KIMGUARD ONE-STEP* Sterilization Wrap is intended to be used to enclose another medical device that is to be sterilized by a healthcare provider in the Amsco V-PRO 1 Low Temperature Sterilization System's Cycle or the Amsco V-PRO 1 Plus Low Temperature Sterilization System's Lumen (identical to the V-PRO 1 Cycle) and Non Lumen Cycles. The wrap is intended to allow sterilization of the enclosed medical device(s) and also to maintain sterility of the enclosed device(s) until opened. The wrap was validated to be effectively aerated during the pre-programmed V-PRO 1 and V-PRO 1 Plus Sterilization Cycles.
Device Story
KIMGUARD ONE-STEP* Sterilization Wrap consists of two sheets of polypropylene SMS (spunbond-meltblown-spunbond) fabric ultrasonically seamed on two sides; designed for simultaneous two-sheet wrapping. Used by healthcare providers in clinical settings to enclose medical instruments prior to sterilization in Amsco V-PRO 1 or V-PRO 1 Plus Low Temperature Sterilization Systems. Wrap allows sterilant penetration and aeration during cycles; maintains sterility of contents post-sterilization. Healthcare providers select appropriate wrap model (KC100-KC600) based on instrument weight and configuration. Benefits include protection of sterile integrity of instruments until point of use.
Clinical Evidence
No clinical data. Bench testing performed in accordance with FDA guidance for medical sterilization packaging systems. Testing included biocompatibility (ISO 10993), sterilant penetration, and physical integrity. Sterility maintenance validated for up to 30 days post-sterilization. All results met acceptance criteria.
Technological Characteristics
Material: Polypropylene three-layer SMS (spunbond-meltblown-spunbond) fabric. Construction: Two sheets ultrasonically seamed on two sides. Form factor: Square or rectangular sheets. Models KC100-KC600 vary by weight capacity. Sterilization compatibility: Amsco V-PRO 1 and V-PRO 1 Plus Low Temperature Sterilization Systems.
Indications for Use
Indicated for healthcare providers to enclose medical devices for sterilization in Amsco V-PRO 1 and V-PRO 1 Plus Low Temperature Sterilization Systems and to maintain sterility until opened.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{0}------------------------------------------------
# NOV - 2 2009
2092167
# 510(k) Summary for the Kimberly-Clark* Corporation KIMGUARD ONE-STEP* Sterilization Wrap (Models KC100, KC200, KC300, KC400, KC500, and KC600) for the Additional Indication for Use with the Amsco® V-PRO 1 and V-PRO 1 Plus Low Temperature Sterilization Systems
Date Summary was Prepared: 510(k) Submitter and Primary Contact:
October 23, 2009
Sterilization Wrap
FRG
Thomas Kozma, PhD Director, Regulatory Affairs Kimberly-Clark Health Care 1400 Holcomb Bridge Road Roswell, GA 30076 Ph: 770.587.8393 FAX: 920.225.3408 Email: thomas.kozma@kcc.com
Device Common Name: Classification Name:
Sterilization Wrap (21 CFR 880.6850)
Product Code:
Additional Intended Use Subject of this 510(k) Submission: KIMGUARD ONE-STEP* Sterilization Wrap is intended to be used to enclose another medical device that is to be sterilized by a healthcare provider in the Amsco V-PRO 1 Low Temperature Sterilization System's Cycle or the Amsco V-PRO 1 Plus Low Temperature Sterilization System's Lumen (identical to the V-PRO 1 Cycle) and Non Lumen Cycles. The wrap is intended to allow sterilization of the enclosed medical device(s) and also to maintain sterility of the enclosed device(s) until opened. The wrap was validated to be effectively aerated during the pre-programmed V-PRO 1 and V-PRO 1 Plus Sterilization Cycles.
See Wrap Model Recommendations on Page 2 of this Summary.
The KIMGUARD ONE-STEP* Sterilization Wrap (Models KC100, KC200, KC300, KC400, KC500, and KC600) for the additional indication for use with the Amsco V-PRO 1 and V-PRO 1 Plus Low Temperature Sterilization Systems is substantially equivalent to the predicate KIMGUARD ONE-STEP* Sterilization Wrap (Models KC100,
KC200, KC300, KC400, KC500, and KC600) (K082177).
