The hip prosthesis is designed for cementless use in total hip arthroplasty in primary or revision surgery. The patient should be skeletally mature. The patient's condition should be due to one or more of: · Severely painful and/or disabled joint: as a result of osteoarthritis, post-traumatic arthritis, rheumatoid arthritis or psoriactic arthritis, Congenital hip dysplasia, Ankylosing spondylitis. · Avascular necrosis of the femoral head. · Acute traumatic fracture of the femoral head or neck. · Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement where sufficient bone stock is present.
Device Story
Versafitcup CC Trio - Additional Liners are acetabular components for total hip arthroplasty; designed for use with Medacta Total Hip Prosthesis System (femoral stems, modular heads). Liners consist of HighCross highly crosslinked ultra-high molecular weight polyethylene (HXUHMWPE). Components are supplied sterile for single-use. Surgeons implant these devices during hip replacement procedures to restore joint function and alleviate pain in patients with degenerative or traumatic hip conditions. Device provides bearing surface for femoral head; facilitates articulation within acetabular shell. Benefits include improved joint mobility and pain reduction.
Clinical Evidence
Bench testing only. Mechanical performance evaluated via comparison to worst-case predicate liners, specifically assessing range of motion, liner-to-shell connection stability, and wear characteristics.
Technological Characteristics
Acetabular liners composed of HighCross (HXUHMWPE). Designed for cementless fixation in total hip arthroplasty. Supplied sterile for single-use. No electronic components or software.
Indications for Use
Indicated for skeletally mature patients requiring primary or revision total hip arthroplasty due to osteoarthritis, post-traumatic/rheumatoid/psoriatic arthritis, congenital hip dysplasia, ankylosing spondylitis, avascular necrosis, acute traumatic fracture, or failed previous hip surgery (reconstruction, fixation, arthrodesis, hemiarthroplasty, surface replacement, or total hip replacement) with sufficient bone stock.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
{0}------------------------------------------------
K120531
Page 1/3
APR 1 8 2012
Image /page/0/Picture/3 description: The image shows the logo for Medacta International. The logo consists of a stylized letter 'M' formed by two triangles, followed by the word 'edacta' in a simple, sans-serif font. A horizontal line underlines the word 'edacta', and below that is the word 'International' in a smaller font, followed by a symbol resembling a plus sign within a square.
## 510(k) Summary
Applicant/Sponsor: Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66
Mr. Adam Gross Contact Person: Director of Regulatory and Quality Medacta USA 4725 Calle Quetzal, Unit B Camarillo, CA, 93012 Phone: (805)437-7085 Fax: (805)437-7553 Email: AGross@medacta.us.com
Date Prepared: February 10, 2012
# DEVICE INFORMATION
Trade/Proprietary Name: Versafitcup CC Trio - Additional Liners Common Name: Acetabular Liners Classification Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis
> 21 CFR 888.3353 Class II Device Product Codes: ŁZO, MEH
Predicate Devices: K103352 Versafitcup CC Trio, Medacta International K103721 Mpact Acetabular System, Medacta International K092265 Versafitcup Double Mobility Highly Crosslinked, Medacta International
Versafitcup CC Trio - Additional Liners
Section 5 - Page 2 of 4
{1}------------------------------------------------
120531
Page 2/3
### Product Description
The Versafitcup® CC Trio family of acetabular components is designed to be used with the Medacta Total Hip Prosthesis System. The Medacta Total Hip Prosthesis system includes the Quadra S, H, R, and C Stems and CoCrMo and ceramic ball heads (K072857, K073337, K080885, K082792, K083558, and K112115). The AMStem femoral stems also work with the Medacta Total Hip Prosthesis System (K093944, K103189). The Medacta Total Hip Prosthesis System is a total hip replacement system consisting of the femoral stem made of metal, a modular femoral head made of metal or ceramic, and acetabular components. The Versafitcup® CC Trio acetabular components that are the subject of this 510(k) consist of two new flat fixed liners that are made of HighCross® highly crosslinked ultra-high molecular weight polyethylene (HXUHMWPE).
All the Versafitcup® CC Trio components are supplied sterile in single-use individual packages.
## Indications for Use
The hip prosthesis is designed for cementless use in total hip arthroplasty in primary or revision surgery.
The patient should be skeletally mature.
The patient's condition should be due to one or more of:
· Severely painful and/or disabled joint: as a result of osteoarthritis, post-traumatic arthritis. rheumatoid arthritis or
psoriactic arthritis. Congenital hip dysplasia, Ankylosing spondylitis.
· Avascular necrosis of the femoral head.
· Acute traumatic fracture of the femoral head or neck.
· Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement where sufficient bone stock is present.
#### Comparison to Predicate Devices
The Versafitcup CC Trio - Additional Liners have the same intended use as the previously cleared Versafitcup CC Trio (K103352). The Versafitcup CC Trio - Additional Liners have the same material and are similar in size to the liners cleared in the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image contains a sequence of characters and numbers. The sequence starts with the letter 'K', followed by the numbers '120', then '531'. The characters and numbers are written in a handwritten style, with varying stroke thicknesses.
Page 3/3
# Performance Testing
The Versafitcup CC Trio - Additional Liners were compared to the worst case liners of the predicate devices in regards to the mechanical tests applicable to these products including range of motion, instability of connection between liner and acetabular shell, and wear. Since the Versafitcup CC Trio - Additional Liners are less critical than the worst case of the predicate devices, the Versafitcup CC Trio - Additional Liners do not introduce any new issues in regards to safety and effectiveness.
#### Conclusion:
Based on the above information, the Versafitcup CC Trio - Additional Liners can be considered substantially equivalent to its predicate devices.
## Versafitcup CC Trio - Additional Liners
Section 5 - Page 4 of 4
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure, composed of three curved lines that suggest wings or feathers. The logo is black and white.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Medacta International % Medacta USA Mr. Adam Gross Director of Regulatory and Quality 4725 Calle Quetzal, Unit B Camarillo, California, 93012
APR 1 8 2012
Re: K120531
Trade/Device Name: Versafitcup CC Trio - Additional Liners Regulation Number: 21 CFR 888.3353
Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis.
Regulatory Class: Class II Product Code: LZO, MEH Dated: March 21, 2012 Received: March 22, 2012
Dear Mr. Gross:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
## Page 2 -- Mr. Adam Gross
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information.on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
FOR Pete O
Mark N. Melkerson Dep Lenda
Director Division of Surgical, Orthopedic
and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K120531
Page 1/1
# Indications for Use
510(k) Number (if known):
Device Name: Versafitcup CC Trio - Additional Liners
Indications for Use:
The hip prosthesis is designed for cementless use in total hip arthroplasty in primary or revision surgery.
The patient should be skeletally mature.
The patient's condition should be due to one or more of:
· Severely painful and/or disabled joint: as a result of osteoarthritis, post-traumatic arthritis, rheumatoid arthritis or
psoriactic arthritis, Congenital hip dysplasia, Ankylosing spondylitis.
· Avascular necrosis of the femoral head.
· Acute traumatic fracture of the femoral head or neck.
· Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement where sufficient bone stock is present.
Prescription Use × (21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
D. Ouel
vision - ign-Off) Versafitsion OCSTirigica Additions of Spers Febandines of 2012 Devices
ວ່າຈາລາວປາກO ໄ (1)O-ugil SuoisiviO)
510(k) Number
510(k) Number K120531
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.