TRACER IMPLANTATION KIT, INDEX IMPLANTATION KIT

K110971 · Navotek Medical, Ltd. · IYE · Jan 10, 2012 · Radiology

Device Facts

Record IDK110971
Device NameTRACER IMPLANTATION KIT, INDEX IMPLANTATION KIT
ApplicantNavotek Medical, Ltd.
Product CodeIYE · Radiology
Decision DateJan 10, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5050
Device ClassClass 2

Indications for Use

The Tracer Implantation Kit is indicated for use to radiographically and radioactively mark soft tissue for future therapeutic procedures. The Tracer is indicated for permanent implantation in the prostate. The Tracer is intended to be used in conjunction with the RealEye system as an adjunct in treatment planning and radiation therapy, to align and monitor the patient's position relative to the isocenter of a linear accelerator. The IndeX Implantation Kit is indicated for use to radiographically mark soft tissue for future therapeutic procedures. The IndeX is intended to be used as a radiographic fiducial marker for visualization using x-ray, CT, or Cone Beam CT.

Device Story

Tracer/IndeX Implantation Kits provide fiducial markers for soft tissue. Tracer: Platinum/Iridium radioactive source; implanted in prostate; emits photons for localization by Navotek RealEye system; used during radiotherapy to align/monitor patient position relative to linear accelerator isocenter. IndeX: non-radioactive version; used as radiographic marker for x-ray, CT, or Cone Beam CT visualization. Both kits include pre-loaded implantation device for single-use sterile delivery. Physician implants marker in clinic/OR; RealEye system tracks radioactive Tracer to assist radiation therapy planning and delivery; improves precision of radiation targeting; reduces potential for healthy tissue damage.

Clinical Evidence

Bench, animal, and clinical testing conducted. Biocompatibility per ISO 10993-1; platinum leakage and radioactive leakage testing per ISO 2919 (classification ISO/99/C53211(X)). Performance testing confirms device functions as intended and is substantially equivalent to predicates.

Technological Characteristics

Materials: Platinum/Iridium. Tracer: radioactive source emitting photons. IndeX: non-radioactive. Form factor: pre-loaded in implantation device. Sterilization: gamma radiation. Single-use. Connectivity: Tracer interacts with RealEye system for localization.

Indications for Use

Indicated for soft tissue marking for therapeutic procedures. Tracer: permanent prostate implantation for radiotherapy alignment/monitoring. IndeX: radiographic fiducial marker for x-ray, CT, or Cone Beam CT visualization.

