VASOMEDICAL BIOX CB-1305-C, 3 CHANNEL ECG HOLTER MONITOR, MODEL CB-1305-C
Applicant
Vasomedical, Inc.
Product Code
DSH · Cardiovascular
Decision Date
Apr 1, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2800
Device Class
Class 2
Attributes
Software as a Medical Device, Pediatric
Indications for Use
The Ambulatory ECG Recorder described in this 510(k) Premarket Notification is a non-invasive device intended to acquire, record and store ambulatory three (3) channel ECG signals from the upper body surface of adults and children over the age of six years old for a minimum of 24 hours. It works in concert with the CB Series Analysis Software to playback, review, analyze, edit and print ECG data. The software does not perform diagnostics. These data are used by physicians for diagnostic evaluations.
Device Story
The Vasomedical Biox 1305 is a portable, microprocessor-based Holter monitor worn by patients to acquire 3-channel ECG signals via 5 or 7 electrodes. Data is stored on an SD card and transferred to a Windows-based PC running CB Series Analysis Software. The software processes the ECG data to provide QRS classification (ventricular, supra-ventricular, normal, artifact), event reporting (e.g., tachycardia, bradycardia, atrial fibrillation), ST-segment analysis, HRV, and pacemaker detection. Physicians review the generated reports, trend charts, and histograms to perform diagnostic evaluations. The device facilitates long-term cardiac rhythm screening in clinical or home settings.
Clinical Evidence
No clinical data provided; substantial equivalence is based on technological and functional comparison to predicate devices.
Technological Characteristics
Portable, microprocessor-based 3-channel ECG recorder; 10-bit resolution (12-bit for SAECG); 256 Hz sampling (10,000 Hz for pacemaker); 1 GB+ SD card storage; LED display; 1 AA alkaline battery; Windows-based PC software for analysis.
Indications for Use
Indicated for adults and children >6 years old requiring ambulatory 3-channel ECG monitoring for at least 24 hours to screen cardiac rhythms.
Regulatory Classification
Identification
A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.
Huntleight Healthcare Cardiology Products Division Medilog Darwin (K053369)
Submission Summary (Full Text)
{0}------------------------------------------------
## 510(k) Summary
# Vasomedical, Incorporated Ambulatory ECG Recorder Ambulatory ECG Recorder Analysis Software
APR - 1 2009
| 1. | Date Prepared: | 20 February 2009 | |
|----------------|------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| 2. | Submitter's Name:<br>and Address | Vasomedical, Inc.<br>180 Linden Ave.<br>Westbury, NY 11590 | |
| 3. | Contact Person: | Melissa M. Purpura<br>Director, Regulatory and Quality Affairs<br>Vasomedical, Incorporated<br>Telephone: (516) 997-4600<br>Facsimile: (516) 997-2299<br>E-mail: mpurpura@vasomedical.com | |
| 4. | Recorder<br>Device Name:<br>Proprietary Name:<br>Common Name:<br>Classification Name: | Medical Magnetic Tape Recorder<br>Vasomedical Biox 1305, 3 Channel ECG Holter Monitor<br>1305 Biox Holter<br>870.2800 Medical Magnetic Tape Recorder | |
| | Analysis Software<br>Device Name:<br>Proprietary Name:<br>Common Name:<br>Classification Name: | Computer, Diagnostic, Programmable<br>Vasomedical Biox ECG Analysis System Software<br>CB Series Software<br>870.1425 Computer, Diagnostic, Programmable | |
| 5. | Predicate Device: | The AECG Recorder listed above is substantially equivalent to the<br>Huntleight Healthcare Cardiology Products Division Recorder<br>Models MedilogAR4 and MedilogAR12. The AECG Analysis<br>Software listed above is substantially equivalent to the Huntleight<br>Healthcare Cardiology Products Division, Medilog Darwin. FDA<br>granted 510(k) clearances for this predicate device on April 6,<br>2007 under K070818 and December 22, 2005 under K053369. | |
| 6. | Device Description: | Vasomedical's Ambulatory ECG Recorder Model 1305 is<br>intended to be used as a Holter ambulatory electrocardiograph<br>monitor for the purpose of screening ECG rhythms for periods un | |
| Lead | 3 | | |
| Electrode | 5 or 7 | | |
| Lead wire type | Lead-7C/5C | | |
| Display | LED | | |
| Batteries | 1 "AA" alkaline | | |
| Dimensions | 3.3" x 2.3" x 0.8" | | |
| Storage Medium | SD Card | | |
| Memory | 1 GB or more | | |
| Sample Rate | 256 Hz/channel; 10,000 Hz/channel for pacemaker; 1,000 Hz/channel with SAECG option | | |
| Resolution | 10 bit, 12 bit for SAECG option | | |
{1}------------------------------------------------
to 72 hours. Cardiac rhythm is acquired by 3 channel ECG signals. The recorders are intended for adults and children over the age of six years old.
