GE LOGIQ I/E & VIVID E COMPACT DIAGNOSTIC ULTRASOUND
K102256 · Ge Medical Systems China Co., Ltd. · IYN · Oct 5, 2010 · Radiology
Device Facts
Record ID
K102256
Device Name
GE LOGIQ I/E & VIVID E COMPACT DIAGNOSTIC ULTRASOUND
Applicant
Ge Medical Systems China Co., Ltd.
Product Code
IYN · Radiology
Decision Date
Oct 5, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
The LOGIQ ile & Vivid e is a general purpose ultrasound system intended for use by a qualified physician for evaluation by ultrasound imaging or fluid flow analysis of the human body. Specific clinical applications and exam types include: Fetal/OG; Abdominal (GYN & Urology); Pediatric; Small Organ (breast, testes, thyroid); Neonatal and Adult Cephalic; Cardiac (adult & pediatric); Peripheral Vascular; Intra-operative (abdominal, thoracic and PV), Musculo-skeletal Conventional & Superficial, Transesophageal, Transrectal and Transvaginal, and Thoracic/Pleural for motion/sliding and fluid detection.
Device Story
Compact, portable diagnostic ultrasound system (LOGIQ i, LOGIQ e, Vivid e) featuring integrated keyboard, LCD display, and interchangeable electronic-array transducers. Device acquires, processes, and displays ultrasound data; supports B-mode, M-mode, PW/CW Doppler, Color/Power Doppler, and harmonic imaging. Used in clinical settings by physicians for real-time imaging and fluid flow analysis. Output displayed on integrated GUI; assists clinicians in diagnostic evaluation of anatomy and pathology. Benefits include portable, point-of-care diagnostic capability for diverse clinical applications.
Clinical Evidence
No clinical data required. Substantial equivalence established via non-clinical bench testing, including acoustic output, biocompatibility, cleaning/disinfection effectiveness, electromagnetic compatibility, and thermal/electrical/mechanical safety testing.
Technological Characteristics
Portable ultrasound system with integrated keyboard, LCD, and electronic-array transducers. Supports B, M, PW/CW Doppler, Color/Power Doppler, and harmonic imaging modes. Conforms to 21 CFR 892.1560, 892.1550, and 892.1570. Design conforms to 21 CFR 820, ISO 9001:2000, and ISO 13485. Connectivity includes digital acquisition and display.
Indications for Use
Indicated for qualified physicians to perform ultrasound imaging or fluid flow analysis. Patient population includes fetal, pediatric, and adult patients across various clinical applications: OB/GYN, urology, small organ, cephalic, cardiac, vascular, musculoskeletal, and thoracic/pleural. Contraindications not specified.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
GE LOGIQ-ile & Vivid-e Compact Ultrasound (K091374)
Submission Summary (Full Text)
{0}------------------------------------------------
**OCT** 5 2010
142
# Attachment B
510(k) Summary of Safety and Effectiveness Prepared in accordance with 21 CFR Part 807.92(c).
#### GE Healthcare GE Medical Systems (China) Co., Ltd No. 19 Changjiang Road, National Hi-Tech Development Zone Wuxi, Jiangsu Province, CHINA 214028 Section a): 1. Submitter: GE Medical Systems (China) Co., Ltd. No. 19 Changjiang Road, National Hi-Tech Development Zone, Wuxi, Jiangsu Province, CHINA 214028 Contact Person: Yalan Wu, Manager, Safety and Regulatory Telephone: 86-510-85278652; Fax: 86-510-85227347 Date Prepared: June 20, 2010 2. Device Name: GE LOGIQ ille, Vivid e Ultrasound Ultrasonic Pulsed Echo Imaging System, 21 CFR 892.1560, 90-IYO Ultrasonic Pulsed Doppler Imaging System, 21 CFR 892.1550, 90-IYN Diagnostic Ultrasonic Transducer, 21 CFR 892.1570, 90-ITX 3. Marketed Device: GE LOGIQ-ile & Vivid-e Compact Ultrasound, 510(k) No: K091374 (90-IYO/IYN/ITX) A device currently in commercial distribution.
