K110950 · Consensus Orthopedics, Inc. · MBH · Jun 27, 2011 · Orthopedic
Device Facts
Record ID
K110950
Device Name
CONSENSUS KNEE SYSTEM
Applicant
Consensus Orthopedics, Inc.
Product Code
MBH · Orthopedic
Decision Date
Jun 27, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3565
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CONSENSUS® KNEE SYSTEM Primary Knee is designed as a system and is not intended for substitution of components from other systems. - A. Primary intervention of rheumatoid arthritis, osteoarthritis, post-traumatic arthritis, or degenerative arthritis. - B. Failed osteotomy or unicompartmental replacements. - C. Replacement of unsatisfactory cemented or press-fit knee components when sufficient bone stock exists. - D. The porous coated (CoCr beads with Titanium) components may be used with or without cement.
Device Story
Consensus Knee System (CKS) is a primary fixed-bearing total knee prosthesis; restores knee function by replicating natural anatomy. System includes femoral, tibial, and patellar components; available in posterior cruciate ligament (PCL) retaining or substituting (posterior stabilized) configurations. Femoral components feature large distal radius to optimize contact areas and reduce stress; trochlear groove allows even load distribution. Metallic components available in non-porous or porous-coated variants (CoCr beads with Titanium) for cemented or uncemented use. Implantation performed by orthopedic surgeons in clinical settings. Device restores joint kinematics; provides stability; reduces contact stress. Line extensions include Reduced Lateral Profile (RLP) femoral components, porous-coated PS femoral components, size 0 tibial baseplates/inserts, 20/22mm thick poly inserts, and porous-coated patella.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing only, including microstructure analysis, corrosion resistance (ASTM F746 & G61), static tensile/shear strength, shear fatigue strength, rotating beam fatigue strength, abrasion testing, contact area/stress distribution analysis, and push-in/push-out force testing.
Technological Characteristics
Materials: CoCr beads with Titanium coating, UHMWPE. Sensing/Actuation: Mechanical prosthesis. Energy: None. Connectivity: None. Sterilization: Not specified. Software: None. Standards: ASTM F746, ASTM G61. Design: Fixed-bearing, PCL-retaining or posterior stabilized, porous-coated or non-porous.
Indications for Use
Indicated for patients requiring primary knee intervention due to rheumatoid arthritis, osteoarthritis, post-traumatic arthritis, degenerative arthritis, failed osteotomy, failed unicompartmental replacement, or revision of unsatisfactory cemented/press-fit components with sufficient bone stock.
Regulatory Classification
Identification
A knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device is designed to achieve biological fixation to bone without the use of bone cement. This identification includes fixed-bearing knee prostheses where the ultra high molecular weight polyethylene tibial bearing is rigidly secured to the metal tibial base plate.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's guidance: “Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses; Guidance for Industry and FDA.” See § 888.1 for the availability of this guidance.
Gender Solutions™ Natural Knee® Flex System (K073286)
Gender Solutions™ Natural Knee® Flex System (K070214)
Submission Summary (Full Text)
{0}------------------------------------------------
KI10950 Page 1 of 4
## JUN 2 7 2011
#### 2. 510(k) SUMMARY
| Sponsor Name: | Consensus Orthopedics, Inc.<br>1115 Windfield Way, Suite 100<br>El Dorado Hills, CA 95762 |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Contact: | Matthew M. Hull, RAC<br>Phone: (916) 355-7156/ Fax: (916) 355-7190<br>mhull@consensusortho.com |
| Date Prepared: | 1 April 2011 |
| Trade Name: | Consensus® Knee System, Line Extensions |
| Common Name: | Total knee prosthesis for cemented or uncemented use |
| Classification Name: | Knee joint patellofemorotibial metal/polymer/metal semi-constrained<br>cemented prosthesis is a class II device per 21 CFR 888.3560 (Product<br>Code JWH)<br>Knee joint patellofemorotibial metal/polymer porous-coated<br>uncemented prosthesis is a Class II device per 21 CFR 888.3565<br>(Product Code MBH) |
| Review Panel: | Orthopedic Devices |
#### Device Description:
The Consensus Total Knee System (CKS) is a primary fixed-bearing total knee system that has been on the market since the mid-1990's.
