LUTRONIC CORPORATION ECO2 PLUS, DENTA III AND III+ AND SP III LASER SYSTEMS
K100610 · Lutronic Corporation · ONG · Jan 20, 2011 · General, Plastic Surgery
Device Facts
Record ID
K100610
Device Name
LUTRONIC CORPORATION ECO2 PLUS, DENTA III AND III+ AND SP III LASER SYSTEMS
Applicant
Lutronic Corporation
Product Code
ONG · General, Plastic Surgery
Decision Date
Jan 20, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The eCO2 Plus Laser System with fractionated handpieces is indicated for use in dermatological procedures requiring ablation (removal), resurfacing and coagulation of soft tissue. Additionally, the 120 micron and 300 micron spot sizes are used in the treatment of wrinkles; rhytides, furrows, fine lines, textural irregularities, pigmented lesions and vascular dyschromia. The eCO2 Plus Laser System using the non-fractionated handpieces (F100, F50, and Zoom) is also indicated for use in surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in medical specialties including: dermatology, plastic surgery, podiatry, neurosurgery, gynecology, otorhinolaryngology (ENT), arthroscopy, (knee), and open endoscopic general surgery. The DENTA III Laser System using the non-fractionated handpieces (F100, F50, and Zoom) is indicated for use in soft tissue dental indications including periodontic procedures such as, but not limited to, removal of diseased or inflamed soft tissue in the periodontal pocket (sulcular debridement), vaporization, gingivectomy-removal of hyperplasias, gingivoplasty, papillectomy, vestibuloplasty, fibroma (nonmalignant tumor, mucosa, tongue), epulis, incision and excision, removal of soft tissue, cysts, and tumors, and laser assisted new attachment procedure (cementum-mediated periodontal ligament new attachment to the root surface in the absence of long junctional epithelium); Oral surgery such as frenectomy, frenum release, drainage (abscess), flap surgery, incisional and excisional biopsy, incision of aphthous ulcers, incision of infection when used with antibiotic therapy, excision and ablation of benign and malignant lesions, oral cavity tumors and hemangiomas, salivary gland pathologies, preprosthetic gum preparation, leukoplakia: partial glossectomy, periodontal gum resection, homeostasis, operculectomy, and crown lengthening. The DENTA III+ Laser System using fractionated handpieces is indicated for use in dermatological procedures requiring ablation (removal), resurfacing and coagulation of soft tissue. Additionally, the 120 micron and 300 micron spot sizes are used in the treatment of wrinkles; rhytides, furrows, fine lines, textural irregularities, pigmented lesions and vascular dyschromia. The DENTA III+ Laser System using non-fractionated handpieces (F100, F50, and Zoom) is indicated for use in soft tissue dental indications including periodontic procedures such as, but not limited to, removal of diseased or inflamed soft tissue in the periodontal pocket (sulcular debridement), vaporization, gingivectomy-removal of hyperplasias, gingivoplasty, papillectomy, vestibuloplasty, fibroma (nonmalignant tumor, mucosa, tongue), epulis, incision and excision, removal of soft tissue, cysts, and tumors, and laser assisted new attachment procedure (cementum-mediated periodontal ligament new attachment to the root surface in the absence of long junctional epithelium); Oral surgery such as frenectomy, frenum release, drainage (abscess), flap surgery, incisional and excisional biopsy, incision of aphthous ulcers, incision of infection when used with antibiotic therapy, excision and ablation of benign and malignant lesions, oral cavity tumors and hemangiomas, salivary gland pathologies, preprosthetic gum preparation, leukoplakia; partial glossectomy, periodontal gum resection, homeostasis, operculectomy, and crown lengthening. The SP III Laser System using the non-fractionated handpieces (F100, F50, and Zoom) is indicated for use in surgical applications requiring the ablation, vaporization, excision, and coagulation of soft tissue in medical specialties including: dermatology, plastic surgery, podiatry, neurosurgery, gynecology, otorhinolaryngology (ENT), arthroscopy, (knee), and open endoscopic general surgery.
Device Story
CO2 RF module generates 10.6 um wavelength laser beam; system utilizes interchangeable handpieces (fractionated vs. non-fractionated) to deliver energy. Physician controls laser pulse power to optimize tissue effects. Used in clinical settings (dermatology, plastic surgery, dentistry, etc.) by trained professionals. Output is laser energy for ablation, vaporization, excision, incision, or coagulation of soft tissue. Enables precise tissue removal or resurfacing; benefits include reduced recovery time or targeted treatment of lesions/pathologies compared to traditional methods.