4. Device Description: The GE Compact Ultrasound is a very compact and portable diagnostic ultrasound system having three variations: LOGIQ e and Vivid e, each with options and features suited for its It has an integrated keyboard, LCD display and several interchangeable electronic-array market niche. transducers with an approximate size of 34 cm wide, 29 cm deep and 6 cm high in transport configuration and provides digital acquisition, processing and display capability. The user interface includes a computer keyboard, an intuitive layout of specialized controls, color GUI display and Doppler audio.
5. Indications for Use: The subject modified device is a general purpose ultrasound system intended for use by a qualified physician for evaluation by ultrasound imaging or fluid flow analysis of the human body. Specific clinical applications and exam types include: Fetal/OB; Abdominal (GYN & Urology); Pediatric; Small Organ (breast, testes, thyroid); Neonatal and Adult Cephalic; Cardiac (adult & pediatric); Peripheral Vascular; Intraoperative (abdominal, thoracic and PV), Musculo-skeletal Conventional & Superficial, Transesophageal, Transrectal and Transvaginal, and Thoracic/Pleural for motion/sliding and fluid detection.
6. Companison with Predicate Device: The modified Compact Ultrasound devices are virually identical having the same design, construction, materials, brand names and intended uses. All technological characteristics and safety and effectiveness features are equivalent. The modified device has additional Transesophageal indication for use and Intima Media Thickness (IMT) measurement with Vascular.
#### Section b):
Non-clinical Tests: The device has been evaluated for acoustic output, biocompatibility, cleaning and 1. disinfection effectiveness, electromagnetic compatibility, as well as themal, electrical and mechanical safety, and has been found to conform with applicable medical device safety standards.
- 2. Clinical Tests: None required.
3. Conclusion: Intended uses and other key features are consistent with traditional clinical practice, FDA guidelines, and established methods of patient examination. The design and development process of the manufacturer conforms with 21 CFR 820, ISO 9001:2000 and ISO 13485 quality systems. The device conforms to applicable medical device safety standards and compliance is verified through independent evaluation with ongoing factory surveillance. Diagnostic ultrasound has accumulated a long history of safe and effective performance. Therefore, it is the opinion of GE Healthcare that the GE LOGIQ ile Vivid e Ultrasound imaging device is substantially equivalent with respect to safety and effectiveness to devices currently cleared for market.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
GE Medical Systems, Ultrasound and Primary Care Diagnostics, LLC % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 2510 Street NW BUFFALO MN 55313
OCT 5 2010
Re: K102256
Trade/Device Name: GE LOGIQ i, LOGIQ e and Vivid e Diagnostic Ultrasound Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasound pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, and ITX Dated: September 27, 2010 Received: September 28, 2010
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the GE LOGIQ i, LOGIQ e and Vivid e Diagnostic Ultrasound, as described in your premarket notification:
| | Transducer Model Number | |
|--------|-------------------------|--------|
| 4C-RS | 12L-RS | 6S-RS |
| 8C-S | 16L-RS | P2D |
| E8C-RS | i12L-RS | 6Tc-RS |
| 8L-RS | i/t739-RS | |
| 9L-RS | 3S-RS | |
{2}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
If you have any questions regarding the content of this letter, please contact Shahram Vaezy at (301) 796-6242.
Sincerely yours,
Mach. A.D.O. Then for
David G. Brown, Ph.D. Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure(s)
{3}------------------------------------------------
# Indications for Use Form
510(k) Number (if known):
Device Name: __GE LOGIQ i, LOGIQ e and Vivid e Diagnostic Ultrasound
#### Indications For Use:
The LOGIQ ile & Vivid e is a general purpose ultrasound system intended for use by a qualified physician for evaluation by ultrasound imaging or fluid flow analysis of the human body. Specific clinical applications and exam types include: Fetal/OG; Abdominal (GYN & Urology); Pediatric; Small Organ (breast, testes, thyroid); Neonatal and Adult Cephalic; Cardiac (adult & pediatric); Peripheral Vascular; Intra-operative (abdominal, thoracic and PV), Musculo-skeletal Conventional & Superficial, Transesophageal, Transrectal and Transvaginal, and Thoracic/Pleural for motion/sliding and fluid detection.
