K103634 · Ascom (Us), Inc. · MSX · Feb 23, 2011 · Cardiovascular
Device Facts
Record ID
K103634
Device Name
ASCOM CLINICALCONNEX CARDIOMAX
Applicant
Ascom (Us), Inc.
Product Code
MSX · Cardiovascular
Decision Date
Feb 23, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2300
Device Class
Class 2
Indications for Use
The intended use of the Ascom ClinicalConneX Cardiomax (Cardiomax) is to provide an interface with clinical systems to forward information associated to the particular event to the designated display device(s). For medical, near real time alarms, the Ascom Cardiomax is intended to serve as a parallel, redundant, forwarding mechanism to inform healthcare professionals of particular medical related events. Ascom Cardiomax does not alter the behavior of the primary medical devices and associated alarm annunciations. The display device provides a visual, and/or audio and/or vibrating mechanism upon receipt of the alert. The Ascom Cardiomax is intended for use as a secondary alarm. It does not replace the primary alarm function on the monitor.
Device Story
Ascom ClinicalConneX Cardiomax is an on-site messaging integration solution; functions as middleware connecting patient monitoring systems to mobile display devices. Input: patient monitor status and alarm data via wired ethernet. Processing: software acquires data, formats information based on user configuration, and transmits alerts wirelessly (e.g., DECT technology) to display devices. Output: near real-time voice, text, or alarm notifications on mobile displays. Used in clinical environments by healthcare professionals to maintain awareness of patient status when away from the bedside. Provides redundant, secondary alarm notification; does not alter primary monitor behavior. Benefits: improves clinical workflow and patient monitoring responsiveness. Messaging activities are logged in a central server for audit trails.
Clinical Evidence
Bench testing only. No clinical data provided. Performance verified through software validation, electrical safety testing, and electromagnetic compatibility (EMC) testing in accordance with applicable standards.
Technological Characteristics
Embedded network appliance; wired ethernet input; wireless output via DECT technology. Software-based middleware for data acquisition, formatting, and routing. Not sterile; no patient contact. Complies with electrical safety and EMC standards.
Indications for Use
Indicated for healthcare professionals to receive secondary, near real-time notifications of patient monitor status and alarm conditions via mobile display devices. Not for use as a primary alarm system; does not replace primary monitor alarm functions.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
Predicate Devices
Spacelabs Medical Clinical Event Interface (CEI), Model 91847 (K062278)
Submission Summary (Full Text)
{0}------------------------------------------------
## KI03634
## 510(k) Summary
FEB 2 3 2011
.
| Submission Date: | 10 December 2010 | | | |
|--------------------------------------|-----------------------------------------------------------------------------------|----------------------------|--------------------------------------------------------------------------------------------------|--|
| Submitter: | Ascom (US), Inc.<br>598 Airport Boulevard. Suite 300<br>Morrisville, NC 27560 | | | |
| Submitter and<br>Official Contact: | Mr. Thomas Kroenke<br>Speed To Market, Inc.<br>PO Box 3018<br>Nederland, CO 80466 | | | |
| | +1 (303) 956-4232<br>tkroenke@speedtomarket.net | | | |
| Manufacturing Site: | Ascom (US), Inc.<br>598 Airport Boulevard, Suite 300<br>Morrisville, NC 27560 | | | |
| Trade Name: | Ascom ClinicalConneX Cardiomax | | | |
| Common Name: | Network and Communication Middleware | | | |
| Classification Name: | System, Network And Communication, Physiological Monitors | | | |
| Classification<br>Regulation: | 21 CFR §870.2300 | | | |
| Product Code: | MSX | | | |
| Substantially<br>Equivalent Devices: | Ascom Model | Predicate 510(k)<br>Number | Predicate Manufacturer<br>and Model | |
| | Ascom<br>ClinicalConneX <br>Cardiomax | K062278 | Spacelabs Medical, Inc. /<br>Spacelabs Medical Clinical<br>Event Interface (CEI), Model<br>91847 | |
.
