Straumann Temporary Abutments VITA CAD-Temp are indicated for use with Straumann Bone Level and Tissue Level implants for temporary crown and bridge restorations, and to maintain, stabilize and shape the soft tissue during the healing phase for up to six months, and should be placed out of occlusion.
Device Story
Straumann Temporary Abutments VITA CAD-Temp are dental components for temporary crown and bridge restorations. Devices attach to Straumann Bone Level or Tissue Level implants to support soft tissue healing. Constructed of PMMA with Titanium Alloy (Ti6Al7Nb) inlays. Used in dental clinics by clinicians. Placed out of occlusion to prevent premature loading. Benefits include soft tissue stabilization and shaping during the healing phase. Performance verified via fatigue testing (compressive and shear forces) per FDA guidance for root-form endosseous dental implants.
Clinical Evidence
No clinical data. Bench testing only. Fatigue testing performed in accordance with FDA guidance for root-form endosseous dental implants and abutments, evaluating compressive and shear (lateral) forces. Sterilization validation performed per ISO 17665-1 and ISO 17665-2 (moist heat, overkill method).
Technological Characteristics
Materials: Polymethyl methacrylate (PMMA) and Titanium Alloy (Ti6Al7Nb). Sensing/Actuation: None (mechanical component). Energy: None. Connectivity: None. Sterilization: Moist heat (ISO 17665-1/2). Form factor: Temporary abutment for dental implants.
Indications for Use
Indicated for patients requiring temporary crown and bridge restorations on Straumann Bone Level or Tissue Level implants; used to maintain, stabilize, and shape soft tissue during healing phase (up to 6 months); must be placed out of occlusion.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
Straumann NNC Post for Temporary Restoration (K111357)
Submission Summary (Full Text)
{0}------------------------------------------------
K122192
Straumann Temporary Abutments, PMMA Traditional 510(k) Section 5: 510(k) Summary
straumann
NOV 2 2012
### 1. Applicant's Name and Address
Straumann USA, LLC (on behalf of Institut Straumann AG) 60 Minuteman Rd. Andover, MA 01810 Telephone Number: 800-448-8168, ext 2513 Fax Number: 978-747-0023 Contact Person: Elaine Alan Requlatory Project Manager
- Date of Submission: 2. July 23, 2012
### 3. Name of the Device
| Trade Name: | Straumann Temporary Abutments VITA CAD-Temp |
|----------------------|---------------------------------------------|
| Common Name: | Temporary Abutments |
| Classification Name: | Endosseous Dental Implant Abutment |
| Regulation Number: | §872.3630 |
- 4. Legally Marketed Device to which Equivalence is Claimed (Predicate Device) K072679, Straumann NC Temporary Abutment K070478, Straumann RC Temporary Abutment
K051717, Straumann RN synOcta Temporary Abutment K111357, Straumann NNC Post for Temporary Restoration
### ર. Description of the Device
Straumann Temporary Abutments VITA CAD-Temp are temporary abutments intended for placement on Straumann Bone Level and Straumann Tissue Level implants of corresponding diameter. The temporary abutments are made of polymethyl methacrylate (PMMA), with a Titanium Alloy inlay.
### 6. Intended Use of the Device
Straumann Temporary Abutments VITA CAD-Temp are indicated for use with Straumann Bone Level and Tissue Level implants for temporary crown and bridge restorations, and to maintain, stabilize and shape the soft tissue during the healing phase for up to six months, and should be placed out of occlusion.
{1}------------------------------------------------
122192
# 7. Technological Characteristics
The proposed devices are substantially equivalent to the currently marketed devices. They share the same indication for use, prosthetic platforms, implant/abutment connections, fundamental operating principles, and contact duration of up to six months. Changes to the proposed devices are listed below followed by a comparison table, Table 1, listing each proposed and predicate device and the respective changes:
- · Material change from PEEK to PMMA, and from Titanium Alloy to PMMA,
- · Design changes of the coronal aspects of the abutments including length and diameter.
