TRULY AUTOMATIC ARM BLOOD PRESSURE MONITORDB21,DB31,DB22,DB23,DB71M,DB32,DB61M, DB62M, DB63M
Applicant
Truly Instrument Co., Ltd.
Product Code
DXN · Cardiovascular
Decision Date
Sep 4, 2009
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 870.1130
Device Class
Class 2
Indications for Use
Truly Automatic Arm Blood Pressure Monitor DB series, Models DB21, DB22, DB23, DB31, DB32, DB61M, DB62M, DB63M, DB71M are a series devices intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The devices features include irregular pulse rhythm detection during measurement, and display a warning signal with the reading once the irregular heartbeat is detected.
Device Story
Automatic arm blood pressure monitor; uses oscillometric method to measure systolic/diastolic blood pressure and pulse rate. Input: pressure signals from inflatable cuff via electronic pressure sensor. Processing: microprocessor analyzes pressure alterations to calculate BP and pulse; compares pulse wave time intervals to detect irregular rhythm (threshold >25% difference). Output: BP/pulse readings and irregular heartbeat warning on LCD. Used in home or clinical settings by adults. Benefits: non-invasive monitoring of BP and detection of irregular heartbeats.
Clinical Evidence
No clinical data provided; substantial equivalence based on bench testing and comparison of technological characteristics to predicate devices.
Technological Characteristics
Oscillometric measurement; electronic pressure sensor; automatic pump and deflation valve; LCD display; ABS housing; nylon cuff (220-340mm circumference); powered by 4x AAA or AA alkaline batteries. Microprocessor-based control. Software includes irregular heartbeat detection algorithm.
Indications for Use
Indicated for adult individuals requiring non-invasive measurement of systolic and diastolic blood pressure and pulse rate via an upper arm cuff. Includes irregular pulse rhythm detection.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
Digibio Digital Blood Pressure Monitor, Model D11 (K014141)
Microlife Blood Pressure Monitor, Model BP3BT0-AP (K041411)
Submission Summary (Full Text)
{0}------------------------------------------------
# K091434
Truly Instrument Limited.
# 510(k) Summarv
SEP - 4 2009
Date of Summary Preparation: 4.20.2009
# 1. Submitter's Identifications
Submitter's Name: Truly Instrument Limited
Truly Industrial Area, Shanwei City, Guangdong Province, Address: China Contact Person: Manager Yang Jian-Hao Telephone: 86-0660-3380070 Fax: 86-0660-3380377
## 2. Name of the Device
Device Classification Name: System, Measurement, Blood-Pressure,
Non-invasive
Models: DB21, DB22, DB23, DB31, DB32, DB61M, DB62M, DB63M, DB71M Classification Panel: cardio-vascular Common/Usual Name: Automatic Arm Blood Pressure Monitor Product Code: DXN Device Classification: Class II Contraindications : N/A
#### 3. The Predicate Devices
a. Digibio Digital Blood Pressure Monitor, Model D11, K014141 b. Microlife Blood Pressure Monitor, Model BP3BT0-AP, K041411
## 4. Device Description
Truly Automatic Arm Blood Pressure Monitor DB series, Models DB21, DB22, DB23, DB31, DB32, DB61M, DB62M, DB63M, DB71M are designed to measure the systolic and diastolic blood pressure and pulse rate of an individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. Our method to define systolic and diastolic pressure is similar to the auscultatory method but uses an electronic pressure sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alterations in cuff pressure to electrical signals, by analyzing those signals to define the systolic and diastolic blood pressure and calculating pulse rate, which is a well-known technique in the market called the "oscillometric method".
The main components of the Truly Automatic Arm Blood Pressure Monitor DB series are the main unit and cuff unit. ABS is used to outer housing of the main unit. The preformed cuff unit, which is applicable to arm circumference approximately between 220 and 340 mm, includes the inflatable bladder and nylon shell. All models of the arm blood pressure monitor use a single size of
20
{1}------------------------------------------------
Truly Instrument Limited.
cuff. The device consists of the microprocessor, the pressure sensor, the operation keys, the pump, the electromagnetic deflation control valve and the LCD. The subject devices are powered by four AAA or AA alkaline batteries.
The device also compares the longest and the shortest time intervals of detected pulse waves to mean time interval and displays a warning signal with the reading to indicate the detection of irregular pulse rhythm when the difference of the time intervals is over 25%.
