K092448 · Hivox Biotek, Inc. · GZJ · Mar 30, 2010 · Neurology
Device Facts
Record ID
K092448
Device Name
HIVOX ELECTRIC STIMULATOR TENS & EMS, HD2
Applicant
Hivox Biotek, Inc.
Product Code
GZJ · Neurology
Decision Date
Mar 30, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
For TENS programs: The device is an electrical nerve stimulator indicated for use for pain relief by applying an electrical current to electrodes on a patient's skin to treat pain. In particular, this device is indicated for use for (1) Symptomatic relief of chronic (long term) intractable pain, and (2) adjunctive treatment in the management of post-surgical and post-traumatic pain problems. For EMS programs: The device is an electrical powered muscle stimulator indicated for medical purpose to repeatedly contract muscles by passing electrical currents through electrodes contacting the affected body area. In particular, this device is indicated for use for (1) Relaxing muscle spasms, (2) Increasing local blood circulation, (3) Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis, (4) Muscle re-education, (5) Maintaining or increasing range of motion, and (6) Preventing or retarding disuse atrophy.
Device Story
HIVOX HD2 is a portable TENS/EMS stimulator; generates small electrical pulses delivered via lead cables to skin-surface electrodes. Operates by passing current through skin to activate underlying nerves (TENS) or induce muscle contraction (EMS). Used by patients or clinicians in clinical/home settings. Device features adjustable frequency (2-120.2 Hz) and fixed pulse width (250 µs). Powered by 3 AAA batteries. Provides symptomatic pain relief and muscle therapy. Healthcare providers use device output to manage pain and physical rehabilitation. Benefits include non-invasive pain management and muscle function maintenance.
Clinical Evidence
No clinical data. Bench testing only, including electrical safety, electromagnetic compatibility, and waveform testing.
Technological Characteristics
Portable TENS/EMS stimulator; DC 4.5V (3x AAA batteries). Constant voltage output. Max output 24.0V/48.0mA (@ 500 Ω). Fixed pulse width 250 µs; frequency 2-120.2 Hz. Plastic housing. Connectivity via lead cables to carbon silicon electrodes (45x45 mm). Standards: IEC/EN 60601-1, IEC/EN 60601-1-2, IEC/EN 60601-2-10.
Indications for Use
Indicated for patients requiring pain relief (chronic intractable, post-surgical, post-traumatic) or muscle stimulation (spasm relaxation, circulation improvement, venous thrombosis prevention, muscle re-education, range of motion maintenance, disuse atrophy prevention).
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
Predicate Devices
MEDIHIGHTEC Medical Co., Ltd. TENS/EMS Combo, EMS, TENS (K082514)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the word "HIVOX" in a bold, sans-serif font, followed by the registered trademark symbol. To the right of the word is a stylized graphic that resembles a plus sign or cross, composed of interconnected squares or blocks. The graphic has a three-dimensional appearance due to the shading on some of the blocks.
### HIVOX BIOTEK INC
8/F, No.98 Shingde Road Sanchong City Taipei 241 Taiwan ROC Fax: +886 2 8511 2669 Phone: +886 2 8511 2668
# SUMMARY OF SAFETY AND EFFECTIVENESS for TENS & EMS
#### 510(k): K092448
DATE OF SUBMISSION
i
MAR 3 0 2010
| SUBMISSION: | July 27, 2009 |
|-------------|----------------------------------------|
| SUBMITTER: | HIVOX BIOTEK INC. |
| | 8F, No.98, Shinde Road, |
| | Sanchong City, Taipei, 24158, TAIWAN |
| | TEL: 886-2-85112668 FAX:886-2-85112669 |
