SMITH & NEPHEW, INC. TOTAL KNEE SYSTEM INSTRUMENTATION
K121393 · Smith & Nephew, Inc. · MBH · Aug 7, 2012 · Orthopedic
Device Facts
Record ID
K121393
Device Name
SMITH & NEPHEW, INC. TOTAL KNEE SYSTEM INSTRUMENTATION
Applicant
Smith & Nephew, Inc.
Product Code
MBH · Orthopedic
Decision Date
Aug 7, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3565
Device Class
Class 2
Indications for Use
Smith & Nephew Total Knee Instruments are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew Total Knee Systems and their cleared Indications for Use.
Device Story
Smith & Nephew Total Knee System Instruments are surgical accessory devices used to assist in the implantation of Smith & Nephew knee prostheses. The instrument set includes trials, cutting instruments, cutting guides, alignment and sizing instruments, impactors, handles, clamps, extraction tools, torque instruments, guides, and protective covers. These instruments are used by orthopedic surgeons in an operating room setting to facilitate the precise placement and fixation of knee implants. They do not provide diagnostic output but serve as mechanical aids to ensure proper alignment and fit of the prosthetic components during primary or revision total knee arthroplasty. By assisting in the accurate preparation of the bone and positioning of the implant, these instruments help restore joint function and alleviate pain for patients suffering from degenerative or inflammatory joint disease.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Orthopedic surgical instrumentation including trials, cutting guides, alignment/sizing instruments, impactors, handles, clamps, and extraction tools. Materials and design are consistent with previously cleared Smith & Nephew instrument systems. No energy source or software components.
Indications for Use
Indicated for patients with rheumatoid arthritis, post-traumatic arthritis, osteoarthritis, degenerative arthritis, or failed previous knee surgeries (osteotomies, unicompartmental, or total knee replacement). Specific systems are indicated for primary and revision surgery where cruciate or collateral ligaments are incompetent or absent.
Regulatory Classification
Identification
A knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device is designed to achieve biological fixation to bone without the use of bone cement. This identification includes fixed-bearing knee prostheses where the ultra high molecular weight polyethylene tibial bearing is rigidly secured to the metal tibial base plate.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's guidance: “Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses; Guidance for Industry and FDA.” See § 888.1 for the availability of this guidance.
{0}------------------------------------------------
Summary of Safety and Effectiveness Total Knee System Instruments Smith & Nephew, Inc.
K12/393
Date of Summary: May 8, 2011
AUG 7 2012
Contact Person and Address Gino Rouss, MS Group Manager, Regulatory Affairs Smith & Nephew, Inc. Orthopaedic Division 7135 Goodlett Farms Parkway Memphis, Tennessee 38016 Tel: (901) 399-6707 FAX: (901) 566-7080
Name of Device: Total Knee System Instruments Common Name: Orthopaedic Surgical Instrumentation Device Classification Name and Reference:
| CFR Number | Description | Classification |
|------------|-------------------------------------------------------------------------------------------|----------------|
| 888.3510 | Knee joint femorotibial metal/polymer constrained cemented prosthesis | II |
| 888.3520 | Knee joint femorotibial metal/polymer non-constrained cemented prosthesis | II |
| 888.3530 | Knee joint femorotibial metal/polymer semi-constrained cemented prosthesis | II |
| 888.3540 | Knee joint patellofemoral polymer/metal semi-constrained cemented prosthesis | II |
| 888.3560 | Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis | II |
| 888.3565 | Knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis | II |
Device Class: Class II Panel Code: Orthopaedics/87
{1}------------------------------------------------
Summary of Safety and Effectiveness Total Knee System Instruments Smith & Nephew, Inc.