Clinical Evidence
No clinical data presented; bench testing only.
Technological Characteristics
CO2 RF laser module; 10.6 um wavelength; modular handpiece system (fractionated and non-fractionated); power-controlled laser pulse; Class II device.
Indications for Use
Indicated for patients requiring soft tissue ablation, vaporization, excision, incision, or coagulation. Specialties include dermatology, plastic surgery, podiatry, neurosurgery, gynecology, ENT, arthroscopy, general surgery, and dentistry (periodontics/oral surgery). Specific conditions include wrinkles, rhytids, furrows, pigmented/vascular lesions, skin tags, scars, keloids, hemangiomas, tattoos, carcinomas, warts, and various dental/periodontal pathologies.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Lutronic Corporation Spectra SP II and Spectra DENTA II Laser Systems (K091320)
Submission Summary (Full Text)
{0}------------------------------------------------
100610 pg 1 6
## 510(k) Summary for the Lutronic Corporation eCO2 Plus, DENTA III, SP III Laser Systems
JAN 2 0 2011
This 510(k) Summary is being submitted in accordance with the requirements of the SMDA 1990, 21 CFR 807.92, and 21 CFR 878.4810.
### 1. General Information
Submitter:
Lutronic Corporation #403-2,3,4, Ilsan Technotown 1141-1 Baeksok-Dong, Ilsan-Gu Goyang-Si, Gyeonggi-Do, 410-722 Republic of Korea
Contact Person:
Maureen O'Connell O'Connell Regulatory Consultants, Inc. 5 Timber Lane North Reading, MA 01864 Telephone: 978-207-1245 Fax: 978-824-2541
#### Summary Preparation Date:
2. Names
Device Name:
Classification Name:
eCO2 Plus Laser System DENTA III Laser System DENTA III+ Laser System SP III Laser System
January 19, 2011
Powered Laser Surgical Instrument With Microbeam\Fractional Output Laser Instrument, Surgical, Powered Product Code: ONG GEX Panel: General & Plastic Surgery Regulation number: 21 CFR 878.4810
#### 3. Predicate Devices
The eCO2 Plus, DENTA III, DENTA III+ and SP III Laser Systems are substantially equivalent to a combination of the Lutronic Corporation eCO2 Laser System (K091115) and the Lutronic Corporation Spectra SP II and Spectra DENTA II Laser Systems (K091320).
{1}------------------------------------------------
K100610
2 of 5
## 4. Device Description
The eCO2 Plus, SP III, DENTA III and DENTA III+ Laser Systems utilize a CO2 RF module to generate a laser beam with a wavelength of 10.6 um and use different handpiece for different indications for use. The physician can optimize the effect for different applications by controlling the power of the laser pulse and using a different handpiece. The Laser Systems are supplied with different handpieces depending upon the device configuration.
## 5. Indications for Use
The eCO2 Plus Laser System with fractionated handpieces is indicated for use in dermatological procedures requiring ablation (removal), resurfacing and coagulation of soft tissue. Additionally, the 120 micron and 300 micron spot sizes are used in the treatment of wrinkles; furrows, fine lines, textural irregularities, pigmented lesions and vascular dyschromia.
The eCO2 Plus Laser System using the non-fractionated handpieces (F100, F50, and Zoom) is also indicated for use in surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in medical specialties including: dermatology, plastic surgery, podiatry, neurosurgery, gynecology, otorhinolaryngology (ENT), arthroscopy, (knee), and open endoscopic general surgery.
## Dermatology & Plastic Surgery
The ablation, vaporization, excision, incision, and coagulation of soft tissue in dermatology and plastic surgery in the performance of:
- Laser skin resurfacing .
- . Treatment of wrinkles, rhytids and furrows
- Ablation and/or vaporization of soft tissue in dermatology and plastic surgery for the . reduction, removal, and/or treatment of actinic keratosis, skin tags, solar/actinic elastosis, actinic cheilitis, lentigines, uneven pigmentation/dyschromia, acne scars, surgical scars, keloids, hemangiomas (including buccal hemangiomas), tattoos, telangiectasia, squamous and basal cell carcinoma, spider and epidermal naevi, xanthelasma palpebrarum, syringoma, and verrucae and seborrhoecae vulgares (warts); laser derm-ablation; and laser burn debridement.