Prescription Use × AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics
MuhaMD.D.M. David 6, Brown
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K102056
Page 1 of 1
{4}------------------------------------------------
### GE Compact Ultrasound System
LOGIQ i, LOGIQ e, Vivid e
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application<br>Anatomy/Region of Interest | Mode of Operation | | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|-----------------|-------|--|
| | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse* | Other | |
| Ophthalmic | | | | | | | | | | | | |
| Fetal / Obstetrics | P | P | P | P | P | P | P | P | P | P | P | |
| Abdominal[1] | P | P | P | P | P | P | P | P | P | P | P | |
| Pediatric | P | P | P | P | P | P | P | P | P | P | P | |
| Small Organ [2] | P | P | P | P | P | P | P | P | P | P | P | |
| Neonatal Cephalic | P | P | P | P | P | P | P | P | P | P | P | |
| Adult Cephalic | P | P | P | P | P | P | P | P | P | P | P | |
| Cardiac [3] | P | P | P | P | P | P | P | P | P | P | P | |
| Peripheral Vascular | P | P | P | P | P | P | P | P | P | P | P | |
| Musculo-skeletal Conventional | P | P | P | P | P | P | P | P | P | P | P | |
| Musculo-skeletal Superficial | P | P | P | P | P | P | P | P | P | P | P | |
| Thoracic/Pleural (specify)[4] | P | P | P | P | P | P | P | P | P | P | P | |
| Other [5] | P | P | P | P | P | P | P | P | P | P | P | |
| Exam Type, Means of Access | | | | | | | | | | | | |
| Transesophageal | N | N | N | N | N | N | N | N | N | N | N | |
| Transrectal | P | P | P | P | P | P | P | P | P | | | |
| Transvaginal | P | P | P | P | P | P | P | P | P | | | |
| Transuretheral | | | | | | | | | | | | |
| Intraoperative | P | P | P | P | P | P | P | P | P | P | | |
| Intraoperative Neurological | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes GYN and Urological
[2] Small organ includes breast, testes, thyroid.
[3] Cardiac is Adult and Pediatric.
{4] For detection of fluid and pleural motion/sliding;
[5] Other use includes Urology/Prostate
[*] Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWD.
["] Coded Pulse is for digitally encoded harmonics.
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics (OIVD)
Nuhal D. O'Kee for David C. Brown
Division Sign Off
510(k) K102256
Office of In Vitro Diagnostic Device Evaluation and Safety
{5}------------------------------------------------
# GE Compact Ultrasound with 4C-RS Transducer
#### LOGIQ i, LOGIQ e, Vivid e
| | Mode of Operation | | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|-------|--|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other | |
| Ophthalmic | | | | | | | | | | | | |
| Fetal / Obstetrics | P | P | P | | P | | P | P | P | P | | |
| Abdominal[1] | P | P | P | | P | | P | P | P | P | | |
| Pediatric | P | P | P | | P | | P | P | P | P | | |
| Small Organ [2] | | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Cardiac[3] | | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | | |
| Thoracic/Pleural (specify) [4] | P | P | P | | P | | P | P | P | P | | |
| Other[5] | P | P | P | | P | | P | P | P | P | | |
| Exam Type, Means of Access | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes GYN and Urological
[2] Small organ includes breast, testes, thyroid.
[3] Cardiac is Adult and Pediatric.
[4] For detection of fluid and pleural motion/sliding;
[5] Other use includes Urology/Prostate
[*] Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWD.
["] Coded Pulse is for digitally encoded harmonics.
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics (OIVD)
Mudand D. O'K. Go. David b. Brown
Division Sign-Off
Prescription Use (21 CFR 801 Subpart D)
Office of In Vitro Diagnostic Device Evaluation and Safe
510(k) K102256
{6}------------------------------------------------
### GE Compact Ultrasound with 8C-RS Transducer
#### LOGIQ i, LOGIQ e, Vivid e
intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|-------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | | | | |
| Abdominal[1] | P | P | P | | P | | P | P | P | | |
| Pediatric | P | P | P | | P | | P | P | P | | |
| Small Organ [2] | P | P | P | | P | | P | P | P | | |
| Neonatal Cephalic | P | P | P | | P | | P | P | P | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac [3] | P | P | P | | P | | P | P | P | | |
| Peripheral Vascular | P | P | P | | P | | P | P | P | | |
| Musculo-skeletal Conventional | P | P | P | | P | | P | P | P | | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Thoracic/Pleural (specify) [4] | P | P | P | | P | | P | P | P | | |
| Other[5] | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes GYN and Urological
[2] Small organ includes breast, testes, thyroid.
