KimVent* Microcuff* Subglottic Suctioning Endotracheal Tubes are indicated for airway management by oral intubation of the trachea and for removal of secretions that accumulate in the subglottic space.
Device Story
Endotracheal tube for airway management; facilitates oral intubation and subglottic secretion removal. Features include Murphy Eye, separate suction lumen with dorsal opening, and suction valve with integrated rinse port for sterile saline or air bolus administration. Device is disposable, sterile, and single-use. Used in clinical settings by healthcare professionals. Suctioning mechanism allows removal of accumulated secretions above the cuff, potentially reducing risk of complications. Barrel-shaped polyurethane cuff provides seal against tracheal wall.
Clinical Evidence
Bench testing only. Performance testing confirmed suctioning efficiency and conformance to ISO 5356-1:2004 and ISO 5361:1999. Biocompatibility testing performed per ANSI/AAMI/ISO 10993 parts 3, 5, 6, and 10. Sterilization validated per ANSI/AAMI/ISO 11135-1:2007 and 11607-1:2006.
Technological Characteristics
PVC tube with barium sulphate radiopaque line; barrel-shaped polyurethane inflatable cuff; suction system with ABS valve; polypropylene vent connector. Sterilized via Ethylene Oxide (EO). Complies with ISO 5356-1 and ISO 5361. Biocompatibility per ISO 10993.
Indications for Use
Indicated for airway management via oral tracheal intubation and subglottic secretion removal in patients requiring endotracheal intubation.
Regulatory Classification
Identification
A tracheal tube is a device inserted into a patient's trachea via the nose or mouth and used to maintain an open airway.
{0}------------------------------------------------
# 23 Kimberly-Clark Corporation
# Section 3 - 510(K) Summary
K120985
DEC 1 9 2012
### Date Summary was Prepared:
December 19, 2012
510(k) Submitter:
Marcia Johnson, RAC Technical Leader, Regulatory Affairs Kimberly-Clark Health Care 1400 Holcomb Bridge Road Roswell, GA 30076 Ph: 770.587.8566 FAX: 920.380.6351 Email: marcia.johnson@kcc.com
Primary Contact for this 510(k) Submission:
Marcia Johnson, RAC Technical Leader, Regulatory Affairs Kimberly-Clark Health Care 1400 Holcomb Bridge Road Roswell, GA 30076 Ph: 770.587.8566 FAX: 920.380.6351 Email: marcia.johnson@kcc.com
Device Trade Name: KimVent* Microcuff* Subglottic Suctioning Endotrachea! Tubes
Device Common names:
Endotracheal Tubes
Device Product Codes and Classification Names:
BTR Class II Endotracheal Tubes (21 CFR 868.5730)
Predicate Devices:
The Kimberly-Clark* KimVent* Microcuff* Subglottic Suctioning Endotracheal Tubes are substantially equivalent to the predicate device, TaperGuard Evac™ Endotrachael Tubes (K090352).
Device Description:
KimVent* Microcuff* Subglottic Suctioning Endotracheal Tubes are available in adult sizes 7.0, 7.5, 8.0, 8.5, and 9.0mm with a Murphy Eye. They include a separate lumen with a dorsal opening above the cuff to provide access to the subglottic space. The subglottic space is accessed via a normally open suction valve that includes a one-way port for rinsing the subglottic space with sterile saline (0.9% Sodium Chloride solution) or administering an air bolus to assist in maintaining a patent suction lumen patency. These devices are sold as disposable, sterile, single-use, devices.
Section 3 - 510(K) Summary
Page 1 of 4 ·
{1}------------------------------------------------
# දිනු Kimberly-Clark Corporation
Intended Use:
KimVent* Microcuff* Subglottic Suctioning Endotracheal Tubes are indicated for airway management by oral intubation of the trachea and for removal of secretions that accumulate in the subglottic space.
