GENESIS II POSTERIOR STABILIZED HIGH FLEXION ARTICULAR INSERT
K032295 · Smith & Nephew, Inc. · JWH · Aug 21, 2003 · Orthopedic
Device Facts
Record ID
K032295
Device Name
GENESIS II POSTERIOR STABILIZED HIGH FLEXION ARTICULAR INSERT
Applicant
Smith & Nephew, Inc.
Product Code
JWH · Orthopedic
Decision Date
Aug 21, 2003
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Genesis II Posterior Stabilized High Flexion Articular Insert is indicated for: 1. Rheumatoid arthritis. 2. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis in older patients whose age, weight, and activity level are compatible with an adequate long-term result. 3. Failed osteotomies, unicompartmental replacement, or total knee replacement. 4. Posterior stabilized knee systems are designed for use in patients in primary and revision surgery, where the anterior and posterior cruciate ligaments are absent or incompetent and the collateral ligaments remain intact. The Genesis II Posterior Stabilized High Flexion Articular Insert is used with existing cemented femoral, tibial tray, and patellar components of the Genesis II Total Knee System cleared via K951987 and K953274. The Genesis II Posterior Stabilized High Flexion Articular Insert is for single use only. (Note: A second version of this statement exists for uncemented applications, specifying use with components cleared via K030612).
Device Story
Genesis II Posterior Stabilized High Flexion Articular Insert is a UHMWPE tibial component designed for use within the Genesis II Total Knee System. It enables increased flexion range for patients with appropriate anatomical capability. The device is intended for single use in primary or revision knee arthroplasty. It functions as part of a semi-constrained prosthesis, replacing damaged joint surfaces in patients with absent or incompetent cruciate ligaments. Surgeons implant the device in an operating room setting. The insert integrates with existing cemented or uncemented femoral, tibial tray, and patellar components. By providing a stable articulation surface, it restores joint function and mobility, potentially reducing pain and improving quality of life for patients with degenerative joint disease.
Clinical Evidence
Bench testing only. Mechanical test data demonstrated the device is equivalent to currently used clinical devices and capable of withstanding expected in vivo loading.
Indicated for patients with rheumatoid arthritis, post-traumatic arthritis, osteoarthritis, or degenerative arthritis requiring primary or revision total knee replacement. Suitable for patients with absent or incompetent anterior/posterior cruciate ligaments and intact collateral ligaments. Age, weight, and activity level must be compatible with long-term results.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
**AUG 21 2003**
K032295
p. 1/2
Smith & Nephew, Inc. Summary of Safety and Effectiveness Genesis II P/S High Flexion Articular Insert
Date of Summary: August 20, 2003
Contact Person and Address Jason Sells Regulatory/Clinical Affairs Specialist I Smith & Nephew, Inc. Orthopaedic Division 1450 East Brooks Road Memphis, TN 38116
Name of Device: Genesis II Postcrior Stabilized High Flexion Articular Insert Common Name: Articular insert
#### Device Description
(901) 399-5520
The Genesis II Posterior Stabilized High Flexion Articular Inserts are URMWPE tibial components which provide the ability for greater flexion to those patients who bave the anatomical capability to allow a greater flexion range. The insect is used with existing cemented femoral, tibial tray, and patellar components of the Genesis II Total Knee System cleared via K951987 and K953274 or with the system's porous, uncemented femoral and tibial tray components cleared in K030612.
| Identification of<br>Device | Product Code | Classification Name | Code | Predicate 510(k) | Indications for Use | | Predicate 510(k) |
|-------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|--------------------|--------------------|-----------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------|
| Genesis II PS<br>High Flex Insert<br>for use with<br>Genesis II Total<br>Knee System<br>Components in<br>cemented<br>applications | JWH -<br>Orthopedics<br>Panel/87 | Knee joint<br>patellofemorotibial<br>polymer/metal/polymer<br>semi-constrained<br>cemented prosthesis -<br>Class II | 21 CFR<br>888.3560 | K951987<br>K953274 | Identification of Device | Indications for Use | |
| Genesis II PS<br>High Flex Insert<br>for use with<br>Genesis II Total<br>Knee System<br>Components in<br>uncemented<br>applications | MBH -<br>Orthopedics<br>Panel/87 | Knee joint<br>patellofemorotibial<br>metal/polymer porous-<br>coated uncemented<br>prosthesis - Class II | 21 CFR<br>888.3565 | K030612 | Genesis II PS<br>High Flex Insert<br>for use with<br>Genesis II Total<br>Knee System<br>Components in<br>cemented<br>applications | The Genesis II Posterior Stabilized High Flexion Articular Insert<br>is indicated for:<br><br>1. Rheumatoid arthritis.<br>2. Post-traumatic arthritis, osteoarthritis, or degenerative<br>arthritis in older patients whose age, weight, and activity<br>level are compatible with an adequate long-term result.<br>3. Failed osteotomies, unicompartmental replacement, or<br>total knee replacement.<br>4. Posterior stabilized knee systems are designed for use in<br>patients in primary and revision surgery, where the<br>anterior and posterior cruciate ligaments are absent or<br>incompetent and the collateral ligaments remain intact.<br><br>The Genesis II Posterior Stabilized High Flexion Articular Insert<br>is used with existing cemented femoral, tibial tray, and patellar<br>components of the Genesis II Total Knee System cleared via<br>K951987 and K953274. The Genesis II Posterior Stabilized High<br>Flexion Articular Insert is for single use only. | K951987<br>K953274 |
