CMRTOOLS AND PLUG-INS VENTRICULAR TOOLS & THALASSAEMIA TOOLS
Applicant
Cardiovascular Imaging Solutions , Ltd.
Product Code
LLZ · Radiology
Decision Date
Feb 14, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
CMRtools™ and plug-ins VentricularTools™ & ThalassaemiaTools™ is a software device used for viewing and analyzing cardiovascular magnetic resonance images. It allows high performance image manipulation, 3D visualization, and advanced image processing on a standard PC without dedicated hardware. All software plugins have comprehensive quantification facilities with full analysis session recording for providing comprehensive audit trails. It contains an image viewer for importing DICOM images, browsing through patient datasets, viewing cine images and performing region-of-interest analysis. Plug-ins are used for specialist cardiac assessment including VentricularTools for left ventricular assessment and ThalassaemiaTools for T2* image quantification. It is important that CMRtools is only used to view and analyze human cardiac and or liver images acquired with a cardiovascular magnetic resonance scanner. CMRtools must not be used to view or analyze images acquired with any other type of imaging device. The software must not be used to view or analyze images of any part of the body except the heart and or liver or the vascular system. Furthermore, the software must not be used to view or analyze images acquired from animals. Only DICOM images will be used for presentation, display and diagnosis.
Device Story
CMRtools is a standalone software package for visualization and analysis of cardiovascular magnetic resonance (CMR) images. Input: DICOM images from CMR scanners. Operation: Software provides image loading, viewing, 3D visualization, and quantification via plug-ins. VentricularTools calculates ventricular volume/mass indices; ThalassaemiaTools calculates T2* values for iron loading. Context: Used in clinical settings on standard PCs; post-scan analysis. Output: Quantitative metrics and visual analysis. Clinicians use output to assist in cardiac and liver assessment. Benefits: Enables advanced image processing and quantification without dedicated hardware; provides audit trails for analysis sessions.
Clinical Evidence
No clinical data provided. Bench testing only, conducted according to a documented Software Test Plan as part of the company's Product Development Process.
Technological Characteristics
Standalone software for Windows XP/Vista. Requirements: 1GHz processor, 512MB RAM, 128MB graphics card, 2GB disk space. Workflow-oriented GUI. DICOM-compliant. No patient contact. Non-networked/standalone deployment.
Indications for Use
Indicated for viewing and analyzing human cardiovascular magnetic resonance (CMR) images of the heart, liver, and vascular system. Used for left ventricular assessment and T2* iron loading quantification in humans. Contraindicated for non-human (animal) images and non-CMR imaging modalities.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
MRI- MAGNETIC RESONANCE ANALYTICAL SOFTWARE SYSTEM (K994283)
Submission Summary (Full Text)
{0}------------------------------------------------
Ko 73194
10/2
## 510(k) Summary of Safety and Effectiveness
## Date Prepared
November 10, 2007
## Submitter's Information
Professor Guang-Zhong Yang, Director Cardiovascular Imaging Solutions Ltd. 53 Cavendish Road, London SW12 OBL, UK Tel: 011 44(0)7771 888 886
## Contact Person
Mr. Robert Merrifield Cardiovascular Imaging Solutions Ltd. 53 Cavendish Road, London SW12 OBL, UK Tel: 011 44(0)7771 888 886
## Device Name
| Common/Usual Name: | Picture Archiving Communications System |
|-----------------------|------------------------------------------------------------|
| Proprietary Name: | CMRtools and plug-ins VentricularTools & ThalassaemiaTools |
| Classification Names: | System, image processing, radiological 892.2050 90-LLZ |
## Substantially Equivalent to both:
| 510(k) Number | K060941 | K994283 |
|--------------------------------------|-----------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| Device Classification Name | system, image processing,<br>radiological | system, nuclear magnetic resonance<br>imaging |
| Device Name | CAAS MRV VERSION 3.0 | MRI- MAGNETIC RESONANCE<br>ANALYTICAL SOFTWARE SYSTEM |
| Applicant | PIE MEDICAL IMAGING B.V.<br>7450 Flying Cloud Dr.<br>Eden Prairie, MN 55344 | MEDIS MEDICAL IMAGING<br>SYSTEMS, B.V.<br>Poortgebouw Rijnsburgerweg 10<br>Leiden, NL 2333 AA |
| Regulation Number | 892.2050 | 892.1000 |
| Classification Product Code | LLZ | LNH |
| Decision Date | 04/19/2006 | 08/30/2000 |
| Decision | substantially equivalent (SE) | substantially equivalent (SE) |
| Classification Advisory<br>Committee | Radiology | Radiology |
## Device Description
CMRtools™ is a software package for the visualization and analysis of cardiovascular magnetic resonance images (CMR). CMRtools™ is comprised of an image loader, an image viewer and a set of optional plug-ins for specialist cardiac analysis, including VentricularTools and ThalassaemiaTools.
- VentricularTools provides dedicated functionality for calculating volume and mass indices of the . ventricles of the heart from magnetic resonance images.
- ThalassaemiaTools allows the calculation of a property called T2* that characterizes iron loading . in the heart and liver.
