PRESTO BREAST BIOPSY DEVICE

K120440 · Sontina Medical, LLC · KNW · May 17, 2012 · Gastroenterology, Urology

Device Facts

Record IDK120440
Device NamePRESTO BREAST BIOPSY DEVICE
ApplicantSontina Medical, LLC
Product CodeKNW · Gastroenterology, Urology
Decision DateMay 17, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1075
Device ClassClass 2

Indications for Use

The Presto Breast Biopsy Device is intended for diagnostic sampling of breast tissue during breast biopsy procedures. It is to be used for diagnostic purposes only and is not intended for therapeutic uses. The Presto Breast Biopsy Device is indicated to provide breast tissue samples for diagnostic sampling of breast abnormalities. It is designed to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality. The extent of histologic abnormality cannot be reliably determined from its mammographic appearance. Therefore, the extent of removal of the imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality (e.g., malignancy). When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.

Device Story

Sterile, single-use percutaneous biopsy device; stainless steel coring cannula with razor edge; stationary internal coil. Handle contains actuation button, sample collection chamber, DC motor. Powered by external reusable medical-grade AC/DC power supply (12V). Operator-controlled; button press rotates cannula to core and transport tissue to collection chamber. Compatible with coaxial introducer. Used in clinical settings for diagnostic sampling of breast abnormalities; provides tissue for histologic examination. Benefits patient by enabling minimally invasive tissue acquisition for diagnosis.

Clinical Evidence

Bench testing only. Studies included in-vitro device performance, electrical and product safety (IEC 60601-1), tensile strength, fatigue, biocompatibility, and simulated use testing.

Technological Characteristics

Stainless steel coring cannula; echogenic working-end. Powered by external 12V AC/DC power supply. Single-use disposable device with reusable power source. Sterile. No software/algorithm.

Indications for Use

Indicated for diagnostic sampling of breast tissue abnormalities in patients requiring histologic examination. Not for therapeutic use.

Regulatory Classification

Identification

A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Presto Breast Biopsy Device (K120440) 510(k) Notification # 2. 510(k) Summary pg 1 of 2 This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92. Date Prepared: February 6, 2012 510(k) number: K120440 Applicant Information: Sontina Medical, LLC 111 Sutro Heights Ave San Francisco, CA 94121 | Contact Person: | Robert Peliks | |-----------------|------------------| | Phone Number: | (415) 873 - 3831 | | Fax Number: | (415) 668 - 4884 | ### Device Information: Classification: Trade Name: Common Name: Classification Name: Product Code: Class II Presto Breast Biopsy Device Biopsy Instrument Biopsy Instrument (21 CFR 876.1075) KNW ### Predicate Device Information: The subject device is substantially equivalent in intended use and/or method of operation to the devices listed in Table 2.1. · | Device Name | Manufacturer | 510(k) # | |------------------------------------|------------------------------------------|----------| | Rubicor Magic Breast Biopsy Device | Encapsule Medical<br>(San Francisco, CA) | K071048 | | Bard Monopty | Bard Biopsy Systems<br>(Tempe, AZ) | K922939 | ## Table 2.1 Predicate Device Information #### Device Description: The Presto Breast Biopsy Device is a sterile, single-use percutaneous biopsy device. The working end of the device includes a stainless steel coring cannula with a razor edge and a stationary coil located within the coring cannula. The handle of the device contains an actuation button, a sample collection chamber, a drive mechanism for rotating the coring cannula and a DC power jack for a 12V input. A reusable, medical grade AC/DC power supply provides 12V to the disposable device. Depressing the button on the handle rotates the coring cannula - allowing the operator to core and transport tissue samples to the collection chamber. The device may be used with a coaxial introducer. {1}------------------------------------------------ Presto Breast Biopsy Device (K120440) ### Intended Use: The Presto Breast Biopsy Device is intended for diagnostic sampling of breast tissue during breast biopsy procedures. It is to be used for diagnostic purposes only and is not intended for therapeutic uses. The Presto Breast Biopsy Device is indicated to provide breast tissue samples for diagnostic sampling of breast abnormalities. It is designed to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality. The extent of histologic abnormality cannot be reliably determined from its mammographic appearance. Therefore, the extent of removal of the imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality (e.g., malignancy). When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures. ## Device Comparison: The Presto Breast Biopsy device shares many similarities with one or all of the predicate devices, including: - . Similar indications for use and anatomical target site; - Similar features, including: . - Single, Insertion -Multiple Sample functionality 0 - One-button user-interface 0 - Compatibility with coaxial introducer 0 - 0 Echogenic working-end; - Similar materials the working end of all devices is predominantly stainless steel; and ● - . Similar device functionality - a reusable, medical grade AC-DC power supply powers a DC motor which spins a forward advancing, spinning cannula to core & severe the targeted specimen. #### Testing Summary: The Presto Breast Biopsy Device was evaluated in the following non-clinical studies: in-vitro device performance, electrical & product safety (IEC 60601-1), tensile strength & fatigue, biocompatibility, predicate device comparison and simulated use testing. Results of the testing demonstrate that the materials, manufacturing process and design of the Presto Breast Biopsy Device meet the established specifications necessary for consistent performance during its intended use #### Conclusion: Based on the intended use, product, and performation provided in this notification, the subject device has been shown to be substantially equivalent to the currently marketed and unmodified predicate devices. {2}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image contains the text 'Public Health Service'. The text is in a simple, sans-serif font and is horizontally aligned. The words are stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 MAY 17 2012 Sontina Medical, LLC % Mr. Robert Peliks 111 Sutro Heights Avenue San Francisco, California 94121 Re: K120440 Trade/Device Name: Presto Breast Biopsy Device Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-urology biopsy instrument Regulatory Class: Class II Product Code: KNW Dated: April 24, 2012 Received: April 26, 2012 Dear Mr. Peliks: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must {3}------------------------------------------------ Page 2 - Mr. Robert Peliks comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, for Pat Mar V.r Mark N. Malkerson Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Presto Breast Biopsy Device KI20440 510(k) Notification ## 1. Indications for Use Statement 510(k) Number (if known): __ Presto Breast Biopsy Device Device Name: N/A #### Indications for Use: The Presto Breast Biopsy Device is intended for diagnostic sampling of breast tissue during breast biopsy procedures. It is to be used for diagnostic purposes only and is not intended for therapeutic uses. The Presto Breast Biopsy Device is indicated to provide breast tissue samples for diagnostic sampling of breast abnormalities. It is designed to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality. The extent of histologic abnormality cannot be reliably determined from its mammographic appearance. Therefore, the cxtent of temoval of the imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality (e.g., malignancy). When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation Neif. Rpo. Oisler for \$m \times n\$ (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K120440 Sontina Medical, LLC CONFIDENTIAL 1-1 of 1-1
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...