K113690 · GE Healthcare · IYN · Dec 29, 2011 · Radiology
Device Facts
Record ID
K113690
Device Name
GE LOGIQ I, LOGIQ E AND VIVID E
Applicant
GE Healthcare
Product Code
IYN · Radiology
Decision Date
Dec 29, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The LOGIQ i/e & Vivid e is a general purpose ultrasound system intended for use by a qualified physician for evaluation by ultrasound imaging or fluid flow analysis of the human body. Specific clinical applications and exam types include: Fetal/OB; Abdominal (GYN & Urology); Pediatric; Small Organ (breast, testes, thyroid); Neonatal and Adult Cephalic; Cardiac (adult & pediatric); Peripheral Vascular; Intra-operative (abdominal, thoracic and PV), Musculo-skeletal Conventional & Superficial, Transesophageal, Transrectal and Transvaginal, and Thoracic/Pleural for motion/sliding and fluid detection.
Device Story
Compact, portable ultrasound system; main console chassis with integrated keyboard, color LCD display, interchangeable electronic-array transducers. Digital acquisition, processing, display; battery or external power. Used in OR, ER, outpatient care, hospital rounds, nursing homes. Physician-operated. Provides ultrasound imaging/fluid flow analysis for clinical evaluation. Supports various modes: B, M, PW/CW Doppler, Color/Power Doppler, Harmonic Imaging, Coded Pulse. Aids clinical decision-making via real-time visualization of anatomy and hemodynamics. Benefits include portability and convenience for point-of-care diagnostics.
Clinical Evidence
Bench testing only. Quality assurance measures included risk analysis, requirements/design reviews, module verification, system integration testing, performance/safety verification, and simulated use validation. No clinical data presented.
Technological Characteristics
Compact ultrasound system; electronic-array transducers. Modes: B, M, PW/CW Doppler, Color/Power Doppler, Harmonic Imaging, Coded Pulse. Materials: Biocompatible patient contact materials (including needle guidance kits). Power: Integrated battery or external charger. Connectivity: Digital acquisition/processing. Software: System-level configurations for specific clinical packages (Nerve, ED, GI).
Indications for Use
Indicated for qualified physicians to perform ultrasound imaging or fluid flow analysis of the human body. Clinical applications include Fetal/OB, Abdominal (GYN/Urology), Pediatric, Small Organ (breast, testes, thyroid), Neonatal/Adult Cephalic, Cardiac (adult/pediatric), Peripheral Vascular, Intra-operative (abdominal, thoracic, PV), Musculo-skeletal (conventional/superficial), Transesophageal, Transrectal, Transvaginal, and Thoracic/Pleural (motion/sliding/fluid detection).
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
K113690
DEC 2 9 2011
GE Healthcare 510(k) Premarket Notification Submission
#### 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided: December 14, 2011 Date: Submitter: GE Healthcare 9900 W. Innovation Dr. Wauwatosa, WI 53226 Primary Contact Person: Brvan Behn Regulatory Affairs Manager GE Healthcare T:(414)721-4214 F:(414)918-8275 Secondary Contact Person: Yalan Wu Regulatory Affairs Manager GE Healthcare T: +86 510 8527 8652 F: +86 510 8527 7347 Device: Trade Name: LOGIQ i, LOGIQ e, Vivid e Common/Usual Name: LOGIQ i, LOGIQ e, Vivid e Classification Names: Class II Ultrasonic Pulsed Doppler Imaging System. 21CFR 892.1550 90-Product Code: IYN Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560. 90-IYO Diagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX Predicate Device(s): LOGIQ i, LOGIQ e, Vivid e, K102256 LOGIO E9 K110943 Device Description: The LOGIQ ife & Vivid e is a full-featured, compact and extremely portable ultrasound system consisting of a main console chassis with integrated keyboard, a color video LCD type display and several interchangeable electronic-array transducers. It has digital acquisition, processing and display capability and operates from an integrated battery or a separate power supply/charger. It is used primarily where portability, size and convenience are essential and is available in three variations: LOGIQ e is for general purpose radiology imaging having a variety of options to tailor it for use by the medical specialist
for use in various departments and patient care areas (e.g., OR,
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a black and white logo of General Electric (GE). The logo consists of the letters 'G' and 'E' intertwined within a circular border. The design is stylized and recognizable as the brand symbol for the company.
