K113369 · Zimmer, Inc. · JWH · Mar 27, 2012 · Orthopedic
Device Facts
Record ID
K113369
Device Name
ZIMMER PERSONA KNEE SYSTEM
Applicant
Zimmer, Inc.
Product Code
JWH · Orthopedic
Decision Date
Mar 27, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
This device is indicated for patients with severe knee pain and disability due to: - Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis. - Collagen disorders, and/or avascular necrosis of the femoral condyle. - Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy. - Moderate valgus, varus, or flexion deformities. - The salvage of previously failed surgical attempts or for a knee in which satisfactory stability in flexion cannot be obtained at the time of surgery. This device is intended for cemented use only.
Device Story
Zimmer Persona Personalized Knee System is a semi-constrained modular knee prosthesis; designed to resurface articulating surfaces of femoral, tibial, and patellar bones. System utilizes modular design between tibial plates and articular surfaces. Device is intended for cemented use only. Used in clinical settings by orthopedic surgeons to treat severe knee pain and disability. Components include femoral, tibial, and patellar implants. Device restores joint function and stability; benefits patients by alleviating pain and improving mobility. Performance verified through non-clinical bench testing, including fatigue, wear, constraint, and locking mechanism evaluations.
Clinical Evidence
No clinical data was required or provided for this device. Substantial equivalence was established through non-clinical bench testing, including fatigue testing of the tibia keel/stem, cantilever fatigue, wear testing per ISO 14243, spine fatigue, tibiofemoral constraint evaluation, lateral subluxation force testing, contact area/pressure analysis, and locking mechanism strength testing.
Technological Characteristics
Semi-constrained modular knee prosthesis; metal/polymer construction. Components include femoral, tibial, and patellar implants. Modular design for tibial plates and articular surfaces. Intended for cemented fixation. Testing performed per ISO 14243 for wear. No software or electronic components.
Indications for Use
Indicated for patients with severe knee pain and disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral condyle, post-traumatic joint loss, or moderate valgus/varus/flexion deformities. Also indicated for salvage of failed surgical attempts or knee instability in flexion. Intended for cemented use only.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Zimmer. The logo consists of a circle with a stylized letter Z inside. Below the circle, the word "zimmer" is written in a serif font.
K113369
MAR 2 7 2012
P.O. Box 708 Warsaw, IN 46581-0708 574 267-6131
### Summary of Safety and Effectiveness
Sponsor:
Contact Person:
Date:
Trade Name:
Product Codes / Device:
Regulation Numbers / Description:
Predicate Device:
Zimmer, Inc. P.O. Box 708 Warsaw, IN 46581-0708
Kelli Anderson Project Manager, Regulatory Affairs Telephone: (574) 371-8087 Fax: (574) 372-4605
March 16, 2012
: Zimmer® Persona™ Personalized Knee System
JWH
21 CFR § 888.3560 - Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented
Natural Knee II System, manufactured by Zimmer, Inc. (K936159, cleared May 22, 1995)
Gender Solutions™ Natural Knee® Flex System, manufactured by Zimmer, Inc. (K070214, cleared March 16, 2007 and K073286, cleared March 7, 2008)
NextGen Complete Knee System, manufactured by Zimmer, Inc. (K933785, cleared January 30, 1995)
. NexGen Complete Knee Solution Legacy Posterior Stabilized (PS) and Constrained Condylar Knee (CCK), manufactured by Zimmer, Inc. (K960279, cleared April 26, 1996)
NexGen® Complete Knee Solution Legacy® Posterior Stabilized (LPS); LPS-Flex Fixed Bearing Femoral and Articular Surface Components, manufactured by Zimmer, Inc. (K991581, cleared July 30, 1999)
ge 1 of
31
{1}------------------------------------------------
# K113369
NexGen® Complete Knee Solution Cruciate Retaining Flex Femoral (CR Flex) Components, manufactured by Zimmer, Inc. (K023211, cleared October 17, 2002)
NexGen® Knee Gender Solutions Female (GSF) Femoral Components, manufactured by Zimmer, Inc. (K060370, cleared April 28, 2006)
The Zimmer® Persona™ Personalized Knee System is a semi-constrained modular knee prosthesis designed to resurface the articulating surface of the femoral, tibial and patellar bones. The Persona Knee System utilizes a modular design between the tibial plates and articular surfaces.
This device is indicated for patients with severe knee pain and disability due to:
- Rheumatoid arthritis, osteoarthritis, traumatic । arthritis, polyarthritis.
- Collagen disorders, and/or avascular necrosis of । the femoral condyle.
- Post-traumatic loss of joint configuration, । particularly when there is patellofemoral erosion, dysfunction or prior patellectomy.
- Moderate valgus, varus, or flexion deformities. ।
- . I The salvage of previously failed surgical attempts or for a knee in which satisfactory stability in flexion cannot be obtained at the time of surgery.
This device is intended for cemented use only.
The proposed Zimmer® Persona™ Personalized Knee System components are similar or identical in intended use, materials, sterility, and performance characteristics to the predicate devices.
Clinical Performance and Conclusions:
Clinical data and conclusions were not needed for this device.
