Truly Automatic Arm Blood Pressure Monitor DB series, Models DB33, DB81, DB82, DB83, DB91, DB91, DB93 are a series devices intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The devices features include irregular pulse rhythm detection during measurement, and display a warning signal with the reading once the irregular heartbeat is detected.
Device Story
Device measures systolic/diastolic blood pressure and pulse rate using oscillometric method; utilizes inflatable upper arm cuff and electronic pressure sensor. Sensor converts cuff pressure changes to electrical signals; microprocessor analyzes signals to calculate blood pressure and pulse. Includes irregular pulse rhythm detection feature, which compares longest/shortest pulse wave intervals to mean interval; displays warning if difference exceeds 25%. Designed for home use by adult patients. Powered by 4 AA batteries; features LCD display, pump, and electromagnetic deflation valve. Output provides blood pressure readings and irregular heartbeat alerts to assist users in monitoring cardiovascular health.
Clinical Evidence
Clinical study included 90 subjects, including 15 hypertensive patients. Compared test device against mercury sphygmomanometer reference standard. Results: Mean difference for systolic was 3.3 mmHg (SD 5.2) and diastolic 4.5 mmHg (SD 4.6), meeting ANSI/AAMI SP10 criteria (mean difference ≤5 mmHg, SD ≤8 mmHg). Between-observer agreement for systolic/diastolic readings within 10 mmHg was 100%; within 5 mmHg was 95% and 99% respectively.
Technological Characteristics
Oscillometric measurement; electronic pressure sensor (MSP-2107). ABS plastic housing; nylon cuff (220-340 mm circumference). Powered by 4x 1.5V AA batteries. Connectivity: None. Standards: ISO-10993 (biocompatibility), IEC60601-1/1-2 (electrical safety).
Indications for Use
Indicated for adult individuals to measure systolic and diastolic blood pressure and pulse rate via non-invasive upper arm cuff. Includes irregular pulse rhythm detection. Intended for Over-The-Counter use.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
Truly Automatic Arm Blood Pressure Monitor, Model DB61M (K091434)
Submission Summary (Full Text)
{0}------------------------------------------------
K113083
Image /page/0/Picture/1 description: The image shows the word "TRULY" in bold, black letters. There is a line underneath the word. A circled "R" is to the right of the word.
Truly Instrument Limited
# 510(k) Summary
JAN 1 9 2012
Date of Summary Preparation: August. 16.2011
## 1. Submitter's Identifications
| Submitter's Name: | Truly Instrument Limited |
|-------------------|---------------------------------------------------------------------------|
| Address: | Site 2, Truly Industrial Area, Shanwei City,<br>Guangdong Province, China |
| Contact Person: | Manager Yang Jian-Hao |
| Telephone: | 86-0660-3380070 |
| Fax: | 86-0660-3380377 |
## 2. Name of the Device
| Device Classification Name: | System, Measurement, Blood-Pressure, Non-invasive |
|-----------------------------|---------------------------------------------------|
| Trade Nam: | Truly Automatic Arm Blood Pressure Monitor |
| Models: | DB33, DB81, DB82, DB83, DB85, DB91, DB92, DB93 |
| Classification Panel: | cardio-vascular |
| Common/Usual Name: | Automatic Arm Blood Pressure Monitor |
| Product Code: | DXN |
| Device Classification: | Class II |
| Contraindications : | N/A |
### 3. The Predicate Devices
Truly Automatic Arm Blood Pressure Monitor, Model DB61M, K091434
### 4. Device Description
Truly Automatic Arm Blood Pressure Monitor DB series, Models DB33, DB81, DB83, DB85, DB91, DB92, DB93 are designed to measure the systolic and diastolic blood pressure and pulse rate of an individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. Our method to define systolic and diastolic pressure is similar to the auscultatory method but uses an electronic pressure sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alterations in cuff pressure to electrical signals, by analyzing those signals to define the systolic and diastolic blood pressure and calculating pulse rate, which is a well-known technique in the market called the "oscillometric method".
The main components of the Truly Automatic Arm Blood Pressure Monitor DB series are the main unit and cuff unit. ABS is used to outer housing of the main unit. The preformed cuff unit, which is applicable to arm circumference approximately between 220 and 340 mm, includes the inflatable bladder and nylon shell. All models of the arm blood pressure monitor use a single size of cuff. The device consists of the microprocessor, the pressure sensor, the operation keys, the pump, the electromagnetic deflation control valve and the LCD. The subject devices are powered by four AA alkaline batteries.
The device also compares the longest and the shortest time intervals of detected pulse waves to mean time interval and displays a warning signal with the reading to indicate the detection Page 1 of 5
{1}------------------------------------------------
K||3083
#### TRULY Truly Instrument Limited
of irregular pulse rhythm when the difference of the time intervals is over 25%.
#### 5. Intended use of device
Truly Automatic Arm Blood Pressure Monitor DB series, Models DB33, DB81, DB82, DB83. DB91, DB91, DB93 are a series devices intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm.
