NEXGEN COMPLETE KNEE SOLUTION LEGACY POSTERIOR STABILIZED (LPS); LPS-FLEX FIXED BEARING FEMORAL AND ARTICULAR SURFACE CO
Applicant
Zimmer, Inc.
Product Code
JWH · Orthopedic
Decision Date
Jul 30, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
This device is intended to reduce or relieve pain and restore function and motion to the knee joint. Total knee replacement is indicated for patients with severe knee pain and disability due to rheumatoid arthritis, osteoarthritis, primary and secondary traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral condyle or pseudogout, posttraumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy, moderate valgus, varus, or flexion deformities. This device may also be indicated in the salvage of previously failed surgical attempts if the knee can be satisfactorily balanced and stabilized at the time of surgery. The LPS-Flex Fixed Bearing Knee provides increased flexion capability for patients who have both the flexibility and desire to increase their flexion range. The LPS-Flex femoral, when used with LPS-Flex articular surfaces, is designed for use with both cruciate ligaments excised and when load bearing range of motion (ROM) is expected to be less than or equal to 155 degrees. This device is intended for cemented use only.
Device Story
Semiconstrained, condylar total knee prosthesis; replaces knee joint surfaces to relieve pain and restore function. System comprises LPS-Flex femoral component and articular surface; requires excision of cruciate ligaments. Designed for increased flexion (up to 155 degrees). Femoral component features modified sagittal plane and cam geometry to improve stability, reduce subluxation, and minimize bending load on polyethylene spine during high-flexion activities (e.g., squatting). Implanted by orthopedic surgeons in clinical/hospital settings. Provides mechanical stability and kinematics for patients with severe knee pathology. Cemented fixation only; secondary screw fixation used on specific articular surfaces to prevent anterior lift-off.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by non-clinical bench testing, including constraint analysis per ASTM F1223 and contact area/conformity assessment.
Indicated for patients with severe knee pain and disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis, or pseudogout; posttraumatic joint loss; patellofemoral erosion/dysfunction; prior patellectomy; or moderate valgus, varus, or flexion deformities. Also indicated for salvage of failed surgical attempts if knee can be balanced/stabilized. Designed for patients desiring increased flexion range (up to 155 degrees) with cruciate ligaments excised. Cemented use only.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
Predicate Devices
NexGen® Complete Knee Solution Legacy® LPS Knee
Submission Summary (Full Text)
{0}------------------------------------------------
K99,581
Image /page/0/Picture/1 description: The image shows the logo for Zimmer. The logo consists of a large, bold letter "Z" enclosed in a circle. Below the circle, the word "zimmer" is written in lowercase letters. The number 30 is in the upper left corner of the image.
. (1))
P.O. Box 708 Warsaw, IN 46581-0708 219 267-6131
## 510(k) Summary of Safety and Effectiveness NexGen® Complete Knee Solution Legacy® Posterior Stabilized (LPS); LPS-Flex Fixed Bearing Femoral and Articular Surface Components
- Submitted by: I.
Zimmer, Inc. P.O. Box 708 Warsaw, Indiana 46581-0708
- II. Contact Person:
Stephen H. McKelvey Senior Regulatory Affairs Associate Regulatory Affairs Telephone: (219) 372-4944 Facsimile: (219) 372-4605
- Ш. Date Prepared:
May 6, 1999
- IV. Name of Device:
| A. | Trade Name: | Femoral and Articular Surface Knee Components |
|----|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| B. | Proprietary Name: | NexGen® Complete Knee Solution Legacy® Posterior<br>Stabilized (LPS); LPS-Flex Fixed Bearing Femoral and<br>Articular Surface Components |
| C. | Common Name: | LPS-Flex Fixed Bearing Knee |
| D. | Classification Name<br>and Reference: | Knee joint patellofemorotibial<br>polyethylene/metal/polyethylene semiconstrained cemented<br>total knee prosthesis - 21 CFR 888.3560 |
Image /page/0/Picture/12 description: The image shows a geometric design. The design is composed of six identical triangular shapes arranged around a central point. Each triangle is further divided into three smaller triangles, creating a star-like pattern within each larger triangle. The overall design has a symmetrical and radial appearance.
