ARAGON SURGICAL RF SYSTEM - 5MM LAPAROSCOPIC INSTRUMENT
K110824 · Aragon Surgical, Inc. · GEI · Jul 11, 2011 · General, Plastic Surgery
Device Facts
Record ID
K110824
Device Name
ARAGON SURGICAL RF SYSTEM - 5MM LAPAROSCOPIC INSTRUMENT
Applicant
Aragon Surgical, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Jul 11, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Aragon Surgical Caiman 5 Laparoscopic Instrument is a dedicated bipolar electrosurgical instrument intended for use in general surgical and gynecologic laparoscopic procedures where ligation and division of vessels is desired. The instrument creates a seal by the application of bipolar electrosurgical RF energy to vascular structures (vessels) interposed between the jaws of the device. A cutting blade is actuated for the division of tissue. Indications for use include general laparoscopic procedures, including urologic, vascular, thoracic, and thoracsocopic, and gynecological laparoscopic procedures where ligation and division of vessels is performed. These procedures include: laparoscopically assisted vaginal hysterectomies, Nissen fundoplication, colectomy, adhesiolysis, oophorectomy etc. The Aragon Surgical Caiman 5 Laparoscopic Instrument can be used on vessels up to and including 7mm, and tissue bundles as large as will fit in the jaws of the instrument. The Aragon Surgical Lektrafuse RF System has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures and should not be used for these procedures.
Device Story
Sterile, single-use bipolar electrosurgical instrument; used in laparoscopic/open surgery for vessel sealing, blunt dissection, grasping, and tissue division. Operates with Aragon Surgical Lektrafuse RF Generator; applies bipolar RF energy to vascular structures between jaws; features integrated manual cutting blade. Used by surgeons in OR settings. Output is sealed/divided tissue. Benefits include efficient hemostasis and tissue management during minimally invasive procedures.
Clinical Evidence
Bench and pre-clinical evaluations performed. No clinical trial data provided.
Technological Characteristics
Bipolar electrosurgical instrument; single-use; sterile. Utilizes bipolar RF energy for vessel sealing and mechanical cutting blade for tissue division. Designed for use with Aragon Surgical Lektrafuse RF Generator. Dimensions accommodate vessels up to 7mm.
Indications for Use
Indicated for patients undergoing general, gynecologic, urologic, vascular, thoracic, and thoracoscopic laparoscopic procedures requiring vessel ligation and tissue division. Compatible with vessels up to 7mm and tissue bundles fitting within device jaws. Contraindicated for tubal sterilization or tubal coagulation.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Aragon Surgical-Caiman 5 Laparoscopic Instrument Premarket Notification
JUL 1 1 2011
## SECTION 5
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
# REGULATORY AUTHORITY
Safe Medical Devices Act of 1990, 21 CFR 807.92
#### COMPANY NAME/CONTACT
Alan Curtis Aragon Surgical, Inc. 1810 Embarcadero Road, Suite B Palo Alto, CA 94303
### NAME OF DEVICE
| Trade Name: | Aragon Surgical RF System<br>Caiman 5 Laparoscopic Instrument |
|------------------------|-------------------------------------------------------------------------------------|
| Common Name: | Electrosurgical System |
| Device Product Code: | GEI |
| Classification Name: | Electrosurgical Cutting and Coagulation Device and<br>Accessories (21 CFR 878.4400) |
| Device Panel: | General Surgery/Restorative Devices |
| Device Classification: | Class II |
#### PREDICATE DEVICES
- Aragon Surgical RF System (K090306) .
- ValleylabLigaSure™ LigaSure 5mm Laparoscopic Sealer . and Divider (K031011)
- ValleylabLigaSure™ LigaSure 5mm Blunt Tip Laparoscopic Sealer . and Divider (K092879)
- ValleylabLigaSure Vessel Sealing System (K981916) .
- ValleyLab ForceTriad RF Generator (K070612) .
CONFIDENTIAL .
{1}------------------------------------------------
### DEVICE DESCRIPTION
The Aragon Surgical Caiman 5 Laparoscopic Instrument is provided as a sterile, single use device. The device is capable of vessel sealing, blunt dissection, grasping and dividing tissue enclosed within its iaws during open and laparoscopic procedures. The device is designed to be used with the Aragon Surgical RF Generator (K093075) and creates vessel ligation by the application of bipolar electrical RF energy and tissue division with a cutting blade.
