The Hi-Lo, Intermediate Hi-Lo, and TaperGuard™ cuffed endotracheal tubes are indicated for oral/nasal intubation of the trachea for anesthesia and for general airway management. Tubes that are pre-cut to special lengths are intended for oral intubation only. The TaperGuard Evac™ Endotracheal tube is indicated for airway management by oral/nasal intubation of the trachea and for evacuation or drainage of the subglottic space. The TaperGuard™ Endotracheal tube is a device inserted into a patient's trachea via the nose or mouth to maintain an open airway.
Device Story
TaperGuard and TaperGuard Evac endotracheal tubes are sterile, single-use devices for airway management; inserted via nose or mouth to maintain open airway. Evac models include secondary lumen for subglottic drainage via capped Luer connector. Features include Magill curve, hooded tip with Murphy Eye, and Tip-To-Tip radiopaque line for visualization. Tapered PVC cuff design reduces micro-aspiration by average of 90% compared to barrel-shaped cuffs; associated with decreased nosocomial pneumonia (VAP) rates. Used in clinical settings by healthcare providers; output is physical airway maintenance and drainage capability. Benefits include improved seal against micro-aspiration and reduced VAP risk.
Clinical Evidence
No clinical data provided. Claims regarding 90% reduction in micro-aspiration and decreased nosocomial pneumonia rates are based on comparative performance against predicate devices.
Technological Characteristics
Sterile, single-use PVC endotracheal tubes with tapered cuff. Features standard 15mm connector, Magill curve, Murphy Eye, and radiopaque line. Evac models include secondary lumen for subglottic drainage. No electronic components or software.
Indications for Use
Indicated for patients requiring oral or nasal tracheal intubation for anesthesia, general airway management, and subglottic drainage (Evac models).
Regulatory Classification
Identification
A tracheal tube is a device inserted into a patient's trachea via the nose or mouth and used to maintain an open airway.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains a sequence of handwritten characters, which appear to be a combination of letters and numbers. The sequence reads as 'K090352'. The characters are written in a bold, black ink, which contrasts with the white background. The style of writing is cursive, with some characters connected to each other.
# Special 510(k) Summary
#### 1. Company Identification
Covidien, (formerly Nellcor Puritan Bennett, Inc.) 6135 Gunbarrel Avenue Boulder, CO 80301
APR - 3 2009
Establishment Registration: 2936999
#### 2. Contact Person
Ted Kuhn Regulatory Affairs Associate II Phone: (303) 305-2272 Fax: (303) 305-2212 Email: ted.kuhn@covidien.com
#### 3. 510(k) Preparation Date
February 10, 2009
#### 4. Device Name
TaperGuard™ Endotracheal Tubes and TaperGuard Evac™ Endotracheal Tubes. Trade Name: Endotracheal Tube, Tracheal Tube Common Name:
#### 5. Device Classification
Class II Tracheal Tube (21 CFR 868.5730), classification code 73 BTR
#### 6. Intended Use
The Hi-Lo, Intermediate Hi-Lo, and TaperGuard™ cuffed endotracheal tubes are indicated for oral/nasal intubation of the trachea for anesthesia and for general airway management. Tubes that are pre-cut to special lengths are intended for oral intubation only.
#### 7. Indications for Use
The TaperGuard Evac™ Endotracheal tube is indicated for airway management by oral/nasal intubation of the trachea and for evacuation or drainage of the subglottic space.
The TaperGuard™ Endotracheal tube is a device inserted into a patient's trachea via the nose or mouth to maintain an open airway.
{1}------------------------------------------------
#### 8. Claims
TaperGuard™ Endotracheal Tubes;
- > The TaperGuard™ PVC Tapered Cuff Endotracheal tubes reduce micro aspiration by an average of 90%, in comparison to the Hi-Lo Basic Barrel Shaped PVC Cuff Endotracheal tube.
