BENCOX ID STEM, BENCOX METAL HEAD, BENCOX BIPOLAR CUP
Applicant
Corentec Co., Ltd.
Product Code
KWY · Orthopedic
Decision Date
Oct 12, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3390
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Bencox Bipolar Cup system is intended for cementless use in partial or total hip arthroplasty in primary or revision surgery for the following conditions: a. Non-inflammatory degenerative joint disease, such as avascular necrosis, osteoarthritis, traumatic arthritis; b. Inflammatory degenerative joint disease, such as rheumatoid arthritis; c. Treatment of non-union, femoral neck fracture and trochantric fractures of the proximal femur with head involvement, unmanageable using other techniques; d. Patients with failed previous surgery where pain, deformity, or dysfunction persists; e. Revision of previously failed hip arthroplasty.
Device Story
Cementless metal-on-polyethylene hip prosthesis system; components include femoral stem (BENCOX ID Stem), modular femoral head (BENCOX Metal Head), and acetabular system (BENCOX Bipolar Cup with shell, PE liner, and locking ring). Used in orthopedic surgery for hip arthroplasty; implanted by surgeons. System provides mechanical articulation for hip joint; self-centering with positive eccentricity. Benefits include restoration of joint function and pain relief in patients with degenerative disease or fractures. Sterilized via gamma irradiation.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing including ISO 7206-4, ISO 7206-9, ISO 7206-10, ISO 14242-1, ISO 14242-2, ASTM F1147, ASTM F1044, ASTM F1160, ASTM F1978, ASTM F1875, ASTM 2582, and ASTM F1820.
Indicated for patients requiring partial or total hip arthroplasty (primary or revision) due to non-inflammatory/inflammatory degenerative joint disease (e.g., avascular necrosis, osteoarthritis, rheumatoid arthritis), non-union, femoral neck/trochanteric fractures, or failed previous hip surgery.
Regulatory Classification
Identification
A hip joint femoral (hemi-hip) metal/polymer cemented or uncemented prosthesis is a two-part device intended to be implanted to replace the head and neck of the femur. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a snap-fit acetabular component made of an alloy, such as cobalt-chromium-molybdenum, and ultra-high molecular weight polyethylene. This generic type of device may be fixed to the bone with bone cement (§ 888.3027) or implanted by impaction.
{0}------------------------------------------------
K112019 (pg. 1 of 4)
OCT 1 2 2011
510(k) Premarket Notification
BENCOX Bipolar Cup System
く
# 510(K) SUMMARY
# Corentec Co., Ltd.
# BENCOX® Bipolar Cup System
July 13, 2011
#### ADMINISTRATIVE INFORMATION
| Manufacturer | Corentec Co., Ltd. |
|--------------|---------------------------------------------------|
| | 247 Giro-ri, Ipjang-myeon, Seobuk-Gu |
| | Cheonan-si, Chungchongnam-do, 331-822 South Korea |
| | Telephone: +82-41-585-7114 |
| | Fax: +82-41-585-7113 |
| Official Contact | J.S. Daniel<br>Project Manager - QA/RA<br>Corentec Co., Ltd<br>11th Chungho Tower, 748-1 Banpo 1 Dong<br>Seocho Gu, Seoul, Korea 137-040<br>Ph: +82 2 3445 5492<br>Fax: +82 2 3445 5497<br>Email: jsdaniel@corentec.com |
|------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
## DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name: | BENCOX Bipolar Cup System |
|-----------------------------|-----------------------------|
| Common Name: | Total Hip Prosthesis System |
| Classification Regulations: | 21 CFR 888.3358, 888.3390 |
| Class | II |
| Product Codes: | LPH, KWY |
| Classification Panel: | Orthopedic Products Panel |
| Reviewing Branch: | Orthopedic Devices Branch |
#### INTENDED USE
The Bencox Bipolar Cup system is intended for cementless use in partial or total hip arthroplasty in primary or revision surgery for the following conditions:
a. Non-inflammatory degenerative joint disease, such as avascular necrosis, osteoarthritis, traumatic arthritis;
b. Inflammatory degenerative joint disease, such as rheumatoid arthritis;
{1}------------------------------------------------
#### 510(k) Premarket Notification
c. Treatment of non-union, femoral neck fracture and trochantric fractures of the proximal femur with head involvement, unmanageable using other techniques; d. Patients with failed previous surgery where pain, deformity, or dysfunction persists; e. Revision of previously failed hip arthroplasty.
#### DEVICE DESCRIPTION
The BENCOX Bipolar Cup System is a cementless, metal-on-polyethylene bearing hip system for hip arthroplasty. It consists of components, Femoral stem -BENCOX ID Stem; Femoral head - BENCOX Metal Head; Acetabular system -BENCOX Bipolar Cup (Shell; Insert; Lock Ring & PE ring) and BENCOX Hip Instrumentation.
The Bencox ID Stem is manufactured from titanium alloy (Ti-6AI-4V ELI) conforming to ASTM F136, Standard Specification for Wrought Titanium-6 Aluminum-4 Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401) and has a proven titanium plasma spray coating with pure titanium conforming to ASTM F1580-07, Standard Specification for Titanium and Ti-6 Al-4 V Alloy Powders for Coatings of Surgical Implants with about two decades of clinical experience
Bencox Metal Head components are manufactured from cobalt-chromiummolybdenum alloy conforming to ASTM F1537, Standard Specification for Wrought Cobalt-28-Chromium-6-Molybdenum Alloys for Surgical Implants (UNS R31537, UNS R31538, and UNS R31539).
