The Cryotop®CL is a cryopreservation storage device that is intended for use in vitrification procedures to contain and maintain human 4-8 cell and blastocyst stage embryos.
Device Story
CryotopCL is a sterile, single-use cryopreservation storage device for human embryos. It consists of a PET plastic shaft with a fine tip featuring five depressed areas for embryo loading, a protective straw enclosure, and a pushing tool. During vitrification, embryos are loaded onto the tip, which is then inserted into the straw enclosure and heat-sealed by the user. The straw is weighted for alignment in storage containers. The device is used in clinical laboratory settings by embryologists. It facilitates the rapid cooling and rewarming of embryos during cryopreservation, helping to maintain embryo viability for future use.
Clinical Evidence
No clinical data was required. Substantial equivalence is supported by non-clinical bench testing, including cooling/warming rate verification, mechanical tensile strength testing (5N), dimensional analysis, endotoxin testing (≤0.5 EU/device), and Mouse Embryo Assay (≥80% development of 1-cell control embryos at 96 hours). Sterility was validated to SAL 10^-6.
Technological Characteristics
Materials: PET and Polypropylene. Form factor: Three-piece assembly (shaft with fine tip, straw, pushing tool). Cooling rate: 3,000°C/min. Rewarming rate: 40,000°C/min. Sterilization: Radiation. Single-use, sterile.
Indications for Use
Indicated for use in vitrification procedures to contain and maintain human 4-8 cell and blastocyst stage embryos.
Regulatory Classification
Identification
Assisted reproduction labware consists of laboratory equipment or supplies intended to prepare, store, manipulate, or transfer human gametes or embryos for in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures. These include syringes, IVF tissue culture dishes, IVF tissue culture plates, pipette tips, dishes, plates, and other vessels that come into physical contact with gametes, embryos or tissue culture media.
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, and clinical testing). The device, when it is a dish or plate intended for general assisted reproduction technology procedures, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
{0}------------------------------------------------
pg. 1 of 4
# 510(k) Summary
# DEC 1 2 2012
for
# Kitazato's Cryotop®CL, K112695
### 1. Submission Sponsor
KITAZATO BioPharma Co., Ltd. 81 Nakajima, Fuji Shizuoka 416-0907 JAPAN Phone: +(81) 545-66-2202 Fax: +(81) 545-60-5772 Contact: Futoshi Inoue, President
#### Submission Correspondent 2.
Emergo Group 611 West 5th Street, Third Floor Austin, TX 78701 Cell Phone: (508) 838.9139 Office Phone: (512) 327.9997 Fax: (512) 327.9998 Contact: Richard Vincins, Vice President, QA Email: project.management@emergogroup.com
#### 3. Date Prepared
7 December 2012
#### 4. Device Name
Cryotop®CL Trade/Proprietary Name: Cryotop Closed Common/Usual Name: Classification Name: Assisted Reproduction Labware Classification Regulation: 884.6160 Obstetrics/Gynecology Classification Panel: Product Code: MQK Device Class: ll
### 5. Predicate Devices
Cryo Bio System, HSV Straw, K092398
{1}------------------------------------------------
Pg. 2 of 4
K112695
#### Device Description 6.
CryotopCL is a cryopreservation storage device that is intended for use in vitrification procedures to contain and maintain human 4-8 cell and blastocyst stage embryos.
The CryotopCL device is composed of PET plastic shaft with a fine tip and a straw cap. The fine tip has 5 rounded depressed areas for loading embryos. The CryotopCL device has a "straw" enclosure system for when the unit is placed in the liquid nitrogen. The straw cap is designed to be heat sealed by the user. The protective straw cap has a weighted end to allow proper alignment in the storage container. The CryotopCL device is provided sterile and is for single use only. The device is also provided with a pushing straw to aid in loading the sample holding component of the device into the "straw" enclosure system.
The CryotopCL device conforms to product quality test specifications of our company: appearance, dimension, durability, tensile strength, colorfastness, endotoxin and Mouse Embryo Assay. The sterilization dose of CryotopCL is validated by sterilization validation to maintain the sterility of the device through anticipated storage and handling.
#### 7. Indication for Use
The Cryotop®CL is a cryopreservation storage device that is intended for use in vitrification procedures to contain and maintain human 4-8 cell and blastocyst stage embryos.
# 8. Technological Characteristics and Substantial Equivalence
The following table compares Kitizato's CryotopCL to the predicate device with respect to intended use, technological characteristics, and principles of operation.