| | Email: ceirs.jen@msa.hinet.net |
#### ESTABLISHMENT REGISTRATION NO: 9611558
OFFICIAL
CONTACT:
Dr. JEN, KE-MIN NO 58, Fu-Chiun Street, Hsin-Chu City, 30067, TAIWAN TEL: 886-3-5208829 FAX:886-3-5209783 Email: ceirs.jen@msa.hinet.net
HIVOX Electric Stimulator TENS & EMS, HD2 TRADE NAME:
COMMON/USUAL Transcutaneous Nerve Stimulator & NAME: Electric Muscle Stimulator
Transcutaneous Nerve Stimulator CLASSIFICATION
NAME:
GZJ, Class II, 882.5890 REGULATION
NUMBER:
ID# Class II 800 59850
SECONDARY PRODUCT CODE: IPF, Class II, 890.5850
PREDICATED DEVICE:
MEDIHIGHTEC Medical Co., Ltd. TENS/EMS Combo, EMS, TENS K082514
{1}------------------------------------------------
| HIVOX<br>Image: [logo] | HIVOX BIOTEK INC | |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| | 8/F, No.98 Shingde Road Sanchong City Taipei 241 Taiwan ROC<br>Phone: +886 2 8511 2668 Fax: +886 2 8511 2669 | |
| Intended Use: | For TENS programs: The device is an electrical nerve stimulator<br>indicated for use for pain relief by applying an electrical current to<br>electrodes on a patient's skin to treat pain. In particular, this device<br>is indicated for use for (1) Symptomatic relief of chronic (long<br>term) intractable pain, and (2) adjunctive treatment in the<br>management of post-surgical and post-traumatic pain problems.<br>For EMS programs: The device is an electrical powered muscle<br>stimulator indicated for medical purpose to repeatedly contract<br>muscles by passing electrical currents through electrodes<br>contacting the affected body area. In particular, this device is<br>indicated for use for (1) Relaxing muscle spasms, (2) Increasing<br>local blood circulation, (3) Immediate post-surgical stimulation of<br>calf muscles to prevent venous thrombosis, (4) Muscle<br>re-education, (5) Maintaining or increasing range of motion, and<br>(6) Preventing or retarding disuse atrophy. | |
| Description of Device: | The HIVOX Electric Stimulator TENS & EMS, HD2 generates<br>small pulses of electrical current. Delivered along lead cables to<br>electrodes placed on your skin, these pulses pass through the skin<br>and activated underlying nerves. The relief from chronic and<br>acute pain that the HD2 can provide results from this electrical<br>stimulation. | |
| Performance Tests<br>Submitted: | The relevant standards including:<br>1. IEC/EN 60601-1 : Medical electrical equipment Part 1.<br>General requirements for safety, 1996.<br>2. IEC/EN 60601-1-2 : Medical electrical equipment, Part 2.<br>Electromagnetic compatibility - Requirements and tests, 2004.<br>3. IEC/EN 60601-2-10 : Medical electrical equipment Part 2.Safety<br>of Nerve and Muscle Stimulators, 1987. | |
| Non-Clinical Tests<br>Submitted: | The HIVOX Electric Stimulator TENS & EMS, HD2 has been<br>tested in accordance with applicable standards for medical device<br>electrical safety, electromagnetic compatibility, and the particular<br>requirements for safety of nerve stimulators.<br>Accessories also meet safety requirements: 510(k) electrodes are<br>specified, and the patient cable utilizes shrouded connectors to<br>meet lead wire safety requirements.<br>System level testing including waveform testing was performed in<br>combination the HIVOX Electric Stimulator TENS & EMS, HD2. | |
| Clinical Tests<br>Submitted: | None | |
| Conclusion: | As the product description and tests as above, the new device:<br>HIVOX Electric Stimulator TENS & EMS, HD2 is as safe and<br>effective as, and the function in a manner equivalent to the<br>predicate device: MEDIHIGHTEC TENS/EMS Combo, EMS,<br>TENS, K082514. | |
ﺗ
. .
devices in this aspect.
