K112148 · Valmed, SA · IPF · Apr 30, 2012 · Physical Medicine
Device Facts
Record ID
K112148
Device Name
POWERED MUSCLE STIMULATOR
Applicant
Valmed, SA
Product Code
IPF · Physical Medicine
Decision Date
Apr 30, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Indications for EMS use are: - Prevention or retardation of muscle disuse atrophy - Relaxation of muscle spasms - Increasing local blood circulation - Muscle re-education - Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis - Maintaining or improving range of motion of extremities Indications for TENS use are: - Symptomatic relief and management of chronic, intractable pain, and - Adjunctive treatment for post-surgical and post-trauma acute pain
Device Story
Battery-powered (4x AAA) transcutaneous nerve and neuromuscular stimulator; delivers synchronous, low-voltage, monophasic rectangular pulses via two independent channels. Operated by medical professionals or patients under supervision; user selects from preset factory programs (4 EMS, 10 TENS) and adjusts intensity. Device features automatic current density limitation to prevent skin burns; no user-accessible controls for frequency or pulse duration. Used in clinical or home settings to mitigate pain and support rehabilitation. Output affects muscle contraction or nerve stimulation; benefits include atrophy prevention, spasm reduction, and pain management. Safety features include minimal electrical charge (<50 microcoulombs) and connector design preventing AC power connection.
Clinical Evidence
Bench testing only. Performance verified via oscilloscope analysis of output waveforms (voltage/current) across varying loads (100-10,000 ohms) and electromagnetic compatibility (EMC) testing per EN 60601-1-2 and EN 60601-2-10 standards. No clinical trials conducted.
Technological Characteristics
Battery-powered (4x AAA) stimulator; plastic enclosure. Output: monophasic rectangular pulses, 1-120 Hz, 20-1000 us duration, max 50V/100mA. Features automatic current density control. Complies with IEC 601-2-10, ISO 9001, ISO 13485. Connectivity: standalone, no AC power. Sterilization: N/A (non-sterile).
Indications for Use
Indicated for patients requiring muscle stimulation for atrophy prevention, spasm relaxation, blood circulation, re-education, post-surgical calf stimulation, or range of motion maintenance. Also indicated for symptomatic relief of chronic intractable pain and adjunctive treatment for post-surgical/post-trauma acute pain.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
APR 3 0 2012
'age
<112148
# A. 510(k) Summary as Required by Section 807.92(c)
### 1. Owner
Valmed, S.A.
### 2. Address
Avenue du Tourbillon 34, Sion 1950, Valais, Switzerland
### 3. Phone
+41 27 203 65 81
### 4. Fax
+41 27 203 65 87
### 5. Website
www.valmed.ch
### 6. Contact Persons
Switzerland: Vincent Tellenbach - Vincent.tellenbach@valmed.ch United States: Carl Magnell comagnell@gmail.com
### 7. Date Prepared/Finalized
22 July 2011/Finalized 23 April 2012
### 8. Name of Device
SWISSSTIM Trade Name: Battery-powered device for transcutaneous nerve and Common Name: neuromuscular electro-stimulation
Classification Name: Panel/Product Code/Regulation: 89 IPF CFR 890.5850 and subsequent Panel/Product Code/Regulation: 88 GZJ, 882.5890
### 9. Predicate Devices
This device is substantially equivalent to the following legally marketed devices:
P4-PHYSIO (K022175), and Styline 882 (K092448)
{1}------------------------------------------------
Page 2
### 10. Indications for Use
The Indications for Use for the SWISSSTIM are shown below:
Indications for EMS use are:
- > · Prevention or retardation of muscle disuse atrophy
- Relaxation of muscle spasms
- Increasing local blood circulation
- Muscle re-education
- > Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis
- Maintaining or improving range of motion of extremities
Indications for TENS use are:
- > Symptomatic relief and management of chronic, intractable pain, and
- > Adjunctive treatment for post-surgical and post-trauma acute pain
### 11. Device Summary
SWISSsme is a battery-powered device for transcutaneous nerve and neuromuscular electro-stimulation. It has two channels (outputs) of stimulation, each with independent intensity control. The unit delivers synchronous, low voltage stimulation impulses at both of its outputs. The stimulation output voltage waveforms are in form of trains of monophasic, rectanqular impulses. Such waveform character is true, when the outputs are connected to anv resistor in a range from 100 Ω to 10,000 Ω.