| Manufacturer | KITAZATO BioPharma Co., Ltd. | Cryo Bio System |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | Cryotop®CL | HSV Straw |
| 510(k) Number | K112695 | K092398 |
| Product Code | MQK | MQK |
| Regulation Number | 884.6160 | 884.6160 |
| Regulation Name | Assisted Reproduction Labware | Assisted Reproduction Labware |
| Indications for Use: | The CryotopCL is a<br>cryopreservation storage device<br>that is intended for use in<br>vitrification procedures to contain<br>and maintain human 4-8 cell and<br>blastocyst stage embryos. | The HSV Straw is a<br>cryopreservation storage device<br>that is intended for use in<br>vitrification procedures to contain<br>and maintain human 4-8 cell and<br>blastocyst stage embryos. |
| Overall Design | The device consists of a three<br>piece assembly comprised of the<br>main part containing the fine tip,<br>the protective "straw," and<br>pushing tool to assist in the proper<br>placement of the main part. The<br>straw is designed to be sealed. | The device consists of 3 parts: a<br>resin straw, a capillary tube with a<br>pre-formed gutter attached to a<br>colored handling rod, and a plastic<br>pushing tool to assist in the<br>proper placement of the capillary<br>tube. The straw is designed to be |
{2}------------------------------------------------
Pg. 3 of 4
K112695
| Manufacturer | KITAZATO BioPharma Co., Ltd. | Cryo Bio System |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | Cryotop®CL | HSV Straw |
| | The straw is weighted to allow<br>proper alignment in the storage<br>container. The CryotopCL device<br>is packaged in a single peel off<br>blister pack that is terminally<br>sterilized; 10 (ten) units are placed<br>in a secondary paper pouch for<br>storage purposes. | sealed. The HSV Straw is packaged<br>in a peel off blister pack. |
| Method of Action | Vitrification Method | Vitrification Method |
| Sterile | Radiation | Radiation |
| Cooling Rate | 3,000°C/min | 2,900°C/min |
| Rewarming Rate | 40,000°C/min | 25,000°C/min |
| Material Composition | PET, Polypropylene | Medical Grade Styrene-Butadiene<br>Copolymer |
| Contact with<br>Warming Medium | The tip (film) and the shaft of<br>CryotopCL are taken out from<br>straw. Directly immerse the tip<br>(film) into warming medium. | The curved spatula, containing the<br>cryopreserved embryo, is<br>immersed in warming medium<br>where thawing and dilution in the<br>warming medium occur<br>simultaneously. |
| Performance Testing<br>of Device | Passed | Passed |
| Mouse Embryo Test<br>Passed | ≥80% of 1-cell control embryos | ≥80% of 1-cell control embryos |
| Endotoxin Testing | ≤0.5 EU/device | ≤0.5 EU/device |
| Sterility Validation<br>Passed | No microbial growth, SAL 10-6 | No microbial growth, SAL 10-6 |
## 9. Non-Clinical Testing
The CryotopCL device has undergone cooling/warming rate testing, mechanical testing, The Cryolopce device has andergone obsing, wouse embryo assay testing, and endotoxin testing, dimensional testing, stermsy cosing, ""
Results support that all the specifications have met the acceptance criteria for the device.
- Cooling Rate Testing: Cooling rate of 3,000 °C/min passed �
- Cooling Rate Testing: Warming rate of 40,000 °C/min passed �
- warming nate Testing: Passes outer diameter, length, width of shaft, and spacing of .
- the embryo holding locations according to specifications
- Mechanical Tensile Testing: Tensile strength to withstand 5N .
- Mechanied Terror: Endotoxin values conform to the value ≤0.5 EU/device ●
- Sterility Testing: No microbial growth from sterility testing .
- Stenity Yesting 10 % of 1-cell control embryos develop at 96 hours .
The CryotopCL device passed all testing and supports the claims of substantial equivalence.
The CryotopCL device complies with the applicable voluntary standards for sterilization. The The Clyotopee de nee complise with national and international standards.
device passed all the testing in accordance with national and international standards.
{3}------------------------------------------------
Pg. 4 of 4
R112695
#### 10. Clinical Testing
There was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate devices. The substantial equivalence of the device is supported by the non-clinical testing. The validation testing of sterility testing and mouse embryo testing was found to acceptable and supports the claims of substantial equivalence.
#### 11. Conclusion
By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device. Or the device has the same intended use and different technological characteristics that can be demonstrated that the device is substantially equivalent to the predicate device, and that the new device does not raise different questions regarding its safety and effectiveness as compared to the predicate device.
It has been shown in this 510(k) submission that the difference between the CryotopCL device and the predicate devices do not raise any new types of questions regarding its safety and effectiveness. Kitazato's Cryotop®CL, as designed and manufactured, is determined to be substantially equivalent to the referenced predicate devices.
{4}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird with three stylized wing segments.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
## December 12, 2012
KITAZATO BioPharma Co., Ltd. % Mr. Richard Vincins, CQA, CBA, RAC (US, EU) Vice President, Quality Assurance Emergo Group 611 West 5th Street, Third Floor AUSTIN TX 78701
Re: K112695
Trade/Device Name: Cryotop®CL Regulation Number: 21 CFR§ 884.6160 Regulation Name: Assisted reproduction labware Regulatory Class: II Product Code: MQK Dated: November 21, 2012 Received: November 23, 2012
Dear Mr. Vincins:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{5}------------------------------------------------
Page 2 - Mr. Richard Vincins, CQA, CBA, RAC (US, EU)
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Section 4 - Indications for Use Statement
510(k) Number (if known): K112695
Device Name: Cryotop®CL
Indications for Use:
The Cryotop®CL is a cryopreservation storage device that is intended for use in vitrification procedures to contain and maintain human 4-8 cell and blastocyst stage embryos.
(Part 21 CFR 801 Subpart D) Prescription Use × 1 AND/OR Over-The-Counter Use ______________________________ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Benjamin R. Fisher -S 2012.12.12 11/110:49 -05'00'
(Division Sign-Off) ision of Reproductive, Gastro-Renal, and logical Devices (k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.