Predicate Devices:
Substantial Equivalence: The KIMGUARD ONE-STEP* Sterilization Wrap is identical to the predicate in intended use, design, materials, and performance. The KIMGUARD ONE-STEP* Sterilization Wrap is comprised of two sheets of KIMGUARD* Sterilization Wrap ultrasonically seamed on two sides. This allows for convenient wrapping with two sheets simultaneously. The sheets of sterilization wrap are square or rectangular fabric produced using a polypropylene three-layer SMS (spunbondmeltblown-spunbond) process.
Page 1 of 2 - 510(k) Summary
{1}------------------------------------------------
K092167
page 2 of 2
### Summary of Testina:
KIMGUARD ONE-STEP* Sterilization Wrap performance has been tested in accordance with the applicable requirements recommended in Premarket Notification [510(k)] Submissions for Medical Sterilization Packaging Systems in Health Care Facilities; Draft Guidance for Industry and FDA (March 7, 2002). Testing included biocompatibility in compliance with the methods of ISO 10993, sterilant penetration, and physical integrity. The Wrap has also been tested for the ability to maintain sterility of pack contents after sterilization for up to 30 days under standard conditions. All results of testing met acceptance criteria.
## Wrap Model Recommendations for Amsco V-PRO 1 and V-PRO 1 Plus Low Temperature Sterilization1
| KIMGUARD<br>ONE-STEP*<br>Sterilization<br>Wrap Models | Intended Loads | Maximum<br>Wrapped<br>Package<br>Content<br>Weights Used in<br>Sterility<br>Maintenance<br>Validation<br>Study2 | Descriptions of Loads Used in Sterility<br>Maintenance Validation Study2 |
|-------------------------------------------------------|------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| KC100 | Very Light Weight<br>Package (for example<br>batteries) | 3 lbs | • 3 lbs metal mass<br>• 6 forceps |
| KC200 | Light Weight Package<br>(for example telescope<br>with light cord) | 6.5 lbs | • 2.5 lbs metal mass<br>• 6 forceps<br>• V-PRO tray (17" x 10" x 3½") at 4 lbs |
| KC300 | Light to Moderate<br>Weight Package (for<br>example: general use<br>medical instruments) | 9 lbs | • 5 lbs metal mass<br>• 6 forceps<br>• V-PRO tray (17" x 10" x 3½") at 4 lbs |
| KC400 | Moderate to Heavy<br>Weight Package (for<br>example: general use<br>medical instruments) | 10 lbs | • 6 lbs metal mass<br>• 6 forceps<br>• V-PRO tray (17" x 10" x 3½") at 4 lbs |
| KC500 | Heavyweight Package<br>(for example: general<br>use medical<br>instruments) | 10 lbs | • 5 lbs metal mass<br>• 6 forceps<br>• V-PRO tray (21" x 10" x 3½") at 5 lbs |
| KC600 | Very Heavy Weight<br>Package (for example:<br>general use medical<br>instruments) | 10 lbs | • 5 lbs metal mass<br>• 6 forceps<br>• V-PRO tray (21" x 10" x 3½") at 5 lbs |
Individual results may differ due to factors such as variations in handling practices, wrapping techniques, and folding methods. Results may also differ due to the use of irregularly shaped contents, which may put added stress on the wrap. Each healthcare facility should determine for itself which wrap model is most appropriate for each intended use. 1 t is recommended to not exceed the maximum wrapped package content weights indicated for each wrap model.
Furthermore, it is recommended to not exceed the number, weight, and size of individual content types that were validated for the KIMGUARD Sterilization Wraps (i.e.; the weight of the metal mass).
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged in a circular pattern around the bird-like figure. The text is in all capital letters and is evenly spaced around the circle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002
Dr. Thomas Kozma Director of Regulatory Affairs Kimberly-Clark Corporation 1400 Holcomb Bridge Road Roswell, Georgia 30076
# NOV - 2 2009
Re: K092167
Trade/Device Name: KIMGUARD ONE-STEP* Sterilization Wrap (Models KC100, KC200, KC300, KC400, KC500 and KC600) for the Use With the Lumen Cycle of the Amsco V-PRO 1 Low Temperature Sterilization System and With the Lumen and Non Lumen Cycles of the Amsco V-PRO 1 Plus Low Temperature Sterilization System
Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II · Product Code: FRG Dated: October 5, 2009 Received: October 7, 2009
Dear Dr. Kozma:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
## Page 2- Dr. Kozma
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRH Offices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Anthony D. Roberts Jr.