The EMS mode of this device has four different stimulation programs. The medical professional or patient under medical supervision can select any of these four stimulation programs, as appropriate for their treatment objectives. For each stimulation program, the medical professional or patient selects the desired phases of the stimulation programs as well as the stimulation duration; the impulse frequencies and impulse durations are different and all are preset at the factory.
There are no user accessible controls that can change either the frequency or impulse duration during a treatment. The user accessible controls are the ON/OFF button, the SELECT button and the two (2) buttons that control menu access and intensity levels for the 2 independent channels. The stimulus is a low voltage, low frequency, rectangular waveform.
The TENS mode of this device has ten (10) programs. The TENS mode is what has been added to this device and therefore modifies the Indications for Use from the predicate device, K022175. The TENS mode is controlled in the same fashion as the EMS mode of the SWISSSTIM.
{2}------------------------------------------------
The device is housed in a plastic enclosure and is powered by four (4) AAA batteries. Accessories include output cables, batteries, User's Manual, storage case and electrodes. There is no provision for an AC adapter and the unit cannot be connected to any AC electrical power circuit.
From a technical perspective, the SWISSsma delivers stimulation pulses of 50v (peak value) ± 10% onto 500Ω at maximum settings, which corresponds to 100 mA (peak value) current intensity. The impulse frequencies produced by the device are in a range of 1 Hz to 120 Hz, depending on the choice of stimulation program. The voltage waveforms are rectangular monophasic with duration time from 20 us to 1000 us. The current impulses are monophasic on 500 Ω load. The tetanization stimulation signal includes ramp, which slowly increases the contraction intensity during treatments.
The EMS function is used as a therapeutic tool, specifically for rehabilitation purposes as noted in the Indications for Use (please see Section B). An example is prevention of disuse muscle atrophy that may occur after musculoskeletal injuries such as damage to bones, joints, muscles, ligaments and tendons. The TENS functions address pain mitigation, whether chronic or acute pain; TENS is a non-invasive, safe nerve stimulation intended to reduce such pain. A number of systematic reviews or metaanalyses have confirmed TENS effectiveness for postoperative pain, osteoarthritis, and chronic musculoskeletal pain.
### 12. Technological Summary
The SWISS TIM technical characteristics are similar to the predicate devices in design. materials and energy source. A summary of these characteristics is provided below:
Image /page/2/Figure/6 description: The image shows an oscilloscope display with a waveform. The waveform shows a voltage signal that is initially high, then drops to a lower level. The x-axis is labeled "M 100us", indicating a time scale of 100 microseconds per division, and the y-axis is labeled "20.0U/DIV", indicating a voltage scale of 20 volts per division. There are also some menu options on the right side of the screen.
SwissStim Voltage IVI
Load: 500 ohms
Image /page/2/Figure/8 description: The image shows an oscilloscope display with a waveform. The waveform has a sharp rise, a slight decay, and then a sharp drop, followed by a flat line. The vertical scale is 20.0U/DIV, and the horizontal scale is M 100 μs, indicating the voltage and time scales of the waveform.
P4 Physio Voltage [V]
{3}------------------------------------------------
# K112148 PAGE 4 OF 12
Image /page/3/Figure/1 description: The image shows four graphs, each displaying data from a LeCroy device. The top left graph is labeled "Hivex Stiline Voltage [V]", while the top right graph is labeled "Comparison of All Devices". The bottom left graph is labeled "SwissStim Current [mA]", and the bottom right graph is labeled "P4 Physio Current". All four graphs have a similar grid layout and display waveforms, with a scale of 20.0U/DIV and M 100μs.
Hivox Styline Voltage [V]
# Hivox Styline Current [mA]
M 1991
50mA/DIU
Image /page/3/Figure/16 description: The image shows an oscilloscope display with a single pulse waveform. The waveform transitions from a low state to a high state and then returns to the low state. The horizontal scale is labeled as "M 100us", indicating a time scale of 100 microseconds per division, while the vertical scale is labeled as "50mA/DIV", indicating a current scale of 50 milliamperes per division. The oscilloscope is a LeCroy brand.
