K111994 · Smith & Nephew, Inc. · HWC · Oct 11, 2011 · Orthopedic
Device Facts
Record ID
K111994
Device Name
SMITH & NEPHEW CANNULATED SCREWS AND WASHERS
Applicant
Smith & Nephew, Inc.
Product Code
HWC · Orthopedic
Decision Date
Oct 11, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
Bone plates and screws from the Smith and Nephew Bone Plate System are used for adult and pediatric patients as indicated for pelvic, small, and long bone fracture fixation. Indications for use include, fractures of the tibia, fibula, femoral condyle, pelvis, acetabulum, metacarpals, metatarsals, humerus, ulna, middle hand and middle foot bones; treatment of the calcaneus; hip arthrodesis, and provisional hole fixation. The 4.0mm Cannulated Screws and associated washers are additionally intended for arthrodesis and osteotomies of small bones and small joints, including scaphoid and other carpal bones, metacarpals, tarsals, metatarsals, patella, ulnar styloid, capitellum, radial head and radial styloid. The 5.5mm, 6.5mm, 7.0mm, and 8.0mm Cannulated Screws and associated washers are additionally intended for reconstruction, osteotomy, and arthrodesis of various bones and bone fragments appropriate for the size of the device including joint fusions (arthrodesis) in the foot and ankle.
Device Story
Cannulated screws and washers provide internal bone fixation. Screws feature fluted tips for self-tapping/drilling and hexagonal heads; cannulated design allows placement over guidewires. Used in clinical settings by orthopedic surgeons for fracture fixation, osteotomies, and arthrodesis. Washers extend surface area of screw heads. Device provides mechanical stability to bone fragments during healing, potentially improving patient recovery outcomes. Operation involves manual insertion by surgeon.
Clinical Evidence
Bench testing only. Performance testing included three-point bend fatigue, static torsional, and axial pullout tests for 4.0mm screws compared to 3.0mm SS predicate. Four-point bend fatigue, static torsional, and axial pullout tests were performed for 5.5mm screws compared to Synthes 4.5mm Ti predicate. Results confirmed subject devices perform equal to or better than predicate devices.
Technological Characteristics
Metallic bone fixation fasteners (screws and washers). Materials: Titanium alloy and stainless steel. Design: Cannulated, fluted tips, hexagonal heads, varying diameters (4.0mm-8.0mm) and lengths. Mechanical performance validated via fatigue, torsional, and axial pullout testing. Non-software device.
Indications for Use
Indicated for adult and pediatric patients requiring fixation of pelvic, small, and long bone fractures, including tibia, fibula, femoral condyle, pelvis, acetabulum, metacarpals, metatarsals, humerus, ulna, hand/foot bones, and calcaneus. Also indicated for hip arthrodesis, provisional hole fixation, and reconstruction, osteotomy, and arthrodesis of small bones/joints (e.g., scaphoid, carpal bones, patella, ulnar styloid, capitellum, radial head/styloid) and foot/ankle joint fusions.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
K111994 # '1/4
OCT
-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------
1572
## 510(k) Summary of Safety and Effectiveness Smith & Nephew Cannulated Screws and Washers
| Submitted By: | Smith & Nephew, Inc.<br>Orthopaedic Division<br>1450 East Brooks Road<br>Memphis, Tennessee 38116 |
|----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date of Summary: | October 6, 2011 |
| Contact Person: | Maureen Whitson, Senior Regulatory Affairs Specialist<br>Tel: 901-399-6845, Fax 901-566-7568 |
| Proprietary Name: | Smith & Nephew Cannulated Screws and Washers |
| Common Name: | Cannulated Screws and Washers |
| Device Classification Name and<br>Reference: | 21 CFR 888.3040 Smooth or threaded metallic bone<br>fixation fastener<br>and<br>21 CFR 888.3030 Single/multiple component<br>metallic bone fixation appliances and accessories |
| Device Class: | Class II |
| Panel Code: | Orthopedics/ 87 - HWC and NDG |
#### Device Description
The subject of this Traditional 510(k) premarket notification is to expand the indications of the already existing, cleared Smith & Nephew Cannulated Screws and Washers by adding osteotomies and arthrodesis. The subject devices are part of the Smith and Nephew Bone Plate System cleared under 510(k) number K993106 and K060736 with the same original indications but the subject 4.0mm, 5.5mm, 6.5mm, 7.0mm and 8.0mm cannulated screws and associated washers have additional indications independent of the Bone Plate System.