The Aragon Surgical Lektrafuse RF Generator is designed to be used with Aragon Surqical RF instrumentation only. The Aragon Surgical Lektrafuse RF Generator is not compatible with any other handheld RF surgical instrumentation.
## INDICATION FOR USE STATEMENT
The Aragon Surgical Caiman 5 Laparoscopic Instrument is a dedicated bioolar electrosurgical instrument intended for use in general surgical and gynecologic laparoscopic procedures where ligation and division of vessels is desired. The instrument creates a seal by the application of bipolar electrosurgical RF energy to vascular structures (vessels) interposed between the jaws of the device. A cutting blade is actuated for the division of tissue.
Indications for use include general laparoscopic procedures, including urologic, vascular, thoracic, and thoracsocopic, and gynecological laparoscopic procedures where ligation and division of vessels is performed. These procedures include: laparoscopically assisted vaginal hysterectomies, Nissen fundoplication, colectorny, adhesiolysis, oophorectomy etc.
The Aragon Surgical Caiman 5 Laparoscopic Instrument can be used on vessels up to and including 7mm, and tissue bundles as large as will fit in the iaws of the instrument.
The Aragon Surgical Lektrafuse RF System has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures and should not be used for these procedures.
### SUBSTANTIAL EQUIVALENCE COMPARISON
### Comparison to Predicate Devices
The technological characteristics and indications for use of the Aragon Surgical Caiman 5 Laparoscopic Instrument are similar to those of the cited predicate electrosurgical devices, as well as the similar RF Systems distributed commercially by other manufacturers. These devices are equivalent in terms of design, materials, principal of operation, and product specifications. Any differences between the Aragon Surgical device and the predicate devices do not raise new issues regarding safety or effectiveness.
### PERFORMANCE DATA
Results of bench and pre-clinical evaluations were used to demonstrate that the Aragon Surgical Lektrafuse RF Generator and Caiman 5 Laparoscopic Instrument are substantially equivalent to the predicate devices and meets design, safety, and effectiveness criteria.
CONFIDENTIAL
{2}------------------------------------------------
Aragon Surgical-Caiman 5 Laparoscopic Instrument Premarket Notification
## CONCLUSION
Based on the design, materials, function, intended use, and pre-clinical evaluation, the Aragon Surgical Lektrafuse RF Generator and Caiman 5 Laparoscopic Instrument are substantially equivalent to the devices currently marketed under the Federal Food, Drug and Cosmetic Act. Sufficient data were obtained to demonstrate that the device is substantially equivalent to the predicate devices and raises no new safety or effectiveness issues. Therefore, safety and effectiveness are reasonably assured, justifying 510(k) clearance for commercial sale.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Aragon Surgical. Inc. % Mr. Alan Curtis 3617 Scott Street San Francisco, California 94123
# JUL 1 1 2011
Re: K110824
Trade/Device Name: Aragon Caiman 5 Laparoscopic Instrument, Model 50136 Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: June 29, 2011 Received: July 5, 2011
Dear Mr. Curtis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may ... publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{4}------------------------------------------------
Page 2 - Mr. Alan Curtis
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### SECTION 4
#### INDICATIONS FOR USE STATEMENT
510(k) Number:
Device Name:
Aragon Surgical Lektrafuse RF System-Caiman 5 Laparoscopic Instrument
Indications for Use:
The Aragon Surgical Caiman 5 Laparoscopic Instrument is a dedicated bipolar electrosurgical instrument intended for use in general surgical and gynecologic laparoscopic procedures where ligation and division of vessels is desired. The instrument creates a seal by the application of bipolar electrosurgical RF energy to vascular structures (vessels) interposed between the jaws of the device. A cutting blade is actuated for the division of tissue.
Indications for use include general laparoscopic procedures, including urologic, vascular, thoracic, and thoracsocopic, and gynecological laparoscopic procedures where ligation and division of vessels is performed. These procedures include: laparoscopically assisted vaginal hysterectomies, Nissen fundoplication, colectomy, adhesiolysis, oophorectomy etc.
The Aragon Surgical Caiman 5 Laparoscopic Instrument can be used on vessels up to and including 7mm, and tissue bundles as large as will fit in the jaws of the instrument.
The Aragon Surgical Lektrafuse RF System has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures and should not be used for these procedures.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Prescription Use X = _ ________________________________________________________________________________________________________________________________________________________________________ (Per 21 CFR 801.109)
Over-The-Counter Use
### Concurrence of CDRH. Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K110824
CONFIDENTIAL
000015
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.