### Dermatology, Plastic Surgery & General Surgery
Laser incision and/or excision of soft tissue in dermatology, plastic and general surgery, including the performance of blepharoplasty and for the creation of recipient sites for hair transplantation, treatment of hemorrhoids, atheroma, cysts, abscesses, and all other soft tissue applications.
## Podiatry
Laser ablation, vaporization, and/or excision of soft tissue in podiatry for the reduction, removal, and/or treatment of verrucae vulgares, and matrixectomy.
{2}------------------------------------------------
100610 pg 3 of 5
#### Otorhinolaryngology (ENT)
Laser incision, excision, ablation and/or vaporization of soft tissue in otorhinolaryngology for the treatment of choanal atresia, leukoplakia of larynx, nasal obstruction, UPP, rhinophyma, adult and juvenile papillomatosis polyps, rhinophyma and verrucae vulgares.
#### Gynecology
Laser incision, excision, ablation and/or vaporization of soft tissue in gynecology for the treatment of cervical intraepithelial neoplasia, condyloma acuminata, leukoplakia (vulvar dystrophies) and vulvar and vaginal intraepithelial neoplasia.
#### Neurosurgery
Laser incision, excision, ablation and/or vaporization of soft tissue in neurology for the treatment of basal turnor-meningioma, posterior fossa tumors, peripheral neurectomy, and lipomas/large tumors.
The DENTA III Laser System using the non-fractionated handpieces (F100, F50, and Zoom) is indicated for use in soft tissue dental indications including periodontic procedures such as, but not limited to, removal of diseased or inflamed soft tissue in the periodontal pocket (sulcular debridement), vaporization, gingivectomy-removal of hyperplasias, gingivoplasty, papillectomy, vestibuloplasty, fibroma (nonmalignant tumor, mucosa, tongue), epulis, incision and excision, removal of soft tissue, cysts, and tumors, and laser assisted new attachment procedure (cementum-mediated periodontal ligament new attachment to the root surface in the absence of long junctional epithelium); Oral surgery such as frenectomy, frenum release, drainage (abscess), flap surgery, incisional and excisional biopsy, incision of aphthous ulcers, incision of infection when used with antibiotic therapy, excision and ablation of benign and malignant lesions, oral cavity tumors and hemangiomas, salivary gland pathologies, preprosthetic gum preparation, leukoplakia: partial glossectomy, periodontal gum resection, homeostasis, operculectomy, and crown lengthening.
The DENTA III+ Laser System using fractionated handpieces is indicated for use in dermatological procedures requiring ablation (removal), resurfacing and coagulation of soft tissue. Additionally, the 120 micron and 300 micron spot sizes are used in the treatment of wrinkles; rhytides, furrows, fine lines, textural irregularities, pigmented lesions and vascular dyschromia.
The DENTA III+ Laser System using non-fractionated handpieces (F100, F50, and Zoom) is indicated for use in soft tissue dental indications including periodontic procedures such as, but not limited to, removal of diseased or inflamed soft tissue in the periodontal pocket (sulcular debridement), vaporization, gingivectomy-removal of hyperplasias, gingivoplasty, papillectomy, vestibuloplasty, fibroma (nonmalignant tumor, mucosa, tongue), epulis, incision and excision, removal of soft tissue, cysts, and turnors, and laser assisted new attachment procedure (cementum-mediated periodontal ligament new attachment to the root surface in the absence of long junctional epithelium); Oral surgery such as frenectomy, frenum release, drainage (abscess), flap surgery, incisional and excisional biopsy, incision and excision of aphthous ulcers, incision of infection when used with antibiotic therapy, excision and ablation of benign and malignant lesions, oral cavity tumors and hemangiomas, salivary gland pathologies, preprosthetic gum
{3}------------------------------------------------
# (100610
preparation, leukoplakia; partial glossectomy, periodontal gum resection, homeostasis, operculectomy, and crown lengthening.
The SP III Laser System using the non-fractionated handpieces (F100, F50, and Zoom) is indicated for use in surgical applications requiring the ablation, vaporization, excision, and coagulation of soft tissue in medical specialties including: dermatology, plastic surgery, podiatry, neurosurgery, gynecology, otorhinolaryngology (ENT), arthroscopy, (knee), and open endoscopic general surgery.