TaperGuard Evac™ Endotracheal Tubes;
- The TaperGuard PVC Tapered Cuff Evac™ Endotracheal tubes reduce micro aspiration by an ﺎﺭ average of 90%, in comparison to the Hi-Lo Evac Barrel Shaped PVC Cuff Endotracheal tube.
In specific clinical circumstances, use of the TaperGuard Evac™ Endotracheal Tubes has been associated with a decrease in the rate of nosocomial pneumonia (VAP).
#### 9. Description of Device
TaperGuard™ Endotracheal Tube (5.0, 5.5, 6.0, 6.5, 7.0, 7.5, 8.0, 8.5, 9.0, 9.5, 10mm)
#### TaperGuard Evac™ Endotracheal Tube - Oral (6.0, 6.5, 7.0, 7.5, 8.0, 8.5, 9.0mm)
The TaperGuard™ and the TaperGuard Evac™ Tracheal Tubes are sterile, single-use devices supplied with a standard 15mm connector. On the Evac tubes, in addition to the main lumen, the tube has a separate Evac lumen which has a dorsal opening above the cuff. Access to the lumen is accomplished via a clear connecting tube with a capped Luer connector. The tube incorporates a Magill curve, a hooded tip with a Murphy Eye and a Tip-To-Tip™ radiopaque line to assist in radiographic visualization.
#### 10. Substantial Equivalence
The predicate devices to which we claim equivalence are the Hi-Lo Cuffed Tracheal Tubes (K871204), Hi-Lo Evac™ Endotracheal Tubes (K965132), SealGuard Endotracheal Tubes and SealGuard Evac™ Endotracheal Tubes (K082520).
The TaperGuard™ Endotracheal tubes and the TaperGuard Evac™ Endotracheal tubes maintain the same intended use as the predicate devices. It is a device inserted into the trachea through the mouth or nose to facilitate breathing.
The TaperGuard™ Endotracheal tubes, TaperGuard Evac™ Endotracheal tubes and the predicate devices consist of the same fundamental technology.
The TaperGuard™ device differs from the predicate device, SealGuard, in that the tapered cuff is made of PVC for TaperGuard™ and PU for SealGuard. Both devices have a tapered cuff.
The TaperGuard™ device differs from the predicate device, Hi-Lo, in that the TaperGuard™ has a tapered cuff and the Hi-Lo has a barrel shaped cuff. The cuff material for both devices is PVC.
The TaperGuard™ and TaperGuard Evac™ device in all clinical applications will be identical to the use of predicate devices. From a Human Factors perspective none of the changes in any way affect the operation or usability of the tube or cuff.
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three overlapping bars representing its wings. The eagle faces right and is positioned to the right of a circular text element. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" and is arranged around the circumference of the circle.
Food and Drug Administration 200 Corporate Boulevard Rockville MD 20850
Mr. Ted Kuhn Regulatory Affairs Associate II Respiratory & Monitoring Solutions Covidien 6135 Gunbarrel Avenue Boulder, Colorado 80301
APR - 3 2009
Re: K090352
> Trade/Device Name: TaperGuard™ Trachael Tubes and TaperGuard Evac™ Trachael Tubes Regulation Number: 21 CFR 868.5730
Regulation Name: Trachael Tube Regulatory Class: II
Product Code: BTR Dated: March 20, 2009 Received: March 23, 2009
#### Dear Mr. Kuhn:
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 – Mr. Kuhn
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Susan Cuomo
Ginette Y. Michaud. M.D. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
**510(k) Number:** K090352
Device Name: TaperGuard™ Tracheal Tubes and TaperGuard Evac™ Tracheal Tubes
Indications for Use:
The TaperGuard Evac™ Endotracheal tube is indicated for airway management by oral/nasal intubation of the trachea, and for evacuation or drainage of the subglottic space.
The TaperGuard™ Endotracheal tube is a device inserted into a patient's trachea via the nose or mouth to maintain an open airway.
Prescription Use (Part 21 CFR 801 Subpart D) C)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) (Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Infection Control, etc.
Confidential & 510(K) Number: 120810352 - Covidien
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.