{1}------------------------------------------------
Kó 73/94
20f2
## 510(k) Summary of Safety and Effectiveness
CMRtools™ allows high performance image manipulation, 3D visualization, and advanced image processing on a standard PC without dedicated hardware. All software plug-ins have comprehensive quantification facilities with full analysis session record for providing comprehensive audit trails.
#### Indications for Use
CMRtools™ and plug-ins VentricularTools™ & ThalassaemiaTools™ is a software device used for viewing and analyzing cardiovascular magnetic resonance images. It allows high performance image manipulation, 3D visualization, and advanced image processing on a standard PC without dedicated hardware. All software plug-ins have comprehensive quantification facilities with full analysis session recording for providing comprehensive audit trails. It contains an image viewer for importing DICOM images, browsing through patient datasets, viewing cine images and performing region-of-interest analysis. Plug-ins are used for specialist cardiac assessment including VentricularTools for left ventricular assessment and ThalassaemiaTools for T2* image quantification. It is important that CMRtools is only used to view and analyze human cardiac and or liver images acquired with a cardiovascular magnetic resonance scanner. CMRtools must not be used to view or analyze images acquired with any other type of imaging device. The software must not be used to view or analyze images of any part of the body except the heart and or liver or the vascular system. Furthermore, the software must not be used to view or analyze images acquired from animals. Only DICOM images will be used for presentation, display and diagnosis.
## Technological Characteristics
CMRtools and plug-ins is a stand-alone software package which can be used on more than one hardware platform, as long as minimum hardware requirements are met. The device does not contact the patient, nor does it control any life-sustaining devices. A physician providing ample opportunity for competent human intervention interprets images and information being displayed and/or printed.
CMRtools and its plug-ins require a PC or a laptop with a mouse, keyboard and display running Microsoft Windows XP/Vista with a 1GHz processor, 512MB RAM, 128MB graphics card and 2GB of free hard disk space. The user interacts with CMRtools and its plug-ins using a desktop or laptop computer with a keyboard, mouse and display. The Software utilizes the standard point-and-click approach to user interface design that is ubiquitous on modern windowing operating systems.
CMRtools provides a user interface with individual items for selecting images, changing the brightness/contrast and accessing the different aspects of functionality. The plug-ins for CMRtools make use of a workflow orientated graphical user interface that guides the User through the analysis process by presenting a set of simple tasks. Each aspect of the software is clearly documented with context-sensitive Interactive Help. CMRtools and its plug-ins are designed to be used after the scanning of the patient is complete. The images will typically be transferred via a CD, network drive or though a PACS transfer from the scanner onto the clinician's local machine. Image viewing and analysis can then be performed without connection to the internet or a network.
#### Testing
CMRtools and plug-ins have been tested according to the specifications that are documented in a Software Test Plan. Testing is an integral part Cardiovascular Imaging Solutions Ltd. software development process as described in the company's Product Development Process.
### Conclusion
The 510(k) pre-market notification for the CMRtools and plug-ins VentricularTools & ThalassaemiaTools contains adequate information and data to enable FDA-CDRH to determine substantial equivalence to the predicate device.
- The device has been and will continue to be manufactured according to the voluntary standards 1. list in the Voluntary Standards section of the submission.
- The submission contains the results of a hazard analysis and the "Level of Concern for potential 2. hazards has been classified as "Minor".
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes extending from its back, representing the department's commitment to health, human services, and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Public Health Service
2000 ।
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Cardiovascular Imaging Solutions Ltd. c/o Mr. Carl Thomas, Consultant OTech, Inc. 1600 Manchester Way CORINTH TX 76210
K073194 Re:
> Trade/Device Name: CMRtools" and plug-ins VentricularTools & Thalassaemia Tools Regulation Number: 21 CFR §892.2050 Regulation Name: Picture archiving and communications system
Regulatory Class: II Product Code: LLZ
Dated: November 12, 2007 Received: November 13, 2007
Dear Mr. Thomas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
TAB2
510(k) Number:
Device Name:
CMRtools™ and plug-ins VentricularTools™ & ThalassaemiaTools™
Indications for Use:
CMRtools™ and plug-ins VentricularTools™ & ThalassaemiaTools™ is a software device used for viewing and analyzing cardiovascular magnetic resonance images. It allows high performance image manipulation, 3D visualization, and advanced image processing on a standard PC without dedicated hardware. All software plugins have comprehensive quantification facilities with full analysis session recording for providing comprehensive audit trails. It contains an image viewer for importing DICOM images, browsing through patient datasets, viewing cine images and performing region-of-interest analysis.
Plug-ins are used for specialist cardiac assessment including VentricularTools for left ventricular assessment and ThalassaemiaTools for T2* image quantification. It is important that CMRtools is only used to view and analyze human cardiac and or liver images acquired with a cardiovascular magnetic resonance scanner. CMRtools must not be used to view or analyze images acquired with any other type of imaging device.
The software must not be used to view or analyze images of any part of the body except the heart and or liver or the vascular system. Furthermore, the software must not be used to view or analyze images acquired from animals.
Only DICOM images will be used for presentation, display and diagnosis.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 807 Subpart C)
hogu Whha
Division of Reproductive, Abdominal and
Radiological Devices
510(k) Number Ko73194
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.