The devices features include irregular pulse rhythm detection during measurement, and display a warning signal with the reading once the irregular heartbeat is detected.
## 6. Summary of Substantial Equivalence
| Parameter | Predicate Devices DB61M | DB<br>33 | DB<br>81 | DB<br>82 | DB<br>83 | DB<br>85 | DB<br>91 | DB<br>92 | DB<br>93 | Result |
|---------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|----------|----------|----------|----------|----------|----------|----------|--------|
| Intended use | Measuring systolic and diastolic blood pressure and pulse rate of adult individual | Measuring systolic and diastolic blood pressure and pulse rate of adult individual | | | | | | | | Same |
| Indications<br>for use | Measuring systolic and diastolic blood pressure and pulse rate of adult individual, Including irregular pulse rhythm detection. Over-The-Counter Use | Measuring systolic and diastolic blood pressure and pulse rate of adult individual, Including irregular pulse rhythm detection. Over-The-Counter Use | | | | | | | | Same |
| Target<br>Population | Adult | Adult | | | | | | | | Same |
| Anatomical<br>sites | Upper Arm | Upper Arm | | | | | | | | Same |
| Where used<br>(hospital,<br>home,<br>ambulance.<br>etc) | Home | Home | | | | | | | | Same |
| Energy used<br>and / or<br>delivered | 4x 1.5V AA Battery | 4x 1.5V AA Battery | | | | | | | | Same |
| Human<br>factors | Blood pressure | Blood pressure | | | | | | | | Same |
| design | Refer to Table-2 | | | | | | | | | Same |
| performance | Measuring systolic and diastolic blood pressure and pulse rate of adult individual, Including irregular pulse rhythm detection | Measuring systolic and diastolic blood pressure and pulse rate of adult individual, Including irregular pulse rhythm detection | | | | | | | | Same |
| materials | Refer to Table-2 | | | | | | | | | Same |
| biocompatibi | Cuff | Cuff | | | | | | | | Same |
Table-1: The comparison table
Page 2 of 5
{2}------------------------------------------------
K//3083
| TRULY® Truly Instrument Limited | | | | | | | | | | |
|------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|----------|----------|----------|----------|----------|----------|----------|--------|
| Parameter | Predicate Devices DB61M | DB<br>33 | DB<br>81 | DB<br>82 | DB<br>83 | DB<br>85 | DB<br>91 | DB<br>92 | DB<br>93 | Result |
| lity | According to ISO-10993 | According to ISO-10993 | | | | | | | | |
| Compatibility<br>with the<br>environment<br>and other<br>devices | Operation Environment:<br>10℃~ 40℃, 15%~90%RH<br>Storage Environment:<br>-20℃~ 60℃, 10%~95%RH | Operation Environment:<br>10℃~ 40℃, 15%~90%RH<br>Storage Environment:<br>-20℃~ 60℃, 10%~95%RH | | | | | | | | Same |
| sterility | N/A | N/A | | | | | | | | Same |
| Electrical<br>safety | According to IEC60601-1-2<br>According to IEC60601-1 | According to IEC60601-1-2<br>According to IEC60601-1 | | | | | | | | Same |
| Mechanical<br>safety | Same | Same | | | | | | | | Same |
| Chemical<br>safety | N/A | N/A | | | | | | | | Same |
| Thermal<br>safety | N/A | N/A | | | | | | | | Same |
## Table-2: The comparison table in Design and Materials
| Parameter | Predicate<br>Devices<br>DB61M | DB<br>33 | DB<br>81 | DB<br>82 | DB<br>83 | DB<br>85 | DB<br>91 | DB<br>92 | DB<br>93 | |
|-------------------------------------|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------|----------|----------|----------|----------|----------|----------|--|
| Measurement<br>algorithm<br>Method | Oscillometric<br>method | No change,all same | | | | | | | | |
| Measurement<br>site of body | Arm | No change,all same | | | | | | | | |
| Pressure<br>Sensor | MSP-2107 | No change,all same | | | | | | | | |
| Cuff | | No change,all same | | | | | | | | |
| Software | | No change,all same | | | | | | | | |
| Irregular<br>heartbeat<br>detection | | More than ±25% to the mean interval of pulse intervals.<br>About the more detailed description of the IH detection algorithm, please<br>refer to "Software validation report I-5. Algorithm description 4.<br>Determination method of irregular heartbeat". | | | | | | | | |
| Memory Size | 4 x 99 | 2X60 | 1X40 | 1X40 | 4X99 | 4X99 | 4X99 | 4X99 | 4X99 | |
| Measurement<br>Pressure<br>Range | 20 ~ 280 mmHg | No change,all same | | | | | | | | |
| Measurement<br>Pulse Range | 40 ~ 195 beats/min | No change,all same | | | | | | | | |
| Mesauring<br>resolution | 1 mmHg | No change,all same | | | | | | | | |
| Accuracy | ±3mmHg | No change,all same | | | | | | | | |
Page 3 of 5
{3}------------------------------------------------
K//3083
| Parameter | Predicate<br>Devices<br>DB61M | DB<br>33 | DB<br>81 | DB<br>82 | DB<br>83 | DB<br>85 | DB<br>91 | DB<br>92 | DB<br>93 | |
|--------------------------|-------------------------------|---------------------------------------------|----------|----------|----------|----------|----------|----------|----------|--|
| Pressure | | | | | | | | | | |
| Accuracy<br>Pulse | $\pm5%$ | No change,all same | | | | | | | | |
| Pressurization Source | Automatic<br>internal pump | No change,all same | | | | | | | | |
| Cuff<br>Deflation | Automatic<br>deflation | No change,all same | | | | | | | | |
| Operating<br>Environment | 10~40℃<br>15~90%RH | No change,all same | | | | | | | | |
| Power<br>Voltage | 4X 1.5V | No change,all same | | | | | | | | |
| Hardware<br>circuit | | No change,all same | | | | | | | | |
| Electronic<br>element | | No change,all same | | | | | | | | |
| PCB | | No change,all same | | | | | | | | |
| Display Type | Liquid crystal<br>display | Liquid crystal display,Only difference size | | | | | | | | |
| Cover | | Difference | | | | | | | | |
## Truly Instrument I imited
## 7. Summary of Clinical study
l ). Subjects:
Ninety subjects including 15 hypertensive patients in the hospital were participated in clinical study.
