K112918 · Cardinal Health200, LLC · FRG · Jan 20, 2012 · General Hospital
Device Facts
Record ID
K112918
Device Name
DURABLUE STERILIZATION (AMSCO V-PRO) WRAP
Applicant
Cardinal Health200, LLC
Product Code
FRG · General Hospital
Decision Date
Jan 20, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
Cardinal Health DuraBlue™ Sterilization Wrap is intended to enclose another medical device that is to be sterilized by a health care provider in the Lumen, and Flexible Cycles in the Amsco® V-PRO® 1, Amsco® V-PRO® 1 Plus and Amsco® V-PRO® maX Low Temperature Sterilization Systems. The wrap is intended to allow sterilization of the enclosed medical device(s) and to maintain sterility of the enclosed device(s) for 30 days. The DuraBlue™ Sterilization Wrap was validated to be effectively aerated during the pre-programmed Amsco® V-PRO® sterilization cycles.
Device Story
DuraBlue™ Sterilization Wrap consists of double-layer 100% polypropylene SMS (spunbond-meltblown-spunbond) fabric; used by healthcare providers to enclose medical devices prior to sterilization in Amsco® V-PRO® 1, 1 Plus, and maX low-temperature systems. Wrap design facilitates simultaneous double-wrapping technique and aseptic opening. Device maintains sterility of enclosed contents for at least 30 days post-sterilization. Healthcare facilities select appropriate wrap model (CH100-CH600) based on package weight and content geometry. Benefits include protection of sterilized instruments during storage and transport, ensuring sterility until point of use.
Clinical Evidence
Bench testing only. Performance validated per FDA guidance for sterilization packaging. Testing included sterilization efficacy, event-related maintenance of package sterility, physical property testing (pre/post-sterilization), and biocompatibility (ISO 10993). No clinical data presented.
Technological Characteristics
Material: 100% polypropylene spunbond-meltblown-spunbond (SMS) trilaminate nonwoven fabric. Form factor: Six models (CH100-CH600) with varying basis weights. Sterilization compatibility: Amsco® V-PRO® 1, 1 Plus, and maX low-temperature systems. Biocompatibility: Compliant with ISO 10993.
Indications for Use
Indicated for healthcare providers to enclose reusable metal and non-metal medical devices, including flexible endoscopes and bronchoscopes, for sterilization in Amsco® V-PRO® low-temperature systems. Supports specific lumen dimensions and load configurations.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{0}------------------------------------------------
K112918
JAN 2 0 2012
CardinalHealt
1430 Waukegan Road McGraw Park, IL 60085
www.cardinal.com
## SMDA REQUIREMENTS
| | <b>510(k) SUMMARY OF SAFETY AND EFFECTIVENESS</b><br>DuraBlue™ Sterilization Wrap<br>Amsco® V-PRO® 1, Amsco® V-PRO® 1 Plus and Amsco® V-PRO® maX |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer: | Cardinal Health 200, LLC<br>1430 Waukegan Road<br>McGaw Park, IL 60085 |
| Regulatory Affairs Contact: | Lavenia Ford<br>1430 Waukegan Road<br>McGaw Park, IL 60085 |
| Telephone Number: | (847) 887-3323 |
| Date summary Prepared: | December 21, 2011 |
| Trade Name: | DuraBlue™ Sterilization Wrap |
| Classification: | Class II per 21 CFR § 880.6850 |
| Classification Name: | Sterilization Wrap |
| Predicate Device: | K092167 - KIMGUARD ONE-STEP Sterilization Wrap |
| Description: | |
Cardinal Health DuraBlue™ Sterilization Wraps are double layer sterilization wraps made from 100% polypropylene spunbond-meltblown-spunbond (SMS) fabric. They are intended to be used to enclose another medical device that is to be sterilized by a health care provider in the the Lumen, or
Flexible Cycles in the Amsco V-PRO® 1, Amsco V-PRO® 1 Plus and Amsco "V-PRO" ma Temperature Sterilization Systems. This wrap design allows for use of the simultaneous doublewrapping technique and also allows for a sterilized pack to be opened aseptically.
This submission covers six different models of Cardinal Health DuraBlue™ Sterilization Wrap. Each model is made from material of a different basis weight, though all models utilize the same material technology.
