The SomnoGuard Series of mandibular advancement devices (SomnoGuard,2.0,AP and AP Pro) is intended for the treatment of nighttime snoring and mild to moderate obstructive sleep apnea in adults.
Device Story
SomnoGuard series are mandibular advancement devices (MADs) used to treat snoring and obstructive sleep apnea by advancing the lower jaw to open the upper airway. The family includes 'boil-and-bite' models (SomnoGuard, 2.0, AP) and a custom lab-fabricated model (AP Pro). Boil-and-bite models are heated in water, molded to the patient's teeth by a clinician, and cooled to set. The AP and AP Pro models are two-part devices featuring an adjustable stainless steel coupling mechanism allowing protrusion up to 12 mm and lateral jaw movement. Used in clinics or sleep labs by dentists or sleep specialists; patients use them at home during sleep. The device provides a non-invasive mechanical intervention to improve airway patency, potentially reducing breathing arrests and snoring. The AP Pro is constructed in dental labs using plaster models and standard acrylic/elastomeric materials.
Clinical Evidence
No clinical data provided. The submission relies on bench testing and the established safety/effectiveness profile of the predicate device class for mandibular advancement therapy.
Technological Characteristics
Materials: Thermoflexible copolymeric material, medical grade polycarbonate, and stainless steel (AISI 304/1.4301 per DIN EN 10088-1). Design: Two-tray mandibular advancement system; boil-and-bite or custom lab-fabricated. Adjustment: Mechanical screw-based protrusion (0-12 mm) on AP/AP Pro models. Connectivity: None. Sterilization: Non-sterile. Software: None.
Indications for Use
Indicated for treatment of nighttime snoring and mild to moderate obstructive sleep apnea (OSAS) in adults.
Regulatory Classification
Identification
Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.
Special Controls
*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”
{0}------------------------------------------------
TOMED
GT
SEP - 8 2006
TOMED Dr. Toussaint GmbH · Lindberghstr. 3a · D - 64625 Bensheim
Dr. Toussaint GmbH
TOMED Dr. Toussaint GmbH Medical Products Lindberghstr. 3a D - 64625 Bensheim Germany
Tel.: +49-6251-98 33 44 Fax: +49-6251-98 33 45 eMail: w.toussaint@tomed.com http://www.tomed.com
## 510(k) SUMMARY of SAFETY and EFFECTIVENESS
#### A. General Information
- 1. Submitter's Name: TOMED Dr. Toussaint, GmbH
- 2. Address:
Lindberghstr. 3A Bersheim, Germany D-64625
- 3. Telephone: 49-6251-983344
- Dr. Winfried Toussaint 4. Contact Person:
- May 1, 2006 5. Date Prepared:
- 6. Registration Number: Pending
- B. Device
| 1. Name: | SomnoGuard Series (SomnoGuard, 2.0, AP, and AP Pro) |
|-------------------------|-----------------------------------------------------|
| 2. Trade Name: | Anti-Snoring / Sleep Apnea Device |
| 3. Common Name: | Anti-Snoring / Sleep Apnea Device |
| 4. Classification Name: | Device, Anti-Snoring |
| 5. Product Code: | LRK |
| 6. Class: | II |
| 7. Regulation Number: | 872.5570 |
{1}------------------------------------------------
TOMED Dr. Toussaint GmbH
TOMED Dr. Toussaint GmbH . Lindberghstr. 3a . D - 64625 Bensheim
TOMED Dr. Toussaint GmbH Medical Products Lindberghstr. 3a D - 64625 Bensheim Germany
Tel .: +49-6251-98 33 44 Fax: +49-6251-98 33 45 eMail: w.toussaint@tomed.com http://www.tomed.com
### C. Identification of Legally Marketed Devices
| 1. Names: | QuietKnight, Silencer (Version I. and II.), TheraSnore, TAP, Silencer, MDSA) |
|------------------|-------------------------------------------------------------------------------------------------------------------------|
| 2. K Number: | K962516, K033822, 033823, 926382, 972061, 954530, 042161 |
| 3. Date Cleared: | January 24, 1997, February 6, 2004, February 6, 2004, July 8, 1993, August 21, 1997, October 30, 1995, October 27, 2004 |
## D. Description of the Device
The whole family of SomnoGuard mandibular advancement devices is used for treating snoring and obstructive sleep apnea. The main effect of all appliances is the advancement of the lower jaw, thereby opening the upper airway and reducing snoring and the breathing arrests due to obstructive sleep apnea.