2). Method:
A standard mercury sphygmomanometer was used as a reference standard. Simultaneous and blinded blood pressure determinations were performed by two doctors.
#### 3). Criteria:
The ANSI/AAMI SP10 Standard recommended :
- a mean difference of x5mmHg, with standard deviation of differences of x8 mmHg A. between test device and reference method.
- Between-observer agreement should be:95% or more of readings made B. simultaneously by observers agree to within x10mmHg and 85% or more agree to within x5 mmHg.
4). Result
Through clinical research, we can convinced that the clinical device is safe and effective. The results of the clinical data refer the follow two tables.
| Table-1 Between test device and reference method | | |
|--------------------------------------------------|-------------------|--------------------|
| Criteria | Test Result | |
| | Systolic Pressure | Diastolic Pressure |
blog 1 Pativan toot daving and reference a
{4}------------------------------------------------
K//3083
Image /page/4/Picture/1 description: The image shows the word "TRULY" in bold, black letters. To the right of the word is a circled "R", indicating that the word is a registered trademark. The word is simple and clear, making it easily readable.
Truly Instrument Limited
| Mean differences | x 5 mmHg | 3.3 mmHg | 4.5 mmHg |
|------------------|----------|----------|----------|
| SD differences | x 8 mmHg | 5.2 mmHg | 4.6 mmHg |
| Table-2 Betwee-observer agreement | | | |
|-----------------------------------|--|--|--|
|-----------------------------------|--|--|--|
| Criteria | Test Result | |
|------------------------------------------------------|-------------------|--------------------|
| | Systolic Pressure | Diastolic Pressure |
| At least 95% of readings agree to<br>within x 10mmHg | 100% | 100% |
| At least 85% of readings agree to<br>within x 5mmHg | 95% | 99% |
## 8. Conclusions
The new subject series devices of Truly Automatic Arm Blood Pressure Monitor continue to follow principles of hardware and software design of the predicate device DB61M(K091434), and the feature, safety, effectiveness are also as same as DB61M., just only in case and LCD''s size changes. Thus, the subject devices are substantially equivalent to the predicate devices.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing the body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Truly Instrument Co., Ltd. c/o: Mr. Jack Lee Truly (USA) INC 2620 Concord Avenue, Suite 106 Alhambra, CA 91803
JAN 1 9 2012
Re: K113083
Truly Automatic Arm Blood Pressure Monitor (DB33, DB81, DB82, DB83, DB91, DB92, DB93) - 8 models Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive blood pressure measurement system Regulatory Class: II Product Code: DXN Dated: January 13, 2012 Received: January 13, 2012
Dear Mr. Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for ass based of to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{6}------------------------------------------------
Page 2 - Mr. Jack Lee
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic forth in the quant) of brovisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 17 you active pov.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Bram D. Zuckerman, M.D.
Director
Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
K113083
# Truly Instrument Limited
# Indications for Use
510(K) Number (if Known)
Device Name: Truly Automatic Arm Blood Pressure Monitor DB Series:
Models DB33, DB81, DB82, DB83, DB85, DB91, DB92, DB93 Indication For Use:
Truly Automatic Arm Blood Pressure Monitor, Models DB33, DB81, DB82, DB83, DB85, DB91, DB92, DB93 are a series device intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm.
The devices features include irregular pulse rhythm detection during measurement, and display a warning signal with the reading once the irregular heartbeat is detected.
AND/OR -Prescription Use _______________ (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) |
|------------------------------------|
| Division of Cardiovascular Devices |
| 510(k) Number | K113083 |
|---------------|---------|
|---------------|---------|
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.