Extensive performance testing has been completed on Cardinal Health DuraBlue™ Sterilization Wrap. Successful completion of the sterilization performance tests listed below demonstrated that the wrap both allows for sterilization of the enclosed contents and maintains sterility of the enclosed contents for at least 30 days.
{1}------------------------------------------------
Indications for Use:
Cardinal Health DuraBlue™ Sterilization Wrap is intended to enclose another medical device that is to be sterilized by a health care provider in the Lumen, and Flexible Cycles in the Amsco® V-PRO® 1, Amsco® V-PRO® 1 Plus and Amsco® V-PRO® maX Low Temperature Sterilization Systems. The wrap is intended to allow sterilization of the enclosed medical device(s) and to maintain sterility of the enclosed device(s) for 30 days. The DuraBlue™ Sterilization Wrap was validated to be effectively aerated during the pre-programmed Amsco® V-PRO® sterilization cycles.
All models of DuraBlue™ Sterilization Wrap have been validated for use with the following Amsco® V-PRO® cycles:
| Amsco® V-PRO®<br>Cycle | Maximum<br>Recommended<br>Chamber Load | Intended Load | Maximum Recommended<br>Wrapped Package<br>Content Weight by Wrap<br>Model¹ |
|------------------------|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Lumen Cycle | 19.65 lbs | Reusable metal and non-metal medical<br>devices, including up to 20 lumens of the<br>following dimensions per chamber load:<br>an inside diameter of 1 mm or larger<br>and a length of 125 mm or shorter an inside diameter of 2 mm or larger<br>and a length of 250 mm or shorter an inside diameter of 3 mm or larger<br>and a length of 400 mm or shorter | Wrap Model Max.<br>Package Weight<br>CH100 3 lbs<br>CH200 6.5 lbs<br>CH300 9 lbs<br>CH400 9.1 lbs<br>CH500 9.1 lbs<br>CH600 9.1 lbs |
| Non Lumen Cycle | 19.65 lbs | Non-lumened reusable metal and non-metal<br>medical devices | |
| Flexible Cycle | 24 lbs | Single or dual lumen surgical flexible<br>endoscopes and bronchoscopes in either of<br>two load configurations:<br>1. Two trays, each containing a flexible<br>endoscope with a light cord (if not<br>integral to endoscope) and mat with no<br>additional load<br>2. One tray containing a flexible<br>endoscope with a light cord (if not<br>integral to endoscope) and mat and an<br>additional tray containing non-lumened<br>medical devices<br><br>The flexible endoscope(s) may contain<br>either:<br>a single lumen with an inside<br>diameter of 1 mm or larger and<br>a length of 1050 mm or shorter two lumens, with one lumen<br>having an inside diameter of 1<br>mm or larger and a length of<br>998 mm or shorter and the<br>other lumen having an inside<br>diameter of 1 mm or larger and<br>a length of 850 mm or shorter | |
Individual results may differ due to factors such as variations in handling practices, wrapping techniques, and folding methods. Results may also differ due to the use of irregularly shaped contents, which may put added stress on the wrap. Each healthcare facility should determine for itself which wrap model is most appropriate for each intended use.
{2}------------------------------------------------
## Substantial Equivalence
The DuraBlue ™ Sterilization Wrap is substantially equivalent to the predicate devices.
- Both devices are double layer sterilization wraps which allow for use of the simultaneous double-� wrapping technique and for a sterilized pack to be opened aseptically.
- Both devices are intended to be used with the same V-PRO® sterilization parameters.
- Both devices are available in six comparable models of varying basis weights, which are recommended for use with the same maximum content weights.
- Both devices have the same dimensional specifications.
- Both devices are100% polypropylene spunbond-me!tblown-spunbond (SMS) trilaminate nonwoven fabric.
- Both devices demonstrate maintenance of package sterility for at least 30 days following sterilization by pre-vacuum steam.
- Performance and safety attributes are substantially equivalent to the predicate. The physical properties of all wrap models have been characterized both before and after exposure to V-PRO®
- sterilization. The resulting data supports the conclusion that Cardinal Health DuraBlue™ Sterilization Wrap is substantially equivalent to the predicate, and the DuraBlue™ Sterilization Wraps are compatible with the identified V-PRO® system sterilization parameters.
