K983004 · Hayes Medical, Inc. · JWH · Nov 25, 1998 · Orthopedic
Device Facts
Record ID
K983004
Device Name
CONSENSUS FEMORAL COMPONENT, POROUS, COCR/TI
Applicant
Hayes Medical, Inc.
Product Code
JWH · Orthopedic
Decision Date
Nov 25, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Consensus® CoCr/Ti Femoral Component is intended for use with the Consensus® Knee System and is not intended for substitution of components from other systems. This device is intended for cemented use only. The indications for use are as follows: - Primary intervention of rheumatoid arthritis, osteoarthritis, post-traumatic 1. arthritis, or degenerative arthritis; - Intervention of failed osteotomies and unicompartmental replacements; 2. - 3. Replacement of unsatisfactory cemented or press fit knee components if sufficient bone stock exists.
Device Story
Consensus® Femoral Component is a porous, CoCr/Ti knee prosthesis designed to replicate natural femoral anatomy. Device features a swept-back radius on outer condyle surfaces for smooth articulation and a deepened trochlear groove for anatomic patellar tracking. Inner box geometry allows specific bone resection (10mm distal, 9mm posterior, 6.5mm anterior); anterior flange angled 4° outward to prevent notching. Smooth tapered pegs on distal condyles provide medial-lateral stability. Device is intended for cemented use only; porous coating of sintered CoCr beads with Titanium alloy coating enhances cement fixation. Used by orthopedic surgeons in clinical settings for total knee arthroplasty to restore joint function and alleviate pain associated with degenerative or inflammatory joint disease.
Clinical Evidence
No clinical data provided; substantial equivalence based on design, material, and performance testing.
Technological Characteristics
Materials: Cast CoCrMo alloy (ASTM F75-92) with sintered porous coating of CoCr beads (ASTM F75-92) and Titanium alloy (ASTM F67-95) coating. Configuration: Left/right femoral components with swept-back condyle radius, deepened trochlear groove, and tapered distal pegs. Fixation: Cemented only. Connectivity: None.
Indications for Use
Indicated for patients requiring primary intervention for rheumatoid, osteoarthritis, post-traumatic, or degenerative arthritis; intervention for failed osteotomies or unicompartmental replacements; or replacement of unsatisfactory cemented/press-fit knee components with sufficient bone stock.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
Predicate Devices
Consensus®, Knee System Femoral Component, Porous, CoCr (K932837)
The Natural-Knee® femoral component (Sulzermedica, Inc.)
The Genesis™ Total Knee System (Smith & Nephew Richards, Corp.)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image contains a sequence of handwritten digits and letters. The sequence starts with the letters 'Kq', followed by the numbers '83004'. The handwriting style is cursive, and the characters are connected.
## HAYES MEDICAL, INC Consensus® Knee System CoCr/Ti Femoral Component
## 510(k) Premarket Notification
## 510(k) SUMMARY
# Submitter's Name, Address, Telephone Number, and Contact Person
Hayes Medical, Inc. 819 Striker Ave., Suite 10 Sacramento, CA 95834 Telephone: (916) 646-5431 Facsimile: (916) 646-5432
John B. Stassi Contact Person: Director, Compliance
#### Date Prepared: August 15, 1998
# Name of Device and Name/Address of Sponsor
Trade name: Consensus® Femoral Component, Porous, CoCr/Ti
Hayes Medical, Inc. 819 Striker Ave., Suite 10 Sacramento, CA 95834
#### Classification Name
Knee joint patellofemorotibial polymer/metal/polymer semiconstrained cemented prosthesis (21 C.F.R. § 888.3560)
#### Predicate Devices
- Consensus®, Knee System Femoral Component, Porous, CoCr, marketed by 1. Hayes Medical, Inc. (K932837)
- The Natural-Knee® femoral component marketed by Sulzermedica, Inc. 2.
- The Genesis™ Total Knee System marketed by Smith & Nephew Richards, 3. Corp.
{1}------------------------------------------------
#### Intended Use
The Consensus® CoCr/Ti Femoral Component is intended for use with the Consensus® Knee System and is not intended for substitution of components from other systems. This device is intended for cemented use only. The indications for use are as follows:
- Primary intervention of rheumatoid arthritis, osteoarthritis, post-traumatic 1. arthritis, or degenerative arthritis;
- Intervention of failed osteotomies and unicompartmental replacements; 2.
