ES Complex Software is an optional software accessory for use with the following models with data management capabilities: a) Contec 08A, blood pressure device b) ESO, oximeter c) EIS-GS, galvanic skin response device and d) ES-BC. Analyzer body composition When use Contec 08A and ESO and/or EIS-GS and/or ES-BC, the ES Complex software uploads the data of the devices and displays the data into a computer for enhanced data management. The ES Complex software is intended for use in clinical settings as an aid for health care professionals to review, analyze and evaluate the historical tests results.
Device Story
ES Complex Software is a PC-based software accessory for data management. It interfaces with four specific medical devices: Contec 08A (blood pressure), ESO (oximeter), EIS-GS (galvanic skin response), and ES-BC (body composition analyzer). The software uploads patient data from these devices—or from local backup files—and displays it on a workstation for clinical review. It provides historical tracking and tabular presentation of physiological parameters, including SpO2, heart rate, HRV metrics, skin conductance, body composition (fat, water, BMI), and blood pressure. Used in clinical settings by healthcare professionals, the software facilitates the review and evaluation of historical test results to support clinical decision-making. It does not perform automated data analysis; it serves as a centralized repository and display tool for data generated by the connected hardware.
Clinical Evidence
Bench testing only. No clinical data provided. Evidence consists of software verification and validation (D-11SRS/SDS/D-12 STD/STR D-33 and 34) and risk management documentation (D-28) to ensure the software correctly uploads and displays data from the connected medical devices.
Technological Characteristics
PC-based software running on Windows 7; Intel architecture. Developed in C++ using Microsoft Visual C++ compiler. Requires 2GB free space and 27MB program size. Integrates with off-the-shelf software (MedCalc, MS Office Word 2007/2010). Connectivity via data upload from device memory or local PC backup files. Complies with ISO 14971 (risk management) and ISO 62304 (software life cycle).
Indications for Use
Indicated for use in clinical settings by healthcare professionals to review, analyze, and evaluate historical test results from connected blood pressure, oximetry, galvanic skin response, and body composition devices.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{0}------------------------------------------------
K113264 pg lof5
MAR - 2 2012
#### L.D Technology LLC.
510(k) Premarket Notification Number: K113264 Preparation date: September, 21 2011
#### 510(k) Summary Electro Sensors Complex Software (ES Complex Software)
Name of the device: Electro Sensors Complex Software Common name: ES Complex Software
#### This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR 8807.92
#### 1. Submitter's Identification:
Manufacturer: L.D TECHNOLOGY Quality Control and designer: MAAREK Albert Address: L.D Technology 100 N.Biscayne Blvd, Suite 500 Miami, FL, 33132, USA Tel: 305-379-9900 E mail: albert.ldteck@gmail.com
#### 2. Device Name / Classification
Electro Sensors Complex Software (ES Complex Software) Class II System components: Regulation number: 21CFR 8701130: Non Invasive Blood pressure measurement system, Class II 21 CFR 870 2700: Oximeter, Class II 21 CFR 882 1540: Galvanic Skin responses device Class II 21 CFR 870 2770: Analyzer Body Composition Class II 21 CFR 862 2100: Calculator/data processing module for clinical use, Class I Product Code: NXD, DQA, GZO, MNW Classification: Class II Classification Panel: Cardiology/Neurology
#### 3. Predicate legally marketed devices
- Health care system software K110948. Applicant: TaiDoc Technology . Corporation. Product codes NBW, DXN, FLL
{1}------------------------------------------------
K113264 pg. 2 of 5
#### 4. Device Description
ES Complex Software is an optional software accessory for use with the following models with management capabilities: a) Contec 08A, blood pressure device b) ESO, oximeter c) EIS-GS, galvanic skin response device and d) ES-BC. Analyzer body composition
When use combination Contec 08A and ESO, and/or EIS-GS and/or ES-BC, the ES Complex software uploads the data of the 4 devices and displays the data into a computer for enhanced data management.
#### Description of the features
- Upload data:
- V ESO data: Patient stats, calculation and Historical Tracking of: SpO2 %, Wave value, Signal Strength, b/a, c/a, d/a, e/a, oximeter wave form in coding digits. Heart Rate (HR), Mean values of RR intervals, Maximum values (Mx), Minimum values (Mn), MxDMn HIB, SDNN, RMSSD, NN50 count, pNN50 %, VLF, LF ,HF and LF/HF ratio.
- > EIS-GS data: Patient stats, measurement and Historical Tracking of :the conductance values of 22 skin pathways
- V ES-BC data: Patient stats, calculation and Historical Tracking of: Actual Impedance, Actual Phase Angle (PA), Estimated Body Fat (FAT), Estimated Fat Free Mass (FFM), Estimated Total Body Water (TBW), Estimated Intra-Cellular Water (ICW), Estimated Extra-Cellular Water (ECW), Estimated Basal Metabolic Rate (BMR), Estimated Daily Energy Expenditure (DEE) and actual Body Mass Index (BMI)
- > Contec 08A data: Patient stats, measurement and Historical Tracking of: Systolic and diastolic pressure
- Data management .
#### 5. Intended use and indications for use
ES Complex Software is an optional software accessory for use with the following models with data management capabilities: a) Contec 08A, blood pressure device b) ESO, oximeter c) EIS-GS, galvanic skin response device and d) ES-BC. Analyzer body composition When use Contec 08A and ESO and/or EIS-GS and/or ES-BC, the ES Complex software uploads the data of the devices and displays the data into a computer for enhanced data management. The ES Complex software is intended for use in clinical settings as an aid for health care professionals to review, analyze and evaluate the historical tests results.
{2}------------------------------------------------
K 113264 pg 3 of 5
#### 6. Specifications and performances
#### Software specifications
- Hardware platform : Laptop or PC based workstation (Intel architecture) .