### 5. Intended use of device
Truly Automatic Arm Blood Pressure Monitor DB series, Models DB21, DB22, DB23, DB31, DB32, DB61M, DB62M, DB63M, DB71M are a series devices intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm.
The devices features include irregular pulse rhythm detection during measurement, and display a warning signal with the reading once the irregular heartbeat is detected.
# 6. Summary of Substantial Equivalence
Table 1: The difference between Truly Automatic Arm Blood Pressure Monitor DB series and Digibio Digital Blood Pressure Monitor, Model D11.
| Parameter | Predicate<br>Devices D11 | DB<br>21 | DB<br>22 | DB<br>23 | DB<br>31 | DB<br>32 | DB<br>61M | DB<br>62M | DB<br>63M | DB<br>71M |
|-------------------------------------|--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------|----------|----------|----------|-----------|-----------|-----------|-----------|
| Measurement<br>algorithm | Oscillometric<br>method | No change ,all same | | | | | | | | |
| Method | | | | | | | | | | |
| Measurement<br>site of body | Arm | No change ,all same | | | | | | | | |
| Pressure<br>Sensor | MSP-2107 | No change, all same | | | | | | | | |
| Cuff | | No change ,all same | | | | | | | | |
| Software | | D11 software + Irregular heartbeat detection. | | | | | | | | |
| Irregular<br>heartbeat<br>detection | | More than ±25% to the mean interval of pulse intervals.<br>About the more detailed description of the IH detection algorithm, please<br>refer to "Software validation report I-5. Algorithm description 4. Determination<br>method of irregular heartbeat".<br>DB22, DB23, DB61M, DB62M, DB63M, DB71M have the IH feature. | | | | | | | | |
| Memory Size | 2 x 60 | 2X60 | 2X50 | 4X99 | 2X60 | 1X99 | 4X99 | 4X99 | 4X99 | 4X99 |
| Measurement<br>Pressure<br>Range | 20~280<br>mmHg | No change ,all same | | | | | | | | |
| Measurement<br>Pulse Range | 40~195<br>beats/min | No change ,all same | | | | | | | | |
{2}------------------------------------------------
# Truly Instrument Limited.
| Parameter | Predicate<br>Devices D11 | DB<br>21 | DB<br>22 | DB<br>23 | DB<br>31 | DB<br>32 | DB<br>61M | DB<br>62M | DB<br>63M | DB<br>71M |
|--------------------------|----------------------------|---------------------|----------|----------|----------|----------|-----------------------------------------------------------------------------|-----------|-----------|-----------|
| Mesauring<br>resolution | 1 mmHg | No change, all same | | | | | | | | |
| Accuracy<br>Pressure | $\pm$ 3mmHg | No change, all same | | | | | | | | |
| Accuracy<br>Pulse | $\pm$ 5% | No change, all same | | | | | | | | |
| Pressurization<br>Source | Automatic<br>internal pump | No change, all same | | | | | | | | |
| Ciff Deflation | Automatic<br>deflation | No change, all same | | | | | | | | |
| Operating<br>Environment | 10~40℃<br>15~90%RH | No change, all same | | | | | | | | |
| Power<br>Vovtage | 4X 1.5V | No change, all same | | | | | | | | |
| Hardware<br>circuit | | No change, all same | | | | | | | | |
| Electronic<br>element | | No change, all same | | | | | | | | |
| PCB | | Only | DB22 | | PCB | is | solely, other PCB of model<br>DB21/23/31/32/61M/62M/63M/71M are same to D11 | | | |
| Display Type | Liquid crystal<br>display | No change, all same | | | | | | | | |
| Cover | | Difference | | | | | | | | |
Table 2: The difference between Truly Automatic Arm Blood Pressure Monitor DB series and Microlife Blood Pressure Monitor, Model BP3BT0-AP
| Parameter | Predicate<br>Devices<br>BP3BT0-AP | DB<br>21 | DB<br>22 | DB<br>23 | DB<br>31 | DB<br>32 | DB<br>61M | DB<br>62M | DB<br>63M | DB<br>71M |
|-------------------------|-----------------------------------|----------------------|----------|----------|----------|----------|-----------|-----------|-----------|-----------|
| Measurement<br>Method | Oscillometric<br>Method | No change - the same | | | | | | | | |
| Pressure<br>Sensor | Capacitive | No change - the same | | | | | | | | |
| Measurement<br>Range:BP | 30~280mmHg | 20 ~ 280 mmHg | | | | | | | | |
| Measurement<br>Range:BP | 40~200<br>beats/min | 40 ~ 195 beats/min | | | | | | | | |
| Measuring<br>resolution | 1mmHg | No change - the same | | | | | | | | |
Page 3 of 4
:
22
{3}------------------------------------------------
Truly Instrument Limited.