# SwissStim Voltage [V]
Image /page/3/Figure/5 description: The image shows an oscilloscope display with a square wave signal. The vertical scale is set to 20.0V/DIV, and the horizontal scale is set to M 100μs. The square wave transitions from a low voltage level to a high voltage level and then back down, indicating a pulse or switching event. The right side of the image shows menu options such as Type, Image, Touche Imo, and Enredistre.
# Comparison of All Devices Voltage [V]
Image /page/3/Figure/7 description: The image shows an oscilloscope display with a graph plotted on it. The graph shows a signal that rapidly rises to a high level, remains there for a short period, and then quickly drops to a lower level. The display also shows text labels such as "Type", "Image", "Enr.Image", and "Enregistrer", along with the settings "20.0U/DIV" and "M 100μs".
### P4 Physio Current [mA]
Image /page/3/Figure/9 description: The image shows a screenshot of an oscilloscope display. A square wave is displayed on the grid, showing a transition from a low to a high state and then back to a low state. The vertical scale is labeled as 50mA/DIV, and the horizontal scale is labeled as M 100us. The right side of the image shows a menu with options such as "Type", "Image", and "Enregistrer".
### Comparison of All Devices Current [mA]
Image /page/3/Figure/11 description: The image shows a graph with a square wave. The x-axis is labeled "M 100us" and the y-axis is labeled "50mA/DIV". The square wave starts at 0, rises to approximately 200mA, stays there for about 200us, and then drops back to 0.
# Load: 2000 ohms
P4 Physio Voltage [V]
Image /page/3/Figure/14 description: The image shows an oscilloscope display. The display shows a waveform with a sharp rise and fall, followed by a period of relatively constant voltage. The text on the screen indicates that the vertical scale is 20.0U/DIV and the horizontal scale is M 100μs. There are also menu options visible on the right side of the screen, including "Type Image", "Touche.Impr", and "Enregistrer".
{4}------------------------------------------------
### Hivox Styline Voltage [V]
Image /page/4/Figure/2 description: The image shows an oscilloscope display with a square wave signal. The signal transitions from a high voltage level to a low voltage level. The vertical scale is set to 20.0V per division, and the horizontal scale is set to 100 microseconds per division. The image also shows a menu on the right side with options such as 'Type', 'Image', and 'Enregistrer'.
SwissStim Current [mA]
| Image: LeCroy logo and icons | | | |
|------------------------------|--|--|-------------|
| | | | Type |
| | | | Image |
| | | | Touche.Impr |
| | | | Enr.Image |
| | | | Enregistrer |
| 50mA/DIV | | | M 100μs |
Hivox Styline Current [mA]
Image /page/4/Figure/6 description: The image shows a screenshot of an oscilloscope display. The display shows a single waveform that transitions from a low level to a high level and then back to a low level. The horizontal scale is labeled as "M 100us", indicating a time scale of 100 microseconds per division, and the vertical scale is labeled as "50mA/DIV", indicating a current scale of 50 milliamperes per division.
# Comparison of All Devices Voltage [V]
Image /page/4/Figure/8 description: The image shows an oscilloscope display with a graph plotted on it. The text on the screen indicates settings of "20.0U/DIV" and "M 100ps", suggesting the vertical scale is 20 units per division and the horizontal scale is 100 microseconds. On the right side of the screen, there is a menu with options like "Type", "Image", "Touche Imp", and "Enr.Image".
P4 Physio Current [mA]) Type lmage Touche Imp Enrimade Enteristie 50mA/DIU M 100 ps
# Comparison of All Devices
Current [mA]
Image /page/4/Figure/12 description: The image shows a graph with a horizontal line that has a step change. The x-axis is labeled "M 100μs", and the y-axis is labeled "50mA/DIV". On the right side of the image, there is a menu with the options "Type", "Image", "Touche Impr", and "Enriman". The graph shows a signal that is initially at a higher level, then drops to a lower level.
{5}------------------------------------------------
# K112148 PAGE 6 OF 12
# Load: 10000 ohms
# SwissStim Voltage [V]
Image /page/5/Figure/3 description: The image shows an oscilloscope display with a square wave signal. The vertical scale is set to 20.0V/DIV, and the horizontal scale is set to M 100 us. The signal transitions from a low state to a high state and then back to a low state. The LeCroy logo is visible in the upper left corner of the display.
