ASCOM MOBILE MONITORING GATEWAY (MMG)

K111215 · Ascom (Sweden) AB · MSX · Jul 5, 2011 · Cardiovascular

Device Facts

Record IDK111215
Device NameASCOM MOBILE MONITORING GATEWAY (MMG)
ApplicantAscom (Sweden) AB
Product CodeMSX · Cardiovascular
Decision DateJul 5, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2300
Device ClassClass 2

Indications for Use

The Ascom Mobile Monitoring Gateway (MMG) is intended to interface with the GE Healthcare patient monitoring network and the Ascom Messaging System, in order to provide a secondary means of automated visual and/or audible annunciating and displaying of patient alarm information to healthcare professionals, via display devices. The MMG does not alter the behavior of the monitoring system. Neither is it intended to replace or alter the primary alarm function on the patient monitor. The MMG is not intended to be used for diagnostic purposes. The MMG is intended for use by professional clinical personnel and relies on proper use and operation of both the communication infrastructure in place at the healthcare facility and the display devices used. The MMG software is installed on specified hardware located in a computer hall or similar, where the MMG can't come into physical contact with patients.

Device Story

Ascom Mobile Monitoring Gateway (MMG) acts as middleware between GE Healthcare patient monitoring networks and Ascom messaging systems. Device inputs patient status/alarm data via wired Ethernet; transforms data into formatted alerts; transmits via wireless platforms (DECT, WiFi, Paging) to mobile display devices. Used in clinical facilities; operated by professional clinical personnel. Provides secondary notification when clinicians are away from primary monitors. Records messaging activities in Ascom Unite Connectivity Manager for audit trails. Benefits patient care by ensuring timely awareness of alarm conditions; does not alter primary monitor function or perform diagnostics.

Clinical Evidence

No clinical data. Bench testing only. Device underwent verification and validation of software, electrical safety, and electromagnetic compatibility (EMC) in accordance with applicable standards. Results confirmed compliance with predetermined specifications.

Technological Characteristics

Embedded network appliance; software-based middleware. Connectivity: Wired Ethernet input, wireless output (DECT, WiFi, Paging). Not sterile; no shelf-life; no patient contact. Software-based alarm routing and formatting.

Indications for Use

Indicated for professional clinical personnel to receive secondary automated visual/audible patient alarm notifications from GE Healthcare monitoring networks via mobile display devices. Not for diagnostic use; does not replace primary patient monitor alarms.

