LUCON M-SERIES PATIENT MONITOR, MODELS M20 AND M30
Applicant
Mediana Co., Ltd.
Product Code
MHX · Cardiovascular
Decision Date
Jun 8, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Lucon M-series (M20, M30) is intended to be used to monitor for adult, pediatric and neonatal patients in all areas of a hospital and hospital-type facilities. Monitor users should be skilled at the level of a technician. The device is capable of monitoring: Electrocardiography (ECG) Heart rate (HR) Noninvasive blood pressure (NIBP) Functional arterial oxygen saturation (SpO2) Pulse rate (PR) Respiration rate (RR) Temperature (Temp) Arrhythmia/ST segment (M30 only) Capnography (M30 only) Note: Hospital use typically includes such areas as general care floors, operating rooms, special procedure areas, intensive and critical care area, within the hospital. Hospital type facilities include physician office-based facilities, sleep labs, skilled nursing facilities, surgical centers, and sub acute care centers. Note: The medically skilled and trained user can be clinicians like doctors and nurses who know to take and interpret a patient's vital signs. These clinicians must take direct responsibility for the patient's life. This can include care-givers or medically trained interpreters who are authorized under the appropriate clinical facility procedures to support patient care. Any inappropriate setting, especially the alarm limit or alarm notification settings, can lead to a hazardous situation that injures the patient, harms the patient, or threatens the patient's life. This equipment should only be operated by trained users who can adjust the settings of the patient monitor.
Device Story
Lucon M-series (M20, M30) is a compact, battery/AC-powered patient monitor for hospital/clinical use. Inputs include ECG leads, NIBP cuff, SpO2 sensor, temperature probe, and capnography sensor (M30). Device processes physiological signals to display real-time vital signs, arrhythmia, and ST-segment data. Operated by clinicians/technicians to monitor patient status; output informs clinical decision-making and alarm notifications. Benefits include continuous patient surveillance and early detection of physiological instability.
Clinical Evidence
Bench testing only. System verification and validation performed per plan #MDR-YW071202-02 using production-equivalent units. Risk analysis and mitigation documented. Quality system certified to 21CFR820, ISO 9001, and ISO 13485.
Indicated for adult, pediatric, and neonatal patients in hospital and hospital-type facilities (physician offices, sleep labs, skilled nursing, surgical centers, sub-acute care) for monitoring ECG, HR, NIBP, SpO2, PR, RR, Temp, and (M30 only) Arrhythmia/ST segment and Capnography. Requires operation by medically skilled/trained clinicians or technicians.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Predicate Devices
Omron Health care Co.,Ltd. Vital Signs Monitors, Model HBP-2070 (K082812)
Philips Medical Systems Co.,Ltd. Patient Monitors, SureSigns VM6 series (K080495)
Welch Allyn Protocol Inc. Cardiac Monitors, Propaq Encore 200 series (K012451)
Spacelabs Healthcare Inc. Vital Signs Monitors, Model 91220 (K062095)
{0}------------------------------------------------
JUN - 8 2010
# 510(k) Summary of Safety and Effectiveness
### Submitter
| Mediana Co.,Ltd. | |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Dongwha Medical Instrument Complex, 1650-1, Donghwa-ri, Munmak-eup, |
| Wonju-si, Gangwon-do, Korea | |
| Tel) +8270-7092-9983 | Fax) +822-542-7447 |
| Company Contact: | Amy M.H.KIM, Manager of Regulatory Affairs<br>Won W.N.LEE, Specialist of Regulatory Affairs |
| Date Summary Prepared: | November 2nd, 2009 |
| Device Name | |
| Trade Name: | Lucon M-series (M20, M30) patient monitor |
| Common Name: | Patient Monitor |
| Classification Name: | Cardiac Monitor (21CFR870.2300) also contains<br>Monitor, Physiological, Patient (21CFR870.2300)<br>(including arrhythmia detection or alarms),<br>Arrhythmia Detector and alarm (21CFR870.1025)<br>(including ST-segment measurement and alarm),<br>NIBP measurement system (21CFR870.1130),<br>Electrocardiograph (21 CFR870.2340) |
### Class II Classification:
# Predicate Devices (Legally Marketed Devices)
The predicate devices for Vital signs monitors, Model Lucon M-series (M20, M30) are:
Non-invasive pulse oximetry, SpO2 (21CFR870.2700) and Clinical electronic thermometer (21CFR870.2910)
- Omron Health care Co.,Ltd. Vital Signs Monitors, Model HBP-2070 cleared by FDA through 510(k) No. K082812, and
- Philips Medical Systems Co.,Ltd. Patient Monitors, SureSigns VM6 series . cleared by FDA through 510(k) No. K080495, and
- Welch Allyn Protocol Inc. Cardiac Monitors, Propaq Encore 200 series . cleared by FDA through 510(k) No. K012451, and
- Spacelabs Healthcare Inc. Vital Signs Monitors, Model 91220 . cleared by FDA through 510(k) No. K062095.
