K100567 · M2 Group Holdings, Inc. · LZG · May 10, 2011 · General Hospital
Device Facts
Record ID
K100567
Device Name
PEARL DIABETES MANAGEMENT SYSTEM
Applicant
M2 Group Holdings, Inc.
Product Code
LZG · General Hospital
Decision Date
May 10, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended use of the Pearl Diabetes Management System is for continuous. subcutaneous delivery of insulin at programmable basal and bolus rates for the management of diabetes mellitus in adult patients requiring insulin.
Device Story
Pearl Diabetes Management System is a continuous, programmable insulin delivery system for adult patients with diabetes. System components include a controller, single-use pump body, single-use adapter, and accessories. Controller features a user interface for programming basal and bolus delivery parameters; attaches to pump body. Pump body houses drive mechanism and battery; utilizes pre-filled insulin cartridges. Adapter connects cartridge to infusion set and integrates an optical pressure sensor for occlusion detection. Device operates via a ratchet drive mechanism applying force to cartridge plunger. Dual-microprocessor controller monitors delivery safety. Used in clinical or home settings by patients; healthcare providers use output to manage insulin therapy. Benefits include precise, automated insulin administration.
Clinical Evidence
No clinical trials were conducted. Evidence consists of bench testing and human factors validation. Bench testing included accuracy, environmental stress (drop, altitude, shock), occlusion sensing, power management, and electrical/mechanical safety (EN 60601-1, IEC 60601-2-24, EN 60601-1-2). Biocompatibility testing followed ISO 10993-1, including cytotoxicity, sensitization, irritation, systemic toxicity, and genotoxicity assays. Human factors/usability studies under simulated use conditions verified product understanding, programming, and operation.
Technological Characteristics
Ratchet drive mechanism for insulin cartridge plunger; dual-microprocessor controller; optical pressure sensor for occlusion detection. Materials evaluated per ISO 10993-1. Electrical/mechanical safety per EN 60601-1 and IEC 60601-2-24; EMC per EN 60601-1-2. System is a programmable, battery-powered infusion pump.
Indications for Use
Indicated for continuous, subcutaneous insulin delivery at programmable basal and bolus rates for management of diabetes mellitus in adult patients requiring insulin.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
{0}------------------------------------------------
# 510(k) Summary
## Submitter Information
| Company: | Asante Solutions<br>1012 Stewart Drive<br>Sunnyvale. CA 94085<br>(408) 716-5600 |
|-----------------|---------------------------------------------------------------------------------|
| Contact Person: | Naghmeh Nouri<br>Vice President of Quality Assurance and Regulatory Affairs |
| Summary Date: | May 09, 2011 |
## Name and Classification
| Common Name: | Insulin Infusion Pump |
|----------------------|----------------------------------|
| Proprietary Name: | Pearl Diabetes Management System |
| Classification Name: | Pump, Infusion, Insulin |
| Product Code: | LZG |
| Regulation Number: | 880.5725 |
| Class: | II |
## Predicate Devices
The Pearl Diabetes Management System has the same intended use as a number of commercially available medical devices, operates within an established range of insulin delivery, and uses technology that is substantially equivalent to the identified predicates.
- Disetronic D-TRON and D-TRONplus (K994186, K043000). a)
- b) Animas IR1200 (K032257).
#### Intended Use
The intended use of the Pearl Diabetes Management System is for continuous. subcutaneous delivery of insulin at programmable basal and bolus rates for the management of diabetes mellitus in adult patients requiring insulin.
## Description of Device
The Pearl Diabetes Management System is a continuous, programmable insulin delivery system. It consists of a controller, a single use pump body, a single use adapter and associated accessories. The controller unit has a user interface to program delivery parameters for basal and bolus insulin delivery and attaches to the pump body. The pump body provides the drive mechanism and battery power for pre-filled insulin cartridges. The adapter connects the insulin cartridge to the infusion set and contains the occlusion sensor.
