K120328 · Sound Surgical Technologies, LLC · MUU · Sep 17, 2012 · General, Plastic Surgery
Device Facts
Record ID
K120328
Device Name
ORIGINS LIPOHARVESTING SYSTEM
Applicant
Sound Surgical Technologies, LLC
Product Code
MUU · General, Plastic Surgery
Decision Date
Sep 17, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Origins LipoHarvesting System is used in the aspiration, harvesting, filtering and transferring of adipose autologous tissue for aesthetic body contouring. If harvested fat is to be reimplanted, the harvested fat is only to be used without any additional manipulation.
Device Story
Origins LipoHarvesting System is a reusable, closed-loop tissue collection system; consists of medical-grade canister, drain port, and lid. Used with standard liposuction aspiration pump; collects fatty tissue for aesthetic body contouring. Tissue enters canister via lid port; physician removes waste via base drain port. Harvested fat transferred to syringes for autologous re-injection. Used in clinical settings by physicians. Benefits patient by providing sterile-compatible, closed-loop collection for autologous fat grafting without additional manipulation.
Clinical Evidence
Bench testing only. No clinical data provided. Performance testing verified compliance with predetermined specifications and biocompatibility standards (ISO 10993-1). Sterilization validation conducted per ISO 17665-1.
Technological Characteristics
Reusable closed-loop tissue collection system; medical-grade biocompatible materials; 100-2,000 mL volume capacity. Operates via external aspiration pump. Sterilization via user-performed autoclave. No software, electrical components, or electronic sensors.
Indications for Use
Indicated for patients undergoing aesthetic body contouring procedures requiring the aspiration, harvesting, filtering, and transfer of autologous adipose tissue.
Regulatory Classification
Identification
A suction lipoplasty system is a device intended for aesthetic body contouring. The device consists of a powered suction pump (containing a microbial filter on the exhaust and a microbial in-line filter in the connecting tubing between the collection bottle and the safety trap), collection bottle, cannula, and connecting tube. The microbial filters, tubing, collection bottle, and cannula must be capable of being changed between patients. The powered suction pump has a motor with a minimum of 1/3 horsepower, a variable vacuum range from 0 to 29.9 inches of mercury, vacuum control valves to regulate the vacuum with accompanying vacuum gauges, a single or double rotary vane (with or without oil), a single or double diaphragm, a single or double piston, and a safety trap.
Special Controls
*Classification.* Class II (special controls). Consensus standards and labeling restrictions.
Shippert Medical Technologies / Tissu Trans Filtron (K101713)
human med AG / LipoCollector II Complete Set
Submission Summary (Full Text)
{0}------------------------------------------------
K120328
1083
SEP 1 7 2012
# 510(k) Summary
#### Submission Date:
30 January 2012
Submitter: .
Sound Surgical Technologies LLC 357 McCaslin Boulevard, Suite 100 Louisville, CO 80027
Submitter and Official Contact: Mr. Stephen C. Smith Vice President of RA/QA Sound Surgical Technologies LLC 357 McCaslin Boulevard, Suite 100 Louisville, CO 80027 +1 (720) 240-2970 SSmith@soundsurgical.com
Manufacturing Site:
Sound Surgical Technologies LLC 357 McCaslin Boulevard, Suite 100 Louisville, CO 80027
Trade Name:
Quito, Ecuador
Common Name: Suction lipoplasty system
Classification Name: System, Suction, Lipoplasty
Sound Model
LipoHarvesting
Origins
System
Classification 21 CFR §878.5040 Regulation:
Product Code: MUU
Substantially Equivalent Devices: Predicate 510(k) Number
Predicate Manufacturer and Model
K092482
Sound Surgical Technologies LLC Origins LipoHarvesting System
K101713
Shippert Medical Technologies / Tissu Trans Filtron
human med AG / LipoCollector II Complete Set
{1}------------------------------------------------
## Device Description:
The Origins LipoHarvesting System consists of a reusable, closed loop tissue collection system comprised of a medical grade canister, a drain port and a lid intended to be used with a standard liposuction aspiration pump to collect fatty tissue for aesthetic body contouring. As the tissue is harvested from the patient, it enters the collection canister via a port in the canister lid. The physician removes unwanted waste materials from the collection system via the drain port at the base. This process leaves fatty tissue that can be transferred to syringes for autologous fat re-injection.