### Device Description
The Mediana Lucon M-series (M20, M30) patient monitor is a lightweight and compact device (250 × 210 × 170 (mm) and 3.2 kg) powered by AC mains (100-240VAC, 50-60Hz) and also powered by internal battery. The monitor provides patient data and monitoring status on TFT-LCD displays.
- Electrocardiography (ECG) i
- Heart rate (HR) .
- Noninvasive blood pressure (NIBP) .
- Functional arterial oxygen saturation (SpO2)
- Pulse rate (PR)
- Respiration rate (RR) .
- Temperature (Temp)
- Arrhythmia/ST segment (M30 only) .
- Capnography (M30 only)
{1}------------------------------------------------
## Intended Use
The Lucon M-series (M20, M30) is intended to be used to monitor for adult, pediatric and neonatal patients in all areas of a hospital and hospital-type facilities. Monitor users should be skilled at the level of a technician. The device is capable of monitoring:
- Electrocardiography (ECG)
- Heart rate (HR)
- Noninvasive blood pressure (NIBP) ı
- Functional arterial oxygen saturation (SpO2)
- Pulse rate (PR) .
- Respiration rate (RR) .
- Temperature (Temp)
- Arrhythmia/ST segment (M30 only)
- Capnography (M30 only)
- Note: Hospital use typically includes such areas as general care floors, operating rooms, special procedure areas, intensive and critical care area, within the hospital. Hospital type facilities include physician office-based facilities, sleep labs, skilled nursing facilities, surgical centers, and sub acute care centers.
- Note: The medically skilled and trained user can be clinicians like doctors and nurses who know to take and interpret a patient's vital signs. These clinicians must take direct responsibility for the patient's life. This can include care-givers or medically trained interpreters who are authorized under the appropriate clinical facility procedures to support patient care. Any inappropriate setting, especially the alarm limit or alarm notification settings, can lead to a hazardous situation that injures the patient, harms the patient, or threatens the patient's life. This equipment should only be operated by trained users who can adjust the settings of the patient monitor.
# Summary of Technical Characteristics of the Device Compared to the Predicate Devices (Legally Marketed Devices)
The Mediana patient monitors, Model Lucon M-Series (M20, M30) are substantially equivalent to the HBP-2070 of Omron Healthcare(No. K082812), SureSigns VM6 series of Philips Medical Systems (No. K080495), and Propag Encore Model 200 series of Welch Allyn Inc (No. K012451). and mCARE 300, Model 91220, Vital Signs Monitor of Spacelabs Medical(No. K062095).
- The Lucon M-series (M20, M30) monitors use the Mediana ECG module, MDE-1 which is identical to the module used in the HBP-2070. The Electrocardiegram(ECG) and heart rate specifications for the Lucon M-series (M20, M30) and HBP-2070 are identical. The ECG measurement performance in VM6 is similar to those in the Lucon M-series (M20, M30) monitors. Because those devices are used identical technology to measure the ECG parameter,
- The Lucon M-series (M20, M30) monitors use the Omron NIBP module M3200 which is identical to the module used in the HBP-2070. The Non-invasive blood pressure (NBP) specifications for the Lucon M-series (M20, M30) and HBP-2070 are identical. And these devices and SureSigns VM6 series comply with the AAMI performance standard SP-10 & IEC60601-2-30 and have same technology to measure the NIBP parameter,
{2}------------------------------------------------
- The Lucon M-series (M20, M30) monitors use the Nellcor Sp02 module, NELL-3 which is used in the mCARE 300, 91220 and HBP-2070. The Pulse Oximetry (SpO2) specifications for the model Lucon M-series (M20, M30), HBP-2070 and 91220 are identical. Also, all devices comply with the performance standard ISO9919.
- The Respiration can be obtained from either the Electrocardiogram (ECG) channel or the Capnography (EtCO2) channel. The algorithm of Impedance measurement in HBP-2070 is identical to that in Lucon M-series (M20, M30) and the respiration specification for the Lucon Mseries (M20, M30) and HBP-2070 are identical Because the respiration measurement (Thoracic Impedance) for HBP-2070 and 91220 are from ECG module that are the same module designed and manufactured by Mediana Co.,Ltd.,
- The Capnography (EtCO2) module is installed to Lucon M-series (M30 only), The algorithm of Airway Measurement in 91220 has identical performance of Lucon M-series (M30 only). Because Airway Measurement is from Capnography module which is manufactured by Philips Respironics Inc. The operation theory of Airway Measurement in Propaq Encore 200 series is same as those in Lucon M-series (M30 only) and 91220 for respiration measurement,
- The Lucon M-series (M20, M30) monitors use the thermometry MDT-1 module which is identical to used in HBP-2070 and 91220. The Temperature measurement technology is equivalent for the model Lucon M-series (M20, M30), HBP-2070 and 91220. Also YSI 400 and 700 probes/probe covers are used for all of these device,
- The Pulse rate derived from the non-invasive blood pressure (NIBP) channel is from M3200 NIBP module in Lucon M-series (M20, M30). Also, this M3200 module is used for HBP-2070 . And the pulse rate derived from the Pulse Oximetry (SpO2) channel is from NELL-3 SpO2 modules in Lucon M-series (M20, M30). Also, this NELL-3 module is used for HBP-2070 and 91220 too.
