PASCAL STREAMLINE PHOTOCOAGULATOR

K100019 · Optimedica Corporation · HQF · May 4, 2010 · Ophthalmic

Device Facts

Record IDK100019
Device NamePASCAL STREAMLINE PHOTOCOAGULATOR
ApplicantOptimedica Corporation
Product CodeHQF · Ophthalmic
Decision DateMay 4, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.4390
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Intended for use in the treatment of ocular pathology in the posterior segment. Retinal photocoagulation, panretinal photocoagulation, focal photocoagulation and grid photocoagulation for vascular and structural abnormalities of the retina and choroid including: proliferative and nonproliferative diabetic retinopathy macular edema choroidal neovascularization branch and central retinal vein occlusion age-related macular degeneration lattice degeneration retinal tears and detachments Intended for use in the treatment of ocular pathology in the anterior segment including: iridotomy trabeculoplasty

Device Story

Integrated system comprising solid-state aiming and treatment lasers, control electronics, graphical user interface, slit lamp, and table. Used by ophthalmologists in clinical settings for retinal and anterior segment photocoagulation. Device delivers laser energy to ocular tissues to treat vascular and structural abnormalities. Physician operates system via interface to control laser parameters; output is therapeutic laser energy. Benefits include precise, controlled photocoagulation for various ocular pathologies.

Clinical Evidence

Bench testing only.

Technological Characteristics

Integrated system with solid-state aiming and treatment lasers, control electronics, graphical user interface, slit lamp, and table. Class II device.

Indications for Use

Indicated for patients requiring treatment of ocular pathology in the posterior segment (e.g., diabetic retinopathy, macular edema, choroidal neovascularization, retinal vein occlusion, age-related macular degeneration, lattice degeneration, retinal tears/detachments) and anterior segment (e.g., iridotomy, trabeculoplasty).

Regulatory Classification

Identification

An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## PASCAL® STREAMLINE™ PHOTOCOAGULATOR - ## SECTION 6 510(k) SUMMARY ## 510(k) Notification K100019 ## GENERAL INFORMATION #### Applicant: . . *. OptiMedica Corporation 3130 Coronado Drive Santa Clara, CA 95054 U.S.A. Pone: 408-850-8600 FAX: 408-850-8595 MAY - 4 2010 #### Contact Person: Darlene Crockett-Billig President Experien Group, LLC 155-A Moffett Park Drive, Suite 210 Sunnyvale, CA 94089-1330 U.S.A. Phone: 408-400-0856 ext. 105 FAX: 408-400-0865 Email: dcb(@experiengroup.com ## Date Prepared: April 12, 2010 #### Classification: 21 CFR§878.4810, Class II 21 CFR§886.4390, Class II #### Product Code: GEX, HQF #### Trade Name: PASCAL® Streamline™ Photocoagulator #### Generic/Common Name: Laser instrument, surgical, powered Laser, ophthalmic #### Predicate Device: PASCAL Streamline Photocoagulator (K092621) {1}------------------------------------------------ ### SECTION 6 510(k) Summary ### Intended Use The PASCAL Streamline Photocoagulator indications for use are the following: Intended for use in the treatment of ocular pathology in the posterior segment. Retinal photocoagulation, panretinal photocoagulation, focal photocoagulation and grid photocoagulation for vascular and structural abnormalities of the retina and choroid including: - proliferative and nonproliferative diabetic retinopathy 지 - 트 macular edema - 포 choroidal neovascularization - branch and central retinal vein occlusion 트 - age-related macular degeneration 트 - 트 lattice degeneration - retinal tears and detachments 에 Intended for use in the treatment of ocular pathology in the anterior segment including: - iridotomy ■ - ı trabeculoplasty ## Product Description The PASCAL Streamline Photocoagulator is an integrated system comprising of solid state aiming and treatment lasers, control electronics, graphical user interface, slit lamp and table. It is intended for use in the treatment of ocular pathology in both the posterior and anterior segments. #### Substantial Equivalence The modified PASCAL Streamline Photocoagulator is substantially equivalent to the predicate device with regard to function, intended use and performance. Any differences in the technological characteristics between the two devices do not raise any new issues of safety or efficacy. Thus, the modified PASCAL Streamline Photocoagulator is substantially equivalent to the predicate device. #### Testing in Support of Substantial Equivalence Determination All necessary bench testing was conducted on the modified PASCAL Streamline Photocoagulator to support a determination of substantial equivalence to the predicate device. #### Summary The PASCAL Streamline Photocoagulator is substantially equivalent to the predicate device. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal features a stylized eagle with outstretched wings, symbolizing protection and service. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 MAY ~ 4 2010 OptiMedica Corporation % Experien Group, LLC Ms. Darlene Crockett-Billig President 155-A Moffett Park Drive, Suite 210 Sunnyvale, California 94089-1330 Re: K100019 Trade/Device Name: PASCAL® Streamline™ Photocoagulator Regulation Number: 21 CFR 886.4390 Regulation Name: Opthalmic laser Regulatory Class: Class II Product Code: HQF Dated: April 12, 2010 Received: April 13, 2010 Dear Ms. Crockett-Billig: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must {3}------------------------------------------------ Page 2 - Ms. Darlene Crockett-Billig comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Mark M. Millers Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## SECTION 5 # INDICATIONS FOR USE STATEMENT 510(k) Number (if known): K100019 Device Name: PASCAL Streamline Photocoagulator #### Indications for Use: The PASCAL Streamline Photocoagulator indications for use are the following: Intended for use in the treatment of ocular pathology in the posterior segment. Retinal photocoagulation, panretinal photocoagulation, focal photocoagulation and grid photocoagulation for vascular and structural abnormalities of the retina and choroid including: - l proliferative and nonproliferative diabetic retinopathy - 트 macular edema - י choroidal neovascularization - 피 branch and central retinal vein occlusion - צ age-related macular degeneration - lattice degeneration - retinal tears and detachments Intended for use in the treatment of ocular pathology in the anterior segment including: - 트 iridotomy - M trabeculoplasty Prescription Use X ___________________________________________________________________________________________________________________________________________________________ (21 CFR Part 801 Subpart D) and/Or Over the Counter Use (21 CFR Part 801 Subpart C) ## (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Neil RP Ohle for MXA (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K100019
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...