Regulatory Classification

Identification

A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K111215 P1/4 JUL - 5 2011 ## 510(k) Summary | Submission Date: | 29 April 2011 | | | |--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|-------------------------------------| | Submitter: | Ascom (Sweden) AB<br>P.O. Box 8783<br>SE-402 76 Göteborg<br>Sweden | | | | Submitter and<br>Official Contact: | Ms. Tania Ottebrink<br>Manager of Product Configuration & Conformity<br>P.O. Box 8783<br>SE-402 76 Göteborg<br>Sweden<br>Phone: +46 31 55 93 00<br>Email : Tania.Ottebrink@ascom.se | | | | Manufacturing Site: | Ascom (Sweden) AB<br>Grimbodalen 2<br>SE-417 49 Göteborg<br>Sweden | | | | Trade Name: | Ascom Mobile Monitoring Gateway (MMG) | | | | Common Name: | Network and Communication Middleware | | | | Classification Name: | System, Network And Communication, Physiological Monitors | | | | Classification<br>Regulation: | 21 CFR §870.2300 | | | | Product Code: | MSX | | | | Substantially<br>Equivalent Devices: | Ascom Model | Predicate 510(k)<br>Number | Predicate Manufacturer<br>and Model | | | Ascom Mobile<br>Monitoring<br>Gateway (MMG) | K103634 | Ascom Cardiomax | {1}------------------------------------------------ Device Description: The Ascom Mobile Monitoring Gateway (MMG) is an on-site messaging integration solution which forwards patient monitor status and alarm information to the user via display devices provided by Ascom or third-party mobile device companies. Users receive interactive, time-critical information from patient monitoring devices directly via their display devices as text, alarms or data. The Ascom MMG allows users to be aware of their patients' status and alarm conditions when they are away from the patient and patient monitoring svstem. > The Ascom MMG connects to the information sources through wired Ethernet connections which are part of the customer's infrastructure. The Ascom MMG software acquires patient data from patient monitoring devices. The user configures the Ascom MMG to determine which information, including alarm notifications, is delivered to which users. The Ascom MMG then formats the data for wireless delivery to the display devices. > All messaging activities are recorded in the Ascom Unite Connectivity Manager providing real-time activity logging for audit trail records and reporting. The Ascom MMG hardware consists of small, embedded network appliances, and application-specific software. The Ascom MMG delivers near real-time text messaging alerts and information to text-capable display devices. The Ascom Mobile Monitoring Gateway operates on a wireless communications system platform. The communication platform uses DECT (Digital Enhanced Cordless Telecommunications), WiFi and Paging technology. The Ascom MMG combined with a wireless communication system, is part of an Ascom end-to-end solution designed to provide all the components necessary to optimize work flow, including display devices, gateways and device management. {2}------------------------------------------------ | Intended Use: | The Ascom Mobile Monitoring Gateway (MMG) is intended to<br>interface with the GE Healthcare patient monitoring network and the<br>Ascom Messaging System, in order to provide a secondary means of<br>automated visual and/or audible annunciating and displaying of patient<br>alarm information to healthcare professionals, via display devices. | |------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | The MMG does not alter the behavior of the monitoring system.<br>Neither is it intended to replace or alter the primary alarm function on<br>the patient monitor. The MMG is not intended to be used for diagnostic<br>purposes. | | | The MMG is intended for use by professional clinical personnel and<br>relies on proper use and operation of both the communication<br>infrastructure in place at the healthcare facility and the display devices<br>used. | | | The MMG software is installed on specified hardware located in a<br>computer hall or similar, where the MMG can't come into physical<br>contact with patients. | | Technology<br>Comparison: | The Ascom Mobile Monitoring Gateway (MMG) employs the same or<br>similar technological characteristics as the predicate device. | | Performance Testing: | | | Sterilization and<br>Shelf-Life | The Ascom Mobile Monitoring Gateway (MMG) is not provided sterile<br>and is not intended to be sterilized by the user. Additionally, the Ascom<br>Mobile Monitoring Gateway (MMG) does not have a shelf-life.<br>Therefore, this section is not applicable. | | Biocompatibility | The Ascom Mobile Monitoring Gateway (MMG) does not directly or<br>indirectly contact the patient. Therefore, this section is not applicable. | | Software Testing | Software for the Ascom Mobile Monitoring Gateway (MMG) was<br>designed and developed according to a robust software development<br>process, and was rigorously verified and validated. | | | Test results indicated that the Ascom Mobile Monitoring Gateway<br>(MMG) complies with its predetermined specifications. | | Electrical Safety | The Ascom Mobile Monitoring Gateway (MMG) was tested for<br>electrical safety in accordance with applicable Standards. | | | Test results indicated that the Ascom Mobile Monitoring Gateway<br>(MMG) complies with its predetermined specifications and with the<br>applicable standards. | | Electromagnetic<br>Compatibility Testing | The Ascom Mobile Monitoring Gateway (MMG) was tested for EMC<br>in accordance with applicable Standards. | | | Test results indicated that the Ascom Mobile Monitoring Gateway<br>(MMG) complies with its predetermined specifications and with the<br>applicable standards. | | Performance Testing<br>- Bench | The Ascom Mobile Monitoring Gateway (MMG) was tested for<br>performance in accordance with predetermined specifications and<br>applicable Standards. | | | Test results indicated that the Ascom Mobile Monitoring Gateway<br>(MMG) complies with its predetermined specifications. | | Conclusion | Based upon a comparison with predicate devices and testing results, the<br>Ascom Mobile Monitoring Gateway (MMG) is substantially equivalent<br>to the predicate device. | . " . {3}------------------------------------------------ {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. In the center of the seal is an abstract image of an eagle with its wings spread. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 JUL - 5 2011 Ascom (Sweeden) AB - Wireless Solutions c/o Ms. Tamia Ottebrink Manager of Product Configuration & Conformity Grimbodalen 2. P.O. Box 8783 Goteborg Sweden SE-402 76 Re: K111215 > Trade/Device Name: Ascom Mobile Monitoring Gateway (MMG) Regulatory Number: 21 CFR 870.2300 Regulation Name: Network and Communication, Physiological Monitors Systems Regulatory Class: II (two) Product Code: MSX Dated: April 29, 2011 Received: May 2, 2011 Dear Ms. Ottebrink: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food, Dr to and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your devil ( in lye found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ Page 2 -- Ms. Tamia Ottebrink Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Bram D. Zuckerman, M.D. Director I Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ ## KII1215 ## Indications for Use | 510(k) Number: | K | |-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name: | Ascom Mobile Monitoring Gateway (MMG) | | Indications for Use: | The Ascom Mobile Monitoring Gateway (MMG) is intended to interface with the GE Healthcare patient monitoring network and the Ascom Messaging System, in order to provide a secondary means of automated visual and/or audible annunciating and displaying of patient alarm information to healthcare professionals, via display devices. | | | The MMG does not alter the behaviour of the monitoring system. Neither is it intended to replace or alter the primary alarm function on the patient monitor. The MMG is not intended to be used for diagnostic purposes. | | | The MMG is intended for use by professional clinical personnel and relies on proper use and operation of both the communication infrastructure in place at the healthcare facility and the display devices used. | | | The MMG software is installed on specified hardware located in a computer hall or similar, where the MMG can't come into physical contact with patients. | | Prescription Use<br>(Part 21 CFR 801 Subpart D) | X | | | AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) | (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Cardiovascular Devices 510(k) Number K111215 Page 1 of 1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...