- The Lucon M-series (M30 only) monitors use the Capnostat 5 and Lo Flo C5 which have already been registered in FDA by Philips Respironics Inc. (#K042601 is for Capnostat 5, #K053174 is for Lo Flo C5). The Capnography measurement module in Lucon M-series (M30 only), is from Respironics Capnography module, both modules are identical to the modules used in the 91220. Also the difference of Propaq Encore 200 series in Capnography specification is minor between the identified predicate devices for all measurements. All 3 devices have identical technology and operation of theory for Capnography,
- The model Lucon M-series (M20, M30) monitors as well as the predicate devices have internal power source, a rechargeable lead acid battery and AC power.
{3}------------------------------------------------
# Summary of Performance Testing
The Mediana vital signs monitors, Model Lucon M-series (M20, M30) substantially have been tested The Mediance with the system V & V plan (#MDR-YW071202-02) and summary included with the submission using production equivalent units prior to release to market.
A risk analysis identifying potential hazards and documenting mitigation of the hazards has been A risk analysis lochitrying potential nazards and documenting andigations's quality system confirms to 21CFR820, ISO 9001, ISO13485 and CMDCAS ISO 13485 certified by DNV (Det Norske Veritas).
## Conclusions
As stated above, the Mediana patient monitors, Model Lucon M-Series (M30, M20) are safe and I its survey and comply with the appropriate medical device standards and are substantially equivalent to the earlier identified predicate devices.
- End of Section -
{4}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle symbol. The logo is in black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mediana Co., Ltd. c/o Mr. Charles Mack Principal Engineer International Regulatory Consultants, LLC 77325 Joyce Way Echo, OR 97826
JUN - 8 2010
Re: K100217
Trade/Device Name: Lucon M-series (M20, M30) Patient Monitor Regulatory Number: 21 CFR 870.1025 Regulation Name: Patient Physiological Monitor (with arrhythmia detection or alarms) Regulatory Class: II (two) Product Code: MHX Dated: May 15, 2010 Received: May 19, 2010
Dear Mr. Mack:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Charles Mack
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices . Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# INDICATIONS FOR USE
Applicant: Mediana Co.,Ltd. Dongwha Medical Instrument Complex, 1650-1, Donghwa-ri, Munmak-eup, Wonju-si, Gangwon-do, Korea Telephone: (82) 70 7092 9983 Fax: (82) 33 742 5498
510(k) Number: : _
Device Name: Lucon M-series (M20,M30), patient monitor
### Indications for Use:
The Lucon M-series (M20, M30) is intended to be used to monitor for adult, pediative mass about The Lucon M-Scrites (M20, NE07 is interest and hospital-type facilities. Monitor users should be skilled at the level of a technician. The device is capable of monitoring:
- · Electrocardiography (ECG)
- · Heart rate (HR)
- · Noninvasive blood pressure (NIBP)
- · Functional arterial oxygen saturation (SpO2)
- · Pulse rate (PR)
- · Respiration rate (RR)
- · Temperature (Temp)
- · Arrhythmia/ST segment (M30 only)
- · Capnography (M30 only)
- Note: Hospital use typically includes such areas as general care filosos, operating rooms, special procedure areas, intensive and critical care area, within the hospital. Hospital Special procedure arous, intonian office-based facilities, sleep labs, skilled nursing facilities, surgical centers, and sub acute care centers.
Note: The medically skilled and trained user can be clinicians like doctors and nurses who I he medically skilled and interpret a patient's vital signs. These clinicians must take direct Know now to take and micrpret a pattent's life. This can include care-givers or medically trained interpreters who are authorized under the appropriate clinical facility procedures to interpreters who are authorized under the uppropriate setting, especially the alarm support pattent care. Ally mappropriate setting, coposially and injures the patient, harms the notification settings, can read to a nazzingous should only be operated by trained users who can adjust the settings of the patient monitor.
(Please do not write below this line - continue on another page if needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|------------------------------------|---------|
| Division of Cardiovascular Devices | |
| 510(k) Number | K060214 |
| Prescription Use (Per 21CFR801.109) | <div style="text-align:center;">X</div> | OR Over-The-Counter |
|-------------------------------------|-----------------------------------------|---------------------|
|-------------------------------------|-----------------------------------------|---------------------|
- End of Section -
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.