Intended Use:
The Origins LipoHarvesting System is used in the aspiration, harvesting, filtering and transferring of autologous tissue. If harvested fat is to be re-implanted, the harvested fat is only to be used without any additional manipulation.
Technology Comparison: The Origins LipoHarvesting System employs the same technological characteristics as the predicate devices.
| Characteristic<br>Comparisons: | Shippert Medical<br>Technologies / Tissu Trans<br>Filtron | human med AG /<br>LipoCollector II Complete<br>Set | Origins LipoHarvesting<br>Device |
|--------------------------------|--------------------------------------------------------------|----------------------------------------------------|----------------------------------------|
| Suction source(s) | Aspiration device | Aspiration device | Aspiration device |
| Volume Range | 100 cc to 500 cc<br>100 cc to 1,000 cc<br>100 cc to 2,000 cc | Up to 1,000 mL | 100 to 2,000 mL |
| Shipped sterile | Yes | Yes | No, sterilized by user prior<br>to use |
| Initial Method | Gamma radiation | Ethylene oxide (EtO) | Not applicable |
| Sterility assurance level | 10-6 | 10-6 | 10-6 |
| Disposable or reusable | Single use, disposable | Canister, lid and basket<br>are reusable | Reusable |
| Resterilization Method | Not applicable | Autoclave | Autoclave |
#### Performance Testing:
Sterilization
The Origins LipoHarvesting System is not provided sterile, but is sterilized by the user prior to use. The sterilization of the Origins LipoHarvesting System will be validated prior to commercial distribution in accordance with ISO 17665-1: 2006, Sterilization of health care products – Moist heat – Part 1 : Requirements for the development, validation and routine control of a sterilization process for medical devices.
1120328
7.83
{2}------------------------------------------------
V1120328
30f3.
made from medical grade biocompatible materials. Test results and analyses indicate that the Origins LipoHarvesting System materials comply in accordance with ISO 10993-1: 2009, Biological evaluation of medical devices – Part 1: Evaluation and testing. The Origins LipoHarvesting System does not contain software, and Software Testing therefore this section does not apply. The Origins LipoHarvesting System does not contain electrical Electromagnetic components, and therefore this section does not apply. Compatibility Testing and Electrical Safety The Origins LipoHarvesting System is tested for performance in Performance Testing - Bench
therefore this section does not apply.
accordance with its predetermined specifications as specified in Section 11, Device Description - Performance Specifications, of this submission.
The Origins LipoHarvesting System is not provided sterile, and
The Origins LipoHarvesting System has patient contact materials and is
Test results indicate that the Origins LipoHarvesting System complies with its predetermined specification.
Conclusion
Shelf-Life
Biocompatibility
Verification and validation activities were conducted to establish the performance and safety characteristics of the Origins LipoHarvesting System. The results of these activities demonstrate that the Origins LipoHarvesting System is considered substantially equivalent to the predicate devices.
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines forming its body and wings. The bird is positioned within a circular border that contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in capital letters, arranged around the circumference of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
1 > 2012
Sound Surgical Technologies, LLC
% Mr. Stephen C. Smith
Vice President of Regulatory Affairs and Quality Affairs
357 McCaslin Boulevard, Suite 100
Louisville, Colorado 80027
Re: K120328
Trade/Device Name: Origins LipoHarvesting System Regulation Number: 21 CFR 878.5040 Regulation Name: Suction lipoplasty system Regulatory Class: Class II Product Code: MUU Dated: August 17, 2012 Received: August 28, 2012
Dear Mr. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
# Page 2 - Mr. Stephen C. Smith
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark Al Melleman
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### Indications for Use
# 510(k) Number (if known):
### K120328
Device Name:
Indications for Use:
Origins LipoHarvesting System
The Origins LipoHarvesting System is used in the aspiration, harvesting, filtering and transferring of adipose autologous tissue for aesthetic body contouring. If harvested fat is to be reimplanted, the harvested fat is only to be used without any additional manipulation.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
(21 CFR 807 Subpart C)
Over-The-Counter Use
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number 1/120328
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.