[3] Cardiac is Adult and Pediatric.
[4] For detection of fluid and pleural motion/sliding;
[5] Other use includes Urology/Prostate
(*) Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWD.
] Coded Pulse is for digitally encoded harmonics .
### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics (OIVD)
Mhll.D.O'K. for David C. Brown
Division Sign-Off
Prescription Use (21 CFR 801 Subpart D)
Office of In Vitro Diagnostic Device Evaluation and S
510(k) K102256
{7}------------------------------------------------
## GE Compact Ultrasound with E8C-RS Transducer
#### LOGIQ İ, LOGIQ e, Vivid e
| | | Mode of Operation | | | | | | | | | |
|-----------------------------------------------------|---|-------------------|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|--|
| Clinical Application<br>Anatomy/ Region of interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | P | P | P | | P | | P | P | P | | |
| Abdominal[1] | P | P | P | | P | | P | P | P | | |
| Pediatric | | | | | | | | | | | |
| Small Organ [2] | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Thoracic/Pleural (specify) [4] | | | | | | | | | | | |
| Other[5] | P | P | P | | P | | P | P | P | | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | P | P | P | | P | | P | P | P | | |
| Transvaginal | P | P | P | | P | | P | P | P | | |
| Transuretheral | | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes GYN and Urological
[2] Small organ includes breast, testes, thyroid.
[3] Cardiac is Adult and Pediatric.
[4] For detection of fluid and pleural motion/sliding;
[5] Other use includes Urology/Prostate
[*] Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWD.
[*] Coded Pulse is for digitally encoded harmonics .
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics (OIVD)
Andrew D. Kirk for David G. Brown
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K102256
{8}------------------------------------------------
## GE Compact Ultrasound with 8L-RS Transducer
#### LOGIQ İ, LOGIQ e, Vivid e
| | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|-------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | | | | |
| Abdominal[1] | P | P | P | | P | | P | P | P | P | |
| Pediatric | P | P | P | | P | | P | P | P | P | |
| Small Organ [2] | P | P | P | | P | | P | P | P | P | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac [3] | P | P | P | | P | | P | P | P | | |
| Peripheral Vascular | P | P | P | | P | | P | P | P | P | |
| Musculo-skeletal Conventional | P | P | P | | P | | P | P | P | | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Thoracic/Pleural (specify) [4] | | | | | | | | | | | |
| Other [5] | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | |
| Intraoperative (specify)[5] | P | P | P | | P | | P | P | P | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes GYN and Urological
[2] Small organ includes breast, testes, thyroid.
[3] Cardiac is Adult and Pediatric.
[4] For detection of fluid and pleural motion/sliding;
[5] Other use includes Urology/Prostate
[*] Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWD.
[1] Coded Pulse is for digitally encoded harmonics.
### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics (OIVD)
Andrew D. Thoms for David C. Brown
Division Sign Off
Prescription Use (21 CFR 801 Subpart D)
Office of In Vitro Diagnostic Device Evaluation and Safe
510(k) K102256
{9}------------------------------------------------
## GE Compact Ultrasound with 9L-RS Transducer
LOGIQ İ, LOGIQ e, Vivid e
Mode of Operation Clinical Application PW CW Color Color M B Power Combined M Harmonic Coded Other Anatomy/Region of Interest Doppler Doppler Doppler Doppler Doppler Modes lmaging Pulse® Ophthalmic Fetal / Obstetrics Abdominal [1] P P P P P P P P Pediatric P P P P P P P P Small Organ 같 P P P P P P P P Neonatal Cephalic Adult Cephalic Cardiac (3) Peripheral Vascular P P P P P P P P Musculo-skeletal Conventional P P P P P P P P Musculo-skeletal Superficial Thoracic/Pleural (specify) [4] Other (5) Exam Type, Means of Access Transesophageal Transrectal Transvaginal Transuretheral Intraoperative (specify)[5] P P P P P P P P Intraoperative Neurological Intravascular Laparoscopic
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes GYN and Urological
[2] Small organ includes breast, testes, thyroid.