SomnoGuard and its minor modification SomnoGuard 2.0 are one-part appliances manufactured from thermoflexible copolymeric material. SomnoGuard can be fitted without the need of taking a patient's dental impressions as a prerequisite to construct the ready-to-use device. To prepare fitting of the appliance it is necessary to heat the appliance in hot water that has been boiled for around 30 seconds. Thereafter, when the device was removed from the hot water and has cooled down for about 10 seconds, a physician or his or her trained staff inserts the device into the mouth of the patient. thereby first the lower jaw into the lower arch, then the upper jaw into the upper channel of the device. The patient then puts forward his or her lower jaw commonly at normal bite conditions to approximately half the maximum extension possible, with the plastic still warm and moldable, and then firmly bites the plastic. When doing this, patient sucks in while closing the mouth and pressing the tongue against the inner surface of the front teeth. In parallel, doctor presses his or her fingers on the outer walls of the mouthpiece to
{2}------------------------------------------------
Dedicated to Customer Value
TOMED Dr. Toussaint GmbH
TOMED Dr. Toussaint GmbH . Lindberghstr. 3a . D - 64625 Bensheim
TOMED Dr. Toussaint GmbH Medical Products Lindberghstr. 3a D - 64625 Bensheim Germany
Tel .: +49-6251-98 33 44 Fax: +49-6251-98 33 45 eMail: w.toussaint@tomed.com http://www.tomed.com
make sure that the appliance is properly molded. After fixing the bite impression by cold water the fitting is finished. To raise the molar bite when necessary (e.g. in case of a "deep-bite") the SomnoGuard pack contains a filler strip made of the same copolymeric material as the appliance itself. To raise the bite the strip is heated in hot water and then sticked in parts into the channel of the appliance.
SomnoGuard 2.0 differs from SomnoGuard by a slightly thicker molar biting zone (i.e. 1.5 mm on each site of the biting plate) and an increase of the molar side walls of 1.0 mm. As opposed to the SomnoGuard time consuming use of the filler strip can thus be avoided with the SomnoGuard 2.0 in patients with a "deep bite".
Both SomnoGuard and - 2.0 contain a hole in the front area, large enough for emergency breathing. Both one-part appliances are primarily considered by sleep specialists and dentists for short-term use up to about one year or as a first-line screening devices to find out whether patients suffering from snoring and / or obstructive sleep apnea (OSAS) respond to oral appliance therapy at all what can't be predicted for any appliance available in the marketplace prior to its usage. Thus in case of no therapeutic effect larger investments for the much more expensive custom dental lab made appliances can be avoided.
The SomnoGuard product family comprises another "boil-&-bite" member i.e., SomnoGuard AP (AP = Adjustable Positioner), a two-part infinitely adjustable (titratable) appliance enabling a protrusion of up to 12 mm, allowing lateral lower jaw movement and breathing through the mouth whenever needed. The device consists of two independent trays with each a thermoplastic body, and both parts linked to each other by a coupling protrusion mechanism. The outer tray shells consist of solid clear and transparent medical grade polycarbonate. The inner lining which accommodates the teeth impressions is made of a thermoplastic copolymer as it is similarly used with the SomnoGuard® one-part appliances. After the oral appliance is heated in a hot water bath its thermoplastic body moulds easily to the teeth and jaws allowing any medical doctor to fit very easily the device chair side.
{3}------------------------------------------------
TOMED
Dr. Toussaint GmbH
TOMED Dr. Toussaint GmbH . Lindberghstr. 3a . D - 64625 Bensheim
TOMED Dr. Toussaint GmbH Medical Products Lindberghstr. 3a D - 64625 Bensheim Germany
Tel .: +49-6251-98 33 44 Fax: +49-6251-98 33 45 eMail: w.toussaint@tomed.com http://www.tomed.com
An adjusting screw made of stainless steel allows the anterior adjustment of the lower tray against the upper tray between 0 and about 12 mm or even more depending on the length of the screw used. The adjustment is only possible extra-orally and when the upper and lower trays are disassembling both trays is also necessary for cleaning.
By using the scale on both sides of the thread you can exactly control the adjustment with an accuracy of about 0.5 mm. Upper and lower trays can be moved laterally.
The highly cost-effective device is considered for medium term use up to two years
All "boil-&-bite" appliances are simple to fit by dentists and other medical specialists, taking about 10 minutes and not requiring any special tools. Since with not any existing oral appliance available in the marketplace treatment outcome can be predicted for sure as already mentioned earlier, it always makes sense from a cost and economical point of view to start treatment at first with cheaper boil & bite devices before doing a larger investment for the fabrication of custom made dental appliances such as the SomnoGuard AP Pro which will be described thereafter.