GE Healthcare 510(k) Premarket Notification Submission
ER & outpatient care). Accordingly it has various configurations (Software option/standard and probes supporting) for different packages (Nerve, ED, and GI). LOGIO i is a full-featured premium radiology system suitable for physician rounds in a hospital, nursing home where a broader range of high performance is needed. Vivid e is similar to the LOGIO e with an emphasis on cardiology examinations. The LOGIO i/e & Vivid e is a general purpose ultrasound Intended Use: system intended for use by a qualified physician for evaluation by ultrasound imaging or fluid flow analysis of the human body. Specific clinical applications and exam types include: Fetal/OB; Abdominal (GYN & Urology); Pediatric; Small Organ (breast, testes, thyroid); Neonatal and Adult Cephalic; Cardiac (adult & pediatric); Peripheral Vascular; Intra-operative (abdominal, thoracic and PV), Musculo-skeletal Conventional & Superficial, Transesophageal, Transrectal and Transvaginal, and Thoracic/Pleural for motion/sliding and fluid detection.
The LOGIQ i/e & Vivid e employs the same fundamental Technology: scientific technology as its predicate.
Determination of Summary of Non-Clinical Tests: Substantial Equivalence: The LOGIQ i/e & Vivid e and its applications comply with voluntary standards as detailed in Section 9, 11 and 17 of this premarket submission. The following quality assurance measures were applied to the development of the system:
- Risk Analysis
- Requirements Reviews .
- · Design Reviews
- · Testing on unit level (Module verification)
- · Integration testing (System verification)
- · Performance testing (Verification)
- Safety testing (Verification
- · Simulated use testing (Validation)
Transducer material and other patient contact materials such as needle guidance kits are biocompatible.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular border. The letters are stylized with curved lines and a vintage design, and the border has a decorative pattern.
GE Healthcare 510(k) Premarket Notification Submission
#### Summary of Clinical Tests:
The subject of this premarket submission, LOGIQ i/e & Vivid e, did not require clinical studies to support substantial equivalence.
GE Healthcare considers the LOGIQ i/e & Vivid e to be as safe, Conclusion: as effective, and performance is substantially equivalent to the predicate device(s).
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" written around it. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing. The overall design is simple and conveys a sense of professionalism and authority.
Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993
Mr. Bryan Behn GE Healthcare Regulatory Affairs Manager 9900 W Innovation Drive WAUWATOSA WI 53226
JAN 1 9 2012
Re: K113690
Trade/Device Name: GE LOGIQ I, LOGIQ e and Vivid e Diagnostic Ultrasound Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, and ITX Dated: December 14, 2011 Received: December 15, 2011
Dear Mr. Behn:
This letter corrects our substantially equivalent letter of December 29, 2011.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the GE LOGIQ I, LOGIQ e and Vivid e Diagnostic Ultrasound, as described in your premarket notification:
Transducer Model Number
| 4C-RS | 12L-RS | 6S-RS |
|--------|-----------|-----------|
| 8C-RS | 16L-RS | P2D |
| E8C-RS | i12L-RS | 6Tc-RS |
| 8L-RS | i/t739-RS | L8-18i-RS |
| 9L-RS | 3S-RS | |
{4}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
If you have any questions regarding the content of this letter, please contact Michael O'Hara at (301) 796-0294.
Sincerely Yours,
Michael D'Antoni for
Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure(s)
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a black and white logo of General Electric (GE). The logo consists of the letters 'G' and 'E' intertwined within a circular border. The border is decorated with small dots or circles, giving it a textured appearance. The logo is simple and iconic, representing the well-known multinational conglomerate.
GE Healthcare ー 510(k) Premarket Notification Submission
510(k) Number (if known):
GE LOGIQ I, LOGIQ e and Vivid e Diagnostic Ultrasound Device Name:
Indications for Use:
The LOGIQ i/e & Vivid e is a general purpose ultrasound system intended for use by a qualified physician for evaluation by ultrasound imaging or fluid flow analysis of the human body. Specific clinical applications and exam types include: Fetal/OB; Abdominal (GYN & Urology); Pediatric; Small Organ (breast, testes, thyroid); Neonatal and Adult Cephalic; Cardiac (adult & pediatric); Peripheral Vascular; Intraoperative (abdominal, thoracic and PV), Musculo-skeletal Conventional & Superficial, Transesophageal, Transrectal and Transvaginal, and Thoracic/Pleural for motion/sliding and fluid detection.
