K101713 · Human Med AG · MUU · Jun 29, 2010 · General, Plastic Surgery
Device Facts
Record ID
K101713
Device Name
LIPCOLLECTOR II
Applicant
Human Med AG
Product Code
MUU · General, Plastic Surgery
Decision Date
Jun 29, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5040
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
Harvesting, filtering and transferring of autologous fat tissue
Device Story
LipoCollector II is a closed-system device for harvesting, filtering, and transferring autologous fat tissue during aesthetic body contouring. It consists of an autoclavable polymeric collection container with a stainless steel lid, a pre-filtering basket, and disposable mesh filters. The device connects to standard suction lipoplasty systems via vacuum tubing. During liposuction, aspirated fat and saline are drawn into the container; separation occurs via buoyancy and gravity, with saline continuously suctioned to a waste container. A stainless steel basket pre-filters large connective tissue strands, while mesh filters (200-315 µm) capture fat particles. The closed system maintains sterility, allowing the physician to collect and filter fat in one step without centrifugation or container transfers. Harvested fat is withdrawn via a sterile extraction cannula into a Luer-Lock syringe for immediate re-injection. Used in operating rooms by physicians, the device reduces contamination risk and processing steps compared to open systems.
Clinical Evidence
Bench testing only. Compliance with DIN EN ISO 13485, ISO 10993-1 (biocompatibility), ISO 11135-1 (ETO sterilization), and ISO 11607 (sterile packaging) was demonstrated. Biocompatibility testing included cytotoxicity (ISO 10993-5), irritation/hypersensitivity (ISO 10993-10), and endotoxin/pyrogen testing (ISO 10993-11). No clinical data was required or provided.
Technological Characteristics
Polymeric collection container with stainless steel lid and basket. Mesh filters (200, 250, 315 µm). Operates via external vacuum source. Closed-system design. Sterilization: ETO for disposables; autoclavable for reusable components. Complies with ISO 10079-1, ISO 7153-1, ISO 9626, and ISO 14971.
Indications for Use
Indicated for aesthetic body contouring in patients undergoing autologous fat transfer procedures.
Regulatory Classification
Identification
A suction lipoplasty system is a device intended for aesthetic body contouring. The device consists of a powered suction pump (containing a microbial filter on the exhaust and a microbial in-line filter in the connecting tubing between the collection bottle and the safety trap), collection bottle, cannula, and connecting tube. The microbial filters, tubing, collection bottle, and cannula must be capable of being changed between patients. The powered suction pump has a motor with a minimum of 1/3 horsepower, a variable vacuum range from 0 to 29.9 inches of mercury, vacuum control valves to regulate the vacuum with accompanying vacuum gauges, a single or double rotary vane (with or without oil), a single or double diaphragm, a single or double piston, and a safety trap.
Special Controls
*Classification.* Class II (special controls). Consensus standards and labeling restrictions.
{0}------------------------------------------------
K101713 page 1/6
## 510(k) Summary LipoCollector II complete set
This 510(k) Summary for the LipoCollector II complete set meets the requirements of 21 CFR § 807.92
### Date: 2010-05-28
JUN 2 9 2010
## 1. Submitter Information
human med AG Wilheim-Hennemann-Straße 9 D-19061 Schwerin Germany
Contact Person:
Dr. Inge Matthiesen c/o human med AG Wilhelm-Hennemann-Straße 9 D-19061 Schwerin Germany
Phone: +49(0)385 395 70 0 +49(0)385 395 70 29 Fax:
### 2. Name of Device:
| Common Name: | Suction Lipoplasty System | | |
|----------------------|------------------------------------------------------------------|--|--|
| Proprietary Name: | LipoCollector II complete set | | |
| Classification: | Class II - 21 CFR § 878.5040 | | |
| Product Code: | MUU | | |
| Indications for use: | Aesthetic body contouring | | |
| ntended use: | Harvesting, filtering and transferring of autologous fat tissue, | | |
## 3. Name of predicate device(s)
- Cytori AFT System; Cytori Therapeutics Inc Product Code MUU; 510(k) . number K072587
- Lipokit; Medi-Khan (USA) Inc Product Code MUU; 510(k) number K083455 .
