K993438 · Biomet, Inc. · KWA · May 18, 2000 · Orthopedic
Device Facts
Record ID
K993438
Device Name
METAL ON METAL ACETABULAR SYSTEM
Applicant
Biomet, Inc.
Product Code
KWA · Orthopedic
Decision Date
May 18, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3330
Device Class
Class 3
Attributes
Therapeutic
Indications for Use
The Metal on Metal Acetabular System is indicated for used in patients requiring total hip replacement due to the following: a.) Non-inflammatory degenerative joint disease including avascular necrosis, diastrophic variant, fracture of the pelvis, fused hip, leg perthes, osteoarthritis, slipped capital epiphysis, subcapital fractures, and traumatic arthritis b.) Rheumatoid arthritis c.) Correction of functional deformity d.) Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques. e.) Revision of previously failed total hip arthroplasty.
Device Story
Metal on Metal Acetabular System; modular hip prosthesis. Components: porous-coated titanium alloy shell (Ti-6Al-4V), cobalt-chromium (Co-Cr-Mo) liner, and 28mm Co-Cr-Mo modular femoral head. Implanted by orthopedic surgeons during total hip arthroplasty. System provides articulating surface for hip joint; modular design allows assembly with Biomet femoral stems. Mechanical testing demonstrates reduced wear compared to traditional polyethylene-metal systems. Shell features hemispherical design for minimal bone removal; optional screw fixation; fins on Mallory Head design prevent rotation. Device provides structural replacement for diseased or damaged hip joints; restores joint function; reduces pain.
Clinical Evidence
Bench testing only. No clinical data provided. Mechanical testing demonstrated decreased wear compared to traditional polyethylene and metal acetabular systems. Sphericity and surface roughness measurements met or exceeded ISO standards.
Indicated for patients requiring total hip replacement due to non-inflammatory degenerative joint disease (e.g., avascular necrosis, osteoarthritis, fractures), rheumatoid arthritis, functional deformity, or non-union/fractures of the proximal femur. Also indicated for revision of failed total hip arthroplasty.
Regulatory Classification
Identification
A hip joint metal/metal semi-constrained, with an uncemented acetabular component, prosthesis is a two-part device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that consist of a femoral and an acetabular component, both made of alloys, such as cobalt-chromium-molybdenum. The femoral component is intended to be fixed with bone cement. The acetabular component is intended for use without bone cement (§ 888.3027).
Predicate Devices
Mallory Head Finned Acetabular Cup
Universal Acetabular Cup
Biomet Co-Cr Femoral Components
Sulzer's Inter-Op Metasul
Submission Summary (Full Text)
{0}------------------------------------------------
## MAY 1 8 2000
## SUMMARY OF SAFETY AND EFFECTIVENESS
| SPONSOR: | Biomet, Inc.<br>P.O. Box 587<br>Airport Industrial Park<br>Warsaw, Indiana 46581-0587 |
|----------------------|---------------------------------------------------------------------------------------|
| CONTANT PERSON: | Michelle L. McKinley |
| DEVICE NAME: | Metal on Metal Acetabular System |
| CLASSIFICATION NAME: | Prosthesis, Hip, Semi-constrained (Metal<br>Uncemented Acetabular Component) |
#### INTENDED USE:
The Metal on Metal Acetabular System is indic. ted for used in patients requiring total hip replacement due to the following:
- a.) Non-inflammatory degenerative joint disease including avascular necrosis, diastrophic variant, fracture of the pelvis, fused hip, leg perthes, osteoarthritis, slipped capital epiphysis, subcapital fractures, and traumatic arthritis
- b.) Rheumatoid arthritis
- c.) Correction of functional deformity
- d.) Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques.
- e.) Revision of previously failed total hip arthroplasty.
#### DEVICE DESCRIPTION:
The Metal on Metal Acetabular System consists of a titanium outer shell with a cobalt chromium metallic liner, which articulates with a cobalt chromium modular head. Based on mechanical testing, it has been shown that the Metal on Metal Acetabular System exhibits decreased wear compared to the wear observed in the traditional polyethylene and metal acetabular systems.