Prescription Use x (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use N/A (Part 21 CFR 801 Subpart C)
#### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Mary S. Patil
(Division Sign-Off) Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) Number
**Page 1 of 1**
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined within a circle. The letters are stylized and connected, with the "G" looping around the "E". The logo is black and white.
### GE Healthcare 510(k) Premarket Notification Submission
#### Indications For Use Forms
The following Indication for Use forms are attached.
System: GE LOGIQ ile & Vivid e Transducer: 4C-RS Transducer: 8C-RS Transducer: E8C-RS Transducer: 8L-RS Transducer: 9L-RS Transducer: 12L-RS Transducer: 16L-RS Transducer: i12L-RS Transducer: i/t739-RS Transducer: 3S-RS Transducer: 6S-RS Transducer: P2D Transducer: 6Tc-RS Transducer: L8-18i-RS
The following forms represent indications with clinical applications and exam types along with the modes of operation for the LOGIQ i/e & Vivid e system and for all of its probe/mode combinations. Combinations identified by "N" are new while "P" represents those previously cleared with the unmodified LOGIO i/e & Vivid e. The subject modification does not alter the previously cleared system level indications or clinical applications.
Mary's Protel
DIVISION SIGH:OH
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
510K. K113690
Confidential and Privileged. This document continential and privileged under . information of General Electric Co. and as such may not be disclosed to others not employed by General Electic Co. All rights reserved
{7}------------------------------------------------
# GE Compact Ultrasound System LOGIO i, LOGIO e, Vivid e
Mode of Operation Power Combined Harmonic Coded Other Color M Clinical Application CW Color PW M B Imaging Pulse Doppler Doppler Modes Doppler Doppler Doppler Anatomy/Region of Interest Ophthalmic P P P P P P P P P P Fetal / Obstetrics P P P P Abdomina[1] P P P P P P P P P P P P P P P P Pediatric P P P Small Organ 121 P P P p P P P P P P P P P P P Neonatal Cephalic P P P P P P P P P P Adult Cephalic P P P P P P P P P Cardiaclif P P P P P Peripheral Vascular P P P P P P P P P P P P P Musculo-skeletal Conventional P P P P Musculo-skeletal P P P P Superficial P P P P Thoracic/Pleural(specify)41 P P P P P P P P Other الموار P p P P P P P P Exam Type, Means of Access P P P P P P P P P P Transesophageal Transectal P P P P P P P P P P P P P P Transvaginal Transurethral Intraoperative(specify)(6) P P P P P P P P Intraoperative Neurological Intravascular Laparoscopic
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes GYN and Urological
- [2] Small organ includes breast, testes, thyroid.
- [3] Cardiac is Adult and Pediatric.
- [4] For detection of fluid and pleural motion/sliding;
[5] Other use includes Urology/Prostate
of fluid and pleural thickening,
includes Urology/Prostate
Office of in Ve
[6] Intraoperative includes abdominal, thoracic and peripheral;
[*] Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWD.
["] Coded Pulse is for digitally encoded harmonics.
Marr S/Pratt
(Division Sign-Off)
Division of Radiological Devices Offica of In Vitro Diagnostic Device Evaluation and Safety
810K
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
{8}------------------------------------------------
### GE Compact Ultrasound LOGIQ i, LOGIO e, Vivid e with 4C-RS Transducer
Mode of Operation Color M Power Other Combined Coded Clinical Application CW Color Harmonic PW B ખ Pulse Modes Imaging Doppler Doppler Doppler Doppler Doppler Anatomy/Region of Interest Ophthalmic P P P P P Fetal / Obstetrics P P P P P P P Abdominal!!! P P P P P P P P P P P P Pediatric Small Organ (2) Neonatal Cephalic Adult Cephalic Cardiad II Peripheral Vascular Musculo-skeletal Conventional Musculo-skeletal Superficial P P Thoracic/Pleural(specify) (4) P P P P P P Other 151 P P P P P P P P Exam Type, Means of Access Transesophageal Transrectal Transvaginal Transurethral Intraoperative (specify)[6] Intraoperative Neurological Intravascular Laparoscopic
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes GYN and Urological
- [2] Small organ includes breast, testes, thyroid.