### Dermatology & Plastic Surgery
The ablation, vaporization, excision, and coagulation of soft tissue in dermatology and plastic surgery in the performance of:
- . Laser skin resurfacing
- . Treatment of wrinkles, rhytids and furrows
- . Ablation and/or vaporization of soft tissue in dermatology and plastic surgery for the reduction, removal, and/or treatment of actinic keratosis, skin tags, solar/actinic elastosis, actinic cheilitis, lentigines, uneven pigmentation/dyschromia, acne scars, surgical scars, keloids, hemangiomas (including buccal hemangiomas), tattoos, telangiectasia, squamous and basal cell carcinoma, spider and epidermal naevi, xanthelasma palpebrarum, syringoma, and verrucae and seborrhoecae vulgares (warts); laser derm-ablation; and laser burn debridement.
## Dermatology, Plastic Surgery & General Surgery
Laser incision and/or excision of soft tissue in dermatology, plastic and general surgery, including the performance of blepharoplasty and for the creation of recipient sites for hair transplantation, treatment of hemorrhoids, atheroma, cysts, abscesses, and all other soft tissue applications.
### Podiatry
Laser ablation, vaporization, and/or excision of soft tissue in podiatry for the reduction, removal, and/or treatment of verrucae vulgares, and matrixectomy.
### Otorhinolaryngology (ENT)
Laser incision, excision, ablation and/or vaporization of soft tissue in otorhinolaryngology for the treatment of choanal atresia, leukoplakia of larynx, nasal obstruction, UPP, rhinophyma, adult and juvenile papillomatosis polyps, rhinophyma and verrucae vulgares.
### Gynecology
Laser incision, excision, ablation and/or vaporization of soft tissue in gynecology for the treatment of cervical intraepithelial neoplasia, condyloma acuminata, leukoplakia (vulvar dystrophies) and vulvar and vaginal intraepithelial neoplasia.
### Neurosurgery
Laser incision, excision, ablation and/or vaporization of soft tissue in neurology for the treatment of basal tumor-meningioma, posterior fossa tumors, peripheral neurectomy, and lipomas/large tumors.
{4}------------------------------------------------
100610 Pg 5 of 5
1000
# 6. Performance Data None presented.
.
・ ・ . ・ :
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows a logo for the Department of Health & Human Services. The logo consists of a stylized eagle with three curved lines representing its wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES" are arranged in a semi-circle around the left side of the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-WO66-G609 Silver Spring, MD 20993-0002
Lutronic Corporation % O'Connell Regulatory Consultants, Inc. Ms. Maureen O'Connell 5 Timber Lane North Reading, Massachusetts 01864
JAN 2 0 2011
Re: K100610
Trade/Device Name: eCO2 Plus Laser System Fractionated and Non-Fractionated Handpieces; DENTA III Laser System - Non-Fractionated Handpieces; DENTA III+ Laser System Fractionated and Non-Fractionated Handpieces; SP III Laser System - Non-Fractionated Handpieces Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: ONG, GEX Dated -- January-03-201-1-Received: January 04, 2011
Dear Ms. O'Connell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
{6}------------------------------------------------
Page 2 -- Ms. Maureen O'Connell
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Ai
Dr. B. 2 h
Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
### Enclosure
{7}------------------------------------------------
510(k) Number (if known): K100610
Device Name: eCO2 Plus Laser System-Fractionated Handpieces
## Indications for Use:
The eCO2 Plus Laser System with fractionated handpieces is indicated for use in dermatological procedures requiring ablation (removal), resurfacing and coagulation of soft tissue. Additionally, the 120 micron and 300 micron spot sizes are used in the treatment of wrinkles; rhytides, furrows, fine lines, textural irregularities, pigmented lesions and vascular dyschromia.
Neil R P Ogden for mxm
Division Sign-Of Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number < 100610
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over The Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 8
{8}------------------------------------------------
K100610
510(k) Number (if known):
Device Name: eCO2 Plus Laser System-Non-Fractionated Handpieces
## Indications for Use:
The eCO2 Plus Laser System using the non-fractionated handpieces (F100, F50, and Zoom) is also indicated for use in surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in medical specialties including: dermatology, plastic surgery, podiatry, neurosurgery, gynecology, otorhinolaryngology (ENT), arthroscopy, (knee), and open endoscopic general surgery.
## Dermatology & Plastic Surgery
The ablation, vaporization, excision, incision, and coagulation of soft tissue in dermatology and plastic surgery in the performance of:
- Laser skin resurfacing .
