Vasomedical-Biox 12 - Channel Ambulatory ECG Holter Recorder, Model 1304, is a Non-Invasive device intended to acquire ambulatory 12 - Channel ECG signals from the upper body surfaces. Cardiac rhythm is acquired via ECG signals. Vasomedical-Biox 3 - Channel Miniature Ambulatory ECG Holter Recorder, Model 1303, is also a Non-Invasive device but intended to acquire ambulatory 3 - Channel ECG signals from the upper body surfaces. Cardiac rhythm is also acquired via ECG signals. The Holter Recorders are intended for adults and children who are over the age of six years. The Models 1304 and 1303 work with the CB Series ECG Analysis Software which has been previously cleared under (k)083820. The system is only for measurement, recording and display. It makes no diagnosis.
Device Story
Portable, microprocessor-based Holter recorders (Models 1304, 1303) acquire ambulatory ECG signals from upper body surfaces via electrodes (10 electrodes for 12-channel; 5 for 3-channel). Devices are worn by patients using a carrying case and strap. Data is stored on SD memory cards. Recorders interface with previously cleared CB Series ECG Analysis Software for data processing. Healthcare providers use the software to review recorded ECG rhythms. Devices provide measurement, recording, and display functions only; they do not provide automated diagnosis. Intended for clinical use to assist in cardiac rhythm screening.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Portable, microprocessor-based; 12-channel (Model 1304) or 3-channel (Model 1303) ECG acquisition; SD memory card storage; 256 Hz/channel sample rate; 10-bit resolution; LED (1304) or LCD (1303) display; battery-powered (AA or AAA alkaline); infrared data transfer; works with CB Series ECG Analysis Software.
Indications for Use
Indicated for adults and children >6 years old requiring ambulatory ECG monitoring for up to 72 hours. Used for measurement, recording, and display of cardiac rhythm. No diagnosis provided.
Regulatory Classification
Identification
A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.
{0}------------------------------------------------
14
JUN 2 8 2011
## 510(k) Summary
Vasomedical-Biox 12 - Channel Ambulatory ECG Holter Recorder. Model 1304 Vasomedical-Biox 3 - Channel Miniature Ambulatory ECG Holter Recorder, Model 1303
- April 25, 2011 1. Date Prepared:
- 2. Submitter's Name: Vasomedical, Inc. and Address 180 Linden Ave. Westbury, NY 11590
- 3. Contact Person: Richard Gordon Manager, Regulatory and Quality Affairs Vasomedical, Inc. Telephone: (516) 997-4600 Facsimile: (516) 997-2299 E-mail: rgordon@vasomedical.com
- 4. Device Names: a) 12 - Channel Ambulatory ECG Holter Recorder, Model 1304 b) 3 - Channel Miniature Ambulatory ECG Holter Recorder, Model 1303
- Proprietary Names: a) Vasomedical-Biox 12 Channel Ambulatory ECG Holter Recorder. Model 1304
- b) Vasomedical-Biox 3 Channel Miniature Ambulatory ECG Holter Recorder, Model 1303
- Common Name: a) 12 - Channel Ambulatory ECG Holter Recorder, Model 1304 b) 3 - Channel Miniature Ambulatory, ECG Holter Recorder, Model 1303
Classification Name: 870.2800 Magnetic Tape Recorder, Medical
- 5. Predicate Device: Vasomedical - Biox, 3 - Channel Ambulatory ECG Holter Recorder, Model 1305 was granted FDA 510(k) clearance on April 1, 2009 (k083820).
- 6. Device Description: Vasomedical-Biox 12 - Channel Ambulatory ECG Holter Recorder and Vasomedical-Biox 3 - Channel Miniature Ambulatory ECG Holter Recorders are intended to be used as a Holter Ambulatory Electrocardiograph device for the purpose of screening ECG rhythms for periods up to 72 hours. Cardiac rhythm is acquired by 12 Channel or 3 Channel ECG signals.
{1}------------------------------------------------
The Recorders are intended for adults and children over the age of six years old.
The Models 1304 and 1303 are modified versions of Model 1305 which FDA granted clearance on April 1, 2009 (k083820). Model 1304 increases the number of channels and ECG signals to 12 from 3 as in the Predicate device Model 1305.
Model 1303 is a miniaturized version of Model 1305
The Recorders are portable, microprocessor based devices that are worn by a patient with the use of a supplied Carrying Case and Strap.
