K120492 · K2 Medical GmbH & Co. KG · DXC · Apr 30, 2012 · Cardiovascular
Device Facts
Record ID
K120492
Device Name
K2 MEDICAL VASCULAR CLAMPS
Applicant
K2 Medical GmbH & Co. KG
Product Code
DXC · Cardiovascular
Decision Date
Apr 30, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4450
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
K2 Medical Vascular Clamps are devices intended for temporary or partial occlusion of blood vessels during vascular surgical procedures.
Device Story
K2 Medical Vascular Clamps are reusable surgical instruments used for temporary or partial occlusion of blood vessels. Available in various configurations, including ring-handled clamps with ratchet closures or spring-loaded bulldog clamps. Surgeons select specific clamp styles based on vessel anatomy, size, and desired occlusion level (full or partial) to minimize trauma. Used in vascular surgical procedures by surgeons to isolate vessel segments or stop blood flow. Devices are supplied non-sterile and require sterilization before use.
Clinical Evidence
Bench testing only. Biocompatibility and sterilization studies were successfully completed to demonstrate performance comparable to the predicate device.
Technological Characteristics
Materials: Stainless Steel (ASTM F 899-07) or Titanium Alloy. Mechanism: Mechanical occlusion via spring, cross-action, or ratchet-closure ring handles. Form factor: Various vascular clamp styles (bulldog, ring handle, etc.). Sterilization: Non-sterile (user-sterilized).
Indications for Use
Indicated for temporary or partial occlusion of blood vessels during vascular surgical procedures in patients requiring vascular surgery.
Regulatory Classification
Identification
A vascular clamp is a surgical instrument used to occlude a blood vessel temporarily.
{0}------------------------------------------------
# 510(k) Summary as required by section 807.92(c)
date prepared February 13th, 2012
#### Submission Applicant:
K2 Medical GmbH & Co.KG Unter Buchsteig 3a 78532 Tuttlingen, GERMANY
## Establishment Registration Number:
3007648354
## Official Correspondent:
Mr. Peter Fetzer K2 Medical GmbH & Co.KG Unter Buchsteig 3a 78532 Tuttlingen, GERMANY
Phone: 049-7462-200-49-0
# Trade name:
K2 Medical Vascular Clamp
## Common name:
Various vascular clamps:
Bulldog Clamps, Vessel Clamps, Microvascular Bulldog Clamps, Ring Handle Bulldog Clamps, Neonatal Vascular Clamp, Pediatric Clamps, Occlusion Clamps, Cardio Clamps, Cardio Clamps, Aorta Clamps, Multi-Purpose Clamps, Blood Vessel Forceps, Anastomosis Clamps, Vena Cava Clamps, Peripheral vascular Clamps, Aotic Aneurysm Clamps, Abdominal Clamps, Profunda Clamps, Shunt Clamps, Aortic Clamps, Patent-Ductus Clamps, Fogarty-Type Clamps, Vascular Forceps, Microvascular Forceps, Minimally Invasive Coronary Systems
, , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , ,
## Classification name:
Clamp, Vascular (21 CFR 870.4450, Product code DXC)
## Regulation Description
Vascular Clamp .
## Substantial Equivalence Claims:
K092544 INSTRUMED VASCULAR CLAMPS
> INSTRUMED INTERNATIONAL, INC. 626 Cooper Court Schaumburg, IL 60173
APR 3 0 2012
{1}------------------------------------------------
#### Description of the Device:
The K2 Medical Vascular Clamos are reusable surgical instruments. To ensure the multi-purpose use of this device, many different models are available. The differences can be as follows:
- Some have ring handles with a ratchet closure to adjust the amount of tension applied to . the vessel for occlusion or partial occlusion.
- Bulldog style Vascular Clamps are another type of vascular cross clamp that use a spring or cross . action mechanism to apply tension to the vessel for occlusion.
- The choice of iaw style depends on the surgeon's preference, based on the type and delicacy of . the vessel to be occluded.
The surgeon chooses the vascular clamp based on the anatomy of the site and the type of occlusion desired, based on the type and size of the blood vessels, and surgical techniques. Clamps are either fully or partially occluding an artery or vein with correct tension to produce minimal trauma to vessels. Full occlusion clamps stops blood flow entirely by covering the full vessel. Partial occlusion clamps are placed on part of the vessel to isolate the area to be worked on while allowing blood flow to continue in the rest of the vessel.
K2 Medical vascular clamps are made of the following standardized materials Stainless Steel ASTM F 899-07 or Titanium Alloy.
The instruments are offered in non-sterile condition.
#### Indications for Use:
K2 Medical Vascular Clamps are devices intended for temporary or partial occlusion of blood vessels during vascular surgical procedures.
#### Comparison with Predicate Device:
The results of non-clinical and bench testing indicates that the new device is completely comparable to the predicate devices. Biocompatibility and sterilization studies were successfully completed.
The K2 Medical product is similiar to the predicate device in terms of technical characteristics, design, Indications for Use, Target population, where it is used, performance, biocompatibility, sterilisation method, mechanical safety characteristics as well as sizes and configurations. Therefore it can be deemed substantially equivalent for its indicated use.
#### Conclusion:
The presented data that was conducted on the K2 Medical Vascular Clamps shows in its results and in comparison to the predicate devices that the products are substantially equivalent to the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its body and wing. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
K2 Medical GmbH & Co.KG C/O Mr. Peter Fetzer Unter Buchsteig 3a 78532 Tuttlingen, GERMANY
APR 3 0 2012
Re: K120492
Trade/Device Name: K2 Medical Vascular Clamp Regulation Number: 21 CFR 870.4450 Regulation Name: Vascular Clamp · Regulatory Class: Class II Product Code: DXC Dated: February 13, 2012 Received: February 17, 2012
Dear Mr. Fetzer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{3}------------------------------------------------
# Page 2 - Mr. Peter Fetzer
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Bram D. Zuckerman, MD
Director
Division of Cardiovascular Devices
Center for Devices and Radiological Health
Office of Device Evaluation
Enclosure
{4}------------------------------------------------
# Indications for Use
KIZO 492 510(k) Number (if known):
Device Name:
K2 Medical Vascular Clamps
Indications For Use:
K2 Medical Vascular Clamps are devices intended for temporary or partial occlusion of blood vessels during vascular surgical procedures.
Prescription Use
(Part 21 CFR 801 Subpart D) X AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
1 1
(Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Cardiovascular Device 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.