The currently last family member is the SomnoGuard AP Pro, an always dental lab made two-part titratable mandibular adjustable positioner with its development closely related and derived from the preceeding development of the AP appliance. The SomnoGuard® AP Pro can easilty be constructed from comment acrylic/elastomeric thermoform dental materials in any dental lab after taking impressions of the lower and upper jaws and producing plaster models. The components used to connect the upper and lower travs of the dental appliance and enable the infinite advancement of the lower jaw are made from stainless steel. The component's technology is based on the preceding development of the SomnoGuard® AP. The components are very durable, more or less indestructible, inexpensive and can most often be reused when the oral appliance has to be remade for some reason. The cost-effectiveness of the SomnoGuard AP Pro and the fact that the device can easily be constructed by any dental lab using standard lab equipment are
{4}------------------------------------------------
TOMED
TOMED Dr. Toussaint GmbH · Lindberghstr. 3a · D · 64625 Bensheim
TOMED Dr. Toussaint GmbH Medical Products Lindberghstr. 3a D - 64625 Bensheim Germany
Dr. Toussaint GmbH
Tel .: +49-6251-98 33 44 Fax: +49-6251-98 33 45 eMail: w.toussaint@tomed.com http://www.tomed.com
considered as key benefits compared to the more expensive competitive dental appliances.
All stainless steel components of the two-part appliances correspond to the material no.1.4301 (AISI-no. 304) resp. to DIN EN 10088-1. The chemical composition is commonly as follows:
C < 0,07 weight%, Cr 17,0 - 19,5 w%, Ni 8,0 - 10,5 w% und N < 0,11 w%, Si < 1,0 w%, Mn < 2,0 w% .
### E. Intended Use Statement
Indicated use: the SomnoGuard series of mandibular advancement devices is intended for the treatment of night-time snoring and mild to moderate obstructive sleep apnea (OSAS) in adults.
## F. Technological Characteristics Summary
Similarities between the devices are the following:
- Indications for Use ●
- Single Patient
- . Multi-Use .
- . Prescription Device
- . Non-Sterile
- . Custom Fabricated (Fit)
- Adjustable
- Environment - Home/Sleep Laboratories
- . Components
- Two Trays (Upper and Lower)
- Materials .
- Removable .
{5}------------------------------------------------
Dedicated to Customer Value
TOMED
ਪ
TOMED Dr. Toussaint GmbH • Lindberghstr. 3a • D - 64625 Bensheim
TOMED Dr. Toussaint GmbH Medical Products Lindberghstr. 3a D - 64625 Bensheim Germany
Dr. Toussaint GmbH
Tel.: +49-6251-98 33 44 Fax: +49-6251-98 33 45 eMail: w.toussaint@tomed.com http://www.tomed.com
The differences are minor and do not have a significant effect on the safety or effectiveness of the SomnoGuard Series.
{6}------------------------------------------------
Public Health Service
Image /page/6/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract caduceus or a representation of human figures.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# SEP - 8 2006
Dr. Winfried Toussaint Tomed Dr. Toussint GmbH Medical Products Lindberghstr. 3A D-64625. Bensheim GERMANY
Re: K061688
Trade/Device Name: Anti-Snoring/Sleep Apnea Device Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices for Snoring and Intraoral Devices for Snoring and Obstructive Sleep Apnea Regulatory Class: II Product Code: LRK Dated: July 28, 2006 Received: August 10, 2006
Dear Mr. Toussaint:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreated of the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Cobmette Prove, market the device, subject to the general controls provisions of the Act. The r ou may, diereleve, mains of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it rr your device is exassined controls. Existing major regulations affecting your device can may be subject to satil additions, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
#### Page 2 - Mr. Toussaint
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice-requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
K061688
#### Indications for Use
510(k) Number (if known): K061688
Device Name: SomnoGuard Series
Indications for Use:
The SomnoGuard Series of mandibular advancement devices (SomnoGuard,2.0,AP and AP Pro) is intended for the treatment of nighttime snoring and mild to moderate obstructive sleep apnea in adults.
Prescription Use _X
AND/OR
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH Office of Device Evaluation (ODE)
Susan Quane
Page I of 1
on Sign-Cif) on of Anesthesiology, General Hospital, on Control, Dental Devices
ີມການອກ
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.