K111531 · Annidis Health Systems Corp. · HLI · Jul 8, 2011 · Ophthalmic
Device Facts
Record ID
K111531
Device Name
RHA2020
Applicant
Annidis Health Systems Corp.
Product Code
HLI · Ophthalmic
Decision Date
Jul 8, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1570
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The Annidis RHA2020-U multi-spectral digital ophthalmoscope is intended to capture images of the fundus of the eye which can be used to assist in diagnosis and observation of fundus diseases.
Device Story
RHA2020-U is a tabletop multi-spectral digital ophthalmoscope for retinal imaging. System comprises an Optical Head Unit (OHU) with LED illumination and camera, a Linux-based host computer, a universal power supply, and a touch screen display. Patient head is stabilized via chin-rest and forehead brace. Device captures multi-spectral retinal images (green to near-infrared, 6 bands) and images of the iris and pupil. Practitioners use the system to visualize and track retinal data to assist in diagnosing and observing fundus diseases. Output is displayed on a monitor for clinical review. Benefits include early detection and improved care through high-resolution spectral imaging.
Clinical Evidence
Bench testing only. Performance verified through comparison of digital images captured by the RHA2020-U against those from predicate devices (Optos P200 and Nidek F-10). Testing included images from both non-mydriatic and mydriatic human eyes. Results confirmed image comparability and adherence to functional specifications.
Technological Characteristics
Tabletop ophthalmoscope; LED light source; 6 spectral bands (green to near-infrared); 42-degree field of view; 4+ megapixel resolution. Linux-based host computer. Complies with IEC 60601-1, IEC 60601-1-2, ISO 15004-2, and UL 60601-1 standards.
Indications for Use
Indicated for use by optometrists and ophthalmologists to capture fundus images of the eye to assist in the diagnosis and observation of fundus diseases in patients.
Regulatory Classification
Identification
An ophthalmoscope is an AC-powered or battery-powered device containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, and vitreous) and the retina of the eye.
Special Controls
*Classification.* Class II (special controls). The device, when it is an AC-powered opthalmoscope, a battery-powered opthalmoscope, or a hand-held ophthalmoscope replacement battery, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}------------------------------------------------
.
Image /page/0/Picture/1 description: The image shows the logo for Annidis Health Systems. The logo features a stylized eye symbol to the left of the text "ANNIDIS". Below "ANNIDIS" is the text "HEALTH SYSTEMS" and "Promoting early detection and care" in a smaller font. The logo is black and white.
# RHA SUMMARY
Annidis Health Systems Corp., 2650 Queensview Drive, Suite 245, Ottawa, Ontario, K2B 8H6 Tel: 613.596.1800 Fax: 613.596.9453
**510(k) Number K**//153/
### 1. Sponsor Information
| Applicant Name: | Annidis Health Systems Corp.<br>2650 Queensview Drive, Suite 245,<br>Ottawa, Ontario, K2B 8H6, Canada<br>Tel: 613.596.1800; Fax: 613.596.9453<br>Email: mmcdonnell@annidis.com |
|-----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Michael R McDonnell |
|-----------------|----------------------------------------------------|
| | Annidis Health Systems Corp. |
| | 2650 Queensview Drive, Suite 245, Ottawa, Ontario, |
| | K2B 8H6, Canada |
| | Tel: 613.596.1800 Ext 2019; Fax: 613.596.9453 |
| | Email: mmcdonnell@annidis.com |
Date of Summary: Mar 31st, 2011
### 2. Device Name and Classification
| Common Name: | RHA Multi-spectral Digital Ophthalmoscope |
|--------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| Trade Name: | RHA |
| Model: | RHA2020-U |
| Classification Name (Primary):<br>Classification Name (Secondary): | CFR Classification section 886.1570 (HLI)<br>CFR Classification section 886.1120 (HKI) |
| Classification: | Class II medical device |
## 3. Predicate Devices:
The RHA multi-spectral digital ophthalmoscope is substantially equivalent to a combination of the following predicate devices:
.
| 510(k) No. | Trade Name | Product Code(s) |
|------------|------------------------------|-----------------|
| K983999 | Panoramic 200 Ophthalmoscope | HLI |
| K072259 | Ophthalmoscope F-10 | MYC |
| K092056 | Smartscope M3-1 EY1 | HKI |
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Annidis Health Systems. The logo features a stylized eye inside of a triangle on the left. To the right of the eye is the text "ANNIDIS" in large, bold letters, with the words "HEALTH SYSTEMS" in smaller letters below. Below that is the text "Promoting early direction and care".
Innidis Health Systems Corp. 2650 Queensview Drive, Suite 245, Ottawa, Ontario, K2B 8H6 Tel: 613.596.1800 Fax: 613.596.9453
#### 4. Device Description:
The Annidis RHA2020U multi-spectral digital ophthalmoscope which presents eye care practitioners (optometrists and ophthalmologists) with a series of retinal spectral images from each eye of a patient. The images each have a spatial resolution of over four million pixels spread over a visual range of greater than 40 degrees. The instrument also captures an image of the iris and pupil.