- Operating system : Windows 7 .
- Use of Off-the-Shelf software : Windows 7, MedCalc and Microsoft office Word (v, 2007 . or 2010)
- Language : C++ .
- . Microsoft Visual C++ compiler requirements: 2 Gb free space
- program size requirements 27Mb .
#### Software performances
Image /page/2/Figure/10 description: The image shows a flowchart of the ES Complex software. The flowchart starts with Contec 08A, ESO, EIS-GS, and ES-BC, which all feed into "Back up in DAT.files". The next step is "ES Complex software", which feeds into "Data Management: To Display results in table and historical tracking", and the final step is "Back up ES Complex Software".
#### 7. Substantial equivalence
#### Predicate legally marketed devices
- Health care system software K110948. Applicant: TaiDoc Technology . Corporation. Product codes NBW, DXN, FLL
The substantial equivalence with the Heath care system software K110948 is based upon: The same intended use and
Different uploaded data, and the information submitted;
- a. Does not raise new questions of safety and effectiveness; and
- b. Demonstrates that the device is at least as safe and effective as the legally marketed device.
{3}------------------------------------------------
# KII3264
#### 8. Table of comparison
#### Table of comparison Health care system software K110948 / ES Complex Software
| Name device (510k<br>number) | Health care system software K110948 | ES Complex Software : uploaded data |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use | Health care system software is an optional<br>software accessory for use with the following<br>models with data management capabilities: a)<br>Clever Check blood glucose meter b) Clever<br>Check blood glucose and blood pressure<br>monitors and c) Clever blood pressure<br>monitors. When use with one of these device,<br>Health care system software transfers data<br>from the device's memory into a computer for<br>enhanced data management. The Health Care<br>System Software is intended for use in home<br>and clinical settings as an aid for users and<br>their health care professionals to review,<br>analyze and evaluate the historical test results<br>to support health management effectively | ES Complex Software is an optional software<br>accessory for use with the following models with<br>data management capabilities: a) Contec 08A,<br>blood pressure device b) ESO, oximeter c) EIS-GS,<br>galvanic skin response device and d) ES-BC.<br>Analyzer body composition.<br>When use Contec 08A and ESO and/or EIS-GS<br>and/or ES-BC, ES Complex software uploads the<br>data of the devices and displays the data into a<br>computer for enhanced data management.<br>The ES Complex software is intended for use in<br>clinical settings as an aid for health care<br>professionals to review, analyze and evaluate the<br>historical tests results. |
| Results/Performances | | |
| Data management | YES | YES |
| Historical test results | YES | YES |
| Operating system | Windows | Windows |
| Results screen | PC | PC |
| | Blood glucose ,Blood pressure and body<br>temperature | SpO2%, HRV data analysis, photoelectrical<br>Plethysmography analysis and waveform,<br>conductance values, body composition and Blood<br>pressure. |
| Displayed data | | |
| Data acquisition | Upload from the device memory | upload from backup folder located in the C/Drive<br>of the PC or upload from the device memory |
| Data analysis | NO | NO |
| Indication for use | At home and clinical settings | Clinical settings |
### 9. Performances and Effectiveness
- 1. Software verification & validation (D-11SRS/SDS/ D-12 STD/ STR D-33 and 34)
- 2. Risk management (D-28)
{4}------------------------------------------------
K113264 pg 5 of 5
#### 10. General Safety Concerns
The data come from medical devices with 510k clearance:
- Contec 08A software K110774.Applicant : Contec Medical Systems CO.LDT Product code: . DXN
- Electro Sensor Oxi (ESO) software K102442.Applicant: LD Technology LLC Product Code . DOA
- . EIS-GS Software K102166. Applicant: LD Technology LLC Product Code GZO
- ES-BC Software K103026. Applicant: LD Technology LLC Product Code MNW .
#### 11. Standards
- Guidance for: Industry; FDA staff; and Content of Premarket Submissions for Software . Contained in Medical Devices May 11, 2005
- ISO 14971: Medical devices Application of risk management to medical devices. March . 01 2007
- . ISO 62304: Medical device software -- Software life cycle processes. May 1 2006
#### 12. Conclusions
Electro Sensors Complex Software (ES Complex Software) is equivalent in performances, technology, safety and efficacy to the legally marketed predicate devices.
#### Signature:
**Albert MAAREK**
Premarket notification |510K| Number: K113264
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a representation of the human form.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
MAR - 2 2012
LD Technology, LLC c/o Mr. Albert Maarek 100 N. Biscayne Blvd. Suite 500 Miami, FL 33132
Re: K113264 Trade/Device Name: Electro Sensor Complex Software Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive blood pressure measurement system Regulatory Class: Class II Product Code: DXN, DQA, GZO Dated: February 29, 2012 Received: March 1, 2012
Dear Mr. Maarek:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Mr. Albert Maarek
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21. CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
M. A. Willemsen
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## Indications for Use
510(k) Number: k113264
Device Name: Electro Sensors Complex Software (ES Complex Software) Indications for Use:
ES Complex Software is an optional software accessory for use with the following models with data management capabilities: a) Contec 08A, blood pressure device b) ESO, oximeter c) EIS-GS, galvanic skin response device and d) ES-BC. Analyzer body composition
When use Contec 08A and ESO and/or EIS-GS and/or ES-BC, the ES Complex software uploads the data of the devices and displays the data into a computer for enhanced data management.
The ES Complex software is intended for use in clinical settings as an aid for health care professionals to review, analyze and evaluate the historical tests results.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
M.A. Killebrew
(Division Sign-Off) Division of Cardiovascular Devices
510(k) Number_ K-113264
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.