The Cannulated Screw System by Pioneer will be indicated for use in long and small bone fracture fixation, which may include the following: - Fractures of the tarsals and metatarsals; 1. - Metatarsal and phalangeal osteotomies; 2. - Fractures of the carpals and metacarpals; 3. - Carpal and metacarpal arthrodesis; 4: - Small fragments of the hand and wrist; 5. - Ligament fixation; 6. - Sacroillac ioint disruptions; 7. - Fractures of the distal femur and proximal tibia; 8. - Intracapsular fractures of the hip; 9. - 10. Ankle arthrodesis; - 11. Pelvis and acetabulum fractures; and - 12. Areas where accurate screw placement is vital. This system is not indicated for use in the spine. The Cannulated Screw System by Pioneer will be offered both sterile and non-sterile and is a single use device.
Device Story
Cannulated screw system for bone fracture fixation; includes screws of varying diameters/lengths/thread configurations and optional washers. Screws are self-cutting and self-tapping to facilitate OR efficiency; feature reverse-buttress threads for secure engagement and ease of removal. Used by surgeons in clinical settings for orthopedic procedures. Output is physical fixation of bone fragments. Single-use device; provided sterile or non-sterile.
Clinical Evidence
Bench testing only. Performance evaluated via ASTM standards: F 1691 (Axial Pull-Out Strength), F 117 (Driving Torque), and F 1622 (Torsional Properties). No clinical data provided.
Technological Characteristics
Cannulated bone screws and washers. Design references ASTM F 543 (Metallic Medical Bone Screws) and ASTM F 116 (Medical Screwdriver Bits). Features self-cutting, self-tapping, reverse-buttress thread geometry. Single-use. Sterilization provided for sterile-labeled units.
Indications for Use
Indicated for long and small bone fracture fixation, including tarsal/metatarsal fractures, metatarsal/phalangeal osteotomies, carpal/metacarpal fractures and arthrodesis, small fragments of hand/wrist, ligament fixation, sacroiliac joint disruptions, distal femur/proximal tibia fractures, hip intracapsular fractures, ankle arthrodesis, and pelvis/acetabulum fractures. Contraindicated for use in the spine.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
ZIMMER MAGNA-Fx
ZIMMER Mini MAGNA-Fx
Synthes Cannulated Screw System
Submission Summary (Full Text)
{0}------------------------------------------------
K003496
10F2
#### 510(K) SUMMARY
V
# Pioneer Surgical Technology 510(K) Notification Summary: Cannulated Screw System by Pioneer
### ADMINISTRATIVE INFORMATION
| Manufacturer Identification<br>And Sponsor | Pioneer Surgical Technology<br>375 River Park Circle<br>Marquette, Michigan 49855-1781<br>Telephone: (906) 226-9909<br>Facsimile: (906) 226-9932 |
|--------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact | Amy H. Mommaerts, Manager<br>Regulatory Affairs |
Date Prepared ................................................................................................................................................................
### DEVICE IDENTIFICATION
| Proprietary Name | Cannulated Screw System by Pioneer |
|-----------------------|------------------------------------|
| Common Name: | Cannulated Screw |
| Classification Name | Screw, Fixation, Bone |
| And Reference | Screw, Fixation, Bone |
| Regulation Number | 888.3040 |
| Classification Number | 87HWC |
| | |
| Proprietary Name | Cannulated Screw System by Pioneer |
| Common Name: Screw | Washer |
| Classification Name | |
| And Reference: Screw | Washer, Bolt, Nut, Orthopaedic |
| Regulation Number | 888.3030 |
| Classification Number | 87HTN |
## Devices on Which Substantial Equivalence is Claimed
The Cannulated Screw System by Pioneer is predicated on the ZIMMER MAGNA-Fx; the ZIMMER Mini MAGNA-Fx; and the Synthes Cannulated Screw System.
### Device Description
The Cannulated Screw System by Pioneer consists of cannulated screws of varying diameters, lengths and thread configurations to accommodate variations in surgical technique and severity level of fracture. The system includes correspondingly sized washers, the use of which is optional. The cannulated screws are self-cutting and self-
{1}------------------------------------------------
K003496
2 OF 2
tapping in appropriate bone stock, which maximizes OR efficiency. The threads are tapping in approphato bone otew removal and decreasing the chance of breakage. reverse batting, faoilitating of engle use and will be offered both sterile and nonsterile.
#### Intended Use
The Cannulated Screw System by Pioneer will be indicated for use in long and small bone fracture fixation, which may include the following:
- Fractures of the tarsals and metatarsals; 1.
- Metatarsal and phalangeal osteotomies; 2.
- Fractures of the carpals and metacarpals; 3.
- Carpal and metacarpal arthrodesis; ধ:
- Small fragments of the hand and wrist; 5.
- Ligament fixation; 6.
- Sacroillac ioint disruptions; 7.
- Fractures of the distal femur and proximal tibia; 8.
- Intracapsular fractures of the hip; 9.