{1}------------------------------------------------
Device Description: The Ascom (US), Inc. (Ascom) ClinicalConneX | Cardiomax (Cardiomax) is an on-site messaging integration solution which forwards patient monitor status and alarm information to the user via display devices provided by Ascom or third-party mobile device companies. Users receive interactive, time-critical information from patient monitoring devices directly via their display devices as voice, text, alarms or data. The Ascom Cardiomax allows users to be aware of their patients' status and alarm conditions when they are away from the patient and patient monitoring system. The Ascom Cardiomax connects to the information sources through wired ethernet connections which are part of the customer's infrastructure. The Ascom Cardiomax software acquires patient data from patient monitoring devices. The user configures the Ascom Cardiomax to determine which information, including alarm notifications, is delivered to which users. The Ascom Cardiomax then formats the data for wireless delivery to the display devices. All messaging activities are recorded in the Ascom Communications Server providing real-time activity logging for audit trail records and reporting. The Ascom Cardiomax hardware consists of small, embedded network appliances, and application-specific software. The Ascom Cardiomax delivers near real-time text messaging alerts and information to text-capable display devices. Ascom provides a wireless communications system platform on which the Ascom Cardiomax may operate: DECT (Digital Enhanced Cordless Telecommunications) technology. The Ascom Cardiomax, combined with an Ascom wireless communication system, is part of an Ascom end-to-end solution designed to provide all the components necessary to optimize work flow, including display devices, gateways and device management. The intended use of the Ascom ClinicalConneX | Cardiomax Intended Use: (Cardiomax) is to provide an interface with clinical systems to forward information associated to the particular event to the designated display device(s). For medical, near real time alarms, the Ascom Cardiomax is intended to serve as a parallel, redundant, forwarding mechanism to inform healthcare professionals of particular medical related events. Ascom Cardiomax does not alter the behavior of the primary medical devices and associated alarm annunciations. The display device provides a visual, and/or audio and/or vibrating mechanism upon receipt of the alert. The Ascom Cardiomax is intended for use as a secondary alarm. It does not replace the primary alarm function on the monitor.
{2}------------------------------------------------
| Technology<br>Comparison: | The Ascom Cardiomax employs the same or similar technological<br>characteristics as the predicate device. | |
|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Performance Testing: | | |
| Sterilization and<br>Shelf-Life | The Ascom Cardiomax is not provided sterile and is not intended to be<br>sterilized by the user. Additionally, the Ascom Cardiomax does not<br>have a shelf-life. Therefore, this section is not applicable. | |
| Biocompatibility | The Ascom Cardiomax does not directly or indirectly contact the<br>patient. Therefore, this section is not applicable. | |
| Software Testing | Software for the Ascom Cardiomax was designed and developed<br>according to a robust software development process, and was rigorously<br>verified and validated. | |
| | Test results indicated that the Ascom Cardiomax complies with its<br>predetermined specifications. | |
| Electrical Safety | The Ascom Cardiomax was tested for electrical safety in accordance<br>with applicable Standards. | |
| | Test results indicated that the Ascom Cardiomax complies with its<br>predetermined specifications and with the applicable standards. | |
| Electromagnetic<br>Compatibility Testing | The Ascom Cardiomax was tested for EMC in accordance with<br>applicable Standards. | |
| | Test results indicated that the Ascom Cardiomax complies with its<br>predetermined specifications and with the applicable standards. | |
| Performance Testing<br>- Bench | The Ascom Cardiomax was tested for performance in accordance with<br>predetermined specifications and applicable Standards. | |
| | Test results indicated that the Ascom Cardiomax complies with its<br>predetermined specifications. | |
| Conclusion | Based upon a comparison with predicate devices and testing results, the<br>Ascom Cardiomax is substantially equivalent to the predicate device. | |
. . . . .
.
. . . .
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three swooping lines representing its wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
## MAR 2 1 2011
Ascom, Inc. c/o Mr. Thomas Kroenke Speed to Market, Inc. PO Box 3018 Nederland, CO 80466
Re: K103634
Trade Name: Ascom ClinicalConneX/Cardiomax Regulation Number: 21 CFR 870.2300 Regulation Name: Network and Communication, Physiological Monitors Systems Regulatory Class: Class II (two) Product Code: MSX Dated: December 10, 2010 Received: December 13, 2010
Dear Mr. Kroenke:
This letter corrects our substantially equivalent letter of February 23, 2011.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Thomas Kroenke
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Bram D. Zuckerman
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
| 510(k) Number (if known): | K103634 |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | Ascom ClinicalConneX Cardiomax |
| Indications for Use: | The intended use of the Ascom ClinicalConneX Cardiomax<br>(Cardiomax) is to provide an interface with clinical systems to<br>forward information associated to the particular event to the<br>designated display device(s).<br>For medical, near real time alarms, the Ascom Cardiomax is<br>intended to serve as a parallel, redundant, forwarding mechanism<br>to inform healthcare professionals of particular medical related<br>events. Ascom Cardiomax does not alter the behavior of the<br>primary medical devices and associated alarm annunciations.<br>The display device provides a visual, and/or audio and/or<br>vibrating mechanism upon receipt of the alert.<br>The Ascom Cardiomax is intended for use as a secondary alarm.<br>It does not replace the primary alarm function on the monitor. |
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Cardiovascular Devices
**510(k) Number** K03634
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.