K092398 · Cryo Bio System · MQK · Oct 6, 2010 · Obstetrics/Gynecology
Device Facts
Record ID
K092398
Device Name
HSV STRAW
Applicant
Cryo Bio System
Product Code
MQK · Obstetrics/Gynecology
Decision Date
Oct 6, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.6160
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K092398 · Oct 6, 2010
HSV STRAW
Cryo Bio System
Multi-center clinical study of routine fertility treatment cycles
The study evaluated the performance of the HSV Straw in 473 vitrification cycles across three clinical centers, demonstrating embryo survival rates >80% and successful pregnancies/live births.
Embryo survival; Vitrification; Clinical study; Routine clinical use
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Multi-center retrospective clinical study
Human embryos (4-8 cell and blastocyst stage) undergoing vitrification; Sample Size: 473 cycles (1509 embryos); Number of Sites: 3
Not applicable for this study
Embryo survival rate, pregnancy, and live birth
Indications for Use
The HSV Straw is a cryopreservation storage device that is intended for use in vitrification procedures to contain and maintain human 4-8 Cell and blastocyst stage embryos.
Device Story
HSV Straw is a cryopreservation storage device for human embryos. Input: embryo suspended in 0.5ul vitrification solution. Operation: embryo placed in curved medical-grade styrene-butadiene copolymer spatula; spatula inserted into outer straw; proximal end sealed; straw immersed in liquid nitrogen for storage. Rewarming: distal tip remains in liquid nitrogen while outer tube is cut; curved spatula immersed directly in thawing solution for simultaneous thawing and dilution. Used in assisted reproduction laboratories by embryologists. Benefits: provides protected environment for embryo vitrification and storage, preventing nitrogen contamination. Output: cryopreserved embryo ready for clinical use.
Clinical Evidence
Clinical study conducted across three centers with 473 vitrification cycles and 1509 embryos. Primary endpoint: post-vitrification/rewarming embryo survival rate. Results: >80% survival across all centers; successful pregnancies and live births reported. No device-associated adverse effects. Bench testing included mouse embryo assay (MEA) with 90-100% survival at 96 hours, endotoxin levels <1 EU/unit, and mechanical testing.
Technological Characteristics
Material: Medical Grade Styrene-Butadiene Copolymer. Form factor: Curved spatula inserted into outer straw. Cooling rate: 2,900°C/min. Rewarming rate: 25,000°C/min. Principle: Vitrification via direct immersion in liquid nitrogen. Sterilization: Not specified. Connectivity: None (standalone).
Indications for Use
Indicated for containment and maintenance of human 4-8 cell and blastocyst stage embryos during vitrification procedures.
Regulatory Classification
Identification
Assisted reproduction labware consists of laboratory equipment or supplies intended to prepare, store, manipulate, or transfer human gametes or embryos for in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures. These include syringes, IVF tissue culture dishes, IVF tissue culture plates, pipette tips, dishes, plates, and other vessels that come into physical contact with gametes, embryos or tissue culture media.
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, and clinical testing). The device, when it is a dish or plate intended for general assisted reproduction technology procedures, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
{0}------------------------------------------------
#### Manufacturer and Submitter
Cryo Bio System 29 Rue Tronchet Paris 75008 FRANCE
#### Contact Person
Anne Linda van Kappel, MSc, PhD Phone: 33-1-49-24-05-05 FAX: 33-1-49-24-05-01 e-mail: alvk@cryobiosystem-imv.com
DCT 6 2010
## Date Prepared: 17 September 2010
#### Establishment Registration
Cryo Bio System Owner Operator Number: 9042657
#### Device Trade Name
HSV Straw
#### Device Generic Name
Cryopreservation straw
### Device Classification and Product Code
Class II (Special Controls) per Final Rule, 63FR 48428, Docket number 97N-0335 21 CFR 884, Subpart G, 884.6160 Assisted Reproduction Labware
Product Code: MQK
#### Indications for Use
The HSV Straw is a cryopreservation storage device that is intended for use in vitrification procedures to contain and maintain human 4-8 Cell and blastocyst stage embrvos.