The device is arranged in four major hardware subcomponents and one remote software component used for clinical visualization and tracking purposes. The four hardware components are listed below:
- ア The Optical Head Unit (OHU) contains the camera and the illuminating LEDs and is aligned to the eye of the sitting patient whose head is stabilized using a chin-rest and forehead brace. The OHU contains the optical and electrical systems required to capture images of the eye and the means to align the patient with the device.
- A The Host Computer (HC), a Linux-based host computer that serves as the operator interface.
- The Universal Power Supply (UPS), a custom designed power supply receives A power from the AC mains and supplies power using low voltage DC to the host computer, the display, and the optical head unit (OHU).
The Touch Screen Display, that displays information from the host computer, configures the optical head unit for patient comfort, visualizes the patient data and triggers image capture.
#### 5. Intended Use:
The Annidis RHA2020-U multi-spectral digital ophthalmoscope is intended to capture images of the fundus of the eye which can be used to assist in diagnosis and observation of fundus diseases
#### 6. Predicate Devices:
The Annidis retinal health assessment device is substantially equivalent to a combination of the following predicate devices:
- . The Optos Panoramic 200 (K983999) is a scanning laser ophthalmoscope (SLO). It uses a laser as a light source that is scanned by a deflection system in two axes across the retina of the eye to generate a colour image. The returned light then travels back along the same path to a light detector that converts the light to an electrical signal. The electrical signal is digitized and used to build up an
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for ANNIDIS Health Systems. The logo features a stylized letter "A" with an eye inside of it. To the right of the "A" is the text "ANNIDIS" in bold, followed by "HEALTH SYSTEMS" and "Promoting early detection and care" in smaller fonts.
Annidis Health Systems Corp., 2650 Queensview Drive, Suite 245, Ottawa, Ontario, K2B 8H6 Tel: 613.596.1800 Fax: 613.596.9453
electronic picture in a personal computer and displayed either on a cathode ray tube or on a liquid crystal display.
- . The Nidek Ophthalmoscope F-10 (K072259) is categorized as a Class I confocal laser scanning ophthalmoscope. The F-10 captures and records confocal images of the fundus by laser scanning using a selection of laser colors: IR (infrared). blue, green, and red and affords Indocyanine green (ICG) and fluorescein (FAG) angiography.
- . The Optomed Smartscope M3-1 EY1 (K092056) is a hand-held digital ophthalmoscope used to capture digital images and video of the cornea, aqueous, lens, vitreous and retina of the human eye. M3-1 EY 1 has a LED light source. Image data is stored on the Flash memory card using 6.6 megapixel CMOS sensor and transferred to the PC by using USB connection. The device has rechargeable batteries.
| | Annidis<br>RHA | Nidek<br>F-10 | Optos<br>P200 | Optomed<br>Smartscope<br>3M-1 EY1 |
|----------------------------------|---------------------------------|----------------------------|--------------------------------------------|-----------------------------------|
| field of view (degrees) | 42 | 40 | >90 | 30 |
| physical configuration | tabletop with<br>headrest | tabletop with<br>headrest | stand alone and<br>patient<br>manipulation | handheld |
| spectral range<br>covered | green-near<br>infrared | blue-near<br>infrared | green & red | visible |
| number of spectral<br>bands | 6 | 4 | 2 | ~3 |
| type of source | LED | laser | laser | LED |
| image construction<br>process | imaging | scan | scan | imaging |
| indicated for retinal<br>imaging | yes | yes | yes | yes |
| viewing software | software accessory | medical device<br>(NAVIS) | software accessory | software<br>accessory |
| product classification | CFR 886.1570 | CFR 886.1570 | CFR 886.1570 | CFR 886.1120 |
| product code | HLI (primary)<br>HKI(Secondary) | MYC (could<br>also be HLI) | HLI | HKI |
#### Table 5-1: Table of similarities between the RHA and the predicate devices
#### 7. Comparison with Predicate Devices
Substantial equivalence of the Annidis RHA Multi-Spectral Ophthalmoscope is addressed relative to three legally marketed predicate devices. The Nidek F-10 and the Optos P-200 are scanning laser ophthalmoscopes, and the Optomed Smartscope is a
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Annidis Health Systems. The logo features a stylized triangle shape on the left, with a design resembling an eye within it. To the right of the triangle is the text "ANNIDIS" in bold, uppercase letters, with the words "HEALTH SYSTEMS" below it in a smaller font. Below that is the text "Promoting early direction and care".
Annidis Health Systems Corp., 2650 Queensview Drive, Suite 245, Ottawa, Ontario, K2B 8H6 Tel: 613.596.1800 Fax: 613.596.9453
handheld digital ophthalmoscope, The data set available for the predicate device parameters has been compared to the similar data for the RHA to provide evidence for the substantial equivalence of the RHA to the legally marketed devices. Based on the data and comparisons presented in this section, the RHA is substantially equivalent to the legally marketed predicate devices. It is similar in characteristics, materials, has similar technological features, intended use and indications for use as the predicates, and does not raise any new questions of safety and effectiveness.