Regulatory Classification

Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### 510(k) SUMMARY # JAN 1 0 2012 Navotek Medical Ltd. - Tracer Implantation Kit # Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared Navotek Medical Ltd. POB 201, 5 HaCarmel St. Yokneam 20692, Israel Phone: +972 72 270 9020 Facsimile: +972 72 270 9021 Contact Person: Moshe Solomon Date Prepared: March 30, 2011 ### Name of Device and Name/Address of Sponsor Tracer Implantation Kit and IndeX Implantation Kit Navotek Medical Ltd. POB 201, 5 HaCarmel St. Yokneam 20692, Israel # Common or Usual Name Soft tissue fiducial marker ## Classification Name Medical linear accelerator ### Regulation Number {1}------------------------------------------------ CFR 892.5050 (Medical charged-particle radiation therapy system) #### Product Code IYE #### Device Class = #### Predicate Devices Navotek RealEye System and Calypso 4D Localization System (Tracer) Visicoil soft tissue marker (IndeX) #### Intended Use / Indications for Use #### Tracer: The Tracer Implantation Kit is indicated for use to radiographically and radioactively mark soft tissue for future therapeutic procedures. The Tracer is indicated for permanent implantation in the prostate. The Tracer is intended to be used in conjunction with the RealEye system as an adjunct in treatment planning and radiation therapy, to align and monitor the patient's position relative to the isocenter of a linear accelerator. #### IndeX: The IndeX Implantation Kit is indicated for use to radiographically mark soft tissue for future therapeutic procedures. The IndeX is intended to be used as a radiographic fiducial marker for visualization using x-ray, CT, or Cone Beam CT. #### Technological Characteristics The Tracer is a Platinum/Iridium radioactive soft tissue fiducial for implantation in the prostate, in or near the treatment target for radiotherapy treatments. The Tracer emits {2}------------------------------------------------ photons that enable its localization by the Navotek RealEye system that is designed to track localized gamma-emitting radioactive sources. The implantation of the Tracer is performed using an Implantation Device, which is supplied with the Tracer pre-loaded within it. Therefore, the Tracer is supplied as a 'Tracer Implantation Kit' which includes the Tracer and the Implantation Device. The Tracer lmplantation Kit is provided in a single package, sterile (by gamma radiation) and ready for use. It is intended for single use only. The IndeX is identical to the Tracer in all respects except that it is not radioactive. It is intended for use to radiographically mark soft tissue and to be visible in x-ray, CT, and Cone Beam CT images. # Performance Data Performance testing demonstrates that the Tracer Implantation Device perform according to specifications and function as intended. Performance testing further demonstrates that differences in technology compared to the predicate device do not impact safety or effectiveness. Biocompatibility testing was performed according to ISO 10993- Part 1 (2003) and the FDA Blue book memorandum G95-1. Testing included platinum leakage testing with respect to platinum toxicity and radioactive leakage testing to demonstrate that the Tracer meets the requirements for a sealed source according to ISO 2919 with classification ISO/99/C53211(X). Bench, animal, and clinical testing also was conducted and demonstrated substantial equivalence to the predicate device. All testing that is not directly related to radioactivity applies to both the Tracer and the IndeX. # Substantial Equivalence The Tracer and its predicates have the same intended use and similar indications, and similar technological characteristics and principles of operation. The identified technological differences do not raise any new types of safety or effectiveness questions. Performance testing demonstrates that the identified technological differences between the Tracer and the predicate device do not impact safety or effectiveness. In particular, bench, animal, and clinical testing results demonstrate that the safety and performance of the Tracer are similar {3}------------------------------------------------ to that of its predicate device. Thus, the Tracer is substantially equivalent to its predicate device. The IndeX and its predicate have the same intended use and indications, and similar technological characteristics and principles of operation. The identified technological differences do not raise any new types of safety or effectiveness questions. Performance testing demonstrates that the identified technological differences between the IndeX and the predicate device do not impact safety or effectiveness. In particular, bench, animal, and clinical testing results demonstrate that the safety and performance of the IndeX are similar to that of its predicate device. Thus, the IndeX is substantially equivalent to its predicate device. {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized symbol that resembles a human figure embracing or supporting another, with flowing lines suggesting movement or care. ## DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002 Navotek Medical Ltd. % Mr. Steven B. Datlof, M.D., J.D. Regulatory Counsel Hogan Lovells US LLP 1835 Market Street, 29th Floor PHILADELPHIA PA 19103 JAN 1 0 2012 · Re: K110971 Trade/Device Name: Tracer Implantation Kit and IndeX Implantation Kit Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE, NEU Dated: December 2, 2011 Received: December 2, 2011 Dear Mr. Datlof: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of {5}------------------------------------------------ medical device-related adverse events) (21 CFR 803); and good manufacturing practice medical device related adverse oviality systems (QS) regulation (21 CFR Part 820). This letter requirements as be forth in keting your device as described in your Section 510(k) premarket with anow you to begin finding of substantial equivalence of your device to a legally marketed nonicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 11 you desire specific ad reserved in Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket 5 150. Thise, product 1000 and 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely Yours, Mary Pastel Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ ### Indications for Use Statement 510(k) Number (if known): K110971 Device Name: Tracer Implantation Kit and IndeX Implantation Kit Indications for Use: The Tracer Implantation Kit is indicated for use to radiographically and radioactively mark soft tissue for future therapeutic procedures. The Tracer is indicated for permanent implantation in the prostate. The Tracer is intended to be used in conjunction with the RealEye system as an adjunct in treatment planning and radiation therapy, to align and monitor the patient's position relative to the isocenter of a linear accelerator. The IndeX Implantation Kit is indicated for use to radiographically mark soft tissue for future therapeutic procedures. The IndeX is intended to be used as a radiographic fiducial marker for visualization using x-ray, CT, or Cone Beam CT. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Mary Patel (Division Sign-Off) Division of Radiological Devices Diagnostic Device Evaluation and Safety ffice of In Vitro Diagno K110971 Page _1_ of _1_
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...