The CKS has been designed to replicate the natural anatomy of the knee in order to restore knee function. It has been developed to preserve and utilize healthy ligamentous structures. For cases where the soft tissues are not functional, the PCL substituting tibial inserts or the posterior stabilized system are available for increased stability.
The CKS incorporates femoral, tibial, and patellar components and all associated instrumentation needed for implantation. The CKS can be used for total knee replacement with posterior cruciate ligament (PCL) retaining or substituting.
The femoral components are provided in left and right side versions and are designed to replicate natural kinematic motion between the femur, tibia and patella. The Consensus femoral component is designed to provide uniform contact zones in the coronal plane throughout the range of motion when the knee is properly aligned. The femoral component is also designed with a large distal radius to optimize contact areas and reduce contact stress. The trochlear groove in
{1}------------------------------------------------
the femur is designed to allow the load from the patella to be evenly distributed on the femur with adequate lateral constraint.
The CKS metallic components are available in non-porous and porous coated variants for cemented use and in a porous coated (CoCr beads with Titanium) version for uncemented use.
#### Indications for Use:
The CONSENSUS® KNEE SYSTEM Primary Knee is designed as a system and is not intended for substitution of components from other systems.
- A. Primary intervention of rheumatoid arthritis, osteoarthritis, post-traumatic arthritis, or degenerative arthritis.
- B. Failed osteotomy or unicompartmental replacements.
- C. Replacement of unsatisfactory cemented or press-fit knee components when sufficient bone stock exists.
- D. The porous coated (CoCr beads with Titanium) components may be used with or without cement.
#### Substantial Equivalence:
The CKS components were originally cleared by FDA in K932837* with a posterior stabilizing (PS) version cleared in K954818* & K962215*. The CKS was expanded in K001456** (femoral components) and K983004** (tibial components) and most recently cleared for uncemented use in K102927.
*These 510(k)'s were cleared prior to the purchase of U.S. Medical Products by Hayes Medical in 1996. **The previous 510(k)'s for these devices were cleared prior to the change in our company's name from Hayes Medical, inc. to Consensus Orthopedics, Inc. in 2008.
This submission is to add the following line extensions to the Consensus Knee System:
- 1) Reduced Lateral Profile (RLP) femoral component in standard and PS versions
- 2) Porous coated PS femoral component
- 3) Size 0 tibial baseplate and inserts
- 4) 20 & 22 mm thick poly inserts
- 5) Porous coated patella for uncemented use.
The CKS implants are also substantially equivalent to the Gender Solutions™ Natural Knee® Flex System by Zimmer, Inc. cleared in K073286 & K070214 with the above options.