# Hivox Styline Voltage [V]
| | 1200 @ hummunum | | | | |
|-----------|-----------------|--|---------|--|-------------|
| | | | | | Type |
| | | | | | |
| | | | | | · Image |
| | | | | | |
| | | | | | |
| | | | | | Touche Impr |
| | | | | | |
| | | | | | Enr.Image |
| | | | | | |
| | | | | | |
| | | | | | |
| | | | | | |
| | | | | | |
| | | | | | |
| | | | | | |
| | | | | | |
| | | | | | |
| | | | | | |
| | | | | | |
| | | | | | |
| | | | | | |
| | | | | | |
| | | | | | |
| | | | | | Enregistrer |
| | | | | | |
| | | | | | |
| 20.90/DIU | | | M 188ps | | |
# SwissStim Current [mA]
Image /page/5/Figure/7 description: The image shows an oscilloscope display with a signal trace. The signal appears to be a square wave with some noise present. The vertical scale is labeled as 5mA/DIV, and the horizontal scale is labeled as M 100us. There are also some menu options visible on the right side of the screen, including 'Type', 'Image', and 'Enregistrer'.
# P4 Physio Voltage [V]
Image /page/5/Figure/9 description: The image shows an oscilloscope display with a waveform. The oscilloscope is a LeCroy model. The vertical scale is set to 20.0U/DIV, and the horizontal scale is set to M 100 μs.
# Comparison of All Devices Voltage [V]
Image /page/5/Figure/11 description: The image shows a screenshot of an oscilloscope display. Two waveforms, labeled 1.1.1 and 1.1.2, are displayed on the grid. The vertical scale is set to 20.0V/DIV, and the horizontal scale is set to M 100μs.
Image /page/5/Figure/12 description: The image shows a graph of P4 Physio Current in mA. The graph shows a square wave with a period of approximately 200 microseconds. The y-axis is labeled 5mA/DIV, and the x-axis is labeled M 100ps. There is a menu on the right side of the image with options such as Type, Image, and Touche Imp.
P4 Physio Current [mA]
{6}------------------------------------------------
KII2148 PAGE 7 OF 12
Image /page/6/Figure/1 description: The image shows a graph of the Hivox Styline Current in milliamperes. The graph is displayed on a LeCroy oscilloscope. The x-axis is labeled as M 100us, and the y-axis is labeled as 5mA/DIV. The graph shows a current signal that is initially at a low level, then increases to a higher level for a short period of time, and then returns to the low level.
### Comparison of All Devices Current [mA]
Image /page/6/Figure/3 description: The image shows a graph with a line plot. The x-axis is labeled "M 100μs", and the y-axis is labeled "5mA/DIV". The graph shows a signal that rises sharply, plateaus for a short time, and then drops sharply again. There is also a menu on the right side of the image with options such as "Type", "Image", and "Enr.Image".
Other characteristics such as Indications for Use, safety, efficacy and operational parameters of the predicate devices and the SWISS TIM are shown below.
| Critical Parameter Comparison Table | | | |
|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Parameter | SWISSSTIM | Styline 882 | P4 Physio |
| | K112148 | K092448 | K022175 |
| Indications<br>for Use | EMS:<br>> Prevention or<br>retardation of<br>muscle disuse<br>atrophy<br>> Relaxation of<br>muscle spasms<br>> Increasing local<br>blood circulation<br>> Muscle re-<br>education<br>> Immediate post-<br>surgical stimulation<br>of calf muscles to<br>prevent venous<br>thrombosis<br>> Maintaining or<br>improving range of<br>motion of<br>extremities | EMS:<br>> Prevention or<br>retardation of<br>muscle disuse<br>atrophy<br>> Relaxation of<br>muscle spasms<br>> Increasing local<br>blood circulation<br>> Muscle re-<br>education and<br>strengthening<br>> Immediate post-<br>surgical stimulation<br>of calf muscles to<br>prevent venous<br>thrombosis<br>> Maintaining or<br>improving range of<br>motion of<br>extremities | EMS:<br>> Prevention or<br>retardation of<br>muscle disuse<br>atrophy<br>> Relaxation of<br>muscle spasms<br>> Increasing local<br>blood circulation<br>> Muscle re-<br>education and<br>strengthening<br>> Immediate post-<br>surgical stimulation<br>of calf muscles to<br>prevent venous<br>thrombosis<br>> Maintaining or<br>improving range of<br>motion of<br>extremities |