[3] Cardiac is Adult and Pediatric.
[4] For detection of fluid and pleural motion/sliding;
[5] Other use includes Urology/Prostate
[6] Intraoperative includes abdominal, thoracic and peripheral; 1*1 Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWD.
[""] Combined modes are B/M, B/P WD, B/Color* WD,
Q.S.A.P.L. - Spatially encoded harmonics
[*] Coded Pulse is for digitally encoded harmonics.
Mary Shady
(Division Sign-Off)
Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Sefety
610K K113690
(PLEASE DO NOT WRITE BELQW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
{9}------------------------------------------------
## GE Compact Ultrasound LOGIQ i, LOGIQ e, Vivid e with 8C-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|-------------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|-------|
| Clinical Application<br>Anatomy/Region of<br>Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes' | Harmonic<br>Imaging | Coded<br>Pulse | Other |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | | | | |
| Abdominal[1] | P | P | P | | P | | P | P | P | P | |
| Pediatric | P | P | P | | P | | P | P | P | P | |
| Small Organ [2] | P | P | P | | P | | P | P | P | P | |
| Neonatal Cephalic | P | P | P | | P | | P | P | P | P | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac[3] | P | P | P | | P | | P | P | P | P | |
| Peripheral Vascular | P | P | P | | P | | P | P | P | P | |
| Musculo-skeletal<br>Conventional | P | P | P | | P | | P | P | P | P | |
| Musculo-skeletal<br>Superficial | N | N | N | | N | | N | N | N | N | |
| Thoracic/Pleura(specify) [4] | P | P | P | | P | | P | P | P | P | |
| Other[5] | | | | | | | | | | | |
| Exam Type, Means of<br>Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intraoperative (specify) [6] | | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes GYN and Urological
- [2] Small organ includes breast, testes, thyroid.
[3] Cardiac is Adult and Pediatric.
[4] For detection of fluid and pleural motion/sliding;
[5] Other use includes Urology/Prostate
[6] Intraoperative includes abdominal, thoracic and peripheral;
[*] Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWF30K
{"} Coded Pulse is for digitally encoded harmonics .
Mary S Patil
---
(Division Sign-Off)
Division of Redictories De Office of In Vitro
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
{10}------------------------------------------------
# GE Compact Ultrasound LOGIO i, LOGIO e, Vivid e with E8C-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|----------------------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|
| Clinical Application<br><i>Anatomy/ Region of<br/>Interest</i> | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse |
| Ophthalmic | | | | | | | | | | |
| Fetal / Obstetrics | P | P | P | | P | | P | P | P | |
| Abdominal <sup>(1)</sup> | P | P | P | | P | | P | P | P | |
| Pediatric | | | | | | | | | | |
| Small Organ <sup>[2]</sup> | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Thoracic/Pleura(specify) <sup>(4)</sup> | | | | | | | | | | |
| Other <sup>[5]</sup> | P | P | P | | P | | P | P | P | |
| Exam Type, Means of<br>Access | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | P | P | P | | P | | P | P | P | |
| Transvaginal | P | P | P | | P | | P | P | P | |
| Transurethral | | | | | | | | | | |
| Intraoperative (specify) <sup>(6)</sup> | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes GYN and Urological
[2] Small organ includes breast, testes, thyroid.
[3] Cardiac is Adult and Pediatric.
Mary S Patel
*Division Sign-Off*
*Division of Radiological Devices*
[4] For detection of fluid and pleural motion/sliding;
[5] Other use includes Urology/Prostate
510K
[6] Intraoperative includes abdominal, thoracic and peripheral;
["] Coded Pulse is for digitally encoded harmonics.
(Division Sign-On)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
[*] Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWD.
[*] Coded Pulse is for digitally encoded harmonics.
(PLEASED ON NOT WRITE BELOW THIS LINE _ CONTINUE ON ANOTHER PAGE IE NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
{11}------------------------------------------------
### GE Compact Ultrasound LOGIQ i, LOGIQ e, Vivid e with 8L-RS Transducer
Mode of Operation Other Power Combined Harmonic Coded Clinical Application PW CW Color Color M B M Imaging Pulse Doppler Modes Doppler Doppler Doppler Doppler Anatomy/Region of Interest Ophthalmic Fetal / Obstetrics P P P P Abdominal™ P P P P P P P P P Pediatric P P P P P P Small Organ 121 P P P P P Neonatal Cephalic Adult Cephalic Cardiac [3] P P P P P P P P P P P P P Peripheral Vascular P P P P P P P P P Musculo-skeletal Conventional Musculo-skeletal Superficial Thoracic/Pleural(specify) [4] Other 151 Exam Type, Means of Access Transesophageal Transrectal Transvaginal Transurethral Intraoperative (specify) (6) P P P P P P P Intraoperative Neurological Intravascular Laparoscopic
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes GYN and Urological
- [2] Small organ includes breast, testes, thyroid.