| Genesis II PS<br>High Flex Insert<br>for use with<br>Genesis II Total<br>Knee System<br>Components in<br>uncemented<br>applications | The Genesis II Posterior Stabilized High Flexion Articular Insert<br>is indicated for:<br><br>1. Rheumatoid arthritis.<br>2. Post-traumatic arthritis, osteoarthritis, or degenerative<br>arthritis in older patients whose age, weight, and activity<br>level are compatible with an adequate long-term result.<br>3. Failed osteotomies, unicompartmental replacement, or<br>total knee replacement.<br>4. Posterior stabilized knee systems are designed for use inpatients in primary and revision surgery, where the<br>anterior and posterior cruciate ligaments are absent or<br>incompetent and the collateral ligaments remain intact.<br><br>The Genesis II Posterior Stabilized High Flexion Articular Insert<br>is used with existing cemented patellar and uncemented porous<br>tibial trays and femoral components of the Genesis II Total Knee<br>System cleared via K030612. The Genesis II Posterior Stabilized<br>High Flexion Articular Insert is for single use only. | K030612 | | | | | |
#### Device Classification
{1}------------------------------------------------
K032295
p. 2/2
### Mechanical and Clinical Data
A review of the mechanical test data indicated that the Genesis II Posterior Stabilized High Flexion Articular Insert is equivalent to devices currently used clinically and is capable of withstanding expected in vivo loading without failure.
### Substantial Equivalence Information
The substantial equivalence of the Genesis II Posterior Stabilized High Flexion Articular Insert is based on its similarities in indications for use, design features, operational principles, and material composition to the following prodicate devices - Smith & Nephew's Genesis II Total Knee System (K951987, K953274, K030612) and Zimmer's Nexgen LPS Knee System (K991581).
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 2 1 2003
Mr. Jason Sells Regulatory/Clinical Affairs Specialist I Smith & Nephew, Inc. Orthopaedic Division 1450 East Brooks Road Memphis, Tennessee 38116
Re: K032295
Trade/Device Names: Genesis II PS High Flex Insert for use with Genesis II Total Knee System Components in cemented applications and Genesis II PS High Flex Insert for use with Genesis II Total Knee System Components in uncemented applications Regulation Numbers: 21 CFR 888.3560 and 21 CFR 888.3565 Regulation Names: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis and Knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis Regulatory Class: II Product Codes: JWH and MBH
Dated: July 24, 2003 Received: July 25, 2003
Dear Mr. Sells:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or 10 devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Jason Sells
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable. the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
Sincerely yours,
Mark A. Mcllhenny
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosures
{4}------------------------------------------------
K032295
p. 1/2
# Genesis II Posterior Stabilized High Flexion Articular Insert Indications Statement
The Genesis II Posterior Stabilized High Flexion Articular Insert is indicated for:
- 1. Rheumatoid arthritis.
- 2. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis in older patients whose age, weight, and activity level are compatible with an adequate long-term result.
- 3. Failed osteotomies, unicompartmental replacement, or total knee replacement.
- 4. Posterior stabilized knee systems are designed for use in patients in primary and revision surgery, where the anterior and posterior cruciate ligaments are absent or incompetent and the collateral ligaments remain intact.
The Genesis II Posterior Stabilized High Flexion Articular Insert is used with existing cemented femoral, tibial tray, and patellar components of the Genesis II Total Knee System cleared via K951987 and K953274. The Genesis II Posterior Stabilized High Flexion Articular Insert is for single use only.
Mark N Millenn
ivision Sion-Division of General, Restorative and Neurological Devices 510(k) Number -
## Concurrence of CDRH, Office of Device Evaluation
Prescription Use (Per 21 CFR 801.109) OR
Over-the-Counter Use
{5}------------------------------------------------
K032295
p. 2/2
# Genesis II Posterior Stabilized High Flexion Articular Insert Indications Statement
The Genesis II Posterior Stabilized High Flexion Articular Insert is indicated for:
- 1. Rheumatoid arthritis.
- 2. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis in older patients whose age, weight, and activity level are compatible with an adequate long-term result.
- 3. Failed osteotomies, unicompartmental replacement, or total knee replacement.
- 4. Posterior stabilized knee systems are designed for use in patients in primary and revision surgery, where the anterior and posterior cruciate ligaments are absent or incompetent and the collateral ligaments remain intact.
The Genesis II Posterior Stabilized High Flexion Articular Insert is used with existing cemented patellar and uncemented porous tibial trays and femoral components of the Genesis II Total Knee System cleared via K030612. The Genesis II Posterior Stabilized High Flexion Articular Insert is for single use only.
Mark N Milken
Orvision of General. Restorative and Neurological I
510(k) Number .
## Concurrence of CDRH, Office of Device Evaluation
Prescription Use (Per 21 CFR 801.109) ાર
Over-the-Counter Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.