| | TENS:<br>> Symptomatic relief<br>and management<br>of chronic,<br>intractable pain,<br>and<br>> Adjunctive<br>treatment for post-<br>surgical and post- | TENS:<br>> Symptomatic relief<br>and management<br>of chronic,<br>intractable pain,<br>and<br>> Adjunctive<br>treatment for post- | |
| | trauma acute pain | | |
| Safety | 1- Complies with<br>ISO 9001/2000 and<br>ISO 13485/2003,<br>International Quality<br>Standard, IEC-601-<br>2-10 European<br>Safety Standard<br>and in compliance<br>with all EC medical<br>device directives<br>applicable to Class<br>IIA medical devices.<br>2 - Automatic<br>limitation of<br>stimulation current<br>density<br>3 - Embedded<br>programs cannot be<br>changed; users can<br>only modify the<br>intensity of<br>stimulation<br>4 - All programs<br>begin with minimal<br>electrical intensity;<br>the medical<br>professional or<br>patient must<br>increase the<br>intensity to desired<br>treatment level<br>5 - Maximum<br>electrical impulse<br>charge is 50 micro-<br>coulombs<br>6 - Automatic<br>stimulation current<br>density control<br>precludes excessive<br>current density at<br>the electrode-skin<br>interface, ensuring<br>skin safety<br>7- Lead cables<br>preclude the<br>possibility of<br>accidental<br>connection to a<br>power source such | 1- Complies with<br>ISO 9001/2000 and<br>ISO 13485/2003,<br>International Quality<br>Standard, IEC-601-<br>2-10 European<br>Safety Standard<br>and in compliance<br>with all EC medical<br>device directives<br>applicable to Class<br>IIA medical devices.<br>2 - No automatic<br>limitation of<br>stimulation current<br>density<br>3 - Users can<br>design individual<br>programs as well as<br>modify the intensity<br>of stimulation<br>4 - All programs<br>begin with minimal<br>electrical intensity;<br>the medical<br>professional or<br>patient must<br>increase the<br>intensity to desired<br>treatment level<br>5 - Maximum<br>electrical impulse<br>charge is not<br>provided<br>6 - No automatic<br>stimulation current<br>density control<br>7- Lead cables<br>preclude the<br>possibility of<br>accidental<br>connection to a<br>power source such<br>as an AC power<br>outlet<br>8 - Specific<br>warnings and<br>contraindications in<br>User Manual | 1- Complies with<br>ISO 9001/2000 and<br>ISO 13485/2003,<br>International Quality<br>Standard, IEC-601-<br>2-10 European<br>Safety Standard<br>and in compliance<br>with all EC medical<br>device directives<br>applicable to Class<br>IIA medical devices.<br>2 - Automatic<br>limitation of<br>stimulation current<br>density<br>3 - Embedded<br>programs can not<br>be changed; users<br>can only modify the<br>intensity of<br>stimulation<br>4 - All programs<br>begin with minimal<br>electrical intensity:<br>the medical<br>professional or<br>patient must<br>increase the<br>intensity to desired<br>treatment level<br>5 - Maximum<br>electrical impulse<br>charge is 23 micro-<br>coulombs<br>6 - Automatic<br>stimulation current<br>density control<br>precludes excessive<br>current density at<br>the electrode-skin<br>interface, ensuring<br>skin safety<br>7- Lead cables<br>preclude the<br>possibility of<br>accidental<br>connection to a<br>power source such |
Page 7
{7}------------------------------------------------
# K112148
PAGE 8 OF 12
း
వ
网
Page 8
{8}------------------------------------------------
# KI1214 PAGE 9 OF 12
| Efficacy | The EMS programs<br>are substantially<br>equivalent to the<br>EMS programs in<br>the P4 Physio.<br><br>The 10 TENS<br>programs are<br>designed to parallel<br>the TENS programs<br>in the predicate<br>device. There are<br>variations but these<br>relate to stimulation<br>patterns rather than<br>frequency ranges. | There are 10 EMS<br>and 10 TENS<br>programs | FDA cleared with 4<br>EMS programs that<br>have six (6)<br>Indications for Use,<br>including:<br>• Relaxation of<br>muscle spasms<br>• Prevention or<br>retardation of<br>disuse atrophy<br>• Increasing local<br>blood circulation<br>• Muscle re-<br>education<br>• Immediate post-<br>surgical stimulation<br>of calf muscles to<br>prevent venous<br>thrombosis, and<br>• Maintaining or<br>increasing range of<br>motion |