{1}------------------------------------------------
#### E. Predicate Device:
NexGen® Complete Knee Solution Legacy® LPS Knee
- F. Device Description:
The LPS-Flex Fixed Bearing Knee is a semiconstrained, condylar system for use without the cruciate ligaments when additional stability is required to prevent anterior subluxation of the femur relative to the tibia in flexion. The LPS-Flex Fixed Bearing Knee consists of an LPS-Flex femoral and an LPS-Flex articular surface.
- G. Intended Use:
This device is intended to reduce or relieve pain and restore function and motion to the knee joint. Total knee replacement is indicated for patients with severe knee pain and disability due to rheumatoid arthritis, osteoarthritis, primary and secondary traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral condyle or pseudogout, posttraumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy, moderate valgus, varus, or flexion deformities. This device may also be indicated in the salvage of previously failed surgical attempts if the knee can be satisfactorily balanced and stabilized at the time of surgery.
The LPS-Flex Fixed Bearing Knee provides increased flexion capability for patients who have both the flexibility and desire to increase their flexion range. This device is intended for cemented use only.
- H. Comparison to Predicate Device:
See attached table.
- I. Non-clinical Performance and Conclusions:
- 1. The level of constraint for the proposed device was analyzed per ASTM F1223. The femoral-to-articular surface motion is resisted by the articulating surface geometry, the polyethylene spine and the close conformity between the two components. The simulation exhibited negligible displacement variances between the LPS-Flex and the predicate device. Constraint was not significantly different.
- 2. The contact area was also determined and showed an increase in conformity for the device as compared to the predicate.
{2}------------------------------------------------
- 3. The tibial baseplate interlock mechanism for the device is the same as the predicate device. A screw is added to some LPS-Flex articular surfaces to enhance fixation during high flexion.
- 4. No additional fatigue strength data was needed because the device uses the predicate tibial baseplate.
- J. Clinical Performance and Conclusions:
Clinical data and conclusions were not needed for this device.
{3}------------------------------------------------
# Technological Comparison to Predicate Device
| Parameter | Identical, Similar<br>or Different? | Similarities and Differences |
|------------------|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Design | Similar | Frontal plane conformity between the femoral and articular surface components is identical. The LPS-Flex femoral has a modified sagittal plane as compared to the predicate. This gives the LPS-Flex increased conformity during walking gait angles of flexion and accommodates the extended flexion range. Compatibility between predicate and LPS-Flex components is high (predicate femorals can be used with the LPS-Flex articular surface). The patellar groove design is identical for the predicate and LPS-Flex device. The same patella is used for both devices. One additional posterior femoral bone cut is required to accommodate the extended flexion range of the LPS- Flex, otherwise the box and patellar groove geometries are the same as the predicate and standard NexGen instrumentation may be used. To improve stability of the femoral component on the articular surface and to reduce the moment applied to the polyethylene spine, the shape of the femoral cam was modified and the clearance posterior to the polyethylene spine was deepened. This improves resistance to subluxation beyond the predicate and provides for the correct rollback and kinematics of the LPS-Flex joint. The cam and spine geometries of the LPS-Flex were also designed to provide a low cam/spine contact which reduces the bending load on the spine during high load flexion activities, such as squatting. |
| Materials | Similar | Femoral component is Zimaloy® Cobalt-Chromium-Molybdenum Alloy for both the predicate and LPS-Flex. Articular surface is UHMWPE for both the predicate and LPS-Flex. A Tivanium insert is added to the LPS- Flex. Surface characteristics are identical for both the predicate and LPS-Flex (Uncoated, PMMA coated and Co- nidium hardened). |
