K110954 · Arjo Wiggins Medical, Inc. · FRG · Aug 24, 2012 · General Hospital
Device Facts
Record ID
K110954
Device Name
ARJORAD STERILE WRAP
Applicant
Arjo Wiggins Medical, Inc.
Product Code
FRG · General Hospital
Decision Date
Aug 24, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
Used to enclose another medical device that is to be sterilized. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the device until used. ArjoRad Sterilization Wraps will maintain sterility for a maximum period of thirty (30) days. Intended for use with STERRAD 100S, NX, and 100NX Systems.
Device Story
ArjoRad Sterilization Wrap is a non-sterile, single-use SMS (spunbond-meltblown-spunbond) sheet system designed for double-wrapping medical devices prior to sterilization. The wrap acts as a sterile barrier, allowing sterilant penetration during STERRAD hydrogen peroxide gas plasma cycles while preventing microbial ingress post-sterilization. Used in clinical settings, the wrap is manually applied by healthcare personnel to enclose instruments. It maintains sterility for up to 30 days. By ensuring the integrity of the sterile barrier, the device protects patients from potential contamination of surgical instruments. Performance is validated against specific STERRAD cycle parameters, including lumen dimensions and tray load weights.
Clinical Evidence
Bench testing only. Evaluated via physical property testing (tensile strength, burst, tear) per EN 868-2:2009 and ASTM D737. Microbial aerosol challenge testing confirmed germ resistance post-sterilization. Sterilization efficacy validated using STERRAD hydrogen peroxide gas plasma cycles, achieving a Sterility Assurance Level (SAL) of 10^-6. Shelf-life study confirmed 30-day sterility maintenance.
Technological Characteristics
SMS (spunbond-meltblown-spunbond) non-woven material. Physical properties tested per EN 868-2:2009 and ASTM D737. Designed for hydrogen peroxide gas plasma sterilization compatibility. Form factor: sheet for double-wrapping. Non-sterile supply.
Indications for Use
Indicated for use as a sterile barrier system to enclose medical devices during hydrogen peroxide gas plasma sterilization and to maintain sterility for up to 30 days post-sterilization. Compatible with STERRAD 100S, NX, and 100NX systems for various instrument configurations including stainless steel lumens, PTFE lumens, and flexible endoscopes.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{0}------------------------------------------------
# 510(k) SUMMARY (as required by 807.92(c))
AUG 2 4 2012
ਨ
Regulatory and Marketing Services, Inc. Regulatory Correspondent: 962 Allegro Lane Apollo Beach, FL 33572 Jonathan Ward wardip@ajwtech.com 813-645-2855 813-677-4787
Submitter of 510(k):
Arjo Wiggins Medical 1301 Charleston Regional Parkway, Suite 500 Charleston, SC 29492 Patrick Ritchie PRitchie@arjomedical.com
August 17, 2012 Date of Summary:
Trade/Proprietary Name:
Classification Name:
Product Code:
#### Device Description:
ArjoRad Sterilization Wrap is a sterilization packaging system consisting of one SMS sheet to be double wrapped during use. ArjoRad Sterilization Wrap will be sold in a non-sterile condition in packaged form of various sizes depending on the needs of the end user.
Sterilization Wrap
FRG
ArjoRad Sterilization Wrap
#### Intended Use: -
Used to enclose another medical device that is to be sterilized. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the device until used. ArjoRad Sterilization Wraps will maintain sterility for a maximum period of thirty (30) days. Intended for use with STERRAD 100S, NX, and 100NX Systems.
| Cycle | Recommended<br>Instrument<br>Configuration | Inside Diameter | Length | Recommended<br>Individual<br>Tray Load Weight | Number of Lumens per<br>Validation Load | Cycle Length |
|------------------------------|--------------------------------------------|------------------|------------------|-----------------------------------------------|-----------------------------------------|--------------|
| STERRAD 100S | Stainless Steel Lumens | 3 mm | 400 mm | 9.1 lbs | 2 trays, 5 lumens each | 55 minutes |
| STERRAD NX<br>Standard | Single Channel PTFE<br>Lumens | 1 mm | 150 mm | 10.7 lbs | 10 lumens | 28 minutes |
| STERRAD NX<br>Advanced | Single Channel<br>Flexible Endoscope | 1.2 mm | 845 mm | 10.7 lbs | 1 endoscope | 38 minutes |
| | Stainless Steel<br>Lumens | 1 mm | 500 mm | | 10 lumens | |
| STERRAD<br>100NX<br>Standard | Stainless Steel<br>Lumens | 0.7 mm | 500 mm | 10.7 lbs | 2 trays, 5 lumens each | 47 minutes |
| STERRAD 100<br>NX Flex | Single Channel<br>Flexible Endoscope(s) | 1.2 mm<br>1.2 mm | 830 mm<br>840 mm | 6.4 lbs | 2 trays, 1 endoscope per<br>tray | 42 minutes |
| STERRAD 100<br>NX Express | Stainless Steel<br>Surfaces<br>(2mm x 8mm) | N/A | N/A | 10.7 lbs | N/A | 24 minutes |
510(k) Summary
{1}------------------------------------------------
#### Predicate Device:
K092167 - Kimguard one step sterilization wrap - Kimberly - Clark Corporation
#### Substantial Equivalence:
The subject wraps have the same fundamental technological characteristics as the predicate device(s). Performance testing between the predicate device and the subject device have verified substantial equivalence in design, materials and intended use, and confirmed there are no significant differences between the proposed and predicate devices that raise new questions of safety or efficacy.