|-----------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Operation | Ease of Use: Only<br>selection of<br>programs, phases<br>and intensity levels<br>are possible;<br>intensity levels are<br>set independently<br>for each channel.<br><br>Pre-treatment<br>preparation and<br>post-treatment<br>instructions are the<br>same as those for<br>both predicate<br>devices. | Ease of Use: User's<br>have additional<br>options and can<br>select not only<br>programs and<br>intensity levels but<br>also program<br>duration time and<br>pulse rates (HZ).<br>These features<br>make this predicate<br>device more<br>complex to operate. | Ease of Use: Only<br>selection of<br>programs, phases<br>and intensity levels<br>are possible. |
### 13. Non-Clinical Performance Data
The SWISSsm device (and predicate devices) have also been analyzed in accordance with the AAMI/ANSI standard "AAMI NS4:1986/(R)2009: AAMI Standard for Transcutaneous Electrical Nerve Stimulators" Electrode/Skin Impedance Model (ESIM)
Image /page/8/Figure/4 description: The image shows a graph of SwissStim Voltage in Volts. The graph shows a square wave with a voltage of 20.0U/DIV and a time scale of M 100μs. The image also shows the LeCroy logo and some menu options such as Type, Image, Touche Impr, Enr.Image, and Enregistrer.
Image /page/8/Figure/5 description: The image shows the title of a graph or plot. The title reads "P4 Physio Voltage [V]". The title indicates that the graph or plot displays the voltage of a physiological signal, specifically from location P4, and that the voltage is measured in volts.
| | | Type |
|-----------|-----|-------------|
| | 14. | Image |
| | 15. | Touche.Impr |
| | 16. | Enr.Image |
| | 17. | |
| | 18. | |
| | 19. | |
| | 20. | Enregistrer |
| 20.00/DIV | | |
Page 9
{9}------------------------------------------------
K112|48 PAGE 10 OF 12
TVDE
Hivox Styline (Voltage [V], load: ESIM) LeCroy REGI promoming my Type louche Imr Enrecistre BUIDI P4 Physio Current [mA] 100 m 200 1 ype
Tited! 200mA/DIU M 100ps
SwissStim Current [mA]
BEEL Rama
Image /page/9/Figure/4 description: The image shows an oscilloscope display with a waveform. The waveform has a sharp positive peak followed by a negative peak, with the rest of the waveform close to zero. The display indicates a vertical scale of 200mA/DIV and a horizontal scale of M 100us.
Comparison of All Devices (Voltage [V], load: ESIM)
~ ��
Image /page/9/Figure/5 description: The image shows a graph from a LeCroy oscilloscope. The graph displays a waveform with a sharp initial peak followed by a rapid decay and then a more gradual decline to a steady-state level. The x-axis is labeled as "M 100us", indicating a time scale of 100 microseconds per division, while the y-axis is labeled as "200mA/DIV", indicating a current scale of 200 milliamperes per division.
HiVox Styline Current [mA]
Image /page/9/Figure/7 description: The image shows an oscilloscope display with two waveforms. The vertical scale is labeled as 200mA/DIV, and the horizontal scale is labeled as M 100us. The title at the bottom reads "Comparison of All Devices Current [mA]".
Image /page/9/Figure/8 description: The image shows the title of a graph or chart. The title is "Comparison of All Devices Current [mA]". The title indicates that the graph or chart will compare the current of all devices, measured in milliamperes (mA).
21.
{10}------------------------------------------------
Page 1 1
#### 14. Device Safety
The SWISSSTIM device is as safe and effective, and functions in substantially equivalent manner to the predicate devices cited in Section C9, above.
This device is manufactured in accordance with ISO 9001/2000 and ISO 13485/2003, International Quality Standard, IEC-601-2-10 European Safety Standard and in compliance with all EC medical device directives applicable to Class IIA medical devices.