- 3. Replacement of unsatisfactory cemented or press fit knee components if sufficient bone stock exists.
#### Device Description
The porous femoral knee component will be available in a left and right configuration and is designed to replicate the natural anatomy of the femur. The design of the femoral component incorporates a swept-back radius on the outer condyle surfaces to allow for smooth articulation throughout the range of motion 2. The outer articular surface of the femoral component will also have a deepened trochlear groove to allow for more anatomic tracking of the patellar component and improved range of motion. The material will be cast CoCrMo alloy (ASTM F75-92). The inner box geometry of the femoral component is designed to allow a minimum bone resection of 10mm distal, 9mm posterior, and 6.5mm anterior. The anterior flange of the femoral component is angled 4°outward, posterior to anterior, to reduce the potential for notching the anterior femur. The anterior flange will also
<sup>2</sup> Mensch, et al.; Knee Morphology as a Guide to Knee Replacement, Clinical Orthopaedics and Related Research, No. 112, October 1975.
{2}------------------------------------------------
incorporate a stepped chamfer which allows for positioning the trochlear groove in a more anatomic position. A smooth tapered peg will be located on each distal condyle to provide medial-lateral stability and enhanced fixation. The surfaces of the inner box geometry of the porous femoral component will incorporate a sintered porous coating of CoCr beads (ASTM F75-92), which will subsequently be coated with Titanium alloy (ASTM F67-95). The porous coating is designed to enhance cement fixation.
## Summary Basis for Finding of Substantial Equivalence
The Consensus® CoCr/Ti Femoral Component has substantially the same intended use and indications for use as the predicates as well as similar performance and principles of operation. The technological differences between the Consensus® CoCr/Ti Femoral Component and the predicate devices have been thoroughly documented and tested in order to verify that no new issues of safety and effectiveness are raised by the design.
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo is circular in shape and features the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter of the circle. In the center of the circle is a stylized image of an eagle with three heads.
NOV 2 5 1998
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. John Stassi Director, Compliance Haves Medical, Inc. 819 Striker Avenue, Suite 10 Sacramento, California 95834
K983004 Re : Consensus@ Knee System Femoral Component, Porous, CoCr/Ti Regulatory Class: II Product Code: JWH Dated: Auqust 20, 1998 Received: August 28, 1998
Dear Mr. Stassi:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). This decision is based on this device being equivalent only to similar devices labeled and intended to be fixed within bone with acrylic "bone cement." You may, therefore, market your device subject to the general controls provisions of the Act and the following limitations:
- This device may not be labeled or promoted for non-1. cemented use.
- All labeling for this device, including package label 2. and labeling included within the package, must prominently state that the device is intended for cemented use only.
- Any non-cemented fixation of this device is considered 3. investigational and may only be investigated as a siqnificant risk device in accordance with the investigational device exemption (IDE) regulation under 21 CFR, Part 812. All users of the device for noncemented fixation must receive approval from their respective institutional review boards (IRBs) and the Food and Drug Administration (FDA) to conduct the investigation.
The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practices, labeling, and prohibitions against misbranding and adulteration.
{4}------------------------------------------------
Page 2 - Mr. John Stassi
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਸ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (OS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Requlations.
This letter will allow you to beqin marketinq your device as described in your 510(k) premarket notification immediately. An FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diaqnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D.
Director
Division of General and
Restorative Devices
Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# 510(k) Number (if known):
Device Name: Consensus® Femoral Component, Porous, CoCr/Ti Indications For Use:
The Consensus® CoCr/Ti Femoral Component is intended for use with the Consensus® Knee System and is not intended for substitution of components from other systems. This device is intended for cemented use only. The indications for use are as follows:
- Primary intervention of rheumatoid arthritis, osteoarthritis, post-traumatic 1. arthritis, or degenerative arthritis;
- Intervention of failed osteotomies and unicompartmental replacements; 2.
- Replacement of unsatisfactory cemented or press fit knee components if 3. sufficient bone stock exist.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
(Division Sign-Off)
Division of General Restorative Devices
510(k) Number K983004
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.