{1}------------------------------------------------
KI111994 # 2/4
The screws vary in diameter, length and thread length; have fluted screw tips for self-tapping and self-drilling and have a hexagonal screw heads. The screws may be used with bone washers to extend the surface area of the head if needed. These screws are cannulated to allow the screws to be accurately positioned with the use of a guidewire instrument. The device offering consist of diameters of 4.0mm, 5.5mm, 7.0mm and 8.0mm in titlanium allov and stainless steel.
| Diameter | Length Range | Thread Length Range | Material |
|----------|--------------|---------------------------------|----------|
| 4mm | 20mm - 70mm | 8mm - 15mm | Ti |
| 4mm | 10mm - 70mm | 5mm - 15mm and fully threaded | SS |
| 5.5mm | 20mm - 120mm | 8mm - 32mm and fully threaded | Ti |
| 5.5mm | 20mm - 120mm | 5mm - 15mm and fully threaded | SS |
| 6.5mm | 30mm - 180mm | 22mm - 46mm and fully threaded | Ti |
| 6.5mm | 30mm - 150mm | 22mm - 46mm and fully threaded | SS |
| 7mm | 30mm - 150mm | 16mm - 32mm and fully threaded | Ti |
| 7mm | 30mm - 150mm | 16mm - 32mm and fully threaded | SS |
| 8mm | 40mm - 180mm | 22mm - 46mm and fully threaded | Ti |
| 8mm | 40mm - 180mm | 22mm - 180mm and fully threaded | SS |
#### Smith & Nephew Cannulated Screw Size Ranges
Summary of Cannulated Screw Accessory Washers
| Accessory<br>Device | Type | Size Range | Material |
|---------------------|-----------|-------------------------------------------------------|-----------|
| Washers | Round | 4.0mm to 8.0mm Internal Diameter | Ti and SS |
| Washer | Butterfly | 8.0mm Internal Diameter and Overall<br>Length 21.7mm | Ti and SS |
#### Intended Use
Bone plates and screws from the Smith and Nephew Bone Plate System are used for adult and pediatric patients as indicated for pelvic, small, and long bone fracture fixation. Indications for use include, fractures of the tibia, fibula, femoral condyle, pelvis, acetabulum, metacarpals, metatarsals, humerus, ulna, middle hand and middle foot bones; treatment of the calcaneus; hip arthrodesis, and provisional hole fixation.
The 4.0mm Cannulated Screws and associated washers are additionally intended for arthrodesis and osteotomies of small bones and small joints, including scaphoid and other carpal bones, metacarpals, tarsals, metatarsals, patella, ulnar styloid, capitellum, radial head and radial styloid.
The 5.5mm, 6.5mm, 7.0mm, and 8.0mm Cannulated Screws and associated washers are additionally intended for reconstruction, osteotomy, and arthrodesis of various bones and bone fragments appropriate for the size of the device including joint fusions (arthrodesis) in the foot and ankle.
{2}------------------------------------------------
K111994 #3/4
## Technological Characteristics
The Smith & Nephew Cannulated Screws and Washers are very similar to legally marketed devices cleared under K090675, K080943, K063298, K102903, K021932, K052483, and K003496. When compared to the predicates, the devices share very similar indications for use and intended use, are manufactured from similar materials and incorporate similar technical design characteristics.
A review of the mechanical data indicates that the Smith & Nephew Cannulated Screws are capable of withstanding expected in vivo loading without failure. The performance bench testing consisted of a three point bend fatigue, static torsional and axial pullout bend test, comparing the S&N 4.0mm screws (Ti and SS) to the S&N 3.0mm SS screw and the results show the subject screw would perform equal or better that the predicate screws. The performance bench testing also consisted of a four point bend fatigue, static torsional and axial pullout bend test, comparing the S&N 5.5mm screws (Ti and SS) to the Synthes titanium 4.5mm screw and the results show the subject screws would perform equal or better than the predicate screws.