[3] Cardiac is Adult and Pediatric.
[4] For detection of fluid and pleural motion/sliding;
[5] Other use includes Urology/Prostate
[*] Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWD.
*] Coded Pulse is for digitally encoded harmonics .
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics (OIVD)
Michal D. O'Shea Lisa David G. Brown
Division Sign-Off
Prescription Use (21 CFR 801 Subpart D)
ce of In Vitro Diagnostic Device Evaluation and Safety
510(k) K102256
{10}------------------------------------------------
## GE Compact Ultrasound with 12L-RS Transducer
#### LOGIQ i, LOGIQ e, Vivid e
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse* |
|----------------------------------------------------|---|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|-----------------|
| Ophthalmic | | | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Pediatric | P | P | P | | P | | P | P | P | P |
| Small Organ [2] | P | P | P | | P | | P | P | P | P |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Peripheral Vascular | P | P | P | | P | | P | P | P | P |
| Musculo-skeletal Conventional | P | P | P | | P | | P | P | P | P |
| Musculo-skeletal Superficial | P | P | P | | P | | P | P | P | P |
| Thoracic/Pleural (specify) [4] | P | P | P | | P | | P | P | P | P |
| Other (specify) [5] | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transuretheral | | | | | | | | | | |
| Intraoperative [5] (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes GYN and Urological
[2] Small organ includes breast, testes, thyroid.
[3] Cardiac is Adult and Pediatric.
[4] For detection of fluid and pleural motion/sliding;
[5] Other use includes Urology/Prostate
[*] Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWD.
(*) Coded Pulse is for digitally encoded harmonics.
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics (OIVD)
Andra D'Oliveira for David 6 Brown
Division Sign Off
510(k) K102256
Office of In Vitro Diagnostic Device Evaluation and Safety
{11}------------------------------------------------
## GE Compact Ultrasound with 16L-RS Transducer
### LOGIQ İ, LOGIQ e, Vivid e
Mode of Operation Clinical Application bM CM B Color Power Combined Color M M Harmonic Coded Anatomy/Region of Interest Doppler Doppler Doppler Doppler Modes Doppler Imaging Pulse® Ophthalmic Fetal / Obstetrics Abdominal Pediatric P P P p P P P P Small Organ 121 P P P P P P P P Neonatal Cephalic Adult Cephalic Cardiac Peripheral Vascular P P P P P P P P Musculo-skeletal Conventional P P P P P P P P Musculo-skeletal Superficial P P P P P P P P Thoracic/Pleural (specify) [4] Other (specify) 15) Exam Type, Means of Access Transesophageal Transrectal Transvaginal Transuretheral Intraoperative [3] (specify) Intraoperative Neurological Intravascular Laparoscopic
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes GYN and Urological
[2] Small organ includes breast, testes, thyroid.
[3] Cardiac is Adult and Pediatric.
[4] For detection of fluid and pleural motion/sliding;
[5] Other use includes Urology/Prostate
[*] Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWD.
["] Coded Pulse is for digitally encoded harmonics.
### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics (OIVD)
Aayyad Dothan Lan David 6. Boiso
Division Sign Off
office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K102256
{12}------------------------------------------------
## GE Compact Ultrasound with i12L-RS Transducer
#### LOGIQ i, LOGIQ e, Vivid e
| | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|-------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | | | | |
| Abdominal [1] | P | P | P | | P | | P | P | P | P | |
| Pediatric | P | P | P | | P | | P | P | P | P | |
| Small Organ [2] | P | P | P | | P | | P | P | P | P | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac [3] | P | P | P | | P | | P | P | P | | |
| Peripheral Vascular | P | P | P | | P | | P | P | P | P | |
| Musculo-skeletal Conventional | P | P | P | | P | | P | P | P | | |
| Musculo-skeletal Superficial | P | P | P | | P | | P | P | P | | |
| Thoracic/Pleural (specify) [4] | | | | | | | | | | | |
| Other [5] | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | |
| Intraoperative (specify)[5] | P | P | P | | P | | P | P | P | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes GYN and Urological
[2] Small organ includes breast, testes, thyroid.