- Lipivage; Genesis Biosystems, Inc. 510(k) number unknown ●
つ
{1}------------------------------------------------
## 4. Description of the new device
The LipoCollector II complete set consists of an autoclavable polymeric tissue collection container with a stainless steel lid and pre-fillering basket, and a disposable set including mesh filters. O-rings and inner tubing. The LipoCollector II complete set also includes a stainless steel extraction cannula (single-use) for withdrawal of the collection container. From the extraction cannula, the fat can be directly transferred to a Luer-Locksyringe (not part of the LipoCollector !! complete set) for autologous fat transfer. The intended use of the LipoCollector II complete set is to filter and collect the aspirated fat tissue for subsequent transfer of the autologous fat into subcutaneous fat tissues during the same procedure.
The LipoCollector II complete set is used together with FDA cleared devices (such as vaccuum devices, tubings, and syringes), for example suction lipoplasty devices (Product Code MUU, Regulation Number 21 CFR 878.5040), and syringes (e.g. 50 cc syringes, Luer-Lock, Product Code KYX). The aspirated fat gets transferred from the aspiration device to the collection container by the application of vacuum. The vacuum is generated by the aspiration device (e.g. suction lipoplasty device).
The LipoCollector il complete set serves to filter fatty and adherent tissue from the aspirate which is extracted during liposuction. The separated water (saline) gets transferred to the waste container of the aspiration device via a tube by the application of vacuum which is generated by the aspiration device. From the LipoCollector II collection container, the harvested fat is being removed by the extraction cannula, and then transferred into a syringe for autologous fat transfer into the desired body area(s).
The filtering of the aspirate with the LipoCollector II complete set is done predominantly using buovancy and gravity. A special filter unit, comprising a stainless steel wire fabric with a mesh/hole size from 200 um to 315 um, captures the appropriate fat particles in a sterile collection container in such a way that only the saline rinsing solution ends up in the suction (waste) container provided. All parts are sterile and air contact to the fat from outside is avoided due to the closed system design of the LipoCollector II.
## Function principles
The filtering of the aspirate with the LipoCollector II complete set is done predominantly using buoyancy and gravity. The suction of the liposuction system bypasses the previously collected aspirate via the integrated bypass. This means that the fatty tissue that has been suctioned off is handled as gently as possible.
The main principle behind the function of the LipoCollector II complete set is the initial separation of liguid and lipocytes due to the physical buoyancy. The less dense fat floats on the liquid which is concurrently and continuously suctioned off. In doing so the filter is constantly rinsed with liquid which reduces blockage of the filter mesh.
The LipoCollector II complete set is equipped with a stainless steel basket that acts as a prefilter. By turning the lid of the LipoCollector II, the basket is positioned precisely beneath the opening of the connection labelled PATIENT, ensuring that incoming aspirate first runs through the basket. Larger strands of connective tissue are retained in the grates of the basket, which reduces blockage of the mesh filter and/or cannula.
Mesh filters: Three sizes of mesh filters are supplied with the LipoCollector II (200 µm, 250 µm and 315 um).
{2}------------------------------------------------
K101713
page 3/6
#### Closed system filtration
The LipoCollector II complete set which is to be used under sterile conditions in the operating room represents a closed system. The LipoCollector II remains in a sterile field at all times. thus reducing the risk of infection. Also, the risk of contamination is lowered because the fat never leaves the closed sterile LipoCollection container until re-injection.
Using the LipoCollector II complete set decreases the number of steps a physician must take to process the fat for autologous fat transfer (e.g. there are no extra steps of transferring the harvested fat between different containers or of subsequent centrifugation required). The LipoCollector II complete set allows collection and filtering in one step in one closed system container, and thereby maintaines the sterility of the harvested fat until the fat is transferred .
#### Extraction cannula of the LipoCollector II
The included sterile extraction cannula is used to withdraw the harvested fat from the LipoCollector II. The filtered aspirate is withdrawn from the collection container by the extraction cannula while keeping the lid of the LipoCollector II closed.
## 5. Intended Use
The intended use of the LipoCollector II is the harvesting, filtering and transferring of autologous fat tissue.
## 6. Summary of technological characteristics of the new device in comparison to those of the predicate devices
The LipoCollector II complete set shares indications, intended use and design principles with the following predicate devices which have been determined by FDA to be substantially equivalent to premarket devices: Cytori AFT System (K072587), Lipokit (K083455) and Genesis Biosystems Lipivage (510(k) number unknown).
