RUBICOR MAGIC BREAST BIOPSY DEVICE, MODEL 31537

K071048 · Rubicor Medical, Inc. · KNW · May 15, 2007 · Gastroenterology, Urology

Device Facts

Record IDK071048
Device NameRUBICOR MAGIC BREAST BIOPSY DEVICE, MODEL 31537
ApplicantRubicor Medical, Inc.
Product CodeKNW · Gastroenterology, Urology
Decision DateMay 15, 2007
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 876.1075
Device ClassClass 2

Indications for Use

The Rubicor Breast Biopsy Device is intended for diagnostic sampling of breast tissue during breast biopsy procedures. It is to be used for diagnostic purposes only and is not intended for therapeutic uses. The Rubicor Breast Biopsy Device is indicated to provide breast tissue samples for diagnostic sampling of breast abnormalities. It is designed to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality. The extent of histologic abnormality cannot be reliably determined from its mammographic appearance. Therefore, the extent of removal of the imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality (e.g., malignancy). When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.

Device Story

Rubicor Breast Biopsy Device is a mechanical instrument for diagnostic breast tissue sampling. Device facilitates histologic examination by removing breast tissue samples from imaged abnormalities. Used by clinicians during biopsy procedures. Output is physical tissue sample for pathology. Device does not provide therapeutic treatment. Clinical decision-making relies on subsequent histologic examination of tissue margins to determine completeness of removal for non-benign findings.

Clinical Evidence

Bench testing only. In-vitro testing demonstrated the device is safe and effective for its intended function.

Technological Characteristics

Mechanical biopsy instrument. No electronic, software, or energy-based components described. Class II device under 21 CFR 876.1075.

Indications for Use

Indicated for diagnostic sampling of breast tissue in patients with breast abnormalities to obtain samples for histologic examination; allows for partial or complete removal of imaged abnormalities.

Regulatory Classification

Identification

A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K0.71048 # 510 (k) Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. Date Prepared: April 11, 2007 510(k) number: _______________________________________________________________________________________________________________________________________________________________ Applicant Information: Rubicor Medical, Inc. 600 Chesapeake Drive Redwood City, CA 94063 | Contact Person: | Chris Daniel | |-----------------|----------------| | Phone Number: | (650) 556-1070 | | Fax Number: | (650) 556-1821 | ### Device Information: | Classification: | Class II | |----------------------|-------------------------------------| | Trade Name: | Rubicor Breast Biopsy Device | | Classification Name: | Biopsy Instrument (21 CFR 876.1075) | ### Equivalent Device: The subject device and accessory are substantially equivalent in intended use and/or method of operation to the Rubicor Magic™ Breast Biopsy Device (K053151) ### Intended Use: The Rubicor Breast Biopsy Device is intended for diagnostic sampling of breast tissue during breast biopsy procedures. It is to be used for diagnostic purposes only and is not intended for therapeutic uses. The Rubicor Breast Biopsy Device is indicated to provide breast tissue samples for diagnostic sampling of breast abnormalities. It is designed to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality. The extent of histologic abnormality cannot be reliably determined from its mammographic appearance. Therefore, the extent of removal of the imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality (e.g., malignancy). When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures. ### Test Results: ### Performance Results of in-vitro testing demonstrate that the Rubicor Breast Biopsy Device is safe and effective for its intended function. MAY 1 5 2007 {1}------------------------------------------------ # Summary: Based on the intended use, product, and performance information provided in this notification, the subject device has been shown to be substantially equivalent to the currently marketed and unmodified predicate device. : . _ {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a stylized human figure. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Rubicor Medical, Inc. % Mr. Chris Daniel Vice President, R&D 600 Chesapeake Drive Redwood City, California 94063 MAY 1 5 2007 Re: K071048 Trade/Device Name: Rubicor Breast Biopsy Device Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-urology-biopsy instrument Regulatory Class: II Product Code: KNW Dated: April 11, 2007 Received: April 16, 2007 Dear Mr. Daniel: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set {3}------------------------------------------------ Page 2 - Mr. Chris Daniel forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permitts your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Omall Manufacturers, International and Consumer Assistance at its toll-free number (800) 633-2011 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html . Sincerely yours, Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ #### Indication for Use Statement 510(k) Number (if known): K071048 Device Name: Rubicor Breast Biopsy Device Indications for Use: The Rubicor Breast Biopsy Device is intended for diagnostic sampling of breast tissue during breast biopsy procedures. It is to be used for diagnostic purposes only and is not intended for therapeutic uses. The Rubicor Breast Biopsy Device is indicated to provide breast tissue samples for diagnostic sampling of breast abnormalities. It is designed to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality. The extent of histologic abnormality cannot be reliably determined from its mammographic appearance. Therefore, the extent of removal of the imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality (e.g., malignancy). When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures. ## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign Division of General, Restorative, and Neurological Devices 1671048 510(k) Summary (Per 21 CFR 801.109) Over-the Counter Use Rubicor Medical, Inc. Page 20
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...