Technological Characteristics Both the Kimberly-Clark* KimVent* Microcuff* Subglottic Suctioning Endotracheal Tubes and the predicate TaperGuard Evac™ Endotracheal Tubes have the same basic fundamental technological characteristics. Both are polyvinylchloride tubes with inflatable cuffs and include a suction lumen. The Kimberly-Clark* KimVent* Microcuff* Subglottic Suctioning Endotracheal Tube incorporates a barrel-shaped polyurethane inflatable cuff while the predicate TaperGuard Evac™ Endotracheal Tubes contain a tapered PVC cuff. The Kimberly-Clark* KimVent* Microcuff* Subglottic Suctioning Endotracheal Tubes incorporate a suction valve with integrated rinse port to aid in removing secretions that accumulate in the subglottic space while the predicate TaperGuard Evac™ Endotracheal Tubes contain a single port for access to the subglottic space. Below is a comparison table that summarizes the technological characteristics of the subject and predicate tubes:
| | Predicate Device -<br>TaperGuard Evac™<br>Endotrachael Tubes<br>(K090352) | Subject Device -<br>Kimberly-Clark*<br>KimVent* Microcuff*<br>Subglottic<br>Suctioning<br>Endotracheal Tubes |
|-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Materials | | |
| Endotracheal Tube | Polyvinylchloride Tube<br>(PVC) with Radiopaque<br>line | PVC tube<br>w/Barium Sulphate<br>Radiopaque line |
| Pilot Balloon<br>Assembly | Plasticol or PVC Pilot<br>Balloon<br>One-way valve (Bespak<br>valve) = PVC, Plastic<br>Acetal copolymer, Nitrile<br>Rubber, stainless steel<br>Inflation Tail = PVC | PVC Pilot Balloon<br>One-way valve<br>(Bespak Check Valve)<br>= PVC, Nitizile/Acetal/<br>stainless steel<br>Inflation Tube/Tail =<br>PVC |
| Inflatable Pressure<br>Cuff | PVC with a tapered-<br>shape | Polyurethane with a<br>barrel-shape |
| Suction System | PVC | PVC Tube with<br>Acrylonitrile Butadiene<br>Styrene (ABS) valve |
| Vent Connector | PVC | Polypropylene |
Section 3 - 510(K) Summary
Page 2 of 4
{2}------------------------------------------------
| | Predicate Device - | Subject Device - |
|------------|----------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| | TaperGuard Evac™<br>Endotrachael Tubes<br>(K090352) | Kimberly-Clark*<br>KimVent* Microcuff*<br>Subglottic<br>Suctioning<br>Endotracheal Tubes |
| Ink | Medical Grade Ink | All inks were confirmed<br>to be non-cytotoxic,<br>non-irritating and non-<br>sensitizing through<br>appropriate ISO 10993<br>testing. |
| Adhesives | Solvents used for<br>assembly:<br>Cyclohexanone, Balloon<br>Glue,<br>Cyclohexanone/Scanning<br>compound No. 5. | UV-Cure Adhesives,<br>Cyanoacrylate<br>Adhesive |
| Sizes | 6.0, 6.5, 7.0, 7.5, 8.0,<br>8.5, 9.0 (mm) | 7.0, 7.5, 8.0, 8.5, 9.0<br>(mm) |
| Shelf-life | 5-year | Launch with 2-year,<br>will extend to 5-year<br>using established<br>protocols |
Summary of Testing:
Bench-top performance testing was conducted to confirm suctioning efficiency and additional bench-top testing was conducted to assure conformance to the following standards.
- · ISO 5356-1:2004, Anesthetic and respiratory equipment -Conical connectors: Part 1: Cones and sockets:
- ISO 5361:1999, Anesthetic and respiratory equipment -. Tracheal tubes and connectors.
Biocompatibility testing was conducted to assure conformance to the following standards:
- ANSI/AAMI/ISO 10993-3: 2003 Biological Evaluation of Medical Devices - Part 3: Tests for Genotoxicity, Carcinogenicity, and Reproductive Toxicity,
- ANSI/AAMI/ISO 10993-5:2009, Biological evaluation of medical devices - Part 5: Tests for In Vitro cytotoxicity
- · ANSI/AAMI/ISO 10993-6, 2007 Biological Evaluation of Medical Devices - Part 6: Tests for Local Effects After Implantation
- ANSI/AAMI/ISO 10993-10: 2010: Biological evaluation of medical devices - Part 10:Tests for Irritation and Skin Sensitization
#### Section 3 ~ 510(K) Summary
{3}------------------------------------------------
# නු Kimberly-Clark Corporation
The product is EO sterilized and validated according to the following standards:
- · ANSI/AAMI/ISO 10993-7:2008, Biological evaluation of medical devices, Part 7: Ethylene oxide sterilization residuals, Section 4.3.3 Prolonged Exposure Devices.
- · ANSI/AAMI/ISO 11135-1:2007 Sterilization of health care products - Ethylene oxide - Part 1: Requirements for development, validation, and routine control of a sterilization process for medical devices.
- · ANSI/AAMI/ISO 11607-1 2006 Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems.
Results of all testing met acceptance criteria.
## Conclusion:
The Kimberly-Clark* KimVent* Microcuff* Subglottic Suctioning Endotracheal Tubes are substantially equivalent to the predicate device, TaperGuard Evac™ Endotracheal Tubes (K090352), in intended use, design, performance, principles of operation, and both are intended for single use. Test results confirm that the device is as safe, as effective, and performs as well as or better than the predicate device.
### Section 3 - 510(K) Summary
{4}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle-like symbol with three curved lines representing wings or feathers. The symbol is positioned to the right of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA".
#### Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Public Health Service
## December 19. 2012
Ms. Marcia Johnson. RAC Technical Leader, Regulatory Affairs Kimberly-Clark Corporation 1400 Holcomb Bridge Road ROSWELL GA 30076
Re: K120985
Trade/Device Name: KimVent* Microcuff* Subglottic Suctioning Endotracheal Tubes Regulation Number: 21 CFR 868.5730 Regulation Name: Tracheal Tube Regulatory Class: II Product Code: BTR Dated: November 26, 2012 Received: November 28, 2012
Dear Ms. Johnson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 -- Ms. Johnson
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
**Kwame** Q. Ulmer
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Section 2 - Indications for Use
510(k) Number (if known): 1520985
Device Name: KimVent" Microcuff" Subglottic Suctioning Endotracheal Tubes
Indications for Use:
KimVent" Microcuff" Subglottic Suctioning Endotracheal Tubes are indicated for airway management by oral intubation of the trachea and for removal of secretions that accumulate in the subglottic space.
Prescription Use × (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
### Concurrence of CDRH, Office of Device Evaluation (ODE)
# Lester W. Schultheis Jr 2012.12.19 1,0:59:44 -05′00″
(Division Sign-Off) Division of Anestheslology, General Hospital Infection Control, Dental Devices
510(k) Number;
#### Section 2 - Indications for Use
Page 8 of 63
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.