The Ophthalmoscope F-10 is intended for use in capturing images of the fundus, which can be used for the diagnosis of fundus diseases.
Device Story
The Ophthalmoscope F-10 is a Class 1 confocal laser scanning ophthalmoscope (SLO). It captures and records high-resolution confocal images of the ocular fundus using laser scanning technology. The device utilizes multiple laser wavelengths—infrared (IR), blue, green, and red—to facilitate standard fundus imaging as well as Indocyanine green (ICG) and fluorescein (FAG) angiography. Operated by eye care professionals in a clinical setting, the device transforms reflected laser light into digital images. These images are displayed to the clinician to aid in the diagnosis of various fundus diseases. The device provides diagnostic visualization, enabling clinicians to assess retinal health and pathology, thereby supporting clinical decision-making and patient management.
Clinical Evidence
Bench testing only. The device was verified and validated against functional specifications and performance requirements. Compliance with U.S. and international standards was confirmed, including IEC 60601-1, 60601-1-1, 60601-1-2, 60825-1, and ISO 15004-2:2007. The device also complies with 21 CFR Parts 1010 and 1040 regarding light-emitting products.
Technological Characteristics
Confocal laser scanning ophthalmoscope. Uses IR, blue, green, and red lasers. Complies with IEC 60601-1, 60601-1-1, 60601-1-2, 60825-1, and ISO 15004-2:2007. Complies with 21 CFR 1040.10 and 1040.11 for laser products. Connectivity and software architecture details not specified.
Indications for Use
Indicated for capturing fundus images to assist in the diagnosis of fundus diseases in patients requiring ophthalmic examination.
Regulatory Classification
Identification
An ophthalmoscope is an AC-powered or battery-powered device containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, and vitreous) and the retina of the eye.
Special Controls
*Classification.* Class II (special controls). The device, when it is an AC-powered opthalmoscope, a battery-powered opthalmoscope, or a hand-held ophthalmoscope replacement battery, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}------------------------------------------------
K072259
MAY - 6 2008
## Pre-Market Notification 510(k) Summary
#### 1. Sponsor Information:
l
Company Name & Address: NIDEK Company, Ltd. 34-14 Maehama, Hiroishicho Gamagori, Aichi, 443-0038 Japan Contact Person: Yoneji Mizuno Contact Title: Manager Regulatory Affairs Contact Phone Number: +81-533-67-8901 Contact Fax Number: +81-533-67-6610
Date of Summary:
May 5, 2008
## 2. Device Name and Classification:
| Common and Usual Name: | Ophthalmoscope |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Proprietary Name: | Ophthalmoscope F-10 |
| Classification Name: | Ophthalmoscope, AC-Powered (21 CFR § 886.1570) |
| Product Code: | MYC |
| Class: | II |
| Performance Standards: | No applicable performance standards have been issued under section 514 or under section 513(b) of the Food, Drug and Cosmetic Act. |
Performance Standards issued for Light Emitting Products: 21CFR Part 1040, Section 1040.10 and 1040.11 (Laser Products).
#### 3. Predicate Device(s):
| K971671: | Heidelberg Engineering |
|----------|----------------------------------------------|
| | Heidelberg Retina Angiograph FA/ICGA (HRA/C) |
| K871268: | G. Rodenstock Instrumente, GMBH |
| | Rodenstock Scanning Laser |
Ophthalmoscope F-10 510(k) Summary
NIDEK COMPANY, LTD.
{1}------------------------------------------------
## 4. Description of Device:
The Ophthalmoscope F-10 is categorized as a Class 1 confocal laser scanning ophthalmoscope. The F-10 captures and records confocal images of the fundus by laser scanning using a selection of laser colors: IR (infrared), blue, green, and red and affords Indocyanine green (ICG) and fluorescein (FAG) angiography.
## న్. Indications for Use:
The Ophthalmoscope F-10 is intended for use in capturing images of the fundus which can be used for diagnosis of fundus diseases.
## 6. Comparison with Predicate Device(s):
The Ophthalmoscope F-10 is substantially equivalent to the predicate devices identified and other commercially available SLO products. It is similar in characteristics, materials, features, has similar technological features, intended use and indications for use as the predicates, and do not raise any new questions of safety and effectiveness.
## 7. Non-Clinical Performance Summary:
Nidek Company Ltd. has verified and validated that the Ophthalmoscope F-10 meets its functional specifications, performance requirements and complies with applicable U.S. and international standards (IFC 60601-1, 60601-1-1, 60601-1-2, 60825-1 and ISO 15004-2:2007) for products of its kind. This device complies with 21 CFR Parts 1010 and 1040, with the exception of those allowable performance deviations noted in FDA's Laser Notice 50.
#### 8. Conclusions:
In summary, Nidek Company Ltd., is of the opinion that the Ophthalmoscope F-10 docs not introduce any new potential safety risks, is as effective, and performs as well as devices currently on the market, and concludes that the Ophthalmoscope F-10 is substantially equivalent to the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized emblem featuring three curved lines that resemble a human figure or a symbol representing health and well-being.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY - 6 2008
NIDEK Co., Ltd. c/o Mr. Paul Sumner Vice President of Regulatory, Clinical & & Quality Systems Arkin Consulting Group, LLC 1733 Canton Lane Marietta, GA 30062
Re: K072259
Trade/Device Name: Ophthalmoscope F-10 Regulation Number: 21 CFR 886.1570 Regulation Name: Ophthalmoscope Regulatory Class: Class II Product Code: MYC Dated: April 16, 2008 Received: April 18, 2008
Dear Mr. Sumner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may. therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration,
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2- Mr. Paul Sumner
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
M. B. Egelston, m.d.
Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use Statement
510(k) Number (if known):
K072254
Device Name:
OPHTHALMOSCOPE F-10
Indications For Use:
The Ophthalmoscope F-10 is intended for use in capturing images of the fundus, which can be used for the diagnosis of fundus diseases.
Prescription Use XX (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Derick
Page 1 of 1
(Division Sign-Off)
Division of Ophthalmic Ear,
Nose and Throat Devises
510(k) Number K072259
NIDEK COMPANY, LTD.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.