K111831 · Bio-Med USA, Inc. · ONG · Dec 19, 2011 · General, Plastic Surgery
Device Facts
Record ID
K111831
Device Name
MX-7000 MICROXEL
Applicant
Bio-Med USA, Inc.
Product Code
ONG · General, Plastic Surgery
Decision Date
Dec 19, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
MX-7000 CO2 laser system is intended for use in non-fractionated mode for Incision, Excision, Ablation, Vaporization, and Coagulation of Human body soft Tissues. This system can be used in Dermatology, Plastic surgery, General Surgery, Gynecology, Neurosurgery, and in Podiatry. When used in fractionated mode, the MX-7000 CO2 system is intended to be used for ablative skin re-surfacing, Wrinkle, Fine lines, Ryhides, Furrow in Dermatology, in Plastic surgery, and in General surgery.
Device Story
MX-7000 is a CO2 laser system (10,600 nm) for soft tissue surgery and skin resurfacing. System components include an optic main bench assembly, articulated arm, LCD control panel, laser cooling system, handpieces (scanned/dynamic), and foot pedal. Operated by clinicians in clinical settings. Non-fractionated mode performs incision, excision, ablation, vaporization, and coagulation. Fractionated mode performs ablative skin resurfacing for wrinkles, fine lines, rhytides, and furrows. Clinician controls laser pulse power and selects handpieces to optimize effects for specific applications. Device provides surgical precision and controlled tissue interaction, benefiting patients through versatile surgical and dermatological treatment options.
Clinical Evidence
Bench testing only.
Technological Characteristics
CO2 laser system; 10,600 nm wavelength; fractional and non-fractional modes; articulated arm delivery; includes scanned and dynamic handpieces; cooling system; LCD control panel; foot pedal switch. Complies with IEC 60601-1 (safety) and IEC 60601-1-2 (EMC).
Indications for Use
Indicated for patients requiring soft tissue incision, excision, ablation, vaporization, or coagulation in dermatology, plastic surgery, general surgery, gynecology, neurosurgery, and podiatry; or ablative skin resurfacing for wrinkles, fine lines, rhytides, and furrows in dermatology, plastic surgery, and general surgery.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
K1118,31 bygewtf@
DEC 1 9 2011
# Summary of safety and effectiveness
In accordance with Section 513 (1) of the SMDA as defined in 21CFR part 807.3 this summary is submitted to obtain Pre-Market 510 (K) notification.
# 1. Submitter / Contact person.
Mr. Young Chi. President Bio-Med USA Inc. ( Reg Nr. 2246683) 111 Ellison Street, Paterson, NJ 07505. U.S.A. Tel: 1 973 278 5222 Fax: 201 934 6030 e mail : biomedusa(@msn.com
## 2. Manufacturer
Dae Shin Enterprise. 401/2, 170-5 Gruo-dong, Guro-gu, Seoul, Korea Tel: 2 2025 1151 Fax : 2 2025 1154 Contact: K. Kim President
#### 3. Name of Device
| Trade name | MX-7000 MICROXEL |
|-------------------------|--------------------------------------------------------------------------|
| Classification name | Powered, Laser surgical instrument with Micro-<br>beam/Fractional output |
| Common name | CO2 Laser system |
| Regulation | 880.4810 |
| Class | II |
| Product Code | ONG |
| Subsequent product code | GEX |
#### 4. Legally marketed Predicate Device
| K100610 | eCO2 Plus Laser | Lutronic |
|---------|-----------------|------------|
| K100590 | Edge CO2 Laser, | JM Systems |
### 5. Device Description
MX-7000 Laser system utilize a Co2 module to generate a laser beam with wave length of 10.600 nm and used fractional or non fractional mode for different indication. The Doctor can optimize the effect for different applications by controlling the power of laser pulse and using different hand pieces.
The system are supplied with two hand pieces ( Scanned or/and Dynamic),depend the device configuration
{1}------------------------------------------------
#### This system consist of
K111831 Page @ of @
Optic main Bench assembly, Articulated Arm LCD control panel, Laser Cooling system Hand pieces set, Foot Pedal switch.
Manufactured in accordance with both mandatory and voluntary standard
IEC 60601 Medical Electrical equipment part 1. General requirement for safety amend 2 1995 IEC60601-1-2 Electro magnetic compatibility test ED 2001 EMC test IEC60601-1-22-22: 1996 Particular requirement for the safety and Diagnostic of Therapeutic laser Equipment
#### 6. Intended use
MX-7000 CO2 laser system is indicated for use in non-fractionated mode is Incision, Excision, Ablation, Vaporization, and Coagulation of Human body soft Tissues. This system can be used in Dermatology, Plastic surgery, General surgery, Gynecology, Neurosurgery, and in Podiatry.
When used in fractionated mode, the MX-7000 CO2 system is indicated for use ablative skin re-surfacing, Wrinkle, Fine line, Rhylides, Furrow in Dermatology, Plastic surgery, and in General surgery.
#### 7. Conclusion.
The D.S.E, MX-7000 Co2 system, in this submission, is substantially equivalent to several already cleared predicate device in respect to Intended use, Main function, Technology, Principal operation and performance.
So, it does not raise any additional concerns regarding safety and effectiveness.
Bio-Med USA will update and include in this summary any other information deemed seasonally necessary by the FDA
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines forming its body and wings. The logo is surrounded by a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in capital letters.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
DEC 1 9 2011
Bio-Med USA, Inc. % Mr. Young Chi President 111 Ellison Street Paterson, New Jersey 07505
Re: K111831
Trade/Device Name: MX-7000, Microxel Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and Plastic surgery and in dermatology Regulatory Class: Class II Product Code: ONG, GEX Dated: September 02, 2011 Received: December 13, 2011
Dear Mr. Chi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Young Chi
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Sincerely yours,
Mark N. Melkerson
Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indication for use statement
510 (K) number : K111831
Device name : MX-7000, MICROXEL
Indication for use : MX-7000 CO2 laser system is intended for use in non-fractionated mode for Incision, Excision, Ablation, Vaporization, and Coagulation of Human body soft Tissues. This system can be used in Dermatology, Plastic surgery, General Surgery, Gynecology, Neurosurgery, and in Podiatry.
> When used in fractionated mode, the MX-7000 CO2 system is intended to be used for ablative skin re-surfacing, Wrinkle, Fine lines, Ryhides, Furrow in Dermatology, in Plastic surgery, and in General surgery.
Prescription use or/and Over the Counter use XX (Part 21 CFR 801 Sub part D ) (Part 21CFR 801 Sub part C)
Please do not write below line-continued an another pages if needed
Concurrence of CDRH, office of Device Evaluation ( ODE )
Nel R.P. Quiles for man
(Division Sign-C Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K111831
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.