Valmed, S.A. has never received notification of any adverse safety events related to any Valmed stimulator devices.
The SWISSsma is designed so that even improper or accidental application of the stimulator will not produce cardiac rhythm disturbances (this does not apply to persons who have implanted cardiac pacemakers). This safety factor is due to the minimal electrical charge of the stimulating impulses, which, under all conditions, do not exceed 50 microcoulombs. The effective value of the stimulation voltage a person will experience during treatment with the SWISSsTM is very low, below 7 volts (root mean square) at maximum setting.
A significant safety feature of the SWISSsm (and the P4 Physio) is automatic limitation of stimulation current density on the skin under the electrodes. (i.e., electrical energy delivered per square centimeter) Due to the voltage-source output circuitry, the SWISSSTIM presents minimal risk of skin burns. The skin current densities of Valmed stimulators are, at all times, well below the safe limit of 2 milliamperes (mA) per square centimeter, as required by the IEC 601-2-10 standard. The skin burn hazards caused by many electrostimulators have been widely reported in medical literature and by the FDA. These hazards are minimal in the SWISSsm due to the electronic circuit design incorporated in the output amplifier of the device (a voltage source generator similar to that used in implantable cardiac pacemakers).
Specific safety features include:
- a. It is not possible for users to modify the embedded programs; users can only modify the intensity of stimulation
- b. All programs begin with minimal electrical intensity; the medical professional or patient must increase the intensity to his/her desired treatment level.
- c. Maximum possible electrical impulse charge is 50 microcoulombs.
- d. Automatic control of stimulation current density, thus precluding excessive current density at the electrode-skin interface and ensuring skin safety.
- e. The connector plug precludes the possibility of accidental connection to a power source such as an AC power outlet.
- f. Specific warnings and contraindications in User Manual.
{11}------------------------------------------------
All device accessories also meet safety standards; 510(k) cleared electrodes are specified (K872976) and supplied with this device.
### Electromagnetic Compatibility and Electrical Safety 15.
FDA recommended testing has been completed, as follows:
| Test | Standards | | Result |
|---------------------------------------------------------------|--------------|----------------------------|--------|
| Emission | EN 55011 | EN 60601-1-2/EN 60601-2-10 | Pass |
| Immunity to Electrical<br>Discharge | EN 61000-4-2 | EN 60601-1-2/EN 60601-2-10 | Pass |
| Immunity to Radiated rf/emf | EN 61000-4-3 | EN 60601-1-2/EN 60601-2-10 | Pass |
| Immunity to Conducted<br>Disturbances Induced by rf<br>Fields | EN 61000-4-6 | EN 60601-1-2/EN 60601-2-10 | Pass |
| Immunity to Power<br>Frequency Magnetic Field | EN 61000-4-8 | EN 60601-1-2/EN 60601-2-10 | Pass |
{12}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/12/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal features the department's name arranged in a circular pattern around a stylized emblem. The emblem consists of three curved shapes, resembling a stylized caduceus or a representation of the flow of health and human services.
### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Valmed, S.A. % Mr. Carl Magnell Official Correspondent Ave du Tourbillion 34 1950 Sion, Valis Switzerland
APR 3 0 2012
Re: K112148
Trade/Device Name: Powered Muscle Stimulator Regulation Number: 21 CFR 890.5850 Regulation Name: Powered muscle stimulator Regulatory Class: Class II Product Code: IPF, GZJ Dated: April 16, 2012 Received: April 19, 2012
Dear Mr. Magnell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{13}------------------------------------------------
### Page 2 - Mr. Carl Magnell
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHQffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalToblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours,
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{14}------------------------------------------------
# A. Indications for Use
510(k) Number: K112148
Device Name: Powered Muscle Stimulator
Indications for EMS use are:
- Prevention or retardation of muscle disuse atrophy A
- Relaxation of muscle spasms
- A Increasing local blood circulation
- > Muscle re-education
- > Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis
- Maintaining or improving range of motion of extremities >
Indications for TENS use are:
- Symptomatic relief and management of chronic, intractable pain, and A
- Adjunctive treatment for post-surgical and post-trauma acute pain A
Prescription Use X (Part 21 CFR 801 Subpart D AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
### Concurrence of CDRH Office of Device Evaluation (ODE
signature
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K112148
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.