The SomnoGuard® SP Soft is intended for the treatment of night-time snoring and mild to moderate obstructive sleep apnea (OSAS) in adults.
Device Story
SomnoGuard® SP Soft is a boil-and-bite mandibular advancement device (MAD) used to treat snoring and mild to moderate obstructive sleep apnea. The device consists of two independent trays (upper and lower) made of medical-grade polycarbonate shells with a thermoplastic copolymer lining. The trays are linked by a coupling protrusion mechanism using Polyoxymethylene (POM) connectors of varying lengths, allowing for up to 10 mm of protrusion and lateral jaw movement. A physician or trained staff member fits the device chair-side by heating it in hot water, molding it to the patient's teeth, and cooling it to set the impression. The device works by advancing the lower jaw to open the upper airway. It is intended for medium-term use (up to two years) and serves as a cost-effective alternative to custom-made dental appliances. It is used in clinical settings by dental or sleep specialists.
Clinical Evidence
No new clinical trials were conducted for this 510(k). Evidence relies on historical clinical data from studies in Germany, the U.S., and Belgium, alongside published literature (e.g., Pancer et al., 1998) and American Academy of Sleep Medicine guidelines. These sources confirm that oral appliance therapy is effective for managing snoring and reducing breathing arrests in mild to moderate OSAS patients. Imaging studies cited in literature demonstrate airway improvement with oral airway dilators.
Technological Characteristics
Two-part mandibular advancement device; medical-grade polycarbonate outer shells; thermoplastic copolymer inner lining; Polyoxymethylene (POM) connectors. Protrusion adjustment up to 10 mm via connector selection. Boil-and-bite fitting process. Non-electronic, mechanical device.
Indications for Use
Indicated for adults with night-time snoring and mild to moderate obstructive sleep apnea (OSAS).
Regulatory Classification
Identification
Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.
Special Controls
*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”
{0}------------------------------------------------
SomnoGuard® SP Soft 510(k) Premarket Application
# Section 5: 510(k) Summary
The following information is provided as required by 21 CFR § 807.87 for a 510(k) premarket notification. In response to the Safe Medical Devices Act of 1990, the following is a summary of the safety and effectiveness information upon which the substantial equivalence determination is based.
| Applicant: | Tomed Dr. Toussaint GmbH<br>Wiesenstr. 4<br>Bensheim, Germany D-64625<br>Phone: +49 6251 98 33 44<br>Facsimile: +49 6251 98 33 45 |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment Registration<br>Number: | 3006038700 |
| Manufacturing/<br>Distribution Address: | Tomed Dr. Toussaint GmbH<br>Wiesenstr. 4<br>Bensheim, Germany D-64625 |
| US Contact: | Scott Whitcomb, President/CEO<br>1st Line Medical, Inc.<br>854 US Route 3<br>Holderness, NH 03245<br>Phone: 603-968-3172<br>Facsimile: 603-968-3158<br>E-mail: SWhitcomb@1stLineMedical.com |
| Date submitted: | May 21, 2012 |
| Proprietary Name: | SomnoGuard® SP Soft¹ |
| Common Name: | Anti-Snoring / Sleep Apnea Device |
| Classification Status: | Class II per regulations §872.5570 |
| Product Codes: | LRK, |
| Predicate Devices: | SomnoGuard® AP (K061688), Tomed Dr. Toussaint GmbH |
SEP 2 8 2012
<sup>1</sup> In some countries there is a second model of SomnoGuard® SP, named SomnoGuard® SP Hard. Both models differ in the thermoplastic impression material only. SomnoGuard® SP Hard uses PCL (Polycaprolactone) instead of Elvax, giving a higher retention. SomnoGuard® SP Hard is not intended to be sold in the U.S.
{1}------------------------------------------------
### Device Description
The entire family of SomnoGuard mandibular advancement devices is used for treating snoring and mild to moderate obstructive sleep apnea. The principal effect of all appliances is the advancement of the lower jaw, thereby opening the upper airway and reducing snoring and the breathing arrests due to mild to moderate obstructive sleep apnea.
SomnoGuard can be fitted without the need of taking a patient's dental impressions as a prerequisite to construct the ready-to-use device. SomnoGuard devices are "boil-&-bite" appliances. To prepare fitting of the appliance it is necessary to heat the appliance in hot water that has been boiled for approximately 3:30 seconds. Thereafter, when the device is removed from the hot water and has cooled down for about 10 seconds, a physician or his or her trained staff inserts the device into the mouth of the patient, thereby first the lower jaw into the lower arch, then the upper jaw into the upper channel of the device. The patient then puts forward his or her lower jaw commonly at normal bite conditions to approximately half the maximum extension possible, with the plastic still warm and mouldable, and then firmly bites the plastic. When doing this, patient sucks in while closing the mouth and pressing the tongue against the inner surface of the front teeth. In parallel, doctor presses his or her fingers on the outer walls of the mouthpiece to make sure that the appliance is properly moulded. The process is completed by fixing the bite impression by rinsing the device in cold water.
Boil-&-bite appliances are primarily considered by sleep specialists and dentists for shortterm use up to about one year or as a first-line screening device to determine whether patients suffering from snoring and / or mild to moderate obstructive sleep apnea (OSAS) respond positively to oral appliance therapy, the outcome of which cannot be predicted for any appliance available in the marketplace prior to its usage. Thus in case of no therapeutic effect larger investments for the much more expensive custom dental lab made appliances can be avoided.