## Device Description and Principles of Operation for the HSV Straw
An embryo is vitrified using the HSV Straw by placing an embryo suspended in a 0.5ul All entrification solution solution in a curved polymeric spatula (Medical Grade. Styrene-Butadiene Copolymer). The curved spatula is inserted in an outer straw with a Styrene-Butadiene Coporymer). The open, proximal end of the straw is sealed by the user. The sealed, weighted unsal Chd. The open, promishar effect freezing-vitrification. The weighted straw seated straw is placed in liquid nitrogen for embryo storage. Thawing-rewarming: While the remains in liquid nitrogen, part of the outer tube is cut away. The curved spatial, distal tip tellians in inquir mirogen, part of the outer straw and immediately immersed in thawing solution where thawing and dilution of the remaining vitrification solution occur simultaneously.
{1}------------------------------------------------
| Attribute | HSV StrawTM,<br>CryoBio System | Curently marketed,<br>previously FDA cleared,<br>cryopreservation device.<br>CryoTipTM, Irvine<br>Scientific, Irvine CA | Curently marketed,<br>previously FDA cleared,<br>cryopreservation device.<br>Rapid-iTM, Vitrolife<br>Sweden AB, Gothenberg,<br>Sweden |
|--------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) # | K092398 | K041562 | K090832 |
| Indications for use | The HSV Straw is a<br>cryopreservation device<br>designed to contain, vitrify and<br>maintain 4-8 cell and blastocyst<br>stage human embryos. | The CryotipTM is a<br>cryopreservation device that is<br>intended to be used to contain,<br>freeze and maintain oocytes<br>and/or embryos. | Rapid-iTM is a cryopreservation<br>device indicated to be used to<br>contain, vitrify and maintain 4-8<br>cell stage embryos. |
| Method of action<br>(vitrification) | An embryo, suspended in a<br>0.5ul drops of the final<br>vitrification solution, is placed in<br>the curved spatula. The curved<br>spatula is inserted in the outer<br>straw and the open end of the<br>straw sealed. The sealed<br>straw is placed in liquid<br>nitrogen. | Embryo samples suspended in<br>the final vitrification solution are<br>aspirated into a capillary tube<br>and seals are effected on each<br>side of the sample.<br>Subsequently, the tube is<br>immersed in liquid nitrogen. | Prefreeze a straw with the<br>open end extending from the<br>liquid nitrogen. A 10 nanoliter<br>drop of vitrification solution<br>holding an embryo is placed in<br>a capillary sized through hole in<br>the stick. This, in turn, is<br>inserted in the pre frozen straw<br>in liquid nitrogen to effect<br>freezing of the embryo.<br>Subsequently, the open end of<br>the straw is sealed. |
| Cooling Rate1. See<br>note | 2,900°C/min | 12,000°C/min | 1,220°C/min |
| Method of action<br>(rewarming) | While the distal tip of the outer<br>tube remains in liquid nitrogen,<br>part of the outer tube is cut<br>away. The curved spatula,<br>containing the cryopreserved<br>embryo, is immersed in<br>thawing solution where thawing<br>and dilution in the thawing<br>solution occur simultaneously. | The capillary tube with sample<br>is removed from liquid nitrogen<br>and placed in a temperature<br>controlled solution. After<br>thawing, the tube is cut open<br>and the embryo and vitrification<br>solution are washed out into a<br>dilution solution. | While the distal end of the<br>straw remains in liquid<br>nitrogen, cut the sealed<br>proximal end of the straw.<br>Withdraw the stick and directly<br>immerse the stick/vitrified drop<br>in warming media. |
| Rewarming Rate1<br>See note | 25,000°C/min | 41,098°C/min | 7,700°C/min |
| Materials in contact<br>with tissue<br>(embryos) | Medical Grade Styrene-<br>Butadiene Copolymer. | Medical grade polyvinyl<br>chloride straw. | Polymethyl methacrylate<br>(PMMA) stick and a poly vinyl<br>chloride (PVC) straw. |
| Warming: Contact<br>with the warming<br>medium | Direct immersion of sample in<br>the warming solution for<br>simultaneous thawing and<br>dilution. | The sealed straw is first<br>immersed in solution and<br>warmed/thawed. In a second<br>step, the thawed sample is<br>expelled and diluted in the<br>warming solution. | Direct immersion of sample in<br>the warming solution for<br>simultaneous thawing and<br>dilution. |
Technological Characteristics of the HSV Straw Compared to Predicate Devices
Note: The cooling/heating performance of the CryoTip and HSV devices were determined by laboration in Note: The coolingheating performance of the Cryo rp and 1139 device was interest the publication.