Predicate Devices:
Smith & Nephew Inc. Cemented Total Knee Systems
| Description | 510(k) | Clearance Date |
|---------------------------------------------------------------------------------|---------|----------------|
| GENESIS II TOTAL KNEE SYSTEM | K951987 | 8/22/1995 |
| GENESIS POROUS TIBIAL TRAYS & REVISION<br>KNEE SYSTEM | K953274 | 2/5/1996 |
| PROFIX POSTERIOR STABILIZED KNEE SYSTEM | K954909 | 1/31/1996 |
| GENESIS II POROUS PLUS HA KNEE SYSTEM | K032683 | 10/15/2003 |
| GENESIS II DEEP FLEXION CRUCIATE RETAINING<br>ARTICULAR INSERT | K041825 | 3/11/2005 |
| HIGH PERFORMANCE KNEE | K042515 | 3/14/2005 |
| PROFIX FLEX CRUCIATE RETAINING ARTICULAR<br>INSERT | K051229 | 7/20/2005 |
| JOURNEY CR KNEE SYSTEM | K101499 | 8/26/2010 |
| GENESIS II ZIRCONIUM FEMORAL<br>COMPONENT/PROFIX ZIRCONIUM FEMORAL<br>COMPONENT | K962557 | 12/5/1996 |
| GENESIS II POSTERIOR STABILIZED HIGH FLEXION<br>INSERT | K032295 | 8/21/2003 |
| PROFIX CONFORMING PLUS TIBIAL INSERT | K946236 | 7/6/1995 |
Smith & Nephew Inc. Cementless Total Knee Systems
| Description | 510(k) | Clearance Date |
|----------------------------------------------------------------|---------|----------------|
| GENESIS II POROUS PLUS HA KNEE SYSTEM | K032683 | 10/15/2003 |
| PROFIX TOTAL KNEE SYSTEM | K030623 | 5/22/2003 |
| GENESIS II TOTAL KNEE SYSTEM | K030612 | 5/27/2003 |
| LEGION POROUS + HA TIBIAL BASEPLATES | K100897 | 5/13/2010 |
| LEGION POROUS PRIMARY | K073325 | 12/20/2007 |
| LEGION POROUS PLUS HA PRIMARY FEMORAL<br>COMPONENTS | K091543 | 12/21/2009 |
| PROFIX FLEX CRUCIATE RETAINING ARTICULAR<br>INSERT | K051229 | 7/20/2005 |
| GENESIS II POSTERIOR STABILIZED HIGH FLEXION<br>INSERT | K032295 | 8/21/2003 |
| GENESIS II DEEP FLEXION CRUCIATE RETAINING<br>ARTICULAR INSERT | K041825 | 3/11/2005 |
{2}------------------------------------------------
Summary of Safety and Effectiveness Total Knee System Instruments Smith & Nephew, Inc.
Smith & Nephew Inc. Constrained Total Knee Systems
| Description | 510(k) | Clearance Date |
|--------------------------------------------------------------------------------------|---------|----------------|
| GENESIS II CONSTRAINED KNEE SYSTEM | K962137 | 8/2/1996 |
| REVISION KNEE SYSTEM | K043440 | 2/18/2005 |
| LEGION COBALT CHROME REVISION KNEE<br>SYSTEM | K060742 | 5/3/2006 |
| CROSSLINKED POLYETHYLENE ARTICULAR<br>INSERTS | K071071 | 9/19/2007 |
| LEGION STEM WITH HOLES | K072531 | 12/6/2007 |
| PROFIX PLUS TIBIAL INSERT, PROFIX PS PLUS<br>TIBIAL INSERT, PROFIX P/S TIBIAL INSERT | K963255 | 1/2/1997 |
| REVISION KNEE SYSTEM | K041106 | 7/1/2004 |
Smith & Nephew Inc. Hinged Total Knee System
| Description | 510(k) | Clearance Date |
|--------------------------|---------|----------------|
| LEGION HINGE KNEE SYSTEM | K081111 | 7/23/2008 |
### Device Description
Subject of this Traditional 510(k) Premarket Notification are the Smith & Nephew, Inc. Total Knee System Instruments. The subject devices are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew Total Knee Systems and their cleared Indications for Use. Smith & Nephew Total Knee System Instruments can be organized into instrument families which are categorized as follows: Trials, Cutting Instruments and Cutting Guides; Cutting Blocks, Alignment and Sizing Instruments, Impactors and Handles, Clamps, Extraction, Torque, Instrument Guides, and Covers and Protectors.
#### Intended Use / Indications for Use
Smith & Nephew Total Knee Instruments are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew Total Knee Systems and their cleared Indications for Use.