SomnoGuard® AP (AP = Adjustable Positioner) as well as SomnoGuard® SP Soft (SP = Sagittal Positioner) are two-part infinitely adjustable appliances enabling a protrusion of up to 10 mm, allowing lateral lower jaw movement and breathing through the mouth whenever needed. The devices consist of two independent trays each with a thermoplastic body, and both parts linked to each other by a coupling protrusion mechanism. The outer tray shells consist of solid, clear and transparent medical grade polycarbonate. The inner lining which
{2}------------------------------------------------
accommodates the teeth impressions is made of a thermoplastic copolymer which is similar to that used with the previously FDA cleared SomnoGuard® one-part appliances. After the oral appliance is heated in a hot water bath its thermoplastic body moulds easily to the teeth and jaws allowing any medical doctor to very easily fit the device chair side.
The SomnoGuard® SP Soft two connectors, made out of Polyoxymethylene (POM), connect the lower and the upper part of the appliance. Each device comes with 6 connectors of different length, enabling individual protrusion of up to 10 mm.
With the SomnoGuard® AP an adjusting screw made of stainless steel allows the anterior adjustment of the lower tray against the upper tray between 0 and about 10 mm or even more depending on the length of the screw used. The adjustment is only possible extra-orally and when the upper and lower trays are disassembling both trays is also necessary for cleaning.
The highly cost-effective device is considered for medium term use up to two years.
All "boil-&-bite" appliances are simple to fit by dentists and other medical specialists, taking about 10 minutes and not requiring any special tools. Since compliance and treatment outcomes cannot be predicted with any oral appliance currently available in the marketplace as previously mentioned, it always makes sense from a cost and economical point of view to initiate treatment with a more economical boil & bite device before subjecting patients to a much larger investment for the fabrication of a custom made dental appliances such as the SomnoGuard AP Pro.
#### Intented Use Statement
The SomnoGuard® SP Soft is intended for the treatment of night-time snoring and mild to moderate obstructive sleep apnea (OSAS) in adults.
### Comparison of Predicate Devices
SomnoGuard® SP Soft and SomnoGuard® AP use identical materials for the thermoflexible impression material and the hard shell trays but have different connectors between the upper and lower jaw trays.
{3}------------------------------------------------
SomnoGuard® SP Soft and SilentNite® are made out of different materials and SilentNite® is individually made but they share the same connecting mechanism out of the same materials between the upper and the lower jaw trays.
All devices have an identical intended use, the reduction of night-time snoring and mild to moderate obstructive sleep apnea in adults.
Substantial equivalence is based on non-clinical data as it relates to the historical significance of oral devices that reposition the jaw and reduce or manage snoring. Pancer et al. described this in an article published in the journal CHEST in 1998, where they concluded that mandibular advancement devices (MAD) were 95% successful in reducing and/or controlling snoring. In addition the American Academy of Sleep Medicine (formerly the American Sleep Disorders Association) published in 1995 their Standards of Practice and Guidelines which demonstrated that oral appliances were effective in the management of snoring.
### Clinical Data
Own clinical data from clinical studies in Germany, the U.S. and Belgium, as well a numerous number of articles referring to other MADs repeatedly looked at oral appliances and their use for the treatment of snoring confirm that oral appliance therapy is an effective means by which snoring can be managed and breathing arrests due to mild to moderate obstructive sleep apnea can be reduced. This effectiveness is embraced from a variety of aspects including safety, convenience and cost.
In conclusion, a number of studies have shown improvement of the airway utilizing imaging associated with the use of oral appliances (also referred to as Oral Airway Dilators), which aids in the management and reduction of snoring as well as breathing arrests due to mild to moderate obstructive sleep apnea.
Clinical and nonclinical data indicate that SomnoGuard® SP Soft is safe and effective for its intended use and performs as well as the predicate devices.
{4}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three wavy lines representing its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
SEP 28 2012
Tomed Dr. Toussaint GmbH C/O Mr. Scott Whitcomb President/Chief Executive Officer 1st Line Medical, Incorporated 854 US Route 3 Holderness, New Hampshire 03245
Re: K121761 Trade/Device Name: SomnoGuard® SP Soft Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices for Snoring and Intraoral Devices for Snoring and Obstructive Sleep Apnea Regulatory Class: II Product Code: LRK Dated: August 17, 2012 Received: September 12, 2012
Dear Mr. Whitcomb:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
#### Page 2- Mr. Whitcomb
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
hh for
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
- 510(k) Number (if known): K121761
Device Name: SomnoGuard® SP Soft
Indications For Use: The SomnoGuard® SP Soft is a mandibular advancement device and is intended for the treatment of night-time snoring and mild to moderate obstructive sleep apnea (OSAS) in adults.
Prescription Use _ X (Part 21 CFR 801 Subpart D) Over-The-Counter Use
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
AND/OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
Spom Reard
(Division Sign-Off)
Division of Anesthesiology, General Hospital
Infection Control, Dental, Devices
- 510(k) Number: K/2176/
Page 1 of _ 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.