| Parameter | Predicate<br>Devices<br>BP3BT0-AP | | DB<br>21 | DB<br>22 | DB<br>23 | DB<br>31 | DB<br>32 | DB<br>61M | DB<br>62M | DB<br>63M | DB<br>71M |
|-------------------------------------|-------------------------------------------------------------------------|--|-------------------------------------------------------------|----------|----------|----------|----------|-----------|-----------|-----------|-----------|
| Accuracy<br>Pressure | $\pm 3mmHg$ | | No change - the same | | | | | | | | |
| Accuracy<br>Pulse | $\pm 5%$ | | No change - the same | | | | | | | | |
| Pressurization<br>Source | Automatic<br>internal pump | | No change - the same | | | | | | | | |
| Cuff Deflation | Automatic<br>deflation | | No change - the same | | | | | | | | |
| Memory Size | 99 | | 2X60 | 2X50 | 4X99 | 2X60 | 1X99 | 4X99 | 4X99 | 4X99 | 4X99 |
| Irregular<br>Heartbeat<br>Detection | More than<br>$\pm 25%$ to the<br>mean interval<br>of pulse<br>intervals | | DB22, DB23, DB61M, DB62M, DB63M, DB71M have the IH feature. | | | | | | | | |
| Power Source | 4 X1.5V | | No change - the same | | | | | | | | |
| Operating<br>Environment | 10~40℃<br>15~90%RH | | No change - the same | | | | | | | | |
| Cuff<br>Attachment<br>Method | By plastic hose<br>connected to<br>monitor | | No change - the same | | | | | | | | |
| Display Type | Liquid crystal<br>display | | Liquid crystal display | | | | | | | | |
### 7. Conclusions
:
The subject devices have all features of the predicate device D11 except the new features such as irregular heartbeat detection. These differences do not affect the safety and effectiveness of the subject devices.
Irregular heartbeat detection technology is same as what is used Microlife Blood Pressure Monitor, Model BP3BT0-AP, K041411
Thus, the subject devices are substantially equivalent to the predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle with three bodies, representing the department's mission to protect the health of all Americans and provide essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle symbol.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
# SEP - 4 2009
Truly Instrument Co., Ltd. c/o Mr. Yang Jian-Hao Manager Truly Industrial Area Shanwei, Guangdong 516600 China
Re: K091434
> Trade/Device Name: Truly Automatic Arm Blood Pressure Monitor Models DB21, DB22, DB23, DB31, DB32, DB61M, DB62M, DB63M and DB71M. Regulatory Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II (Two) Product Code: DXN Dated: Undated Received: August 7, 2009
Dear Mr. Yang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
#### Page 2 - Mr. Yang Jian-Hao
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
W.M.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
09 14 34
**TRULY**®
# Truly Instrument Limited
Indication for Use
510(k) Number (if known):
Device Name: Truly Automatic Arm Blood Pressure Monitor DB Series,
Models DB21,DB22,DB23,DB31,DB32,DB61M,DB62M,DB63M,DB71M Indication For Use:
Truly Automatic Arm Blood Pressure Monitor ,Models DB21, DB23, DB31, DB32, DB61M, DB62M, DB63M, DB71M are a series devices intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm.
The devices features include irregular pulse rhythm detection during measurement, and display a warning signal with the reading once the irregular heartbeat is detected.
Prescription Use
Over the Counter Use
(21 CFR Part 801 Subpart D)
(21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
And
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
K.M.D.
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K091434
HongKong: 2/F, Chung Shun Knitting Centre, 1-3 Wing Yip Street, Kwai Chung, N.T., Hong Kong FAX: (852) 2146 5756 TEL: (852) 2487 9803 (20 Lines) E-MAIL: healthcare@truly.com.hk URL: http://www.trulyinstrument.com China: Truly Industrial Area .ShanWei City. Guangdong Province ,China FAX:86-0660-3380377 TEL:86-0660-3380070
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.