- . Treatment of wrinkles, rhytids and furrows
- . Ablation and/or vaporization of soft tissue in dermatology and plastic surgery for the reduction. removal, and/or treatment of actinic keratosis, skin tags, solar/actinic elastosis, actinic cheilitis, lentigines, uneven pigmentation/dyschromia, acne scars, surgical scars, keloids, hemangiomas (including buccal hemangiomas), tattoos, telangiectasia, squamous and basal cell carcinoma, spider and epidermal naevi, xanthelasma palpebrarum, syringoma, and verrucae and seborrhoecae vulgares (warts); laser derm-ablation; and laser burn debridement.
## Dermatology, Plastic Surgery & General Surgery
Laser incision and/or excision of soft tissue in dermatology, plastic and general surgery, including the performance of blepharoplasty and for the creation of recipient sites for hair transplantation, treatment of hemorrhoids, atheroma, cysts, abscesses, and all other soft tissue applications.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over The Counter Use _ (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
*Nytke Ogle* *for mxm*
(Division Sign-Off) Page 2 of 8
Division of Surgical, Orthopedic,
and Restorative Devices
CONFIDENTIAL 510(k) Number K100610
{9}------------------------------------------------
K100610
## Podiatry
Laser ablation, vaporization, and/or excision of soft tissue in podiatry for the reduction, removal, and/or treatment of verrucae vulgares, and matrixectomy.
### Otorhinolaryngology (ENT)
Laser incision, excision, ablation and/or vaporization of soft tissue in otorhinolaryngology for the treatment of choanal atresia, leukoplakia of larynx, nasal obstruction, UPP, rhinophyma, adult and juvenile papillomatosis polyps, rhinophyma and verrucae vulgares.
## Gynecology
Laser incision, excision, ablation and/or vaporization of soft tissue in gynecology for the treatment of cervical intraepithelial neoplasia, condyloma acuminata, leukoplakia (vulvar dystrophies) and vulvar and vaginal intraepithelial neoplasia.
## Neurosurgerv
Laser incision, excision, ablation and/or vaporization of soft tissue in neurology for the treatment of basal tumor-meningioma, posterior fossa tumors, peripheral neurectomy, and lipomas/large tumors.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over The Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nulal dylen for ma Page 3 of 8.
(Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
CONFIDENTIAL
510(k) Number K100610
Page 13
{10}------------------------------------------------
K100610
510(k) Number (if known):
Device Name: DENTA III Laser System-Non-Fractionated Handpieces
#### Indications for Use:
The DENTA III Laser System using the non-fractionated handpieces (F100, F50, and Zoom) is indicated for use in soft tissue dental indications including periodontic procedures such as, but not limited to, removal of diseased or inflamed soft tissue in the periodontal pocket (sulcular debridement), vaporization, gingivectomy-removal of hyperplasias, gingivoplasty, papillectomy, vestibuloplasty, fibroma (nonmalignant tumor, mucosa, tongue), epulis, incision and excision, removal of soft tissue, cysts, and tumors, and laser assisted new attachment procedure (cementum-mediated periodontal ligament new attachment to the root surface in the absence of long junctional epithelium); Oral surgery such as frenectomy, frenum release, drainage (abscess), flap surgery, incisional and excisional biopsy, incision of aphthous ulcers, incision of infection when used with antibiotic therapy, excision and ablation of benign and malignant lesions, oral cavity tumors and hemangiomas, salivary gland pathologies, preprosthetic gum preparation, leukoplakia; partial glossectomy, periodontal gum resection, homeostasis, operculectomy, and crown lengthening.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over The Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| | <div style="text-align:center;"></div> |
|-----------------------------------|----------------------------------------|
| (Division Sign-Off) | Page 4 of 8 |
| Division of Surgical, Orthopedic, | |
| and Restorative Devices | |
CONFIDENTIAL
510(k) Number K100610
{11}------------------------------------------------
K100610
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: DENTA III+ Laser System-Fractionated Handpieces
### Indications for Use:
The DENTA III+ Laser System using fractionated handpieces is indicated for use in dermatological procedures requiring ablation (removal), resurfacing and coagulation of soft tissue. Additionally, the 120 micron and 300 micron spot sizes are used in the treatment of wrinkles; rhytides, furrows, fine lines, textural irregularities, pigmented lesions and vascular dyschromia.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over The Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
for 15 x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x Page 5 of 8
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
CONFIDENTIAL
510(k) Number K100610