### Indications for Cruciate Retaining Cemented Knee Replacement
- 1. Rheumatoid arthritis.
- Post-traumatic arthritis, osteoarthritis, or degenerative arthritis. 2.
- 3. Failed osteotomies, unicompartmental replacement, or total knee replacement.
Cruciate Retaining Cemented Knee components are indicated for use with cement and are single use devices.
{3}------------------------------------------------
Indications for Cruciate Retaining Cementless Knee Replacement
- Rheumatoid arthritis. ﺎ ﻳﺘﻢ ﺍﻟﻤﺴﺘﻘﺒﺔ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤ
- 2. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis.
- Failed osteotomies, unicompartmental replacement, or total knee replacement. 3.
Cruciate Retaining Cementless Knee components are indicated for use without cement and are single use devices.
Indications for Posterior Stabilized Cemented Knee Replacement
- 1. Rheumatoid arthritis.
- 2. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis.
- Failed osteotomies, unicompartmental replacement, or total knee replacement. 3.
- 4. Primary and revision surgery, where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intact.
Posterior Stabilized Cemented Knee components are indicated for use with cement and are single use devices.
# Indications for Posterior Stabilized Cementless Knee Replacement
- 1. Rheumatoid arthritis.
- 2. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis.
- 3. Failed osteotomies, unicompartmental replacement, or total knee replacement.
- 4. Primary and revision surgery, where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intact.
Posterior Stabilized Cementless Knee components are indicated for use without cement and afe single use devices.
# Indications for Constrained Total Knees
- Rheumatoid arthritis. 1.
- Post-traumatic arthritis, osteoarthritis, or degenerative arthritis. 2.
- Failed osteotomies, unicompartmental replacement, or total knee replacement. 3.
- Constrained knee systems are designed for use in patients in primary and revision surgery, 4. where the posterior cruciate ligament and one or both of the collateral ligaments (i.e. medial collateral and/or lateral collateral ligament) are incompetent.
Constrained Total Knee components are indicated for use with cement and are single use devices.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's commitment to health, human services, and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Smith & Nephew, Inc. % Mr. Gino Rouss Group Manager, Regulatory Affairs 1450 E Brooks Road Memphis, Tennessee 38116
Re: K121393
Trade/Device Name: Smith & Nephew Inc. Total Knee System Instrumentation Regulation Number: 21 CFR 888.3565 Regulation Name: Knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: MBH, JWH, HRY, HSX, KRO, KRQ, KRR, NPJ Dated: May 8, 2012 Received: May 9, 2012
Dear Mr. Rouss:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
7 2012 AUG
{5}------------------------------------------------
# Page 2 - Mr. Gino Rouss
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Mark A. Milliken
Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# 510(k) Number (if known):
# Device Name: Smith & Nephew Inc. Hinged Total Knee Systems
# Indications for Use:
Smith & Nephew Hinge Total Knee Instruments are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew Hinge Total Knee Systems and their cleared' Indications for Use.
### Indications for Hinged Total Knees
- 1. Rheumatoid arthritis.
- 2. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis.
- Failed osteotomies, unicompartmental replacement, or total knee replacement. 3.
- Posterior stabilized knee systems are designed for use in patients in primary and revision বা surgery where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intact.
- Hinge knee systems are designed for use in patients in primary and revision surgery, where 5. the posterior cruciate ligament and one or both of the collateral ligaments (i.e. medial collateral and/or lateral collateral ligament) are absent or incompetent.
Hinged Total Knee components are indicated for use with cement and are single use devices.
Over-The-Counter Use x AND/OR Prescription Use (21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Andut
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K121393
Page of
{7}------------------------------------------------
# 510(k) Number (if known):
Device Name: Smith & Nephew Inc. Constrained Total Knee Systems
### Indications for Use:
inditorient to revew Constrained Total Knee Instruments are accessory devices and are intended to Smith & Rophen and and of Smith & Nephew Constrained Total Knee Systems and their cleared Indications for Use.
# Indications for Constrained Total Knees
- Rheumatoid arthritis. 1.
- Post-traumatic arthritis, osteoarthritis, or degenerative arthritis. 2.
- Failed osteotomies, unicompartmental replacement, or total knee replacement. 3.