Models 1304 and 1303 specifications are listed in Table 1 below:
| | Model 1504 | Model 1505 |
|-------------------------|-------------------------------------------------------|-----------------------------------------------|
| Lead | 12 | 3 |
| Electrode | 10 | 5 |
| Lead wire type | 10F/10G | 7E/5E |
| Display | LED | LCD |
| Batteries | 1 x "AA" Alkaline | 1 x "AAA" Alkaline |
| Carrying Case/<br>Strap | Supplied | Supplied |
| Card Reader | Supplied | Supplied |
| Dimensions | 3.3 x 2.3 x 0.88" | 2.6 x 2.1 x 0.7" |
| Weight | 2.45 oz | 1.76 oz |
| Storage | SD Memory Card | SD Memory Card |
| Memory<br>Capacity | 1 GB or more | 1 GB or more |
| Sample Rate | 256 Hz/Channel<br>10,000 Hz<br>Pacemaker<br>Detection | 256 Hz/Channel<br>1,000Hz for SAECG<br>option |
| Resolution | 10 Bit | 10 Bit; 12 Bit for<br>SAECG option |
| Infrared<br>Adaptor | Supplied | Supplied |
Model 1304
Model 1303
Table 1: Model 1304 and 1303 Specifications
{2}------------------------------------------------
## 7. Intended Use:
Vasomedical-Biox 12 - Channel Ambulatory ECG Holter Recorder, Model 1304, is a Non-Invasive device intended to acquire ambulatory 12 - Channel ECG signals from the upper body surfaces. Cardiac rhythm is acquired via ECG signals.
Vasomedical-Biox 3 - Channel Miniature Ambulatory ECG Holter Recorder, Model 1303, is also a Non-Invasive device but intended to acquire ambulatory 3 - Channel ECG signals from the upper body surfaces. Cardiac rhythm is also acquired via ECG signals.
The Holter Recorders are intended for adults and children who are over the age of six years.
The Models 1304 and 1303 work with the CB Series ECG Analysis Software which has been previously cleared under (k)083820
The system is only for measurement, recording and display. It makes no diagnosis.
Refer to Attachment I and II, Vasomedical-Biox Model 1305/1304 AECG Instruction Manual for Users and Model 1303 AECG Instruction Manual for Users, Sections 2.2, Indications for Use, 2.2.1 Intended Use and 2.2.2 for Contraindications.
## 8. Comparison of Technological Characteristics:
Technological and functional characteristics of the devices listed in this Special 510(k) Notification for Modification are essentially the same as those of the Predicate device. The devices listed in this 510(k) Premarket Notification is therefore substantially equivalent to the predicate device.
"The Intended Use of the modified devices as described in its labeling, has not changed as a result of the modifications".
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized graphic of what appears to be an abstract caduceus, a symbol often associated with healthcare.
Food and Drug Administration 1 0903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Vasomedical, Inc. c/o Mr. Richard E. Gordon Manager, Regulatory and Quality Affairs 180 Linden Avenue Westbury, NY 11590
JUN 2 8 2011
Re: K11180
Trade/Device Name: Vasomedical-Biox 12-Channel Ambulatory ECG Holter Recorder (Model 1304) and Vasomedical-Biox 3-Channel Miniature Ambulatory ECG Holter Recorder (Model 1303) Regulatory Number: 21 CFR 870.2800 Regulation Name: Medical Magnetic Tape Recorder Regulatory Class: II (two) Product Code: DSH Dated: June 2, 2011
Received: June 3, 2011
Dear Mr. Gordon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Richard E. Gordon
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm11.5809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours
C. Bram D. Zuckerman, M.D.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
April 25, 2011 '
K111180
なこ :
/
/3
## Indications for Use
| 510(k) Number: | k | |
|-----------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------|
| Device Names: | a) Vasomedical-Biox 12 - Channel Ambulatory ECG Holter Recorder,<br>Model 1304 | |
| | b) Vasomedical-Biox 3 - Channel Miniature Ambulatory ECG Holter<br>Recorder, Model 1303 | |
| Indications for Use: | Vasomedical-Biox 12 - Channel Ambulatory ECG Holter Recorder,<br>Model 1304, is a Non-Invasive device intended to acquire ambulatory<br>12 - Channel ECG signals from the upper body surfaces. Cardiac Rhythm is<br>acquired via ECG signals. | |
| | Vasomedical-Biox 3 - Channel Miniature Ambulatory ECG Holter Recorder,<br>Model 1303, is also a Non-Invasive device intended to acquire ambulatory<br>3 - Channel ECG signals from the upper body surfaces. Cardiac rhythm is<br>also acquired via ECG signals. | |
| | The Holter Recorders are intended for adults and children who are over the<br>age of six years. | |
| | The Models 1304 and 1303 work with the CB Series ECG Analysis Software<br>which has been previously cleared under (k083820) | |
| | "The Intended Use of the modified devices as described in its labeling,<br>has not changed as a result of the modifications".<br>The system is only for measurement, recording and display. It makes no<br>diagnosis. | |
| | | |
| Prescription Use: YES<br>(Part 21 CFR 801 Subpart D) | AND/OR | Over-The-Counter Use: NO<br>(21 CFR 801 Subpart C) |
| (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF<br>NEEDED) | | |
| | Concurrence of CDRH, Office of Device Evaluation (ODE) | |
| (Division Sign-Off)<br>Division of Cardiovascular Devices | | |
510(k) Number_K 111,80
·
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.