| Change | Proposed Device | Predicate Device |
|------------------------|-----------------------------------------------------------------|---------------------------------------|
| | NC Temporary Abutment, PMMA | NC Temporary Abutment, PEEK |
| Material | PMMA, polymethyl methacrylate | PEEK, polyetheretherketone |
| Length | 14.65mm | 14.75mm |
| Change | Proposed Device | Predicate Device |
| | RC Temporary Abutment, PMMA | RC Temporary Abutment, PEEK |
| Material | PMMA, polymethyl methacrylate | PEEK, polyetheretherketone |
| Length | 16.30mm | 16.40mm |
| Change | Proposed Device | Predicate Device |
| | RN Temporary Abutment, PMMA | RN synOcta Temporary Abutment, PEEK |
| Material | PMMA, polymethyl methacrylate | PEEK, polyetheretherketone |
| Change : | Proposed Device | Predicate Device |
| | WN Temporary Abutment, PMMA | RN synOcta Temporary Abutment, PEEK |
| Material | PMMA, polymethyl methacrylate | PEEK, polyetheretherketone |
| Diameter | 10.10mm | 7.0mm |
| Implant | Ø 4.8mm Wide Neck Tissue | Ø3.3, 4.1 and 4.8mm Regular Neck |
| Compatibility | Level Implant | Tissue Level Implants |
| Prosthetic<br>Platform | Ø 6.5mm | Ø 4.5mm |
| Change | Proposed Device | Predicate Device |
| | NNC Temporary Abutment, PMMA | NNC Post for Temporary Restoration, - |
| Material | PMMA, polymethyl methacrylate<br>Titanium Alloy Inlay, Ti6Al7Nb | Titanium Alloy, Ti6Al7Nb |
| Diameter | 5.0mm | 3.70 mm |
Table 1: Comparison Table
## 8. Performance Testing
Verification and validation testing were performed to ensure that the devices subject to this 510(k) Premarket Notification function as intended and that design input matches design output. Testing included:
Revised 510(k) Summary Straumann, USA, LLC
{2}------------------------------------------------
- 1. Performance Testing
- FDA guidance document "Class II Special Controls Guidance ﻨﺖ. Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments."
- 1. Fatigue testing was performed in accordance to the test set-up of the guidance document for systems that include polymeric components. Compressive and shear (lateral) forces were tested; successful testing was measured by the predetermined acceptance criteria which was based on the predicate devices. Angled abutment testing was not performed as no angled devices were proposed. The proposed devices met or surpassed the acceptance criteria.
- 2. Sterilization Validation
- i. Sterilization validation was carried out in accordance with ISO 17665-1 (overkill method, partial cycle): Sterilization of health care products - Moist heat - Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices, and ISO 17665-2: First edition 2009-01-15 Technical Specification Sterilization of health care products - Moist heat - Part 2: Guidance on the application of ISO 17665-1.
- 1. Requirements were met; there were no deviations to the applicable standards.
### 9. Conclusion
The results from the testing conducted demonstrated that the Straumann Temporary Abutments VITA CAD-Temp function as intended and met the predetermined acceptance criteria.
The results of the performance bench testing and risk analysis indicate that the Straumann Temporary Abutments VITA CAD-Temp are substantially equivalent to the named predicate devices.
Revised 510(k) Summary Straumann, USA, LLC
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" around the perimeter. Inside the circle is a stylized symbol that resembles a human figure embracing or protecting another person, represented by three curved lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
NOV 2 2012
Straumann USA, Limited Liability Company Ms. Elaine Alan Regulatory Project Manager 60 Minuteman Road Andover, Massachusetts 01810
Re: K122192
Trade/Device Name: Straumann Temporary Abutments VITA CAD-Temp Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: October 8, 2012 Received: October 9, 2012
Dear Ms. Alan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
### Page 2- Ms. Alan
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
ph foc
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use
510(k) Number (if known):
Device Name: Straumann Temporary Abutments VITA CAD-Temp
Indications for Use:
Straumann Temporary Abutments VITA CAD-Temp are indicated for use with Straumann Bone Level and Tissue Level implants for temporary crown and bridge restorations, and to maintain, stabilize and shape the soft tissue during the healing phase for up to six months, and should be placed out of occlusion.
X Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-the-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Runner
(Division Sign-Off) Division of Anesthesiology, General Hospita Infection Control, Dental Devices
510(k) Number: K122190
Response Letter: K122192 Straumann Temporary Abutments, PMMA Straumann, USA, LLC
October 5, 2012 Page 7 of 44
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.