- r uned restored thee systems are designed for use in patients in primary and revision surgery, 4. where the posterior cruciate ligament and one or both of the collateral ligaments (i.e. medial collateral and/or lateral collateral ligament) are incompetent.
Constrained Total Knee components are indicated for use with cement and are single use devices.
| Prescription Use<br>(Part 21 CFR 801 Subpart D) | X AND/OR |
|-------------------------------------------------|----------|
| Over-The-Counter Use<br>(21 CFR 807 Subpart C) | |
(Please do not WRITE BELOW THIS LINE -- Continue on ANOTHER PAGE IF NEÉDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number_________________________________________________________________________________________________________________________________________________________________
Page_ ੀ
{8}------------------------------------------------
510(k) Number (if known):
Device Name: Smith & Nephew Inc. Posterior Stabilized Cementless Knee Systems
Indications for Use:
Smith & Nephew Posterior Stabilized Cementless Knee Instruments are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew Posterior Stabilized Cementless Knee Systems and their cleared Indications for Use.
#### Indications for Posterior Stabilized Cementless Knee Replacement
- 1. Rheumatoid arthritis.
- Post-traumatic arthritis, osteoarthritis, or degenerative arthritis. 2.
- 3. Failed osteotomies, unicompartmental replacement, or total knee replacement.
- 4. Primary and revision surgery, where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intact.
Posterior Stabilized Cementless Knee components are indicated for use without cement and are single use devices.
Prescription Use x (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Aisul
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K121393
Page _________________________________________________________________________________________________________________________________________________________________________ of
{9}------------------------------------------------
### 510(k) Number (if known):
Device Name: Smith & Nephew Inc. Posterior Stabilized Cemented Knee Systems
# Indications for Use:
Smith & Nephew Posterior Stabilized Cemented Knee Instruments are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew Posterior Stabilized Cemented Knee Systems and their cleared Indications for Use.
# Indications for Posterior Stabilized Cemented Knee Replacement
- Rheumatoid arthritis. 1.
- Post-traumatic arthritis, osteoarthritis, or degenerative arthritis. 2.
- 3. Failed osteotomies, unicompartmental replacement, or total knee replacement.
- 4. Primary and revision surgery, where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intáct.
Posterior Stabilized Cemented Knee components are indicated for use with cement and are single use devices.
AND/OR
Prescription Use x (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 807 Subpart C)
(Please DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
As
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K121393
Page of
{10}------------------------------------------------
### 510(k) Number (if known):
Device Name: Smith & Nephew Inc. Cruciate Retaining Cementless Knee Systems
### Indications for Use:
Smith & Nephew Cruciate Retaining Cementless Knee Instruments are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew Cruciate Retaining Cementless Knee Systems and their cleared Indications for Use.
#### Indications for Cruciate Retaining Cementless Knee Replacement
- 1. Rheumatoid arthritis.
- Post-traumatic arthritis, osteoarthritis, or degenerative arthritis. 2.
- 3. Failed osteotomies, unicompartmental replacement, or total knee replacement.
AND/OR
Cruciate Retaining Cementless Knee components are indicated for use without cement and are single use devices.
Prescription Use × (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K121393
Page of
{11}------------------------------------------------
## 510(k) Number (if known):
Device Name: Smith & Nephew Inc. Cruciate Retaining Cemented Knee Systems
### Indications for Use:
Smith & Nephew Cruciate Retaining Cemented Knee Instruments are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew Cruciate Retaining Cemented Knee Systems and their cleared Indications for Use.
### Indications for Cruciate Retaining Cemented Knee Replacement
- 1. Rheumatoid arthritis.
- 2. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis.
- 3. Failed osteotomies, unicompartmental replacement, or total knee replacement.
Cruciate Retaining Cemented Knee components are indicated for use with cement and are single use devices.
| Prescription Use<br>(Part 21 CFR 801 Subpart D) | X |
|-------------------------------------------------|-------------------------------------------------------|
| | AND/OR Over-The-Counter Use<br>(21 CFR 807 Subpart C) |
(Please DO NOT WRITE BELOW THIS LINE – Continue on ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K121393
Page
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.