The recorders are portable, microprocessor based devices that are worn by a patient with the use of a carrying case and strap.
1305 Biox Holter specifications are noted in table 1
Table 1: Specification of 1305 Biox Holter.
Vasomedical's Ambulatory ECG CB Series Analysis System Software allows transfer of data from the recorder to a Windows based PC via a removable storage card for the purpose of creating printouts and reports. These data are used by physicians for diagnostic evaluations.
Vasomedical's Ambulatory ECG Analysis System Software features are noted in table 2.
| QRS classification | Ventricular, Supra-Ventricular, Normal, Artifact |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Comprehensive event reports | VE, VE pair, VE bigeminy, V-run, R on T, SVE,<br>SVE pair, SVE-run, pause, long RR, tachycardia,<br>bradycardia, and atrial fibrillation |
| ST-segment reports | Deviation, Slope, Duration, Heart rate |
| Analysis capability | Arrhythmia, ST-segment, HRV, pacemaker, artial<br>fibrillation, SAECG, HRT |
| Viewing modes | Standard ECG, ECG thumbnail, Page scanning |
| Data reporting | Trend charts, statistical tables, category specific<br>histograms |
Table 2: Features of CB Series Software.
{2}------------------------------------------------
:
. .- "
ਸੰ
K083820
20 February 2009 p 3/3
| 7. Intended Use: | The Ambulatory ECG Recorder described in this 510(k)<br>Premarket Notification is a non-invasive device intended to<br>acquire, record and store ambulatory three (3) channel ECG<br>signals from the upper body surface of adult and adolescent<br>patients for a minimum of 24 hours. It works in concert with the<br>CB Series Analysis Software to playback, review, analyze, edit<br>and print ECG data. |
|-------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 8. Comparison of<br>Technological<br>Characteristics: | Technological and functional characteristics of the<br>device listed in this 510(k) Premarket Notification are<br>essentially the same as those of the predicate devices. The device<br>listed in this 510(k) Premarket Notification are therefore |
substantially equivalent to the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird with three stylized wings.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
### APR - 1 2009
Vasomedical, Inc. c/o Ms. Melissa M. Purpura Director, Regulatory and Quality Affairs 180 Linden Avenue Westbury, NY 11590
Re: K083820
Trade/Device Name: Vasomedical Biox 1305, 3 Channel ECG Holter Monitor and Ambulatory ECG Analysis System Software CB Series Regulation Number: 21 CFR 870.2800 Regulation Name: Medical Magnetic Tape Recorder Regulatory Class: Class II (two) Product Code: DSH Dated: February 20, 2009 Received: February 24, 2009
Dear Ms. Purpura:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Melissa M. Purpura
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance. please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
07 Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K083820, Amendment #1
### Indications for Use Statement
510(k) Number:
Device Name:
Vasomedical Biox 1305, 3 Channel ECG Holter Monitor Ambulatory ECG Analysis System Software CB Series
Indications for Use:
The Ambulatory ECG Recorder described in this 510(k) Premarket Notification is a non-invasive device intended to acquire, record and store ambulatory three (3) channel ECG signals from the upper body surface of adults and children over the age of six years old for a minimum of 24 hours. It works in concert with the CB Series Analysis Software to playback, review, analyze, edit and print ECG data. The software does not perform diagnostics. These data are used by physicians for diagnostic evaluations.
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
| | B Zuckerman |
|------------------------------------|-------------|
| (Division Sign-Off) | 3/31/09 |
| Division of Cardiovascular Devices | |
| 510(k) Number | K083820 |
| Prescription Use X | OR | Over-The Counter Use ______ |
|----------------------|----|-----------------------------|
| (Per 21 CFR 801.109) | | |
14
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.