Susan Runner, D.D.S., M.A. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### Indications for Use
#### 510(k) Number (if known): K092167
Device Name: _ KIMGUARD ONE-STEP* Sterilization Wrap (Models KC100, KC200, KC300, KC400, KC500, and KC600) for the Additional Indication for Use with the Amsco® V-PRO™ 1 and V-PRO 1 Plus Low Temperature Sterilization Systems
#### Indications for Use:
KIMGUARD ONE-STEP* Sterilization Wrap is intended to be used to enclose another medical device that is to be sterilized by a healthcare provider in the Amsco V-PRO 1 Low Temperature Sterilization System's Cycle or the Amsco V-PRO 1 Plus Low Temperature Sterilization System's Lumen (identical to the V-PRO 1 Cycle) and Non Lumen Cycles. The wrap is intended to allow sterilization of the enclosed medical device(s) and also to maintain sterility of the enclosed device(s) until opened. The wrap was validated to be effectively aerated during the pre-programmed V-PRO 1 and V-PRO 1 Plus Sterilization Cycles.
KIMGUARD ONE-STEP* Sterilization Wrap Recommendations for use with the Amsco V-PRO 1 and V-PRO 1 Plus Low Temperature Sterilization Systems are provided on Page 2.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
<PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
| | Eli sheth P. Clavene-Wilde |
|--------------------------------------------------------|----------------------------|
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
| (Division Sign-Off) | |
| Division of Anesthesiology, General Hospital | |
| Infection Control, Dental Devices | |
| 510(k) Number: | K092167 |
{5}------------------------------------------------
#### Indications for Use
# Wrap Model Recommendations for Amsco V-PRO 1 and V-PRO 1 Plus Low Temperature Sterilization'
| KIMGUARD<br>ONE-STEP*<br>Sterilization<br>Wrap Models | Intended Loads | Maximum<br>Wrapped<br>Package<br>Content<br>Weights Used<br>in Sterility<br>Maintenance<br>Validation<br>Study2 | Descriptions of Loads Used in<br>Sterility Maintenance Validation<br>Study2 |
|-------------------------------------------------------|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| KC100 | Very Light Weight<br>Package (for<br>example batteries) | 3 lbs | 3 lbs metal mass 6 forceps |
| KC200 | Light Weight<br>Package (for<br>example telescope<br>with light cord) | 6.5 lbs | 2.5 lbs metal mass 6 forceps V-PRO tray (17" x 10" x 3½") at 4 lbs |
| KC300 | Light to Moderate<br>Weight Package<br>(for example:<br>general use<br>medical<br>instruments) | 9 lbs | 5 lbs metal mass 6 forceps V-PRO tray (17" x 10" x 3½") at 4 lbs |
| KC400 | Moderate to Heavy<br>Weight Package<br>(for example:<br>general use<br>medical<br>instruments) | 10 lbs | 6 lbs metal mass 6 forceps V-PRO tray (17" x 10" x 3½") at 4 lbs |
| KC500 | Heavyweight<br>Package (for<br>example: general<br>use medical<br>instruments) | 10 lbs | 5 lbs metal mass 6 forceps V-PRO tray (21" x 10" x 3½") at 5 lbs |
| KC600 | Very Heavy Weight<br>Package (for<br>example: general<br>use medical<br>instruments) | 10 lbs | 5 lbs metal mass 6 forceps V-PRO tray (21" x 10" x 3½") at 5 lbs |
Individual results may differ due to factors such as variations in handling practices, wrapping techniques, and folding methods. Results may also differ due to the use of iregularly shaped contents, which may put added stress on the wrap. Each healthcare facility should determine for itself which wrap model is most appropriate for each intended use. 2 It is recommended to not exceed the maximum wrapped package content weights indicated for each wrap model. Furthernore, it is recommended to not exceed the number, weight, and size of individual content types that were validated for the KIMGUARD Sterilization Wraps (i.e .: the weight of the metal mass).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.