INSTRUMED VASCULAR CLAMPS

K092544 · Instrumed International, Inc. · DXC · Oct 26, 2009 · Cardiovascular

Device Facts

Record IDK092544
Device NameINSTRUMED VASCULAR CLAMPS
ApplicantInstrumed International, Inc.
Product CodeDXC · Cardiovascular
Decision DateOct 26, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.4450
Device ClassClass 2
AttributesTherapeutic

Indications for Use

INSTRUMED Vascular Clamps are devices intended for temporary or partial occlusion of blood vessels during vascular surgical procedures.

Device Story

Reusable surgical instruments for temporary or partial occlusion of blood vessels. Available in various configurations including ring-handle with ratchet closure or spring-loaded bulldog style. Surgeon selects clamp based on vessel anatomy, delicacy, and required occlusion type (full or partial). Used in vascular surgical procedures to isolate vessel segments while maintaining blood flow elsewhere or to stop flow entirely. Operates via mechanical tension applied to vessel walls. Benefits include minimal trauma to vessels during surgical access. No electronic or software components.

Clinical Evidence

Bench testing only. No clinical data provided. Biocompatibility and sterilization studies were successfully completed to demonstrate substantial equivalence.

Technological Characteristics

Materials: Stainless Steel (ASTM F 899-07) or Titanium Alloy. Mechanical occlusion via spring-loaded or ratchet-closure mechanisms. Reusable. Non-sterile delivery.

Indications for Use

Indicated for temporary or partial occlusion of blood vessels during vascular surgical procedures in patients requiring vascular intervention.

Regulatory Classification

Identification

A vascular clamp is a surgical instrument used to occlude a blood vessel temporarily.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K092544 # 510(k) Summary ### 510(k) Summary as required by section 807.92(c) date prepared 08/17/2009 #### Submission Applicant: INSTRUMED INTERNATIONAL, INC. 626 Cooper Court Schaumburg, IL 60173 OCT 2 3 2009 #### Establishment Registration Number: 1421101 #### Official Correspondent: Mr. Berndt Fetzer INSTRUMED INTERNATIONAL, INC. 626 Cooper Court Schaumburg, IL 60173 Phone: 847-908-0292 #### Trade name: Instrumed Vascular Clamp #### Common name: Various vascular clamps: Bulldog Clamps, Vessel Clamps, Microvascular Bulldog Clamps, Ring Handle Bulldog Clamps, Neonatal Vascular Clamp, Pediatric Clamps, Occlusion Clamps, Vascular Clamps, Cardio Clamps, Aorna, Roman, Multi-Purpose Clamps, Blood Vessel Forceps, Anastomosis Clarms, Vena Clamps, Torto Clamps, Torto Plan vascular Clamps, Aortic Aneurysm Clamps, Abdominal Clamps, Profunda Clamps, Shunt Clamps, Aortic Clamps, Patent-Ductus Clamps, Fogarty-Type Clamps, Vascular Forceps, Microvascular Forceps, Minimally Invasive Coronary Systems #### Classification name: Clamp, Vascular (21 CFR 870.4450, Product code DXC) Regulation Description Vascular Clamp #### Substantial Equivalence Claims: | K072834 | SIBEL SA VASCULAR CLAMP; SURGICAL INSTRUMENTS BELGIUM SA<br>P.D. K973080 Surgical Vascular Clamp, Walter Lorenz Surgical, Inc | |---------|-------------------------------------------------------------------------------------------------------------------------------| | K992053 | AESCULAP VASCULAR INSTRUMENTS<br>AESCULAP, INC.<br>P.D. Preamendments: Aesculap, Codman, Heartport, V.Mueller, Walter Lorenz | | K070661 | VITALCOR FEATHERWEIGHT VASCULAR CLAMPS; VITALCOR, INC.<br>P.D. K022754 S&T SINGLE USE MICRO ANASTOMOSIS CLAMP; S & T AG | | K982278 | MICRINS MICRO ANASTOMOSIS CLAMP; MICRINS SURGICAL INC | {1}------------------------------------------------ #### Description of the Device: The Instrumed Vascular Clamps are reusable surgical instruments. To ensure the multi-purpose use of this device, many different models are available. The differences can be as follows: - . Some have ring handles with a ratchet closure to adjust the amount of tension applied to the vessel for occlusion or partial occlusion. - Bulldog style Vascular Clamps are another type of vascular cross clamp that use a spring or cross � action mechanism to apply tension to the vessel for occlusion. - The choice of jaw style depends on the surgeon's preference, based on the type and delicacy of ↓ the vessel to be occluded. The surgeon chooses the vascular clamp based on the anatomy of the site and the type of occlusion desired, based on the type and size of the blood vessels, and surgical techniques. Clamps are either fully or partially occluding an artery or vein with correct tension to produce minimal trauma to vessels. Full occlusion clamps stops blood flow entirely by covering the full vessel. Partial occlusion clamps are placed on part of the vessel to isolate the area to be worked on while allowing blood flow to continue in the rest of the vessel. Instrumed vascular clamps are made of the following standardized materials Stainless Steel ASTM F 899-07 or Titanium Alloy, The instruments are offered in non-sterile condition. #### Indications for Use: INSTRUMED Vascular Clamps are devices intended for temporary or partial occlusion of blood vessels during vascular surgical procedures. #### Comparison with Predicate Device: The results of non-clinical and bench testing indicates that the new device is completely comparable to the predicate devices. Biocompatibility and sterilization studies were successfully completed. The Instrumed product is similiar to the predicate device in terms of technical characteristics, design, Indications for Use, Target population, where it is used, performance, biocompatibility, sterilisation method, mechanical safety characteristics as well as sizes and configurations. Therefore it can be deemed substantially equivalent for its indicated use. #### Conclusion: The presented data that was conducted on the Instrumed Vascular Clamps shows in its results and in comparison to the predicate devices that the products are substantially equivalent to the predicate devices. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle, there is a stylized image of an eagle or bird-like figure with outstretched wings. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002 OCT 2 0 2009 Intrumed International, Inc. Michael Massong RA/QA Director 626 Cooper Court Schaumburg, IL 60173 Re: K092544 Trade/Device Name: Intrumed Vascular Clamps Regulation Number: 21 CFR 870.4450 Regulation Name: Vascular Clamp Regulatory Class: Class II Product Code: DXC Dated: September 30, 2009 Received: October 1, 2009 Dear Mr. Massong: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act {3}------------------------------------------------ ### Page 2 - Mr. Massong or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. Duna R. Vuchner Image /page/3/Picture/7 description: The image shows a handwritten letter 'f' in a simple, slightly stylized font. The letter is in black ink and appears to be part of a larger text or word. The stroke of the 'f' is continuous, with a curved top and a descending stem. Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use 510(k) Number (if known): K092544 Device Name: Instrumed Vascular Clamps Indications For Use: INSTRUMED Vascular Clamps are devices intended for temporary or partial occlusion of blood vessels during vascular surgical procedures. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (Please do NOT WRITE BELOW THIS LINE-CONTINUE on another Page if needed) Concurrence of CDRH, Office of Device Evaluation (ODE) Duma R. Vc. Amez Division Sign-C Division of Cardiovascular Devices 10(k) Number_K0925A4 Page 1 of 1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...