#### Summary of Testing
DuraBlue™ Sterilization Wrap performance has been tested in accordance with the applicable requirements recommended in the FDA's Guidance Document Premarket Notification 510(k) Submissions for Medical Sterilization Packaging System in Health Care Facilities; Draft Guidance for Industry and FDA (March 7, 2002). Testing included sterilization efficacy, event related maintenance of package sterility, physical properties, and biocompatibility in compliance with the methods of ISO 10993. Data from testing demonstrates that the performance of the DuraBlue™ Sterilization Wrap is substantially equivalent to that of Kimberly-Clark KIMGUARD ONE-STEP Sterilization Wrap.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three tail feathers, representing the three levels of government: federal, state, and local. The eagle is surrounded by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Cardinal Health 200, LLC C/O Mr. Ned Devine Responsible Third Party Official Underwriters Laboratories, Incorporated 333 Pfingsten Road Northbrook, Illinois 60062
JAN 2 0 2012
Re: K112918
Trade/Device Name: Cardinal Health DuraBlue™ Sterilization Wrap Amsco® V-PRO® 1, Amsco®V-PRO®1 Plus and Amsco® V-PRO® maX Regulation Number: 21 CFR 880.6850
Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: FRG Dated: January 3, 2012 Received: January 6, 2012
Dear Mr. Devine:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register.
{4}------------------------------------------------
# Page 2 - Mr Devine
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the I ou inal of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Austin James
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# CardinalHealth
## Indication for Use
## 510(k) Number (if known): K112918
Device Name: Cardinal Health DuraBlue™ Sterilization Wrap
Amsco® V-PRO® 1, Amsco® V-PRO® 1 Plus and Amsco® V-PRO® max
Indications for Use:
Cardinal Health DuraBlue™ Sterilization Wrap is intended to enclose another medical device that is to be sterilized by a health care provider in the Lumen, and Flexible Cycles in the Amsco® V-PRO® 1, Amsco® V-PRO® 1 Plus and Amsco® V-PRO® maX Low Temperature Sterilization Systems. The wrap is intended to allow sterilization of the enclosed medical device(s) and to maintain sterility of the enclosed device(s) for 30 days. The DuraBlue™ Sterilization Wrap was validated to be effectively aerated during the pre-programmed Amsco® V-PRO® sterilization cycles.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Elizabe H. F. Clarence-Wills
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K112918
{6}------------------------------------------------
All models of DuraBlue™ Sterilization Wrap have been validated for use with the following Amsco "V-PRO® cycles:
| Amsco® V-PRO® Cycle | Maximum Recommended Chamber Load | Intended Load | Maximum Recommended Wrapped Package Content Weight by Wrap Model¹ | | | | | | | | | | | | | | |
|---------------------|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|
| Lumen Cycle | 19.65 lbs | Reusable metal and non-metal medical devices, including up to 20 lumens of the following dimensions per chamber load:<br>• an inside diameter of 1 mm or larger and a length of 125 mm or shorter<br>• an inside diameter of 2 mm or larger and a length of 250 mm or shorter<br>• an inside diameter of 3 mm or larger and a length of 400 mm or shorter | Wrap Model Max. Package Weight CH100 3 lbs CH200 6.5 lbs CH300 9 lbs CH400 9.1 lbs CH500 9.1 lbs CH600 9.1 lbs | | | | | | | | | | | | | | |
| Non Lumen Cycle | 19.65 lbs | Non-lumened reusable metal and non-metal medical devices | | | | | | | | | | | | | | | |
| Flexible Cycle | 24 lbs | Single or dual lumen surgical flexible endoscopes and bronchoscopes in either of two load configurations:<br><br>1. Two trays, each containing a flexible endoscope with a light cord (if not integral to endoscope) and mat with no additional load<br><br>2. One tray containing a flexible endoscope with a light cord (if not integral to endoscope) and mat and an additional tray containing non-lumened medical devices<br><br>The flexible endoscope(s) may contain either:<br>• a single lumen with an inside diameter of 1 mm or larger and a length of 1050 mm or shorter<br>• two lumens, with one lumen having an inside diameter of 1 mm or larger and a length of 998 mm or shorter and the other lumen having an inside diameter of 1 mm or larger and a length of 850 mm or shorter | | | | | | | | | | | | | | | |
Individual results may differ due to factors such as variations in handling practices, wrapping techniques, marriadal rocelle hay also differ due to the use of irregularly shaped contents, which may put added stress on the wrap. Each healthcare facility should determine for itself which wrap model is most appropriate for each intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.