{2}------------------------------------------------
# Non-Clinical Performance Data:
| Porous Coating: CoCr beads with Ti coating (applicable for all CoCr implants) | | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| Specification | Acceptance Criteria | Verification Results |
| Microstructure of the modified surface | N/A | Bead to Bead Neck Diameter 0.33 mm<br>Pore Size 0.432 mm<br>Volume % Porosity 37%<br>Coating Thickness 0.889 mm |
| Corrosion of the modified surface shall be equal of less | Equal to or improved corrosion resistance when | Critical Potential Breakdown Potential<br>CoCr Beads 1290 mV 1200 mV |
| than that measured in a legally marketed device. | compared with CoCr beads using ASTM F746 & G61 | Ti Coated<br>CoCr Beads 1315 mV 1200 mV |
| Modified surface shall exhibit adequate static tensile strength | The static tensile strength will exceed 20 MPa. | Static Tensile Strength of 58.32 MPa |
| Modified surface shall exhibit adequate static shear strength. | The static shear strength will exceed 20 MPa. | Static Shear Strength of 58.32 MPa |
| Modified surface shall exhibit adequate shear fatigue strength. | The shear fatigue strength will exceed 10 million cycles. | 10 million cycles achieved with a strength of 13.78 MPa |
| Modified surface shall exhibit adequate rotating beam fatigue strength. | The rotating beam fatigue strength will exceed 10 million cycles. | 10 million cycles achieved with a strength of 206.7 MPa |
| Modified surface shall not exhibit excessive abrasion. | N/A | 200N load:<br>Avg. mass loss 0.006 g<br>Avg. thickness loss 6%<br><br>1500N load:<br>Avg. Mass loss 0.179 g<br>Avg. thickness loss 23% |
| RLP Femoral Components | | |
| Specification | Acceptance Criteria | Verification Results |
| The articulating surface of the RLP components shall<br>match that of their standard and PS counterparts. | Similar contact area and surface stress distributions. | RLP had the same contact area as the original CKS. The PS<br>RLP matches the articulating surface of the RLP. |
| Size 0 Tibial Base Plate and Insert | | |
| Specification | Acceptance Criteria | Verification Results |
| Size 0 tibial baseplate/insert assembly shall permit<br>adequate push-in/push-out force. | Similar push-in/push-out loads when compared with existing baseplate/insert combinations. | Minimum push-out load was 428 lbs with failure mode being deformation of anterior snap recess in poly insert; Similar to other insert/baseplate. |
| | | combinations. Components<br>can be easily inserted by hand. |
| Thicker Tibial Insert | | |
| Specification | Acceptance Criteria | Verification Results |
| Insert thickness per FDA<br>Guidance Jan. 16, 2003. | Insert thickness must be<br>>6mm. | Insert thickness was greater<br>than 6mm. |
| Porous Coated Metal Backed Patella | | |
| Specification | Acceptance Criteria | Verification Results |
| The articulating surface of the<br>porous coated metal backed<br>patella shall match that of<br>current patellas. | Identical articulating surface. | Articulating surface of the porous<br>coated metal backed patella is<br>identical to existing patella. |
| The mating geometry between<br>the UHMWPE and metal back<br>of the porous coated metal<br>backed patella shall match that<br>of current metal backed<br>patella. | Identical mating geometry. | The mating geometry of the<br>porous coated metal backed<br>patella is identical to the existing<br>metal backed patella. |
Below is a summary of the testing related to the various aspects of the CKS line extension.
{3}------------------------------------------------
:
·
.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Consensus Orthopedics, Inc. % Mr. Matthew Hull 1115 Windfield Way, Suite 100 El Dorado Hills, California 95762
JUN 2 7 2011
Re: K110950
Trade/Device Name: Consensus® Knee System, Line Extensions Regulation Number: 21 CFR 888.3565 Regulation Name: Knee joint patellofemorotibial metal/polymer porous coated uncemented prosthesis Regulatory Class: Class II Product Code: MBH. JWH Dated: April 1, 2011 Received: April 4, 2011
Dear Mr. Hull:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{5}------------------------------------------------
Page 2 - Mr. Matthew Hull
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Erind Keith
Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
### 1. INDICATIONS FOR USE STATEMENT
510(k) Number (if known):
Device Name: Consensus® Knee System
### Indications for Use:
The CONSENSUS® KNEE SYSTEM Primary Knee is designed as a system and is not intended for substitution of components from other systems.
- A. Primary intervention of rheumatoid arthritis, osteoarthritis, post-traumatic arthritis, or degenerative arthritis.
- B. Failed osteotomy or unicompartmental replacements.
- C. Replacement of unsatisfactory cemented or press-fit knee components when sufficient bone stock exists.
- D. The porous coated (CoCr beads with Titanium) components may be used with or without cement.
Prescription Use ※ (21 CFR Part 801 Subpart D)
AND/OR
Over the Counter Use (21 CFR Part 801 Subpart C)
KI10950
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Ar M. Melleron
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K110950
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.