- [3] Cardiac is Adult and Pediatric.
- [4] For detection of fluid and pleural motion/sliding;
[5] Other use includes Urology/Prostate
Mory Spath
(Division Sign-Off)
Division of Rediological Devices Office of In Vitro Diagnostic Device Evaluation and Satety
510K
[6] Intraoperative includes abdominal, thoracic and peripheral; [*] Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWD.
[*] Coded Pulse is for digitally encoded harmonics.
['] Coded Pulse is for digitally encoded harmonics.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
{12}------------------------------------------------
# GE Compact Ultrasound LOGIQ i, LOGIQ e, Vivid e with 9L-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|-------------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|-----------------|-------|
| Clinical Application<br>Anatomy/Region of<br>Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse* | Other |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | | | | |
| Abdominal [1] | P | P | P | | P | | P | P | P | P | |
| Pediatric | P | P | P | | P | | P | P | P | P | |
| Small Organ [2] | P | P | P | | P | | P | P | P | P | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac[3] | | | | | | | | | | | |
| Peripheral Vascular | P | P | P | | P | | P | P | P | P | |
| Musculo-skeletal<br>Conventional | P | P | P | | P | | P | P | P | P | |
| Musculo-skeletal<br>Superficial | N | N | N | | N | | N | N | N | N | |
| Thoracic/Pleural(specify) [4] | N | N | N | | N | | N | N | N | N | |
| Other [5] | | | | | | | | | | | |
| Exam Type, Means of<br>Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intraoperative (specify) [6] | P | P | P | | P | | P | P | P | P | |
| Intraoperative<br>Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
Notes: [1] Abdominal includes GYN and Urological
- [2] Small organ includes breast, testes, thyroid.
- [3] Cardiac is Adult and Pediatric.
- [4] For detection of fluid and pleural motion/sliding;
- [5] Other use includes Urology/Prostate
[6] Intraoperative includes abdominal, thoracic and peripheral; 610K
[*] Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWD.
["] Coded Pulse is for digitally encoded harmonics.
Image /page/12/Figure/12 description: The image shows a signature, presumably of a person. The signature appears to read "Mary Stott". Below the signature is the text "(Division Sign-Off)" in a smaller font size. A horizontal line is drawn beneath the signature and above the text.
Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEFDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
{13}------------------------------------------------
### GE Compact Ultrasound LOGIQ i, LOGIO e, Vivid e with 12L-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Mode of Operation Clinical Application CW Color Color M Power Combined Harmonio Coded PW B M Imaging Pulse Doppler Doppler Dopple Modes Doppler Doppler Anatomy/Region of Interest Ophthalmic Fetal / Obstetrics N N N N N N Abdominal N N P P P P P Pediatric P P P Small Organ 121 P P P P P P P P Neonatal Cephalic Adult Cephalic Cardiac P P P P P Peripheral Vascular P P P P P Musculo-skeleta! P P P P P P Conventional Musculo-skeletal P P P P P P P P Superficial Thoracic/Pleural(specify) (41 P P P P P P P P Other (specify) [3] Exam Type, Means of Access Transesophageal Transrectal Transvaginal Transurethrai Intraoperative(specify) 161 Intraoperative Neurological Intravascular Laparoscopic N = new indication; P = previously cleared by FDA; E = added under Appendix E
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Notes: [1] Abdominal includes GYN and Urological
[2] Small organ includes breast, testes, thyroid.
[3] Cardiac is Adult and Pediatric.
[4] For detection of fluid and pleural motion/sliding;
[5] Other use includes Urology/Prostate
[5] Other use includes Urology/Prostate
Mary Stastil
610K
Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
[6] Intraoperative includes abdominal, thoracic and peripheral;
[*] Combined modes are B/M. B/PWD. B/Color/PWD, B/Power/PWD.