The BENCOX Bipolar Cups are provided as the preassembled cup (CoCr shell + PE Liner) and PE ring is assembled by the user (surgeon). The shell is made of Cobaltchromium-molybdenum alloy conforming to ASTM F75-07, Standard Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Castings and Casting Alloy for Surgical Implants. The PE liner and locking ring is made of Ultra-high molecular weight polyethylene conforming to ASTM F648 Standard Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants.
Bencox Bipolar Cup System components are sterilized by gamma irradiation. EQUIVALENCE TO MARKETED DEVICES
Corentec Co., Ltd. submits the following information in this Premarket Notification to demonstrate that for the purposes of FDA's regulation of medical devices, the BENCOX Bipolar Cup System is substantially equivalent in indications and design principles to the following predicate devices, each of which has been determined by FDA to be substantially equivalent to pre-amendment devices:
{2}------------------------------------------------
K112019 (pg. 3 of 4)
510(k) Premarket Notification
Biomet. Inc.. Taperloc® 12/14 Taper Femoral Components cleared under K043537: Biomet, Inc., Malloy Head® Lateralized Press-Fit Femoral cleared under K003429: Corin Group Ltd., Tri-Fit® Femoral Stem cleared under K010243; Biomet. Inc., Metal On Metal Acetabular System cleared under K993438; Biomet, Inc., M2a 32mm Taper System cleared under K003363; StelKast Inc., Stelkast Bipolar Head cleared under K972961; Biomet, Inc., RingLoc® Bi-Polar Acetabular Component cleared under K051569; Stryker (Osteonics Corp.), UHR Universal hip Replacement cleared under K800207;
The Bencox Bipolar Cup System's Bencox ID Stem and the Taperloc Stem (K043537), Mallory Head (K003429) and the TriFit (K010243) are made of the same material and have the same titanium plasma spray coating. Bencox ID Stem & TriFit (K010243) has the same neck angle. All the stems have a similar overall geometries and ranges of sizes.
The Bencox Bipolar Cup System's Bencox Metal Head and the Metal on Metal Acetabular System (K993438) and M2a™ 32 mm Acetabular System (K003363) all include modular femoral heads made of the same material and a similar tapered interface to the femoral stem. The subject and predicate modular femoral heads all encompass a similar range of head sizes, neck lengths and femoral offsets, and all have similar surface finish and sphericity.
The Bencox Bipolar Cup System's Bencox Bipolar Cup and StelKast (K972961), Ringloc Bi-polar (K051569) and UHR (K800207), all the shell is made with a highly polished Cobalt Chrome alloy (CoCrMo) shell and a polyethylene (UHMWPE) liner with an integrated or separate polyethylene femoral head lock ring. Bencox Bipolar Cup has a liner locking mechanism similar to Ringloc Bi-polar (K051569) using a titanium lock ring. Bencox Bipolar Cup has a femoral head lock ring similar to StelKast (K972961).All the devices are self-centering with positive eccentricity. The devices are offered to fit either 22 mm and/or 28 mm femoral heads. All the bipolar cups shell has a similar range of sizes.
Performance testing was performed to demonstrate substantial equivalence and included methods described in the following standards: ISO 7206-4, ISO 7206-9, ISO 7206-10, ISO 14242-1, ISO 14242-2, ASTM F1147, ASTM F1044, ASTM F1160, ASTM F1978, ASTM F1875, ASTM 2582 and ASTM F1820.
{3}------------------------------------------------
### 510(k) Premarket Notification
Any differences in technological characteristic between the subject and predicate devices do not raise new issues of safety or efficacy of the bipolar cup system components for its intended use.
Overall, the BENCOX Bipolar Cup System has the following similarities to the predicate devices:
· has the same intended use,
· uses the same operating principles,
· incorporates the same basic designs,
· incorporates the same materials, and
· has similar packaging and is sterilized using the same materials and processes.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Corentec Co., Ltd. % Mr. J.S. Daniel Project Manager - QA/RA 1100 Chungho Tower, 748-1 Banpo 1 Dong Seocho Gu, Seoul, Korea 137-040
OCT 1 2 2011
Re: K112019 Trade/Device Name: BENCOX Bipolar Cup System Regulation Number: 21 CFR 888.3390 Regulation Name: Hip joint femoral (hemi-hip) metal/polymer cemented or uncemented prosthesis Regulatory Class: Class II Product Code: KWY, LPH Dated: July 13, 2011 Received: July 14, 2011
Dear Mr. Daniel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{5}------------------------------------------------
Page 2 – Mr. J.S. Daniel
forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safetv/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours,
Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### INDICATIONS FOR USE
510(k) Number (if known): K112019
Device Name: BENCOX Bipolar Cup System
The Bencox Bipolar Cup system is intended for cementless use in partial or total hip arthroplasty in primary or revision surgery for the following conditions:
a. Non-inflammatory degenerative joint disease, such as avascular necrosis, osteoarthritis, traumatic arthritis;
b. Inflammatory degenerative joint disease, such as rheumatoid arthritis;
c. Treatment of non-union, femoral neck fracture and trochantric fractures of the proximal femur with head involvement, unmanageable using other techniques;
d. Patients with failed previous surgery where pain, deformity, or dysfunction persists;
e. Revision of previously failed hip arthroplasty
Prescription Use: X (Per 21 CFR 801 Subpart D)
AND / OR
Over-The Counter Use: -(Per 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Surgical, Orthopedic,
and Restorative Devices
510(k) Number K112019
Corentec Co. Ltd.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.