[3] Cardiac is Adult and Pediatric.
{4] For detection of fluid and pleural motion/sliding;
[5] Other use includes Urology/Prostate
[*] Combined modes are B/M, B/PWD, B/Color/PWD, 8/Power/PWD.
[1] Coded Pulse is for digitally encoded harmonics .
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
nce of CDRH, Office of In Vitro Diagnostics (OIVD)
Auhul 1052K for David b Brown
Division Sign-Off
fice of In Vitro Diagnostic Device Evaluation and Safety
510(k) K102256
{13}------------------------------------------------
## GE Compact Ultrasound with i/t739-RS Transducers
#### LOGIQ i, LOGIQ e, Vivid e
| | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|-------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | | | | |
| Abdominal[1] | P | P | P | | P | | P | P | P | | |
| Pediatric | P | P | P | | P | | P | P | P | | |
| Small Organ [2] | P | P | P | | P | | P | P | P | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac[3] | P | P | P | | P | | P | P | P | | |
| Peripheral Vascular | P | P | P | | P | | P | P | P | | |
| Musculo-skeletal Conventional | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Thoracic/Pleural (specify) (4) | | | | | | | | | | | |
| Other (4) | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | |
| Intraoperative (specify)[5] | P | P | P | | P | | P | P | P | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes GYN and Urological
[2] Small organ includes breast, testes, thyroid.
[3] Cardiac is Adult and Pediatric.
[4] For detection of fluid and pleural motion/sliding;
[5] Other use includes Urology/Prostate
[*] Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWD.
["] Coded Pulse is for digitally encoded harmonics.
### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics (OIVD)
Ardd DDK for David b. Brown
Division Sign Off
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K102256
{14}------------------------------------------------
## GE Compact Ultrasound with 3S-RS Transducer
### LOGIQ i, LOGIQ e, Vivid e
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse* | Other |
|----------------------------------------------------|---|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|-----------------|-------|
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | P | P | P | P | P | P | P | P | P | P | |
| Abdominal[1] | P | P | P | P | P | P | P | P | P | P | |
| Pediatric | P | P | P | P | P | P | P | P | P | P | |
| Small Organ [2] | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | P | P | P | P | P | P | P | P | P | P | |
| Cardiac [3] | P | P | P | P | P | P | P | P | P | P | |
| Peripheral Vascular | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Thoracic/Pleural (specify)[4] | P | P | P | P | P | P | P | P | P | P | |
| Other [5] | P | P | P | P | P | P | P | P | P | P | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes GYN and Urological
[2] Small organ includes breast, testes, thyroid.
[3] Cardiac is Adult and Pediatric.
[4] For detection of fluid and pleural motion/sliding;
[5] Other use includes Urology/Prostate
[*] Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWD.
[*] Coded Pulse is for digitally encoded harmonics .
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics (OIVD)
Nuhad D. Khawam for David 6. Brown
Division Sign-Off
Prescription Use (21 CFR 801 Subpart D)
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K102256
{15}------------------------------------------------
# GE Compact Ultrasound with 6S-RS Transducer
#### LOGIQ İ, LOGIQ e, Vivid e
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|-----------------|-------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse¹ | Other |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | | | | |
| Abdominal[1] | P | P | P | P | P | P | P | P | P | P | P |
| Pediatric | P | P | P | P | P | P | P | P | P | P | P |
| Small Organ [2] | | | | | | | | | | | |
| Neonatal Cephalic | P | P | P | P | P | P | P | P | P | P | P |
| Adult Cephalic | P | P | P | P | P | P | P | P | P | P | P |
| Cardiac[3] | P | P | P | P | P | P | P | P | P | P | P |
| Peripheral Vascular | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Thoracic/Pleural (specify) [4] | | | | | | | | | | | |
| Other [5] | P | P | P | P | P | P | P | P | P | P | P |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | |
| Intraoperative (specify)[5] | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes GYN and Urological
[2] Small organ includes breast, testes, thyroid.
[3] Cardiac is Adult and Pediatric.
[4] For detection of fluid and pleural motion/sliding;
[5] Other use includes Urology/Prostate
["] Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWD.
["] Coded Pulse is for digitally en…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.