["] Coded Pulse is for digitally encoded harmonics.
(19).E.A.SE.D.O.NOT WRITE.BELOW THIS LINE_CONTINUE.ON ANOTHER JAGE (ENEFOD)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
{14}------------------------------------------------
## GE Compact Ultrasound LOGIQ i, LOGIO e, Vivid e with 16L-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|-----------------|
| Clinical Application<br>Anatomy/Region of<br>Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse* |
| Ophthalmic | | | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Pediatric | P | P | P | | P | | P | P | P | P |
| Small Organ [2] | P | P | P | | P | | P | P | P | P |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Peripheral Vascular | P | P | P | | P | | P | P | P | P |
| Musculo-skeletal<br>Conventional | P | P | P | | P | | P | P | P | P |
| Musculo-skeletal<br>Superficial | P | P | P | | P | | P | P | P | P |
| Thoracic/Pleural(specify)[4] | | | | | | | | | | |
| Other (specify) [5] | | | | | | | | | | |
| Exam Type, Means of<br>Access | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intraoperative(specify) [6] | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| N = new indication; P = previously cleared by FDA; E = added under Appendix E<br>Notes: [1] Abdominal includes GYN and Urological<br>[2] Small organ includes breast, testes, thyroid.<br>[3] Cardiac is Adult and Pediatric.<br>[4] For detection of fluid and pleural motion/sliding;<br>[5] Other use includes Urology/Prostate | | | | | | | | | | |
HURST PI
(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)
・・
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
[6] Intraoperative includes abdominal, thoracic and peripheral; [*] Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWD.
[*] Coded Pulse is for digitally encoded harmonics.
{15}------------------------------------------------
# GE Compact Ultrasound LOGIO i, LOGIO e, Vivid e with i12L-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application<br>Anatomy/Region of<br>Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse* | Other |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|-----------------|-------|
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | | | | |
| Abdominal[1] | P | P | P | | P | | P | P | P | P | |
| Pediatric | P | P | P | | P | | P | P | P | P | |
| Small Organ [2] | P | P | P | | P | | P | P | P | P | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac[3] | P | P | P | | P | | P | P | P | | |
| Peripheral Vascular | P | P | P | | P | | P | P | P | P | |
| Musculo-skeletal<br>Conventional | P | P | P | | P | | P | P | P | | |
| Musculo-skeletal<br>Superficial | P | P | P | | P | | P | P | P | | |
| Thoracic/Pleural(specify)[4] | | | | | | | | | | | |
| Other[5] | | | | | | | | | | | |
| Exam Type, Means of<br>Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intraoperative (specify)[6] | P | P | P | | P | | P | P | P | | |
| Intraoperative<br>Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
| N = new indication; P = previously cleared by FDA; E = added under Appendix E | | | | | | | | | | | |
| Notes: [1] Abdominal includes GYN and Urological | | | | | | | | | | | |
| [2] Small organ includes breast, testes, thyroid. | | | | | | | | | | | |
| [3] Cardiac is Adult and Pediatric. | | | | | | | | | | | |
| [4] For detection of fluid and pleural motion/sliding; | | | | | | | | | | | |
| [5] Other use includes Urology/Prostate | | | | | | | | | | | |
| [6] Intraoperative includes abdominal, thoracic and peripheral;[*] Combined modes are B/M, B/PWD,<br>B/Color/PWD, B/Power/PWD.<br>[*] Coded Pulse is for digitally encoded harmonics. | | | | | | | | | | | |
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
{16}------------------------------------------------
## GE Compact Ultrasound LOGIQ i, LOGIQ e, Vivid e with i/t739-RS Transducers
| | Mode of Operation | | | | | | | | | | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|-----------------|
| Clinical Application<br>Anatomy/Region of<br>Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | | | | |
| Abdominal[1] | P | P | P | | P | | P | P | P | P | |
| Pediatric | P | P | P | | P | | P | P | P | P | |
| Small Organ [2] | P | P | P | | P | | P | P | P | P | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac[3] | P | P | P | | P | | P | P | P | P | |
| Peripheral Vascular | P | P | P | | P | | P | P | P | P | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | | |
| Thoracic/Pleural(specify)[4] | | | | | | | | | | | |
| Other[4] | | | | | | | | | | | |
| Exam Type, Means of<br>Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intraoperative (specify)[6]…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.