| Performance | Similar | Interlock mechanism - 17 and 20 mm thick LPS-Flex articular surfaces require a secondary screw fixation, in addition to the dovetail locking system already present in the predicate and LPS-Flex. This screw provides an additional safety factor to prevent anterior lift-off of the articular surface from the tibial baseplate. Tibial Baseplate Fatigue Strength - The LPS-Flex knee uses the predicate tibial baseplate and has the same fatigue strength. Lateral Stability of the Patellofemoral Joint - The patellar groove design and stability of the LPS-Flex femoral component is the same as the predicate. Both are used with all-polyethylene patellar components. Wear Data -The LPS-Flex has a minimum polyethylene thickness of 6.5 mm. Even though the LPS-Flex increases the maximum active flexion angle to 155 degrees, the design has maintained the conformity necessary to minimize or eliminate any new movement mechanisms that could affect wear. |
| Parameter | Identical, Similar<br>or Different? | Similarities and Differences |
| Sterility | Identical | • There are no differences between the predicate and LPS-Flex sterility processes. Both are terminally sterilized<br>by gamma radiation. Gamma radiation processing and dose mapping are conducted according to<br>ANSI/AAMI/ISO 11137-1994. The products are accepted for release as sterile though a validated dosimetric<br>release program designed to provide a sterility assurance level (SAL) of 10° or better (ANSI/AAMI/ISO<br>11137-1994, ANSI/AAMI ST32-1991 and ISO/TR 13409-1996). |
| Biocompatibility | Identical | • Material specifications for both the predicate and LPS-Flex meet or exceed ASTM standards, are common to<br>orthopaedic products today and do not require additional biocompatibility testing. |
| Pyrogenicity | Identical | • Neither the predicate or the LPS-Flex are labeled as nonpyrogenic. Per USP XXIII, NF18 (1995 edition),<br>page 1719, "These requirements do not apply to orthopaedic products." |
{4}------------------------------------------------
.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of a human figure in profile, with three faces overlapping to create a sense of depth and dimension. The figure is positioned to the right of a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in all capital letters.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Jul 30 1999
Mr. Steven H. McKelvey Senior Regulatory Affairs Associate Zimmer Inc. P.O. Box 708 Warsaw, Indiana 46581-0708
K991581 Re:
> Trade Name: NexGen Complete Knee Solution Legacy Posterior Stabilized (LPS)- Flex Fixed Bearing Femoral and Articular Surface Components Regulatory Class: II Product Code: JWH Dated: May 6, 1999 Received: May 7, 1999
Dear Mr. McKelvey:
We have reviewed your Section 510(k) notification of intent to market the device referenced above, and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general control provisions of the Act. The general control provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General Regulation (21 CFR Part 820), and that, through periodic (OS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{6}------------------------------------------------
Page 2 -- Mr. Stephen H. McKelvey
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
Page of
510(k) Number (if known): 991581 ん
Device Name:
NexGen® Complete Knee Solution Legacy® Posterior Stabilized (LPS); LPS-Flex Fixed Bearing Femoral and Articular Surface Components
### Indications for Use:
This device is intended to reduce or relieve pain and restore function and motion to the knee joint. Total knee replacement is indicated for patients with severe knee pain and disability due to rheumatoid arthritis, osteoarthritis, primary and secondary traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral condyle or pseudogout, posttraumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy, moderate valgus, varus, or flexion deformities. This device may also be indicated in the salvage of previously failed surgical attempts if the knee can be satisfactorily balanced and stabilized at the time of surgery.
The LPS-Flex Fixed Bearing Knee provides increased flexion capability for patients who have both the flexibility and desire to increase their flexion range. The LPS-Flex femoral, when used with LPS-Flex articular surfaces, is designed for use with both cruciate ligaments excised and when load bearing range of motion (ROM) is expected to be less than or equal to 155 degrees. This device is intended for cemented use only.
(Please do not write below this line - Continue on another page if needed)
Concurrence of CDRH, Office of Device Eyaluation (ODE)
(Division Sign-Off)
Division of General Restorative Devices
510(k) Number K991581
Prescription Use
(Per 21 CFR 801.109)
OR Over The-Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.