{1}------------------------------------------------
## 510(k) #K100567 Pearl Diabetes Management System
## Summary of Technological Characteristics
The system uses existing established technology, materials, and construction techniques. The system employs a ratchet drive mechanism to apply controlled and accurate force to the " plunger movement of the insulin cartridge. The dual microprocessor based controller allows safety monitoring for basal and bolus insulin delivery. Occlusion is detected by monitoring an optical based pressure sensor located in the adapter.
## Performance Testing
The system has been verified for performance and functionality to provide assurance that the proposed device has been designed and tested to assure conformance to the requirements for its intended use
The following tests were performed on the Pearl Pump System:
- . accuracy
- . free fall drop
- . altitude, altitude shock, and positive and negative elevation
- environmental testing .
- . occlusion sensing
- . power management
- shipping and packaging testing .
- single fault condition testing .
- Mechanical and electrical safety testing was performed in accordance to EN 60601-1 . and the specific requirements of IEC 60601-2-24
- Electromagnetic compatibility testing was performed in accordance to EN 60601-1-2. .
## User EvaluationTesting
User assessment evaluations and human factors validation studies were conducted under simulated use conditions to verify product understanding, programming modalities and pump operation. The users were provided a user manual and were trained by professional diabetic educators. Usability testing demonstrated the Pearl Diabetes Management System performs as designed and intended, and is safe and effective for its intended use.
{2}------------------------------------------------
Asante Solutions. Inc.
#### 510(k) #K100567 Pearl Diabetes Management System
## Biocompatibility Testing
The biocompatibility of the flow path of the Pearl Pump device was assessed in accordance with ISO 10993-1:2009 - Biological evaluation of medical devices, Part 1 - Evaluation and tests within a risk management process.
The biocompatibility tests listed below were conducted in accordance with the provisions of the FDA GLP requlations 21 CFR Part 58.
- Cytotoxicity MEM Elution Test using L-929 Mouse Fibroblast cells (ISO) .
- Sensitization Local Lymph Node Assay (LLNA) .
- Sensitization ISO Guinea Pig Maximization Sensitization Test .
- Irritation Intracutaneous Reactivity (Irritation) Test .
- Systemic Toxicity USP/ISO Systemic Injection Test .
- Systemic Toxicity USP/ISO Systemic Injection .
- Genotoxicity Salmonella Typhimurium Reverse Mutation Assay (Ames Test) .
- Genotoxicity In Vitro Mouse Lymphoma Assay .
- . Implantation -- ISO Intramuscular Implant Test 7-Day Duration in Rabbits
- Bacterial Endotoxin (LAL) USP Test .
The biocompatibility test results confirmed that the flow path materials are non-cytotoxic, non-sensitizing, non-irritating, not systemically toxic, and non-genotoxic when evaluated under the respective test conditions.
The flow path material has also been tested for insulin compatibility and was confirmed to be compatible with Humalog®.
## Conclusion
Based upon the successful safety and performance tests and the similarities to predicate devices, the Pearl Diabetes Management System is substantially equivalent to the predicate devices in design, features, performance, materials, packaging, fundamental scientific technology, and intended use.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight. The symbol is composed of three curved lines that suggest the wings and body of the bird.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Naghmeh Nouri Vice President, Quality and Regulatory Affairs M2 Group Holdings, Incorporated 1012 Stewart Drive Sunnyvale, California 94085
MAY 1 0 201
Re: K100567
વ્
·
Trade/Device Name: Pearl Diabetes Management System Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II * Product Code: LZG Dated: April 11, 2011 Received: April 12, 2011
Dear Ms. Nouri:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements
{4}------------------------------------------------
## Page 2 - Ms. Nouri
of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (2) CFR Part 820)
the Act): 21 CFR 1000 1050 the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiogical Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Wh for
Anthony Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Asante Solutions, Inc.
510(k) #K100567 Pearl Diabetes Management System
## Indications for Use
510(k) Number: K100567 Device Name: Pearl™ Diabetes Management System
For the continuous, subcutaneous delivery of insulin at Indication for Use: programmable basal and bolus rates for the management of diabetes mellitus in adult patients requiring insulin.
Prescription Use
(Part 21 CFR 801 Subpart D)
AND/OR Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Rhd C. Chapman 5/20/11
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K100567
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.