The LipoCollector II complete set and the predicate devices are substantially equivalent with respect to their indication for use and their intended use, as they are all intended for the same procedure of harvesting, filtering and transferring of autologous fat tissue.
#### Design and Materials
The design and materials of the LipoCollector II complete set and the predicate devices (Cytori AFT System (K072587), Lipokit (K083455) and Genesis Biosystems Lipivage (510(k) number unknown)) are substantially equivalent, as they share the major design principles of utilizing vacuum (liposuction devices or house vacuum) to aspirate adipose tissue from the patient and subsequently transport the adjpose tissue through a tube into a collection container that contains a filtering mechanism to allow fluids to pass, but retains the adipose tissue within the container.
{3}------------------------------------------------
K101713 page 4/6
In summary, the design and materials of the LipoCollector II complete set and its predicate devices (Cytori AFT System (K072587), Lipokit (K083455) and Genesis Biosystems Lipivage (510(k) number unknown)) are substantially equivalent, as they are all polymer constructed with surgical stainless steel components for direct patient contact, with collection containers which are operated manually, and utilizing varying sources of vacuum to withdraw and collect saline fluids containing fat tissue, from the body.
The LipoCollector II complete set is substantially equivalent to the above mentioned predicate devices, as they all consist of a polymeric tissue collection container with a filter unit within the chamber, and all have connection ports on the superior ends of the collection container for the attachment of suction tubing, and stainless steel cannulae that contact the patient, and connection ports to connect the vacuum tubing that draws vacuum and carries waste to the waste container.
Like the above mentioned predicate devices, the LipoCollector II complete set decreases the number of steps a physician must take to process the fat for autologous fat transfer. When using the LipoCollector II complete set the fat remains in the sterile container at all times until immediately before reinjection, thus reducing the risk of contamination. The LipoCollector II complete set allows collection and filtering in one step in one collector.
## 7. Summary of compliance tests according to recognized standards
## Design and Manufacturing Process Control Activities
human med AG has implemented a Quality Management System in accordance to DIN EN ISO 13485 and to the Council Directive 93/42/EEC on medical devices.
All design and manufacturing control activities for the LipoCollector II complete set are in accordance to section 7.3 of DIN EN ISO 13485 and to the Code of Federal Regulations (CFR) Title 21, Section 820.30.
The design and manufacturing activities of the LipoCollector II complete set all comply with the requirements of DIN EN ISO 10079-1:2009. CSA Z 168.11-94. ISO 2768-1:1989, ISO 2768-2:1989, ISO 7153-1:1991/Amd.1:1999 and ISO 9626:1991/Amd.1:2001.
For product labeling of the LipoCollector II complete set the requirements according to ISO 15223-1 are applied. The requlations of FDA 21 CFR 801.1. Device Labeling Guidance, and the Guidance for Labeling of the Canadian MDR were examined for special requirements regarding product labeling and product information.
Acceptance criteria for verification and validation testing of the device are based upon the approved product design specifications, including any additional design requirements imposed by Risk Management Process. Acceptance criteria for the LipoCollector II complete set include factors addressing product function, product safety, instructions for use, and packaging. All requirements of ISO 14971:2007 were met.
## Biocompatibility Testing
Biocompatibility of the LipoCollector II complete set was examined for patient contact materials. The evaluation of the scope of biological testing was carried out as per EN ISO 10993-1:2009.
{4}------------------------------------------------
The intended use of the LipoCollector II complete set involves contact with tissue and bodily fluids for a period of less than 24 hours. To evaluate the biocompatibility of the LipoCollector II complete set, the following tests were necessary:
KI01713
- Cytotoxicity test as per EN ISO 10993-5:2009-10, Annex C(metabolic cell activity)
- Cytotoxicity test as per EN ISO 10993-5 (membrane integrity)
- Epicutane test as per EN ISO 10993-10:2002-09,
- Quantitative detection of endotoxine in liquids and eluates as per EN ISO 10993-11:2006-08
- In vitro pyrogen test (IPT) as per EN ISO 10993-11:2006-08.