### Substantial Equivalence Information:
The substantial equivalence of the Smith & Nephew Cannulated Screws and Washers is based on its similarities in indications for use, design features, operational principles, and material composition to the devices listed below.
| Manufacturer | Description | Submission<br>Number |
|----------------|-------------------------------------------------|----------------------|
| Smith & Nephew | Smith & Nephew 3.0 Cannulated Screw, SS | K090675 |
| Synthes | 4.5 Headless Compression Screw (cannulated), Ti | K080943 |
| Osteomed | Osteomed Headless Cannulated Screw System | K063298 |
| Pioneer | Pioneer Cannulated Screw System | K102903 |
| Synthes | Synthes 6.5 mm Cannulated Screw | K021932 |
| Synthes | Synthes Spherical Washers | K052483 |
| Pioneer | Cannulated Screw System, Washer | K003496 |
{3}------------------------------------------------
K 11119944444
Similarities and Difference between the Subject Devices and predicate Devices
| Similarities |
|--------------|
| |
| | Similarities of Subjects to Predicates |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Design | Similar in dimensions (screw diameter, screw length ) |
| Features | Both predicate and subject are cannulated, both have<br>varying thread lengths, both are offered with accessory bone<br>screw washers |
| Material | Both are offered in SS and Ti Alloy |
| Indications for Use | Both are indicated for fractures, osteotomies and arthrodesis<br>of various small bones and joints |
# Differences
| | Differences of Subjects to<br>Predicates | Rationale why this difference does not affect<br>the Safety and Effectiveness |
|-------------------------|------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Design | Screw head dimensions may<br>have slight size differences<br>and some may be smooth<br>and others may be threaded<br>or "headless" | Headless screws are used for counter-<br>sinking to prevent the implanted screw from<br>sitting too proud on the bone surface and are<br>for surgeon preference. The screw head size<br>does not affect the performance of the<br>implanted screw in regard to the mechanical<br>stresses on the shaft of the screws. |
| Design | Varying screw thread length | The thread lengths do not affect the<br>mechanical stresses of the implanted screws.<br>Their application is chosen based on surgeon<br>preference and on what kind of fracture,<br>osteotomy or arthrodesis in which the screw<br>will be used. |
| Chemical<br>Composition | The predicate Synthes<br>4.5mm Ti alloy screw is made<br>from Ti -6Aluminum -7<br>Niobium instead of Ti 6<br>Aluminum- 4Vanadium | The niobium was substituted for the<br>vanadium as the beta stabilizing element in<br>this alloy. According to the manufacturer of<br>the raw material, the mechanical properties<br>are almost identical and it is a common alloy<br>used in orthopedic implants. Our bench<br>testing shows the subject screws performed<br>equal or better than the predicate screw<br>made with this material |
.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular and contains the department's name around the perimeter. In the center of the seal is a stylized image of an eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Smith & Nephew, Inc. % Ms. Maureen Whitson Senior Regulatory Affairs Specialist 7135 Goodlett Farms Parkway Cordova. Tennessee 38106
OCT 1 1 2011
Re: K111994
Trade/Device Name: Cannulated Screws and Washers Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: HWC. NDG Dated: July 12, 2011 Received: July 13, 2011
Dear Ms. Whitson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{5}------------------------------------------------
Page 2 – Ms. Maureen Whitson
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.html for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, international and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Pon Nth
Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Premarket Notification Indications for Use Statement
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name:
Cannulated Screws and Washers
Indications for Use:
Bone plates and screws from the Smith and Nephew Bone Plate System are used for adult and pediatric patients as indicated for pelvic, small, and long bone fracture fixation. Indications for use include, fractures of the tibia, fibula, femoral condyle, pelvis, acetabulum, metacarpals, metatarsals, humerus, ulna, middle hand and middle foot bones, treatment of the calcaneus; hip arthrodesis, and provisional hole fixation.
The 4.0mm Cannulated Screws and associated washers are additionally intended for arthrodesis and osteotomies of small bones and small joints, including scaphoid and other carpal bones, metacarpals, tarsals, metatarsals, patella, ulnar styloid, capitellum, radial head and radial styloid.
The 5.5mm, 6.5mm, 7.0mm, and 8.0mm Cannulated Screws and associated washers are additionally intended for reconstruction, osteotomy, and arthrodesis of various bones and bone fragments appropriate for the size of the device including joint fusions (arthrodesis) in the foot and ankle.
AND/OR
Prescription Use
(Part 21 CFR 801 Subpart D)
×
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
urrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.