Non-Clinical Performance and Conclusions:
Device Description:
Intended Use:
Comparison to Predicate Device:
Performance Data (Nonclinical and/or Clinical):
ge 2 of 4
{2}------------------------------------------------
Page 4
March 16, 2012
# K113369
| Property or<br>Characteristic | Test Results |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Fatigue Test of the Persona Tibia<br>Keel and Stem Extension | Demonstrated that the Persona tibia keel and stem extension<br>taper junction provide sufficient fatigue strength to survive<br>expected worst case loading conditions. |
| Cantilever Fatigue Test of the<br>Persona Cemented Tibia | Demonstrated adequate fatigue strength in the cantilever<br>loading condition. |
| Wear Testing of Persona CR<br>Conventional Articular Surfaces<br>Under Load and Motion Curves<br>From the ISO 14243 Standard | Demonstrated that the wear characteristics of the Persona<br>CR articular surfaces, when articulated against the Persona<br>CR femoral component, are sufficient to survive expected in<br>vivo loading conditions. |
| Wear Testing of Persona PS<br>Conventional UHMWPE Articular<br>Surfaces Under Load and Motion<br>Curves from the ISO 14243<br>Standard | Demonstrated that the wear characteristics of the Persona PS<br>articular surfaces, when articulated against the Persona PS<br>femoral component, are sufficient to survive expected in vivo<br>loading conditions. |
| Wear Testing of Persona UC<br>Conventional UHMWPE Articular<br>Surfaces Under Displacement<br>Control with Load and Motion<br>Curves from the ISO 14243-1<br>Standard | Demonstrated that the wear characteristics of the Persona<br>UC articular surfaces, when articulated against the Persona<br>CR femoral component, are sufficient to survive expected in<br>vivo loading conditions. |
| Spine Fatigue Evaluation of the<br>Persona PS Conventional<br>UHMWPE Articular Surfaces | Demonstrated that the spine of the Persona PS articular<br>surfaces has sufficient strength to survive expected in vivo<br>stress/strain loading conditions. |
| Tibiofemoral Constraint Evaluation<br>of the Persona Conventional<br>UHMWPE Articular Surface | Demonstrated that constraint values for the Persona articular<br>surfaces are comparable to data from similar NexGen<br>articular surfaces. Therefore, the Persona articular surfaces<br>provide adequate constraint through the needed tibiofemoral<br>flexion angles. |
| Property or<br>Characteristic | Test Results |
| Lateral Constraint Evaluation of<br>the Persona Conventional All-Poly<br>Patellar Component on the Persona<br>CR and PS Femoral Components | Demonstrated that the lateral subluxation force of the<br>Persona conventional all-poly patellar component on the<br>Persona CR and PS femoral implants at tibiofemoral flexion<br>angles 0° to 90° was comparable to control testing on<br>NexGen predicate devcies. |
| Tibiofemoral Contact Area and<br>Contact Pressure Evaluation of the<br>Persona CR/UC/PS Conventional<br>Articular Surfaces | Demonstrated that the contact area and contact pressure of<br>the Persona articular surfaces are comparable to data from<br>previous testing on similar NexGen articular surfaces. |
| Contact Area and Contact Stress<br>Evaluation of the Persona<br>Conventional All-Poly Patellar<br>Component on the Persona<br>Primary CR and PS Femoral<br>Components | Demonstrated that, for all flexion angles, the contact areas<br>were similar between the CR and PS femoral components. |
| Anterior Liftoff Testing of the<br>Persona Articular Surfaces | Demonstrated sufficient locking mechanism strength to<br>survive potential worst case anterior liftoff loading<br>conditions during deep flexion. |
| Posterior Liftoff Fatigue Strength of<br>the Persona Articular Surfaces | Demonstrated sufficient locking mechanism strength to<br>survive potential worst case shear loading conditions. |
| Assembly Testing of the Persona<br>Tibia Locking Mechanism | Demonstrated successful assembly of the modular articular<br>surfaces at normal and maximum interference conditions. |
| Static Shear Strength of the Persona<br>Tibia Locking Mechanism | Demonstrated adequate resistance of the modular articular<br>surfaces to disassembly. |
3 of
C
To
33
{3}------------------------------------------------
Page 5
March 16, 2012
# K113369
.
. . . . .
·
・ :
{4}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Zimmer, Inc. % Ms. Kelli J. Anderson Project Manager, Regulatory Affairs P.O. Box 708 Warsaw, Indiana 46581-0708
MAR 2 7 2012
Re: K113369
Trade/Device Name: Zimmer™ Persona® Personalized Knee System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis
Regulatory Class: II Product Codes: JWH Dated: March 16, 2012 Received: March 19, 2012
Dear Ms. Anderson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{5}------------------------------------------------
## Page 2 - Ms. Kelli J. Anderson
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours
P
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and . Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
K113369
# 510(k) Number (if known): Jaknown
Device Name:
Persona™ Knee System
# Indications for Use:
This device is indicated for patients with severe knee pain and disability due to:
- Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis. ー
- Collagen disorders, and/or avascular necrosis of the femoral condyle. -
- Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectorny.
- Moderate valgus, varus, or flexion deformities. -
- The salvage of previously failed surgical attempts or for a knee in which satisfactory stability in flexion cannot be obtained at the time of surgery.
This device is intended for cemented use only.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter. Use (21 CFR 807 Subpart C)
(Please do not write below this line - Continue on another page if needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Dullel
(Dision Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K113364
Page 1 of 1
ા ર
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.