- 10. Ankle arthrodesis;
- 11. Pelvis and acetabulum fractures; and
- 12. Areas where accurate screw placement is vital.
This system is not indicated for use in the spine. The Cannulated Screw System by Pioneer will be offered both sterile and non-sterile and is a single use device.
## Technological Characteristic Compared to Predicate Device
The Cannulated Screw system by Pioneer was designed referencing the following ASTM standards, similar to the predicate Zimmer Magna-FX, Mini Magna-FX and AO Hi Standardo, Similar to the production for Metallic Medical Bone Screws; and 2) F 116, Standard Specification for Medical Screwdriver Bits
#### Performance
r oreemianes
Based on the screw material equivalence in the proposed system an at least equivalent comparison has been logically derived from the information available.
#### Performance Data
The Cannulated Screw System by Pioneer is predicated on the use Zimmer Magna-FX and Synthes Cannulated screws. The Cannulated Screw system by Pioneer will be evaluated and compared to the predicate devices with the following ASTM Test standards: 1) F 1691, Standard Test Method for Determining the Axial Pull-Out Strength of Medical Bone Screws; 2) F 117, Standard Test Method for Driving Torque of Medical Bone Screws; and 3) F 1622, Standard Test Method for Measuring the Torsional Properties in Metallic Bone Screws.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an image of three stylized human profiles facing to the right.
Public Health Service
FEB = 8 2001
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Amy H. Mommaerts Manager, Regulatory Affairs Pioneer Surgical Technology 375 River Park Circle Marquette, Michigan 49855
Re: K003496 Trade Name: Cannulated Screw System Regulatory Class: II Product Codes: HTN and HWC Dated: November 10, 2000 Received: November 13 2000
### Dear Ms. Mommaerts:
We have reviewed your Section 510(k) notification of intent to market the device referenced we have reviewed your Sccubit 910(x) notifically equivalent (for the indications for use above and we have determined the correct in interstate commerce prior to May 28, 1976, the stated in the elieboute) to device Amendments, or to devices that have been reclassified in enactment date of the Medical Dovieo Frinentialized, brug, and Cosmetic Act (Act). The program accordance with the provisions of the reacted control provisions of the Act. The general therefore, market the device, subject to answal registration, listing of devices, control provisions of the fiel merilibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III 11 your device is classified (see acc re) and additional controls. Existing major regulations (Pleniance Apployal), it may of success to sates from the Code of Federal Regulations, Title 21, Parts 800 to 895. allecting your devroe can or remination assumes compliance with the current Good A substantially cquiralent covernent, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to mspections, the Food and Drug Hammesult in regulatory action. In addition, FDA may publish Comply with the GMI Tegatator may readeral Register. Please note: this response to your premarket notification submission does not affect any obligation you might It spollse to your promation houseaus of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket I his letter will anow you to ocgin maxantial equivalence of your device to a legally marketed nontication. The I Driving sification for your device and thus, permits your device to proceed to the market.
{3}------------------------------------------------
# Page 2 - Ms. Amy H. Mommaerts
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and If you desire specific advice for your devices), please contact the Office of Compliance at
additionally 809.10 for in vitro diagnostic devices), please contact the Office of additionally 809.10 for in Vitro diagnostic no nother promotion and advertising of your device, (301) 594-4659. Additionally, for questions of the pressent the regulation please contact the Office of Compliance at (301) ->++0.5: "(21 CFR 807.97). Other general
entitled, "Misbranding by reference to premarks obteined from the Division of Small entitled, "Misbranding by reference up perilance. Inonited from the Division of Small
information on your responsibilities under the Act may be obtained from the Division of information on your responsibilites under (800) 638-2041 or at (301) 443-6597, or at its
Manufacturers Assistance at its toll-free 111 Mandracturers 715siblantes av fda.gov/cdrh/dsmamain.html".
Sincerely yours,
L. Mark N. Milkman
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page of
510(k) NUMBER (IF KNOWN): K003496
DEVICE NAME: Cannulated Screw
INDICATIONS FOR USE:
The Cannulated Screw System by Pioneer will be indicated for use in long and small bone fracture fixation, which may include the following:
- 1. Fractures of the tarsals and metatarsals;
- 2. Metatarsal and phalangeal osteotomies;
- 3. Fractures of the carpals and metacarpals;
- 4. Carpal and metacarpal arthrodesis;
- 5. Small fragments of the hand and wrist;
- 6. Ligament fixation as appropriate;
- 7. Sacroiliac joint disruptions;
- 8. Fractures of the distal femur and proximal tibia;
- 9. Intracapsular fractures of the hip;
- 10. Ankle arthrodesis; and
- 11. Pelvis and acetabulum fractures;
This system is not indicated for use in the spine. The Cannulated Screw System by Pioneer will be offered both sterile and non-sterile and is a single use device.
AhMounacb- 01/22/01
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter-Use
(Optional Format 1-2-96)
for Mark N Melkers
(Division Sign-Off)
Division of General, Restorative
and Neurological Devices
K003496
510(k) Number.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.