number of commercially available medical devices, operates within an established range of insulin delivery, and uses existing established technology also known as Continuous Subcutaneous Insulin Infusion (CSII).
- Pearl Diabetes Management System (#K100567, Asante Solutions) a)
## Reference Devices
- Animas IR1200 (#K032257, Animas Corporation) a)
- D-TRONplus (K043000, Disetronic) b)
{1}------------------------------------------------
#### Intended Use
The Pearl Diabetes Management System is intended for continuous, subcutaneous delivery of insulin at programmable basal and bolus rates for the management of diabetes mellitus in adult patients requiring insulin.
#### Description of Device
The Pearl Diabetes Management System is a continuous, programmable insulin delivery system. It consists of a controller, a single use pump body and an infusion set adapter. The controller unit has a user interface to program delivery parameters for basal and bolus insulin delivery. It attaches to the pump body. The pump body provides the drive mechanism and battery power and holds the pre-filled insulin cartridges. The infusion set adapter is the conduit from the insulin cartridge to the final delivery tubing.
#### Summary of Technological Characteristics
The system uses existing established and unchanged technology and materials, relative to the previously cleared device. The system employs the same drive mechanism as that of the cleared Pearl System (#K100567) to apply controlled and accurate force to move the plunger in the insulin cartridge. The microprocessor based controller is unchanged.
#### Performance Testing
The system has been verified for performance and functionality to provide assurance that the device modifications have been designed and tested to assure conformance to the requirements for its intended use.
The following are the major tests that were performed on the Pearl Pump System to support the modifications:
- . accuracy
- altitude and altitude shock .
- . free fail
- . liquid ingress
- positive and negative elevation .
- system occlusion .
- software validation .
{2}------------------------------------------------
#### Conclusion
Performance testing on the modified device demonstrated that the Pearl Diabetes Management System performs reliably and delivers insulin as intended through the Asante Infusion Sets, and is safe for its intended use. Design verification testing confirmed that no new questions of safety or effectiveness were identified during device testing of the modified device.
Based upon the successful safety and performance tests and the similarities to predicate device, the modified Pearl Diabetes Management System is substantially equivalent to the predicate Pearl System in design, features, performance, fundamental scientific technology, and indications for use.
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines representing its body and wings. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" in a circular arrangement.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 3, 2013
Mr. Robin Bush Project Manager, Regulatory Affairs Asante Solutions, Incorporated 352 East Java Drive SUNNYVALE CA 94089
Re: K122483
Trade/Device Name: Pearl™ Diabetes Management System Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: LZG Dated: December 18, 2012 Received: December 18, 2012
Dear Mr. Bush:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Anthong La m
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Asante Solutions Inc. Pearl" Diabetes Management System
Special 510(k) August 10, 2012
Indications for Use Section 1.1
510(k) Number:
. asante
4122483
Device Name:
Pearl™ Diabetes Management System
Indications for Use: The Pearl Diabetes Management System is indicated for the continuous subcutaneous delivery of insulin at programmable basal and bolus rates for the management of diabetes mellitus in adult patients requiring insulin.
Prescription Use ﮯ (Part 21 CFR 801 Subpart D) Over-The-Counter Use _ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
and/or
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) |
|----------------------------------------------|
| Division of Anesthesiology, General Hospital |
| Infection Control, Dental Devices |
| 510(k) Number: | K122483 |
|----------------|---------|
|----------------|---------|
| Keith G. Marin | <div> Digitally signed by Keith G. Marin <br/> DN: c=US, o=U.S. Government, <br/> ou=HHS, ou=FDA, ou=People, <br/> cn=Keith G. Marin, <br/> 0.9.2342.19200300.100.1.1=00112 <br/> 50397 <br/> Date: 2013.01.03 16:07:51 -05:00 </div> |
|----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
Page 1 of 1
1-3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.