#### Acetabular Shell
The acetabular shells that will initially be available in two designs, the Mallory Head Radial and the Universal Acetabular Component. Both designs are modular two-piece systems consisting of a modular cobalt chromium (Co-Cr-Mo) liner and a porous coated titanium shell. Hemispherical shape of both outer shells closes matches the natural acetabulum, which leads to minimal bone removal in preparation for implantation. The two acetabular shell designs are available with holes or as a solid dome. If necessary, the
Image /page/0/Picture/16 description: The image shows a sequence of numbers, specifically '000358'. The numbers are printed in a bold, stencil-like font, giving them a slightly rugged appearance. The digits are uniformly sized and spaced, creating a clear and easily readable sequence.
{1}------------------------------------------------
shell with holes allows the use of 6.5 mm dome and the 5.0 mm rim screws for optional supplemental fixation. The solid shell configuration without dome holes increases the surface area of the porous plasma spray coating. The Mallory Head acetabular shell features four fins, which aid in preventing rotation.
The outer surface of the shell is covered with a porous coating of titanium alloy (Ti-6Al-4V) powder conforming to ASTM-F136, which ensures immediate component fixation and maximum bone-to-implant contact. The plasma sprayed surface consists of particles which are bonded together to form a random pattern with interconnecting pores.
#### Acetabular Liner
The metallic cobalt chromium bearing liner fits into the outer shell by means of a taper similar to the taper used for the attachment of the modular head. The locking mechanism consists of a three degree taper with a maximum engagement of 0.075 inches. The surface roughness of the articulating surface averages 0.17 micro-inches Rg. The sphericity of the liner components has an average Peak to Valley deviation of 59.8024 micro-inches. This is less than half the allowable deviation set forth in the ISO standard for modular head sphericity.
#### Modular Femoral Head
The Metal on Metal Acetabular System utilizes a 28mm cobalt chromium (Co-Cr-Mo) modular femoral head with seven neck lengths (-6mm to +12mm). Each modular head has a four degree angle included in the bore with a surface roughness of 34 micro-inches. The articulating surface is highly polished with a surface roughness of 0.43 micro-inches.
The modular heads may be used in conjunction with any of Biomet's commercially available femoral components. At the time of surgery, the modular head is assembled with a femoral stem.
#### POTENTIAL RISKS:
The potential risks associated with this device are the same as with any joint replacement device. These include, but are not limited to:
Fracture of the component Cardiovascular disorders Implant loosening/migration Soft tissue imbalance Deformity of the joint Tissue growth failure Delayed wound healing Metal sensitivity
Bone fracture Hematoma Blood vessel damage Nerve damage Excessive wear Infection Dislocation Breakdown of the porous surface
## 000359
{2}------------------------------------------------
## SUBSTANTIAL EQUIVALENCE:
The Metal on Metal Acetabular System is similar to previously marketed devices. Direct comparison was made with the following predicate devices:
Mallory Head Finned Acetabular Cup Universal Acetabular Cup Biomet Co-Cr Femoral Components Sulzer's Inter-Op Metasul
# 000360
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of an eagle with three lines representing its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular fashion around the eagle image.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
### MAY 1 8 2000
Ms. Michelle L. McKinley Regulatory Specialist Biomet. Inc. Airport Industrial Park P.O. Box 587 Warsaw, Indiana 46581-0587
Re: K993438/S1
Trade Name: Metal on Metal Acetabular Component Regulatory Class: III Product Code: KWA Dated: February 18, 2000 Received: February 22, 2000
Dear Ms. McKinley:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general control provisions of the Act. The general control provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
{4}------------------------------------------------
Page 2 - Ms. Michelle L. McKinley
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Russell Sage
Su Celia M. Whitten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
| Page | ____________ |
|------|--------------|
| of | ____________ |
K993438 510(k) Number if Known: ______________________________________________________________________________________________________________________________________________________ Device Name: Metal on Metal Acetabular System
The Metal on Metal Acetabular System is indicated for used in patients requiring total hip replacement due to the following:
- a.) Non-inflammatory degenerative joint disease including avascular necrosis, diastrophic variant, fracture of the pelvis, fused hip, leg perthes, osteoarthritis, slipped capital epiphysis, subcapital fractures, and traumatic arthritis
- b.) Rheumatoid arthritis
- c.) Correction of functional deformity
- d.) Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmar igeable using other techniques.
- e.) Revision of previously failed total hip arthroplasty.
### (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per CFR 801.109)
or
Over the Counter Use (Optional Format 1-2-96)
Venetia Taylor. for term
(Division Sign-Off) (Division of General Restorative Devices 510(k) Number
## 000004
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.