{2}------------------------------------------------
# HIVOXiff
HIVOX BIOTEK INC
8/F . No.98 Shingde Road Sanchong City Taipei 241 Taiwan ROC Fax: +886 2 8511 2669 Phone: +886 2 8511 2668
# Comparison for Predicate Device & Subject Device
| Comparison Areas | NEW DEVICE | PREDICATE DEVICE |
|-----------------------------------------------------|-------------------------------|--------------------------------------------|
| 510(k) number | K092448 | K082514 |
| Manufacturer | HIVOX Biotek Inc | Medihightec Medical Co. Ltd. |
| Device name | HD2<br>TENS & EMS | MH8000/6000<br>TENS/EMS Combo |
| Power source | DC 4.5V<br>(AAA Battery x 3 ) | One DC 9V battery |
| Channels<br>Synchronous<br>Reciprocal | SAME | 2<br>YES<br>YES |
| Computerized | SAME | No |
| Software provided | SAME | No |
| Constant current | No | YES |
| Constant voltage | YES | No |
| Max output voltage<br>(Vp ±20%) | 24.0V ±20% @ 500 Ω load | 20.4V±20% @500 Ω load<br>(40.8 Vp-p) |
| Max output current<br>(mA ±20%) | 48.0mA ±20% ( 500 Ω load ) | 40.8 mA±20% ( 500Ω load)<br>(81.6 mA, p-p) |
| Pulse Width ( µs ) | 250 µs (fixed) | 50~300 µs |
| Maximum Frequency | 2~120.2 Hz (adjustable) | 2~150 Hz (adjustable) |
| Max phase charge | □12.0 µC @ 500 ohms load | 24.0 µC @ 500 ohm |
| Maximum Current<br>Density | 0.59 mA/cm² @500Ω | 1.06 mA/cm² @500Ω |
| Maximum Power<br>Density | 0.014 W/cm² @500Ω | 2.4 m W/cm² @500Ω |
| Comparison Areas | NEW DEVICE | PREDICATE DEVICE |
| Channel isolation | SAME | YES |
| Line current isolation | SAME | N/A |
| Patient override<br>control method | SAME | YES |
| Burst mode | SAME | YES |
| Housing<br>material<br>and construction: | SAME | Plastic material |
| Standards: | SAME | IEC 60601-1:1988<br>EN 60601-1-2:1990 |
| Timer setting<br>Automatic shut off | SAME<br>2 minutes | YES (05~90Min)<br>30 Sec |
| Automatic overload<br>trip current/voltage<br>level | N/A | YES |
| Automatic no load trip | N/A | YES |
| Indicator display<br>Unit functioning: | Power ON | Power ON, Vpp |
| Low battery | Yes | NO |
| indicator:<br>Voltage level: | NO | YES |
| Weight ( unit ) | 105 g w/o Battery | 162 grams w/ battery |
| Dimensions ( unit ) | 120 x 52 x 17 mm | 136 x 70 x 27 mm |
| Size of electrode | Carbon silicon of 45x45 mm | Irregular: 10x15cm |
{3}------------------------------------------------
K092448
Image /page/3/Picture/1 description: The image shows the word "HIVOX" in all caps, followed by a small superscript symbol and a stylized plus sign. The plus sign is made up of several small squares connected in a winding pattern. The text and symbol are all in black and are set against a white background.
# HIVOX BIOTEK INC
8/F, No.98 Shingde Road Sanchong City Taipei 241 Taiwan ROC Phone: +886 2 8511 2668 Fax: +886 2 8511 2669
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three tail feathers, representing the three levels of government: federal, state, and local. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring. MD 20993-0002
Hivox Biotek Inc. c/o Dr. Ke-Min Jen 8F, No. 98, Shinde Rd., Sanchong City Taipei Hsien China (Taiwan) 24158
MAR 3 0 2010
Re: K092448
Trade/Device Name: Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: Class II Product Code: GZJ, IPF Dated: March 12, 2010 Received: March 19, 2010
Dear Dr. Jen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{5}------------------------------------------------
Page 2 - Dr. Ke-Min Jen
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Kevin Alexander
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
#### 510(k) Number:
K092448
Device Name:
#### HIVOX Electric Stimulator TENS & EMS, HD2
#### Indications for Use :
The TENS programs of HIVOX Electric Stimulator TENS & EMS, HD2:
The device is an electrical nerve stimulator indicated for use for pain relief by applying an electrical current to electrodes on a patient's skin to treat pain. In particular, this device is indicated for use for (1) Symptomatic relief of chronic (long term) intractable pain, and (2) adjunctive treatment in the management of post-surgical and post-traumatic pain problems.
The EMS programs of HIVOX Electric Stimulator TENS & EMS, HD2:
The device is an electrical powered muscle stimulator indicated for medical purpose to repeatedly contract muscles by passing electrical currents through electrodes contacting the affected body area. In particular, this device is indicated for use for (1) Relaxing muscle spasms, (2) Increasing local blood circulation, (3) Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis, (4) Muscle re-education, (5) Maintaining or increasing range of motion, and (6) Preventing or retarding disuse atrophy.
Prescription Use _ V (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
Page 1 of 1
510(k) Number K092448
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.