#### 8. Performance Standards:
No applicable performance standards have been issued under the Food, Drug and Cosmetic Act.
The RHA retinal health assessment device has been tested by an accredited third party test laboratory and conforms to the following voluntary standards:
- A IEC 60601-1:1988 Medical Electrical Equipment; Part 1: General Requirements for Safety + A1:1991 and A2:1995, including Appendix - National Deviations -Test Results according to ONLINE CB BULLETIN as of June 2009.
- A IEC60601-1-1:2000 Medical Electrical Equipment - Part 1: General Requirements For Safety 1: Collateral Standard: Safety Requirements for Medical Electrical Systems, 2nd Edition.
- A IEC 60601-1-2:2001 Medical Electrical Equipment Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility -Requirements and Tests + A1:2004
- A ISO 15004-2:2007 Ophthalmic Instruments - Fundamental Requirements and Test Methods - Part 2: Light Hazard Protection.
- A UL 60601-1 (first Edition, 2003): Medical Electrical Requirements Part 1: General requirements for Safety.
- > IEC60950-1 (First Edition 2001) A11:2004: Information Technology Equipment - Safety- Part 1: General Requirements.
- 人 CAN / CSA C22.2 No. 601.1-M90 Medical Electrical Equipment - Part 1: General Requirements for Safety (Adopted IEC 601-1, Amendment 1:1991)
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Annidis Health Systems. The logo features a stylized eye symbol to the left of the word "ANNIDIS" in large, bold letters. Below the name, the words "HEALTH SYSTEMS" are printed in a smaller font, followed by the tagline "Promoting early detection and care".
alth Systems Corp. 2650 Queensview Drive, Suite 245, Ottawa, Ontario, K2B 8H6 Tel: 613.596.1800 Fax: 613.596.9453
### 9. Non-Clinical Performance Summary:
Annidis Health Systems Corp. has verified and validated that the RHA Multi-Spectral Digital Ophthalmoscope meets its functional specifications and performance requirements.
The requirements for the RHA are set out in the Product Requirements Specifications, and addressed by Performance Testing Reports and are illustrated through the use of Performance Traceability Matricies. The RHA is controlled mostly by software and firmware to which performance testing was conducted. The test management process used was an industry off-the-shelf validated management tool.
The RHA2020-U Multi-Spectral Digital Ophthalmoscope was subjected to Performance Testing and the image output of the RHA2020-U device was compared with two predicate devices. The results of Performance Testing demonstrates substantial equivalence between the image output (i.e.. subject digital images captured) of the two selected predicate devices and those of the digital images captured using the RHA2020-U device when reviewed by the Performance Test Investigator. Submitted data included:
- . Digital images evaluated from digital scan output taken of non-mydriatic eyes of human subjects.
- . Digital images evaluated from digital scan output taken of mydriatic eyes of human subjects.
It is concluded that in a direct comparison of images between subject and predicate devices, that the images are comparable and meet performance requirements for acceptance and the intended use.
The RHA Multi-Spectral Ophthalmoscope complies with applicable U.S. and international electrical, electromagnetic compatibility and light safety standards (as outlined above) for products of its kind.
### 10. Conclusions
Based on the information contained in this submission, similarity to the predicate devices (the Optos Panoramic 200, the Nidek Ophthalmoscope F-10, and the Optomed Smartscope M3-1 EY1), and the results of our design control activities and non-clinical performance testing, has been established.
In summary, Annidis Health Systems Corp. believes that the RHA Multi-Spectral Ophthalmoscope is as effective, and performs as well as devices currently on the market, and concludes that the RHA Multi-Spectral Ophthalmoscope is substantially equivalent to the predicate devices.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Annidis Health Systems Corp. c/o William Sammons Third Party Reviewer Intertek Testing Services NA, Inc. 2307 E. Aurora Rd. Unit B7 Twinsburg, OH 44087
JUL - 8 2011
Re: K11531
Trade/Device Name: RHA Multi-spectral Digital Ophthalmoscope Model RHA2020-U Regulation Number: 21 CFR 886.1570 Regulation Name: Opthalmoscope Regulatory Class: Class II Product Code: HLI and HKI Dated: June 1, 2011 Received: June 2, 2011
Dear Mr. Sammons:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Kenia Alexander
Image /page/6/Picture/8 description: The image shows a handwritten cursive "for" on the left side of the image. To the right of the cursive text are three lines of text that are cut off. The text is in a serif font and is black.
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Annidis 510(k)
## INDICATIONS FOR USE
510(k) Number (if known):
Device Name:
RHA2020-U multi-spectral digital ophthalmoscope
Indications for Use:
The Annidis RHA2020-U multi-spectral digital ophthalmoscope is intended to capture images of the fundus of the eye which can be used to assist in diagnosis and observation of fundus diseases.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Thoms Axhm
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number K///53/
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.