with microminature temperature monitoring. Data for the Rapid-I device used with microminature temperature montioning. Data for na map purification and Warming Embryos Using
Tarankonov et al, 2009, Numerical Simulations Tarankonov et al, 2007, Numerical Unitations Deticles of the Conference 2009, Milan.
the Rapid-i™ Device, Proceedings of the COMSOL Conference 2009, Milan.
Premarket Notification Summary, K092398 Page 2 of 3
{2}------------------------------------------------
#### Substantial Equivalence to Marketed Products
Predicate devices:
Cryotip, Irvine Scientific, Irvine CA, K041562
Rapid-i, Virtrolife Sweden AB, K090832
Clinical use of the HSV Straw was studied in three centers representing diverse approaches to assisted fertility (blastomeres vs blastocysts, multiple embryo transfers vs single embryo transfers, hormone induced cycles vs natural cycles, etc). A total of 473 vitrification cycles using 1509 embryos were captured in the study. Each of the three centers had post vitrification/rewarming embryo % survival of greater than 80%. No device associated adverse effects were experienced by the study sites and the centers continue to routinely use the device. The three centers had high levels of embryo survival after thawing and successfully used the vitrified embryos for pregnancies and live births.
The submission contains information demonstrating substantial equivalence between the devices. Test data in the submission include mouse embryo assay (MEA), endotoxin and sterility data, mechanical testing, and heating and cooling rate data. MEA testing with mouse blastocysts at 96 hours yielded 90-100% survival and endotoxin testing consistently demonstrated <1 EU per unit. Mechanical testing demonstrated the performance of the device under use conditions.
Design information included in the submission demonstrated that the curved form of the spatula, surface tension of the embryo containing drop and device dimensions and configuration provide a safe environment whereby the embryos are protected from nitrogen contamination inside a sealed straw.
When the HSV Straw is used in accordance with indications and instructions for use, the differences between the HSV Straw and predicate devices do not raise new questions of safety and effectiveness. The clinical studies of the HSV Straw and engineering and other tests did not identify any adverse effects. This conclusion is based on the performance data and specifications. It can be concluded that the intended use, material composition and scientific technology properties of the HSV Straw demonstrate that the device is as safe, as effective, and performs as well as or better than the identified, legally marketed predicate devices in the United States.
> Premarket Notification Summary, K092398 Page 3 of 3
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features an eagle-like bird with three lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002.
Cryo Bio System % Mr. David Balding Consultant 26552 Tampico Place MISSION VIEJO CA 92691
6 2010
Re: K092398 Trade Name: HSV Straw Regulation Number: 21 CFR \$884.6160 Regulation Name: Assisted reproduction labware Regulatory Class: II Product Code: MOK Dated: September 17, 2010 Received: September 21, 2010
Dear Mr. Balding:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related
{4}------------------------------------------------
Page 2
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Hubert Lenoir MD
Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health.
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K092398
Device Name: HSV Straw
Indications for Use: The HSV Straw is a cryopreservation storage device that is intended for use in vitrification procedures to contain and maintain human 4-8 Cell and blastocyst stage embryos.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Signature
n Sign-Off) of Reproductive, Gastro-Re · Page of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.