{12}------------------------------------------------
Klooglo
510(k) Number (if known):
Device Name: DENTA III+ Laser System-Non-Fractionated Handpiece
#### Indications for Use:
The DENTA III+ Laser System using non-fractionated handpieces (F100, F50, and Zoom) is indicated for use in soft tissue dental indications including periodontic procedures such as, but not limited to, removal of diseased or inflamed soft tissue in the periodontal pocket (sulcular debridement), vaporization, gingivectomy-removal of hyperplasias, gingivoplasty, papillectomy, vestibuloplasty, fibroma (nonmalignant tumor, mucosa, tongue), epulis, incision and excision, removal of soft tissue, cysts, and tumors, and laser assisted new attachment procedure (cementum-mediated periodontal ligament new attachment to the root surface in the absence of long junctional epithelium); Oral surgery such as frenectomy, frenum release, drainage (abscess), flap surgery, incisional and excisional bionsy, incision of aphthous ulcers, incision of infection when used with antibiotic therapy, excision and ablation of benign and malignant lesions, oral cavity tumors and hemangiomas, salivary gland pathologies, preprosthetic gum preparation, leukoplakia; partial glossectomy, periodontal gum resection, homeostasis, operculectomy, and crown lengthening.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over The Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Neiller for more
Page 6 of 6
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
CONFIDENTIAL
| 510(k) Number_________________________________________________________________________________________________________________________________________________________________ | ﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮ | Klooblo | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
Page 16
{13}------------------------------------------------
boo610
510(k) Number (if known):
Device Name: SP III Laser System-Non-Fractionated Handpieces
## Indications for Use:
The SP III Laser System using the non-fractionated handpieces (F100, F50, and Zoom) is indicated for use in surgical applications requiring the ablation, vaporization, excision, and coagulation of soft tissue in medical specialties including: dermatology, plastic surgery, podiatry, neurosurgery, gynecology, otorhinolaryngology (ENT), arthroscopy, (knee), and open endoscopic general surgery.
# Dermatology & Plastic Surgerv
The ablation, vaporization, excision, incision, and coagulation of soft tissue in dermatology and plastic surgery in the performance of:
- Laser skin resurfacing .
- . Treatment of wrinkles, rhytids and furrows
- Ablation and/or vaporization of soft tissue in dermatology and plastic surgery for the . reduction, removal, and/or treatment of actinic keratosis, skin tags, solar/actinic elastosis, actinic cheilitis, lentigines, uneven pigmentation/dyschromia, acne scars, surgical scars, keloids, hemangiomas (including buccal hemangiomas), tattoos, telangiectasia, squamous and basal cell carcinoma, spider and epidermal naevi, xanthelasma palpebrarum, syringoma, and verrucae and seborrhoecae vulgares (warts); laser derm-ablation; and laser burn debridement.
# Dermatology, Plastic Surgery & General Surgery
Laser incision and/or excision of soft tissue in dermatology, plastic and general surgery, including the performance of blepharoplasty and for the creation of recipient sites for hair transplantation, treatment of hemorrhoids, atheroma, cysts, abscesses, and all other soft tissue applications.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over The Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil R. Dahlen for mkn
Page 7 of 8
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
CONFIDENTIAL
510(k) Number K100610
{14}------------------------------------------------
Kbo Glo
## Podiatry
Laser ablation, vaporization, and/or excision of soft tissue in podiatry for the reduction, removal, and/or treatment of verrucae vulgares, and matrixectomy.
## Otorhinolaryngology (ENT)
Laser incision, excision, ablation and/or vaporization of soft tissue in otorhinolaryngology for the treatment of choanal atresia, leukoplakia of larynx, nasal obstruction, UPP, rhinophyma, adult and juvenile papillomatosis polyps, rhinophyma and verrucae vulgares.
## Gynecology
Laser incision, excision, ablation and/or vaporization of soft tissue in gynecology for the treatment of cervical intraepithelial neoplasia, condyloma acuminata, leukoplakia (vulvar dystrophies) and vulvar and vaginal intraepithelial neoplasia.
## Neurosurgerv
Laser incision, excision, ablation and/or vaporization of soft tissue in neurology for the treatment of basal tumor-meningioma, posterior fossa tumors, peripheral neurectomy, and lipomas/large tumors.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over The Counter Use -(Part 21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 8 of 8
Neil R. Ogden for mxm
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K100610
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.