### Results of the biological testings according to 10993-10 and 10993-11
Using the test material as mentioned before no cytotoxic effects could be found. The tests for irritation and delayed-type hypersensitivity according to the DIN EN ISO 10993-10. 2002 were performed with an epicutane test with 10 volunteers. No neaative effects were observed. Also, with the LAL-Test (Limulus - Amebocyte - Test) no endotoxine could be detected in the eluate of the product.
Based on the submitted biocompatibility documentation, the LipoCollector II complete set can be regarded as biocompatible for the intended use of harvesting, filtering and transferring of autologous fat tissue.
## Validation of ETO Sterilisation Cycle
The disposable products of LipoCollector II complete set are ETO sterilized by a CE and FDA registered contract sterilisation company. The sterilisation procedure to be used has been successfully validated by physical and microbiological methods in accordance to ISO 11135-1:2007. The evaluation reveals that the single-use products of the LipoCollector II collection container and the extraction cannula sterilized with the validated cycle and a sufficient sterilization safety level of 10° is guaranteed. The single-use products of the LipoCollector II complete set were evaluated for compliance with ISO 10993-7 for ETO residuals. The test results demonstrate compliance with the specifications.
#### Validation of Sterile Packaging
Packaging for the disposable products of the LiboCollector II complete set was successfully validated, and is compatible with the ETO sterilisation method. The packaging was qualified by testing to assure imperviousness to micro organisms. The validation of sterile packaging is in accordance to ISO 11607-1, 11607-2, DIN 58953 and EN 868-5.
### Validation of Reprocessing
The reprocessing of the reusable parts of the LipoCollector II complete set (e.g. collection container, stainless steel lid and basket) has been successfully validated according to the requirements of AAMI/ANSI ST81:2004 "Sterilization of medical devices - Information to be provided by the manufacturer for the processing of resterilizable medical devices".
## 8. Conclusions
Based upon the intended use, materials, and technological characteristics as presented in this notification, the subject device is shown to be substantially equivalent to the currently marketed predicate devices. There are no significant differences between the LipoCollector II complete set and its predicate devices relevant for safety and effectiveness with respect to the intended use of harvesting, filtering and transferring of autologous fat tissue
{5}------------------------------------------------
K101703 page 6/6
The intended use, the design, the materials and the collection principle of the LipoCollector II complete set are considered as substantially equivalent to its predicate devices (Cytori AFT System (K072587), Lipokit (K083455) and Genesis Biosystems Lipivage (510(K) number unknown)).
Also, the documentation for the LipoCollector II complete set on biocompatibility testing, sterilisation and further compliance testing according to recognized standards, as submitted in this 510(k) application, demonstrate that the LipoCollector II complete set can be regarded as safe and effective for the intended use of harvesting, filtering and transferring of autologous fat tissue.
7
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling an eagle or bird in flight, with three curved lines forming the wings and body.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
## JUN 2 9 2010
Human Med AG % TUV SUD America Inc. Mr. Stefan Preiss 1775 Old Highway 8 NW New Brighton, Minnesota 55112-1891
Re: K101713
Trade/Device Name: LipoCollector II Regulation Number: 21 CFR 878.5040 Regulation Name: Suction lipoplasty system Regulatory Class: II Product Code: MUU Dated: June 11, 2010 Received: June 18, 2010
Dear Mr. Preiss:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{7}------------------------------------------------
Page 2 - Mr. Stefan Preiss
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050,
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
erely yours,
N. Melkerson
Mark N. Melkers Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Indications for Use Statement
510(k) Number (if known): K101713
Device Name: _________________________________________________________________________________________________________________________________________________________________
,
Indications for Use:
: 上一篇:
.
Aesthetic body contouring
Intended Use:
Harvesting, filtering and transferring of autologous fat tissue
| Prescription Use<br>(Part 21 CFR 801 Subpart D) | yes | | Over-the-Counter Use<br>(21 CFR 801 Subpart C) | no |
|-------------------------------------------------|-----|--|------------------------------------------------|----|
|-------------------------------------------------|-----|--|------------------------------------------------|----|
AND/OR
David Krone for MXM
(Division Sign-Off)
Division of Surgical, Orthopedic,
and Restorative Devices
| 510(k) Number | K101713 |
|---------------|---------|
|---------------|---------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.