#### Performance Testing:
The intended use is to contain medical devices and to maintain sterility. Physical protection is consequently needed during the handling of non-sterile product. It is controlled with following performance testing: substance weight, strip tensile strength, Mullen burst and Elmendorf tear. Arjorad developed higher tensile strength and burst and tearing strength than predicate sterilization wrap. In addition ArjoRad meets the acceptance criteria with regard to the minimum requirement for strength properties dedicated to non-woven sterilization wrap. When applied to the minimum performance level, the properties of ArjoRad still meet the criteria of acceptation after sterilization which equals the minimum requirement level for sterilization wrap defined according to EN 868-2:2009 and ASTM D737: Standard Test Method for Air Permeability of Textile Fabrics.
During microbial aerosol challenge test, similarly to predicate, ArjoRad demonstrated germ resistance after hydrogen peroxide sterilization cycle. The microbial aerosol challenge test and the shelf life study demonstrate that, the subject wrap is considered as a sterile barrier system that is capable to maintaining sterility for 30 days after sterilization, as long as the sterile barrier is not compromised in any way.
The ability to let the sterilant penetrate has been validated as well. This was demonstrated by sterilization efficacy studies after different cycles of STERRAD hydrogen peroxide gas plasma sterilization. Studies carried out demonstrated that Arjorad can assure an SAL of 10 after each half cycle tested.
Conclusion: The minor differences between the proposed and predicate devices do not introduce new issues of safety and efficacy. The ArjoRad Sterilization Wrap is substantially equivalent to the predicates in that it meets all of the requirements for use in a STERRAD 100S, NX and 100NX Systems. All testing performed indicates that the ArjoRad Sterilization Wrap will provide a sterile barrier for use in the STERRAD device. There are a series of detailed Comparison Tables for testing results included in Section 18 Performance Testing, which was completed by comparing the ArjoRad Sterilization Wrap against the primary predicate device. The KC300 sterilization wrap chosen as a test predicate has been cleared for low temperature V-Pro sterilization, the comparative performance data demonstrates that the proposed device performs equally or better than the predicate device.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird with outstretched wings, rendered in black. The bird is positioned above the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA", which is arranged in a circular fashion around the bird.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -- WO66-G609 Silver Spring, MD 20993-0002
Arjo Wiggins Medical, Incorporated C/O Mr. Jonathan Ward President Regulatory and Marketing Services, Incorporated 962 Allegro Lane Apollo Beach, Florida 33572
Re: K110954
Trade/Device Name: ArjoRad Sterilization Wrap Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: FRG Dated: August 17, 2012 Received: August 22, 2012
### Dear Mr. Ward:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
AUG 2 4 2012
{3}------------------------------------------------
Page 2- Mr. Ward
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
For
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a sequence of alphanumeric characters, specifically 'K110954'. The characters are rendered in a bold, sans-serif font, giving them a clear and distinct appearance. The arrangement of the characters suggests a code or identification number.
Indications for Use
510(k) Number (if known): K110954
Device Name: ArjoRad Sterilization Wrap
Used to enclose another medical device that is to be sterilized. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the device until used. ArjoRad Of the choose will maintain sterility for a maximum period of thirty (30) days. Intended for use with STERRAD 100S, NX, and 100NX Systems
| Cycle | Recommended<br>Instrument<br>Configuration | Inside<br>Diameter | Length | Recommended<br>Individual<br>Tray Load Weight | Number of Lumens<br>per<br>Validation Load | Cycle Length |
|------------------------------|--------------------------------------------|--------------------|------------------|-----------------------------------------------|--------------------------------------------|--------------|
| STERRAD 100S | Stainless Steel Lumens | 3 mm | 400 mm | 9.1 lbs | 2 trays, 5 lumens each | 55 minutes |
| STERRAD<br>NX<br>Standard | Single Channel<br>PTFE Lumens | 1 mm | 150 mm | 10.7 lbs | 10 lumens | 28 minutes |
| STERRAD<br>NX Advanced | Single Channel<br>Flexible Endoscope | 1.2 mm | 845 mm | 10.7 lbs | 1 endoscope | 38 minutes |
| | Stainless Steel<br>Lumens | 1 mm | 500 mm | | 10 lumens | |
| STERRAD<br>100NX<br>Standard | Stainless Steel<br>Lumens | 0.7 mm | 500 mm | 10.7 lbs | 2 trays, 5 lumens each | 47 minutes |
| STERRAD<br>100 NX Flex | Single Channel<br>Flexible<br>Endoscope(s) | 1.2 mm<br>1.2 mm | 830 mm<br>840 mm | 6.4 lbs | 2 trays, 1 endoscope<br>per tray | 42 minutes |
| STERRAD<br>100 NX<br>Express | Stainless Steel<br>Surfaces<br>(2mm x 8mm) | N/A | N/A | 10.7 lbs | N/A | 24 minutes |
Sterilization Loads
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use X AND/OR (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
Mau' Smith (Attorney)
(Division Sign-Off) Division of Anestheslology, General Hospital Infection Control, Dental Devices
**510(k) Number:** *K110454*
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.